59044E/2015-02
Subject to technical change without notice. Printing and other errors reserved.
TECHNICAL INFORMATION.
COLD CHAIN TEMPERATURE LOGGER
Applications
Monitoring of the cold chain during transport of sensitive
freight such as pharmaceuticals, foods and technical pro-
ducts. The logger is designed for single use only.
TL-CC1
Type of logger Single use
Parameter Temperature
Temperature sensor NTC thermistor
Accuracy ±0.5 °C (-30…70 °C)
Resolution 0.1 °C
Range of application
(temperature)
-30…70 °C / -22...158 °F
Range of application
(humidity)
< 80 %RH
Protection IP65 (when placed in plastic bag)
Weight Approx. 10 g
Battery CR2032 (not replaceable)
Dimensions without bag 80(L) x 43(W) x 2.5(D) mm
Dimensions with bag 105(L) x 55(W) x 2.5(D) mm
Logging interval 0.5/5/10/30/60/90/120 min
Start delay 0.5/5/10/30/60/90/120 min
Alarm ranges -20…-10 °C, -10…0 °C, 2…8 °C, 0…15 °C,
0…25 °C, 15…25 °, freely selectable values
Alarm type Deactivated, single or cumulative
Storage period 12 months
Storage capacity 8192 data points
MARK function Marking of up to 8 points during transport
Operation START/STOP button, MARK button
LED indicators LED « Low » fl ashes red
when value below alarm point
LED « Hi » fl ashes red
when value above alarm point
LED « OK» fl ashes green
during data logging
LED « Battery » fl ashes red
when low battery
Communication USB 2.0 port, type A
Operating system Windows
Conformity GxP, EN 12830, FDA 21 CFR Part 11
and GAMP5
Delivery
In box of 10 with instruction manual: TL-CC1-10
Quick Guide
• Programing: plug into a USB port
• IP65: place into the sleeve and seal closed
• Start logging: press START key for 2 seconds
• Bookmarking: press MARK key
• Downloading of the PDF fi le: remove the logger
from the sleeve, plug into a USB port
PDF Report
5
10
15
20
25
30