RKI Instruments SDM-2009 User manual

www.rkiinstruments.com
SDM-2009 Docking Station
Standalone Configuration
Operator’s Manual
Part Number: 71-0261RK
Revision: B
Released: 2/17/17

Warranty
Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from
defects in materials and workmanship, and performance for a period of one year from
date of shipment from RKI Instruments, Inc. Any parts found defective within that
period will be repaired or replaced, at our option, free of charge. This warranty does
not apply to those items which by their nature are subject to deterioration or
consumption in normal service, and which must be cleaned, repaired, or replaced on
a routine basis. Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough
handling, or repairs procedures not in accordance with the instruction manual. This
warranty indicates the full extent of our liability, and we are not responsible for
removal or replacement costs, local repair costs, transportation costs, or contingent
expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND
ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI
INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO THE WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN
NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT,
INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND
CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS
PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users only by authorized
distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the
operation of this gas monitor and our warranty is limited to replacement of parts or our
complete goods.
Absorbent cartridges Batteries
Pump diaphragms and valves Filter elements
Fuses

Table of Contents
Table of Contents
Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
About the SDM-2009. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
System Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cautions & Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Chapter 2: Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
AC Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Single-Port AC Adapter . . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . . .. . . . . . .. 5
4-Port ACAdapter . . . . . .. . . . . . .. . . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . 5
USB Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Air Filter and Sample Tubing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Instrument Cradle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Power Jack. . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . . .. . . . . . . . 9
Sample Fittings. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . . .. . . . . .. . 9
PC Connection . . . . . . . . .. . . . . . .. . . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . .. 9
Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Chapter 3: Preparing to Use the SDM-2009. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Hardware Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Setting the Operational Parameters in Edit Mode. . . . . . . . . . . . . . . . . . . . . . . . 14
Bump Test &Calibration Parameters. . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . . .. . . . . . . 14
Turningon the SDM-2009 with anInstrument . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . 16
Setting theBump Test Parameters. . . . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . 18
Setting the CalibrationParameters. . . . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . 20
Connecting Calibration Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Table of Contents
Installing the Single Module Data Viewer Software . . . . . . . . . . . . . . . . . . . . . . 23
Chapter 4: Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Bump Testing a GX-2009 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Calibrating a GX-2009. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Charging an Instrument in a Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Recharging theBatteries After Performing a Bump Test
or Calibration . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . 37
Recharging theBatteries WithoutPerforming Any Operations. . . . . .. . . . . . . .. . . . . . . 38
Batteries TooDrained for Standalone Operation . . . . .. . . . . . . . .. . . . . . .. . . . . . .. . . 38
Calibration and Bump Test Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Available Memory inthe SDM-2009. . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . . .. . . . . . . 39
Copying Calibration and BumpTest Records. . . . .. . . . . . .. . . . . . .. . . . . . . . . .. . . . . 39
Clearing theSDM-2009’s Memory . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . . .. . . . . . . 40
Bump Test and CalibrationRecord Files . . . . . . . .. . . . . . .. . . . . . .. . . . . . . .. . . . . . . 40
Bump Testing or Calibrating and Saving Files to a Flash Drive
Multiple Times in OneDay . . . . . . .. . . . . . . .. . . . . . .. . . . . . . . .. . . . . . .. . . . . . .. . . 41
Chapter 5: Single Module Data Viewer Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Launching the Single Module Data Viewer Program . . . . . . . . . . . . . . . . . . . . . 42
Data ViewingWindow. .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . . .. . . . . . . 43
Using the Single Module Data Viewer Program . . . . . . . . . . . . . . . . . . . . . . . . . 43
Importing Files Into the Database. . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . 44
Organizing the Data . . . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . 45
Viewing the Data. . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . 46
Deleting Data . . .. . . . . . .. . . . . . .. . . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . 50
Changingthe Password . . . . . . . . . . . .. . . . . . .. . . . . . .. . . . .. . . . . . .. . . . . . .. . . . . 51
Exiting the Program . . . . .. . . . . . .. . . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . . .. . . . . 52
Spare Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
CAUTION: Read and understand this manual before using the SDM-2009. Also
read and understand the GX-2009 Operator’s Manual.

Overview • 1
Chapter 1: Introduction
Overview
This chapter briefly describes the SDM-2009 Docking Station and the Single Module
Data Viewer Program. This chapter also describes the SDM-2009 Docking Station
Standalone Operation Operator’s Manual (this document). Table 1 at the end of this
chapter lists the SDM-2009’s specifications.
About the SDM-2009
The SDM-2009 Docking Station is an advanced, reliable system that provides
charging, calibration, bump testing, and calibration and bump test records for the
GX-2009 portable gas monitor. It is designed to save the calibration and bump test
records to a USB flash drive (standalone functionality) or to be connected directly to a
computer (PC controlled functionality). If calibration and bump test records are stored
to a USB flash drive while operating in the standalone configuration, the Single
Module Data Viewer Program can then be used with a Windows-based personal
computer to retrieve calibration and bump test data files from the USB flash drive or
from the computer’s hard drive if the files have been transferred to the hard drive from
the flash drive. If you are using the PC Controller Program while operating in the PC
controlled configuration, you may retrieve instrument data, bump test, and calibrate
up to 10 instruments at once. Instrument information and data for each instrument
can be viewed directly using the PC Controller Program and can be printed from the
PC Controller Program. For instructions to use the SDM-2009 with the PC Controller
Program, see the SDM-2009 Docking Station PC Controlled Configuration Operator’s
Manual.
The purpose of this manual is to explain how to set up and use the SDM-2009 in
Standalone configuration. It also explains how to use the Single Module Data Viewer
Program. You will learn how to:
• install and launch the Single Module Data Viewer Program
• prepare the SDM-2009 for use
• perform a bump test
• perform a calibration
• save calibration and bump test records to a USB flash drive
• view, print, and export calibration and bump test records
• use the SDM-2009 to charge an instrument

2 • System Requirements
CAUTION: The GX-2009 detects oxygen deficiency and elevated levels of oxygen,
combustible gases, carbon monoxide, and hydrogen sulfide, all of
which can be dangerous or life threatening. When using the GX-2009,
you must follow the instructions and warnings in the GX-2009
Operator’s Manual to assure proper and safe operation of the
instrument and to minimize the risk of personal injury.
CAUTION: The operator of this instrument is advised that if the equipment is used
in a manner not specified in this manual, the protection provided by the
equipment may be impaired.
System Requirements
To use the Single Module Data Viewer Software, your personal computer must meet
the following requirements:
•Operating Systems: Windows® XP, Windows® Vista, Windows® 7,
Windows® 8, or Windows® 10.
•Processor: IBM® compatible PC running Pentium® 2 processor or equivalent
minimum
•Memory: 32 MB RAM minimum
•Hard Disk Space: 32 MB minimum
• Available USB port

Specifications • 3
Specifications
Table 1: SDM-2009 Specifications
Input Power 12 VDC
NOTE: AC Adapter with 100 - 240 VAC, 50/60 Hz,
0.6A input and 12 VDC, 1.2A output provided as
standard.
Environmental Conditions • For Indoor Use Only
• -10° C to 40° C, below 80% Relative Humidity, Non-
Condensing
Applicable Instruments GX-2009
Memory Capacity 64 KB
Maximum Record Size 256 bytes
Maximum Number of
Records Saved
200
Number of Calibration Gas
Cylinders
One calibration gas cylinder per bump test or
calibration
Standard Accessories • AC Adapter
• USB Flash Drive
• Single Module Data Viewer Program*
• SDM-GX Docking Station PC Controller
Program*
• Inlet Air Filter
• Instruction Manual
• 10 Foot Long Exhaust Tube
• One 3 Foot Long Tube for GAS Fitting
• 10 Foot Long 3/16 Inch Tube for GAS
manifolding (used for PC Controlled
Configuration only)
• 2 T-Fittings (used for PC Controlled
Configuration only)
• Check Valve (used for PC Controlled
Configuration only)
• USB Cable, Type A to Type B
* Not sent with SDM-2009. Download from www.rkiinstruments.com/sdm2009.

4 • About this Manual
About this Manual
The SDM-2009 Docking Station Standalone Configuration Operator’s Manual uses
the following conventions for notes, cautions, and warnings.
NOTE: Describes additional or critical information.
CAUTION: Describes potential damage to equipment.
WARNING: Describes potential danger that can result in injury or death.
Cautions & Safety Information
• Use only polyurethane sample tubing with the SDM-2009. Consult RKI
Instruments, Inc. for other materials.
• Do not subject the SDM-2009 to infrared or intense light. This may cause
communication errors.
• Do not expose the SDM-2009 to water.
• Do not subject the SDM-2009 to any hard impact.

Overview • 5
Chapter 2: Description
Overview
This section describes the SDM-2009 docking station. It is designed to be used on a
table top and consists of the AC adaptor, Type A to Type B USB cable, air filter, check
valve, 2 plastic T-fittings, sample tubing, instrument cradle, back panel, control panel,
status LEDs, and 2 USB ports.
AC Adapter
Single-Port AC Adapter
The single-port AC adapter is a wall plug style adapter with a 5 foot cable. The end of
the cable has a plug that connects to the power jack on the SDM-2009’s back panel.
The AC adapter is rated 100 - 240 VAC input, 12 VDC 1.2 A output.
4-Port AC Adapter
The 4-port AC adapter is a wall plug style adapter with four 5-foot cables. The end of
each cable has a plug that connects to the power jack on the SDM-2009’s back
panel. The AC adapter is rated 100 - 240 VAC input, 12 VDC 2.0 A output.
To Power Jack
on SDM-2009
Back Panel
Figure 1: Single-Port AC Adapter
To Power Jack
on SDM-2009
Back Panels
Figure 2: 4-Port AC Adapter

6 • USB Cable
USB Cable
A Type A to Type B USB cable is provided with the docking station. It is only for use
with the PC Controlled configuration. It is not used in the Standalone configuration.
Figure 3: USB Cable
Air Filter and Sample Tubing
A cylindrical particle filter with a short length of tubing is supplied with the SDM-2009
for installation to the AIR fitting on the back panel. The filter keeps particulate
contamination out of the docking station.
Three types of sample tubes are included with the docking station. One 3 foot length
of 3/16 inch ID polyurethane tubing is provided to connect the regulator on a
calibration cylinder to the GAS fitting on the back panel. One 10 foot length of 3/16
inch ID polyurethane tubing is provided for manifolding the GAS fitting. This tubing is
only needed in the PC Controlled configuration of the SDM-2009. In addition, a 10
foot length of 5/16 inch ID polyurethane tubing is provided for connection to the
exhaust fitting on the back panel to allow routing of the exhaust to a location such as
an open window where the exhaust can disperse.
Exhaust Tubing, 5/16 Inch
ID,10 feet
Particle Filter for Air Inlet
Calibration Gas Sample Tubing, 3/16
Inch ID, 3 feet Manifold Tubing, 3/16 Inch
ID,10 feet
Figure 4: Air Filter & Sample Tubing

Air Filter and Sample Tubing • 7
WARNING: Do not use an exhaust tube that is longer than 10 feet. The increased flow
restriction caused by a longer tube may affect gas response and cause
inaccurate calibration and bump test results.
A check valve is included with the SDM-2009 but is not needed for the Standalone
configuration. It is used for the PC Controlled configuration.
Two T-fittings are included with the SDM-2009. The larger fitting is for the exhaust
tubing. The smaller fitting is for the GAS tubing. Neither T-fitting is needed for the
Standalone configuration. They are used in the PC Controlled configuration.
Figure 5: Check Valve
For Exhaust
Tubing
For GAS
Tubing
Figure 6: T-Fittings

8 • Instrument Cradle
Instrument Cradle
The instrument cradle is a recessed area on the top of the SM-2009 that is designed
to accept the GX-2009. Insert the GX-2009 in the instrument cradle when you
perform a bump test or calibration. The cradle has charging contacts on the bottom
that mate with the contacts on the back of the GX-2009. There are sample ports in
the front of the cradle that match up with the GX-2009’s sensors and route air and
calibration gas to the sensors during a bump test or calibration. An infrared (IR) port
at the back of the cradle lines up with the GX-2009’s IR port when it is inserted in the
cradle and is used to communicate with the GX-2009. A retaining tab that holds the
GX-2009 in the cradle is located to the right of the IR port.
EDIT
ENTER
CAL.
POWER
CHARGE
SDM-2009 1SEC ON
3SEC OFF
COPY
GX-2009 OFF
BUMP
SDM-2009 Without GX-2009
H2SCO%LELO
2
RKI GX-2009
POWER
MODE
AIR
BUMP
GX-2009 OFF
COPY
1SEC ON
3SEC OFF
SDM-2009
CHARGE
POWER
CAL. EDIT
ENTER
Charging Contacts
SDM-2009 With GX-2009
IR Port
Instrument Cradle
Instrument Retaining
Tab
Figure 7: Instrument Cradle

Back Panel • 9
Back Panel
The back panel includes the power jack, sample fittings, and a USB PC connector.
Power Jack
The power jack is located in the bottom left corner of the back panel. The plug on the
end of the AC adapter cable mates to it.
Sample Fittings
Three sample fittings are located on the back of the SDM-2009. The AIR fitting is in
the upper left corner and draws air into the SDM-2009. The GAS fitting is to the right
of the AIR fitting and is used to connect the SDM-2009 to a calibration gas cylinder.
Both fittings accept 3/16 inch ID tubing.
An exhaust fitting is located in the upper right corner. It allows routing of the
exhausted calibration gas to a convenient location. This fitting accepts 5/16 inch ID
tubing. Even though the exhaust gas can be routed to an area to be safely dispersed,
the docking station should still be installed in a well ventilated area.
PC Connection
A type B USB connection is located beneath the exhaust fitting on the SDM-2009’s
back panel. It is only used for the PC Controlled configuration and is not used in the
Standalone configuration.
Gas Fitting
Exhaust Fitting
Back Panel USB
Port (Type B),
For Computer
Connection
Air Fitting
Power Jack
Figure 8: Fittings and Connections

10 • Control Panel
Control Panel
The control panel is used to setup and operate the docking station in the Standalone
configuration. It is located at the front of the docking station. It includes the control
buttons, the control button LEDs, and the CHARGE status LED.
Five control buttons are located on the control panel. From left to right they are BUMP
, CAL , EDIT ENTER, COPY, and POWER. The BUMP , CAL , and COPY
control buttons each have an LED above them that indicates the status of the function
controlled by that button. The CHARGE LED is located above the POWER button
and functions as a pilot LED, a system failure LED, and a charge indication LED.
Table 2: Control Button Functions
Control
Button Control Button Function(s) Control Button LED
Function(s)
BUMP • Initiates a bump test
• Cancels a bump test
• Moves down a list of parameters
• Decreases an adjustable
parameter
Indicates status of a bump test
in progress
CHARGE
1SEC ON
3SEC OFF
COPY
GX-2009 OFF POWER
SDM-2009
BUMP CAL.
Control Buttons
EDIT
ENTER
BUMP LED
CHARGE LED
COPY LED
CAL LED
Figure 9: Control Panel

Control Panel • 11
CAL • Initiates a calibration
• Cancels a calibration
• Clears data from docking station
memory (when used with COPY
button)
• Moves up a list of parameters
• Increases an adjustable parameter
Indicates status of a calibration
in progress
EDIT
ENTER
• Puts docking station into various
edit modes
• Makes a displayed parameter
editable
• Escapes or cancels an operation
• Turns off connected instrument
(when used with POWER button)
n/a
COPY • Copies data to a USB flash drive.
• Clears data from docking station
memory (when used with CAL
button)
• Indicates amount of docking
station memory used
• Indicates status of copying
function
• Indicates the result of a copy
operation
POWER • Turns on the docking station
• Turns off the docking station
• Turns off connected instrument
(when used with EDIT ENTER
button)
n/a
Table 2: Control Button Functions
Control
Button Control Button Function(s) Control Button LED
Function(s)

12 • Front Panel
Front Panel
A type A USB port is located on the front of the docking station. This port can be used
to save calibration and bump test data to a USB flash drive. This USB port is for use
only in the Standalone configuration of the SDM-2009.
NOTE: The SDM-2009 does not support connection of a computer to the front
USB port, only a USB flash drive.
Figure 10:Front Panel

Overview • 13
Chapter 3: Preparing to Use the SDM-2009
Overview
There are four tasks that must be completed before you can begin to use the
SDM-2009: hardware assembly, setting or confirming the bump test and calibration
parameters, connecting calibration gas, and installing the Single Module Data Viewer
Software on your computer. This chapter describes how to assemble the parts that
are shipped with the SDM-2009, how to set or view the bump test and calibration
parameters, and how to install the Single Module Data Viewer Software on a
Windows based personal computer.
Hardware Assembly
The hardware assembly consists of connecting the AC adapter, installing the air filter,
and connecting the sample tubing. Perform the following to complete the hardware
assembly:
1. Place the SDM-2009 on a convenient table top near an AC wall socket in a
well ventilated area. A location near a window that can be opened is best so
that the exhaust can be routed to the window.
2. Connect the AC adapter’s wall plug into a wall AC socket.
3. Insert the round plug on the end of the AC adapter’s cable into the power jack
on the back of the SDM-2009.
NOTE: If you have multiple SDM-2009s and are using a 4-port AC adapter,
plug each of the round plugs on the end of the AC adapter into the
power jack on the back of 4 separate SDM-2009s.
4. Install the air filter so that the arrow on the filter that indicates direction of flow
is pointing towards the AIR fitting. Push the open end of the flexible tube that is
on one end of the filter onto the AIR fitting on the back of the SDM-2009.
5. Install the 10 foot long 5/16 inch ID flexible tube that is included with the SDM-
2009 onto the exhaust fitting located in the upper left corner of the back panel.
Route the tube to an area where the exhaust can be safely dispersed, such as
an open window.
CAUTION: The maximum recommended length for the exhaust tube is 10 feet. Do
not use more than 10 feet of tubing or tubing with an ID of less than
5/16 inch for the exhaust tube or the bump test and calibration
accuracy may be adversely affected. The tube that is shipped with the
SDM-2009 has an ID of 5/16 inch and is 10 feet long.
6. Install the 3 foot long 3/16 inch ID tube that is included with the SDM-2009 on
the GAS fitting.

14 • Setting the Operational Parameters in Edit Mode
Setting the Operational Parameters
in Edit Mode
Once the hardware has been assembled, use Edit Mode to confirm or adjust the
bump test and calibration parameters before using the SDM-2009.
The bump test parameters define how long fresh air and calibration gas are applied to
an instrument during a bump test. They also define the tolerance used in determining
whether an instrument fails or passes a bump test and whether or not a calibration
automatically takes place if a bump test fails. The calibration parameters define how
long fresh air and calibration gas are applied to an instrument during a calibration.
The bump test and calibration parameters are saved in the SDM-2009’s memory. If a
parameter is changed with one particular instrument installed in the SDM-2009, the
change will be in effect for the bump test or calibration of any subsequent instruments
until the parameter is changed again.
Bump Test & Calibration Parameters
There are four bump test parameters and two calibration parameters. The two
calibration parameters, air sample time and calibration gas sample time, are also
bump test parameters. The parameters are described below. Table 3 below shows
the factory settings for the bump test and calibration parameters. If you wish to use
the factory settings, then you do not need to make any parameter adjustments. If you
wish to confirm or change the parameter settings, follow the instructions below in
“Setting the Bump Test Parameters” on page 18 or “Setting the Calibration
Parameters” on page 20.
Table 3: Bump Test & Calibration Parameter
Parameter Display
Tag
Available
Choices
Bump Factory
Setting
Cal Factory
Setting
Air Sample Time AIr • 15 seconds
• 30 seconds
• 45 seconds
• 60 seconds
• 90 seconds
15 seconds 45 seconds
Calibration Gas
Sample Time
GAS • 20 seconds
• 30 seconds
• 45 seconds
• 60 seconds
• 90 seconds
• 120 seconds
20 seconds 90 seconds

Setting the Operational Parameters in Edit Mode • 15
Air Sample Time (AIr)
The air sample time can be set separately for bump testing and calibration. It is the
length of time that the SDM-2009 will draw air through the AIR fitting on the back of
the docking station before an air adjust operation.
Calibration Gas Sample Time (GAS)
The calibration gas sample time can be set separately for bump testing and
calibration. It is the length of time that the SDM-2009 will draw calibration gas through
the GAS fitting on the back of the docking station during a bump test or calibration.
Bump Test Check Tolerance (CHE)
The bump test check tolerance only applies to bump testing. It determines how close
the instrument gas reading must be to the calibration gas concentration for each
channel during a bump test in order to pass the bump test. It is defined as a
percentage of the calibration gas concentration. The amount that the instrument gas
reading differs from the calibration gas concentration must be equal to or less than
this percentage of the calibration gas concentration. For example, if the tolerance is
set to 50%, and the %LEL calibration gas concentration is 50% LEL, then the bump
test gas reading for the LEL channel on the instrument must be 50 %LEL ± 25 %LEL.
The F10, F20, F30, etc. options represent a fast bump tolerance. If one of these “F”
type tolerances is selected and the gas reading for the sensor(s) being tested is
above the lower tolerance and below the upper tolerance within 10 seconds, the
sensor(s) pass bump testing, the gas application will be stopped, and the test will
move on to the next sensor(s) or to the fresh air purge. If the gas reading on any of
the tested channels is below the lower tolerance or above the upper tolerance within
10 seconds, the gas application will continue for the time period defined by the GAS
bump test parameter, and the pass/fail status of the sensor(s) will be determined at
that point. Using an “F” type tolerance allows calibration gas to be saved when the
tested sensors respond quickly and accurately.
Bump Test Check
Tolera nce
CHE • ± 10%
•± 20%
•± 30%
•± 40%
•± 50%
• ± F10%
• ± F20%
• ± F30%
• ± F40%
• ± F50%
F50% n/a
Automatic
Calibration
CAL • On
•Off
Off n/a
Table 3: Bump Test & Calibration Parameter
Parameter Display
Tag
Available
Choices
Bump Factory
Setting
Cal Factory
Setting

16 • Setting the Operational Parameters in Edit Mode
Consider the following scenario as an example.
• 50% LEL methane used for bump testing the combustible gas channel
• GAS bump test parameter set to 20 seconds
• Tolerance set to ± F50%, which means that the acceptable reading range is 25%
LEL - 75% LEL
The table below shows possible readings at 10 seconds and the effect on the gas
application.
Automatic Calibration (CAL)
Automatic calibration only applies to bump testing. It can be set to on or off. If it is set
to on, then the docking station will automatically perform a calibration if a bump test
fails.
Turning on the SDM-2009 with an Instrument
Do the following to turn on the SDM-2009 and establish a connection with an
instrument:
1. Confirm that the AC adapter is connected to the SDM-2009 and to an AC wall
socket.
2. Press and hold the SDM-2009’s POWER button. The LEDs will turn amber.
3. When the BUMP and CAL LEDs turn off, release the POWER button.
4. The COPY LED will be steadily on or off and the CHARGE LED will be blinking
green if the SDM-2009 is operating properly or solid red if there is a system
failure. The amount of free memory in the SDM-2009 will dictate the condition
of the COPY LED (see “Available Memory in the SDM-2009” on page 39).
5. Install an instrument into the instrument cradle.
6. Press and hold the POWER MODE button on the instrument until you hear a
beep, then release it. The instrument will begin its power up sequence (see
Table 4: Example Fast Bump Scenarios
Gas Reading 10 Seconds
Into Fast Bump Outcome
15% LEL (-70% of 50% LEL) • Gas application continues for the full 20 seconds
• Pass/fail determined at end of 20 seconds
30% LEL (-40% of 50% LEL) • Gas application stops
• Sensor passes bump testing
60% LEL (+20% of 50% LEL) • Gas application stops
• Sensor passes bump testing
80% LEL (+60% of 50% LEL) • Gas application continues for the full 20 seconds
• Pass/fail determined at end of 20 seconds
Table of contents
Other RKI Instruments Docking Station manuals

RKI Instruments
RKI Instruments SDM-2012 User manual

RKI Instruments
RKI Instruments SDM-3R User manual

RKI Instruments
RKI Instruments SDM-3R User manual

RKI Instruments
RKI Instruments SDM-6000 User manual

RKI Instruments
RKI Instruments SDM-03 User manual

RKI Instruments
RKI Instruments SDM-2012 User manual

RKI Instruments
RKI Instruments SDM-03 User manual

RKI Instruments
RKI Instruments SDM-2012 User manual