Use Chemstrip® Calibration Strips
(Cat. No. 11379194160).
Calibrate every 7 days or when the message
“REPEAT CALIBRATION” appears on the
display.
1 If the message “REPEAT CALIBRATION”
is in the display, press the START button.
The message “START CALIBRATION” is
displayed.
If the Urisys 1100® analyzer is in Ready-
to-Measure mode, press the left function
key to select “Calibr.” Next, the message
“START CALIBRATION” is displayed.
2 Place the calibration strip, with pads facing
upward, on the test strip tray. About 2 mm
of strip must be held under the clip.
3 Press the START button.
4 If the calibration is valid, the message
“CALIBRATION O.K.” prints.
5 If one of the following messages appears
on the display, refer to the Operator’s
Manual for further action:
n “REFERENCE PAD ERROR!”
n “CALIBRATION INVALID!”
n “CALIBRATION ERROR!”
Calibrating the Urisys 1100® Urine Analyzer
Clean the test strip tray with water and
disinfectant once daily.
1 Power off the instrument.
2 Pull the test strip tray out of the instrument.
3 Rinse under running water.
4 Remove any crystalline deposits, especially
those contaminating the retaining bar
mechanism or the cogs on the underside of
the test strip tray, with a soft brush.
5 Wipe down the test strip tray with 70 %
alcohol or 10 % bleach. Dry with a soft lint-
free cloth.
6 Install the test strip tray, being careful not
to touch the gray reference pad.
7 Power on the instrument to resume testing.
Daily Maintenance
1 Open printer cover by pressing area directly
below and to the right of the printer paper
slot. Remove the old paper roll.
2 Insert the new roll of printer paper with the
loose end coming from the bottom of the
roll. Draw paper forward over the edge of
the paper slot.
3 Close the cover again by pressing until it
locks audibly into position. Make sure that
the paper goes through the paper slot.
ATTENTION: To remove the printed test report,
tear off the paper by pulling it in front across
the edge.
Changing Printer Paper
If you need assistance, contact Roche
Diagnostics Customer Support Center at
1-800-428-4674, Monday-Friday,
7 a.m.-11 p.m. EST/EDT.
Technical Service