Schmitz OPX mobilis 200 User manual

Letzte Änderung: Monday, 4. July 2005 10:16
Operating manual
Operating table OPX mobilis

2Operating manual
This operating manual includes instructions for the
operation of the mobile operating tables
–OPX mobilis 200 (161.201)
–OPX mobilis 200/G (161.205)
–OPX mobilis 300 C (161.311)
–OPX mobilis 300 C/G (161.315)
–OPX mobilis 300 CL (161.321)
–OPX mobilis 300 CL/G (161.325)
–OPX mobilis 300 CE (161.331)
–OPX mobilis 300 CE/G (161.335)
–OPX mobilis 300 CLE (161.341)
–OPX mobilis 300 CLE/G (161.345)

Operating table OPX mobilis –
E
DITION
01-05-GB ID.-Nr.: 02007844
3
Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
About this operating manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Symbols used in the text . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Summary of safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Description of the operating table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Keeping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Transportation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Mobility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Potential equalization socket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Battery charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Adjusting the table-top . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Longitudinal shift function of the table-top . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Integrated kidney bridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Leg plates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Head plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Back section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Arm rest 101.107 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Add-on kidney bridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Shoulder arthroscopy support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Accessories for urology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Further accessories of the Schmitz range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Cleaning and disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Contents
4Operating manual
Check-ups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Electromagnetic Compatibility (EMC). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Electromagnetic Emissions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Electromagnetic Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Electromagnetic Immunity for Non-Life Supporting Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Separation distance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Malfunctions and repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Product designation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Nameplates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Symbols employed on the product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Standards applied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Declaration of Conformity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
After sales service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

Operating table OPX mobilis –
E
DITION
01-05-GB ID.-Nr.: 02007844
5
Introduction
About this operating manual
In this paragraph you will find information about the
layout of this operating manual and explanations
regarding the marks and symbols used.
This operating manual includes instructions for the use of
the operating tables OPX mobilis, also called operating
tables in the following.
Our products are constantly being improved, this is why
constructional changes carried out after printing of this
operating manual could not be taken into consideration.
In case of any questions, we therefore kindly ask you to
contact Schmitz u. Söhne.
This operating manual is to be read and observed by any
person using or handling the operating tables OPX
mobilis.
In addition to the operating manual and to the obligatory
regulations for the prevention of accidents effective in the
user’s country and on the site of use, the acknowledged
rules for safe and professional work are also to be
observed.

Introduction
6Operating manual
Symbols used in the text
In this operating manual following designations or signs
are used for pieces of information of special importance
g
Danger!
This symbol will appear wherever safety
instructions are designed to protect people from
physical harm. The symbol stands for imminent
danger of death or serious injury.
v
Caution!
This symbol will appear where situations are
described, which might be dangerous, and
which might inflict slight injuries.
a
Attention!
This symbol will appear in front of such hints
which shall prevent the table or other
equipment from being damaged.
h
This symbol will appear in front of additional
helpful pieces of advice.
•A dot in front of the text means:
This is what you have got to do.
–A dash in front of the text means:
This is part of a listing.

Operating table OPX mobilis –
E
DITION
01-05-GB ID.-Nr.: 02007844
7
Safety instructions
The operating table OPX mobilis has been constructed
according to the latest state of engineering and according
to the acknowledged rules of safety engineering.
Nevertheless, its use may inflict danger to life or physical
safety of the user or of third parties, or impairment to the
operating table or to other material assets.
Do not use the operating table unless in perfect condition
and only for its intended use, with regard to safety and
possible dangers, and observing the operating manual!
Any malfunction which may affect the safety has to be
eliminated immediately!
Electrically conductive double castors, electrically
conductive pads and a potential equalization socket are
standard features of the operating table OPX mobilis. It
may be used inside rooms where the electrical installation
conforms to German VDE (i. e. the Association of
German Electrical Engineers) 0100-710 standard or to
an equivalent national standard. If the operating table
OPX mobilis is equipped with non-electrically conductive
(coloured) pads, it must not be used in the presence of
flammable anaesthetic mixtures.
Always keep this operating manual at hand at the site of
use of the operating table!
Additionally to the operating manual, observe the
general rules implied by the law and otherwise obligatory
for accident prevention and environmental protection!
Do not carry out any modifications, extensions or
reconstructions of the operating table unless approved by
the manufacturer.
Spare parts have to meet the requirements stipulated by
the manufacturer. This is always guaranteed when using
original spare parts.
Observe the intervals prescribed or stated in this
operating manual for periodical check-ups!
Take care that running and process materials as well as
parts replaced are disposed of safely and with minimum
environmental impact!
Intended use
According to German VDE 0100-710 standard, the
operating table OPX mobilis may be used inside rooms of
the application groups 0, 1 or 2. It is exclusively designed
for purposes of human medicine. The operating table
serves to position patients during an examination or
during surgical interventions. The nursing staff has to take
care to position the patients in such a way as to prevent
any danger to their respiration, to their nervous system or
to their circulation. This is especially important when
patients are under anaesthetics. Any use apart or beyond
these purposes is not intended. The manufacturer is not
liable for any damage resulting of such non-intended
use, which would be entirely at the user’s risk.
The operating table may only be handled by persons who
have been briefed in its professional handling and who
have familiarized themselves with the product by means
of this operating manual.
Intended use also means following the operating manual
and observing the conditions for inspection and
maintenance.

Introduction
8Operating manual
Summary of safety instructions
g
Danger!
Operating tables with an “E” in their type
designation are equipped with rechargeable
batteries. Do not charge the batteries during a
surgical intervention inside the operating
theatre.
g
Danger!
Patients run the risk of falling off the operating
table unless they are secured on the table-top.
Patients lying on the operating table have to be
secured to the table-top according to
professional standards.
g
Danger!
Patients run the risk of getting hurt when the
operating table is moved while he or she is lying
on top in knee-elbow-position. Do not move
the table with the patient in knee-elbow-
position.
g
Danger!
Patients run the risk of getting hurt if the
operating table is moved with an attached
extension device, due to its diminished stability
against tilting. Do not move the operating table
with an attached extension device when a
patient is lying on top.
g
Danger!
If an extension device is attached to the
operating table, its stability against tilting will be
diminished. Always use the props of the
extension device during an operation.
g
Danger!
If patients are not positioned correctly, there
may be danger to their respiration, to their
nerve cords or to their circulation. Always
position patients in such a way as to prevent any
possible danger to their respiration, to their
nerve cords or to their circulation.
g
Danger!
In case the components of the table-top hit
against an obstacle during adjustment, the
operating table may tilt. Remove possible
obstacles before adjusting the operating table
or its individual components. When folding the
leg plate down, make sure it cannot hit the
floor.
g
Danger!
If its castors are released, the operating table
may move unintentionally during patient
transfer. Always secure the operating table by
means of the central brake before transferring
the patient.
g
Danger!
Patients run the risk of getting hurt by
unforeseen movements of the table accessories.
When attaching accessories, make sure that
they are properly fixed. Do not use worn or
damaged accessories.
g
Danger!
Patients run the risk of getting hurt by
unforeseen movements of the operating table.
Always secure the patient to the table-top
according to the recommended practices of
patient positioning.
g
Danger!
When HF surgical equipment or defibrillators
are being used, there is danger of the patient’s
suffering burns if he or she touches the metal
parts of the operating table or of its accessories,
or if he or she is lying on soaked sheets or cloths
covering the electrically conductive pads of the
table-top. Always observe the manufacturer’s
instructions for use.

Operating table OPX mobilis –
E
DITION
01-05-GB ID.-Nr.: 02007844
9
g
Danger!
Patients run the risk of getting hurt if the
longitudinal shift function is released with the
table positioned in Trendelenburg or Reverse
Trendelenburg angle of more than 5°. Reduce
the inclination angle of the table-top before
activating the longitudinal shift function.
g
Danger!
If the operating table OPX mobilis is used in the
presence of flammable anaesthetic mixtures,
the floor must be electrically conductive.
v
Caution!
The operating table is only stable on a level
floor. Choose a horizontal and safe site of
installation for the operating table.
v
Caution!
In case the table-top is adjusted into extreme
positions, there is danger of jamming one’s
finger in certain points. Do not grab below the
operating table or behind the frames of the
table-top elements.
v
Caution!
In order to ensure a gentle patient positioning
when using the add-on kidney bridge, use an
additional pad on top of the break. (This pad
comes together with the add-on kidney bridge.)
v
Caution!
If the operating table is positioned on a moist or
freshly cleaned floor, please keep in mind that
the frictional forces will be diminished. The
operating table may therefore be moved by
external force even if the double castors are
blocked. Make sure before starting an
intervention that the operating table is in a
sufficiently stable position.
g
Danger!
For tables with longitudinal shift feature, the
position of the table-top will bear influence on
the stability of the operating table. Always shift
the table-top into such a position, within the
limits of the specific application, as to improve
the stability of the table.
h
If there is an extension device attached to the
table or if the patient is positioned in knee-
elbow position, the stability of the table will be
diminished. Improve the stability of the table by
moving the table into such a position as to
make the double castors at the foot end of the
table point to the foot end.

Introduction
10 Operating manual
h
An unfavourable position of the double castors
may diminish the stability of the table. It may
help to move the operating table in such a way
as to raise the stability. Please see the examples
given in the illustration.
h
A patient weight of max. 135 kg allows all
surgical interventions to be carried out in either
orientation of the patient (patient’s head lying
at head or foot end of the table), provided the
castors of the operating table are blocked.
h
In the case of normal patient orientation and
table-top in mid-position (for operating tables
with longitudinal shift function) the operating
table may be charged with a patient weight of
max. 185 kg. Upon patient transfer, or when
letting the patient mount the operating table,
make sure that the table-top is loaded mainly
above the table column. Familiarize yourself
with the reactions of the operating table under
this kind of charge before starting a surgical
intervention.

Operating table OPX mobilis –
E
DITION
01-05-GB ID.-Nr.: 02007844
11
Description of the operating table
In the following operating manual the terms left, right,
front, and rear are used as seen by a person sitting on the
operating table in normal position.
Seat section
Back section
Leg plate
Head plate
Double castor
Control panel
for electric version
Pedal for brake
and for
directional
castor
Foot pump lever
Selector lever
Potential equalization socket

Description of the operating table
12 Operating manual
The operating tables OPX mobilis can be equipped
differently, depending on the model.
All operating tables are mobile. They are equipped with
four electrically conductive swivel-type double castors
with central brake. The operating table models with a “C”
in their type designation have got an additional
directional castor.
All operating tables are equipped with mechanically
operated hydraulic pumps. The operating table of the
200 model series have got two hydraulic adjustment
functions: Height adjustment Raise/Lower and
Trendelenburg/Reverse Trendelenburg. The tables of the
300 model series have got an additional lateral
adjustment function.
Operating tables with an “E” in their type designation are
equipped with an additional electrically-driven hydraulic
pump. In this case the operating table is equipped with a
hand-held remote control unit as well, and with
rechargeable batteries installed inside the floor pan.
Operating tables with an “L” in their type designation are
equipped with a table-top with longitudinal shift function.
By means of this feature, such parts of the patient’s body
can be screened which under normal circumstances are
inaccessible to the C-arm equipment.
Operating tables with a “G” in their type designation are
equipped with an integrated kidney bridge. The
integrated kidney bridge is adjusted hydraulically by
means of the foot pump.
All operating tables can be equipped with X-ray cassettes,
which can be inserted below the table-top either from the
head end or from the foot end.
All operating tables can be adapted optimally to various
applications by means of accessories.
According to the regulations of European Standard
60601-2-46, the operating tables are designed to carry
a maximum patient weight of 135kg. In case a higher
charge is intended, please observe the instructions given
below.
h
A patient weight of max. 135 kg allows all
surgical interventions to be carried out in either
orientation of the patient (patient’s head lying
at head or foot end of the table), provided the
castors of the operating table are blocked.
h
In the case of normal patient orientation and
table-top in mid-position (for operating tables
with longitudinal shift function) the operating
table may be charged with a patient weight of
max. 185 kg. Upon patient transfer, or when
letting the patient mount the operating table,
make sure that the table-top is loaded mainly
above the table column. Familiarize yourself
with the reactions of the operating table under
this kind of charge before starting a surgical
intervention.
Pads
The operating tables are normally equipped with black
antistatic pads. Coloured pads, which, however, are not
electrically conductive, are available upon special
request. The antistatic black version fulfils the limits of
resistance according to ISO 2882. Coloured pads are not
electrically conductive, which means that in this case the
operating table is not explosion-proof.
Safety devices
All operating tables are equipped with electrically
conductive swivel-type double castors, with (normally)
electrically conductive pads, and with a potential
equalization socket.

Operating table OPX mobilis –
E
DITION
01-05-GB ID.-Nr.: 02007844
13
Commissioning
Keeping
Inside its original packing the operating table may be
exposed for a period of 15 weeks to environmental
conditions within the limits of:
Transportation
Upon leaving the factory, the operating tables are
positioned fit for transportation. In case of further
transportation at a later date, we recommend to adjust
the operating table into the same positions.
•In order to transport the operating table, adjust seat
and back section into horizontal position.
•Remove head and leg plates, and detach other
accessories.
•Adjust the table-top into minimum height.
Unpacking
Carry the operating table, if possible, in its original
packing to its final site of use. Check the operating table’s
condition.
Any possible damage occurred during transportation
should be reported immediately to Schmitz u. Söhne. You
will find their address and telephone no. on the last page
of this operating manual.
Ambient temperature -5°C to +50°C
Relative atmospheric humidity 10% to 95%
Atmospheric pressure 500 hPa to 1,060 hPa

Commissioning
14 Operating manual
Electrical connection
The operating table can be equipped with an electrically
driven hydraulic pump. In that case it is also equipped
with a hand-held remote control unit activating the
electro-hydraulic pump. The hydraulic pump draws its
energy from the integrated batteries.
If the main switch is switched on, the battery state is
indicated by means of the LEDs at the hand-held remote
control unit.
Hand-held remote control unit
Operating tables with an “E” in their type designation are
equipped with an electro-hydraulic pump and a hand-
held remote control unit.
The hand-held remote control is connected to the
operating table OPX mobilis by means of a plug-type
connector. You will find the connector socket for the hand-
held remote control unit at the top of the height
adjustment column.
The connector plug and socket are marked by dots. The
plug will lock inside the socket when the relative dots are
brought into alignment.
•In order to connect the hand-held remote control unit
to the table, insert the plug into the socket, with the
marking dots in alignment, and plug the plug into the
socket until it locks.
•In order to disengage the hand-held control unit, draw
the front part of the plug back. Then pull the plug out
of the socket.
Operating tables with an “E” in their type designation are
equipped with rechargeable batteries.
Batteries
The batteries have to be recharged as soon as the yellow
LED on the remote control unit is on. You will find details
how to charge the batteries under “Charging the
batteries” on page 18.
The electric drive of the hydraulic pump can be activated
additionally by means of the main switch. You will find the
main switch on the control panel at the left hand side of
the floor pan.
h
The table-top of the operating table can always
be adjusted mechanically, without electric drive,
by means of the foot pump.
•The electrically driven hydraulic pump of the operating
tables with an “E” in their type designation has to be
activated by means of the main switch.
Marking dots

Operating table OPX mobilis –
E
DITION
01-05-GB ID.-Nr.: 02007844
15
Operation
Mobility
Depending on the model, the operating tables
OPX mobilis can equipped in different ways.
All operating tables are mobile. Depending on the
model, they are either equipped with
–four double castors with central brake
–four double castors with central brake and an
additional directional castor.
Moving and locking the operating table
g
Danger!
Patients run the risk of falling off the operating
table unless they are secured on the table-top.
Secure the patient to the table-top according to
professional standards.
•Move the table into the desired position and then turn
the castors outwards in order to improve the stability
of the operating table.
h
The position of the castors will bear influence
on the stability of the operating table.
v
Caution!
If the operating table is positioned on a moist or
freshly cleaned floor, please keep in mind that
the frictional forces will be diminished. The
operating table may therefore be moved by
external force even if the double castors are
blocked. Make sure before starting an
intervention that the operating table is in a
sufficiently stable position.

Operation
16 Operating manual
The operating tables, except the standard model, are
equipped with an additional directional castor. The
directional castor can be activated by means of a foot
pedal for the brake and for the directional castor.
The pedal for the brake and for the directional castor can
have three different positions.
–If the pedal is raised, the operating table will move in
straight run. An additional directional castor
positioned below the height adjustment column is now
touching the ground.
–If the pedal is in mid position, the operating table can
easily be moved in any direction. The directional
castor is retracted.
–If the pedal is pressed down, the operating table is
secured against unintentional movement.
g
Danger!
Patients run the risk of getting hurt when the
operating table is moved while they are lying on
top in knee-elbow-position. Do not move the
table with a patient lying on top in knee-elbow-
position.
g
Danger!
Patients run the risk of getting hurt if the
operating table is moved with an attached
extension device, due to its diminished stability
against tilting. Do not move the operating table
with an attached extension device when a
patient is lying on top.
•In order to move the operating table in straight run,
raise the pedal completely.
•In order to move the operating table easily in any
desired direction, raise the pedal into mid-position.
•Move the operating table into the desired location.
•Secure the operating table afterwards by pressing the
pedal completely down.
Selector lever
Pedal for brake
and for
directional castor
Foot pump lever

Operating table OPX mobilis –
E
DITION
01-05-GB ID.-Nr.: 02007844
17
Potential equalization socket
The operating table is equipped with a potential
equalization socket. You will find the potential
equalization socket at the front side of the floor pan.
g
Danger!
Patients may be exposed to danger by
electrostatic tensions. By means of the potential
equalization socket electrostatic tensions are
conducted to earth. Always connect the
potential equalization socket of the operating
table to earth before use.
•Before putting the operating table into operation,
connect the potential equalization socket to the earth
connection.

Operation
18 Operating manual
Battery
If the operating table is equipped with an electro-
hydraulic pump, this pump draws its energy from
batteries installed inside the floor pan. The batteries have
to be recharged regularly in order to avoid an exhaustive
discharge.
g
Danger!
Plugging the plug of the charging cable in or
out may produce sparks. In order to avoid
danger of explosion, the batteries must not be
charged inside the operating theatre during
surgical interventions.
Charging the batteries
h
The batteries should be recharged regularly, e.
g. overnight. The batteries cannot be
overcharged if you use the battery charger
which comes with the table.
a
Attention!
If the batteries of the operating table are
charged by means of a battery charger other
than the one supplied with the table, the
batteries may get damaged. Always use the
original battery charger for charging the
batteries.
The hand-held remote control unit is equipped with LEDs.
The green LED is on when the batteries are fully charged.
The yellow LED will be on as soon as the batteries have to
be recharged. When the red LED is on, the electro-
hydraulic pump can no longer be activated before the
batteries have been recharged.
•If the red LED is on, switch the main switch at the
control panel off, use the foot pump in order to adjust
the table-top until the operation is finished, and
recharge the batteries afterwards.
h
The fuses for the electro-hydraulic pump (F1 =
10 A, time lag type, 5 x 20 mm) and for the
charging current (F2 = 3.15 A, time lag type, 5
x 20 mm) are also inserted into the control
panel.
Fuse
F1 = 10A
Main switch
Fuse
F2 = 3.15A
Charging
socket

Operating table OPX mobilis –
E
DITION
01-05-GB ID.-Nr.: 02007844
19
Battery charger
The battery charger is fit for a voltage input of between
90V to 264V. It can be equipped with the feed cable
corresponding to the requirements on site.
•In order to charge the batteries, move the operating
table out of the operating area.
h
The batteries cannot be charged as long as the
main switch is on.
h
The charging current is up to 2A. Make sure
that the correct fine-wire fuse (3.15A, slow-
blow type) has been installed inside the fuse
holder F2 for charging current. This is especially
important for the more ancient models of the
OPX mobilis range.
•Switch the main switch at the control panel off.
The charging cable of the battery charger is connected to
the operating table by means of a plug contact. You will
find the socket for the battery charger at the control panel
at the left hand side of the floor pan.
The plug and the charging socket are marked by dots.
The plug will lock inside the charging socket if it is
plugged into the charging socket with the dots in
alignment.
h
The control panel is equipped with a covering
sheet, which is designed to protect the control
electronics against electrostatic discharge
current. Fold this covering sheet up in order to
get access to the control panel.
•Plug the plug of the battery charger into the charging
socket so that the marking dots are in alignment and
press the plug in until it locks inside the charging
socket.
•Connect the battery charger to the mains power
supply.
The operating modes of the battery charger are indicated
by means of a 3-coloured LED.
LED, green, yellow or orange-coloured
Charging
current
Charging diagram
Capacity
Charging voltage
Electric
current (A)
Voltage (V)
Rapid charge
LED=orange
Charge
LED=yellow
Compensation charge
LED=green

Operation
20 Operating manual
If the batteries are drained, it will take approx. 8 to 10
hours to recharge them. The batteries are charged as
soon as the LED is green. The battery charger can then be
disconnected.
•In order to detach the charging cable, pull the front
part of the plug back and pull the plug out of the
socket.
Now the table-top can again be adjusted with the aid of
the electro-hydraulic pump.
h
With fully charged, practically new batteries, the
table can carry out all its functions approx. 60
times when charged by a patient weight.
This manual suits for next models
9
Table of contents
Other Schmitz Medical Equipment manuals
Popular Medical Equipment manuals by other brands

laerdal
laerdal Nursing Baby VitalSim Directions for use

surfacide
surfacide Helios instructions

Accurate Bio-Medical Technology
Accurate Bio-Medical Technology HS20A Operator's manual

CYMEDICA Orthopedics
CYMEDICA Orthopedics QB1 Quick reference guide

Animas
Animas vibe Quick reference guide

Orliman
Orliman THERA GO TGO480 INSTRUCTIONS FOR USE, STORAGE AND WARRANTY