ScienTECH AC2500 User manual

PN9458Q
Vector™ Series
Calorimeter
Laser Power
Measurement
Operating Manual
Thank you for choosing a Scientech Vector™calorimeter detector. Scientech’s employees are pleased to
provide you with a product designed for years of reliable and accurate service. Please read this operating
manual before using your detector and power meter. This reference information will allow you to fully
understand the capabilities of the product. The detector is intended to be used only in the manner outlined
in this manual. Operation not within specifications for the product may cause product damage.

PN9458Q 2
TABLE OF CONTENTS
Calorimeter Operating Parameters............................................................................................................ 3
Vector™ Calorimeter Specifications........................................................................................................ 4
Absorption of HD Material................................................................................................................ 6
Environmental Specifications................................................................................................................... 6
CE Certification................................................................................................................................. 7
Unpacking and Setup................................................................................................................................ 7
Correcting Vector™ HD Calorimeter Operating Parameters for use at Different Wavelengths.............. 8
Vector™ Calorimeter Operation Without Power Meter........................................................................... 8
Cable Requirements........................................................................................................................... 8
Operation of VectorCalorimeters with a Digital Volt Meter......................................................... 9
Calibration of Vector™ Calorimeters Using Electric Substitution Heating............................................. 9
Vector™ Large Aperture Calorimeter Specifications ............................................................................ 12
Unpacking and Set Up ............................................................................................................................ 14
Cable Requirements for Large Calorimeters................................................................................... 14
LARGE Calorimeter Operation Without Powre Meter.......................................................................... 15
Operation of Large Aperture Calorimeters with a Digital Volt Meter............................................ 15
Calibration of Large Aperture Calorimeters.................................................................................... 16
Scientech Calibration Service................................................................................................................. 17
Limited Warranty.................................................................................................................................... 17
Return Material Procedure...................................................................................................................... 18
Disposal of Electronic Equipment.......................................................................................................... 18
HD Absorption vs. Wavelength.............................................................................................................. 20

PN9458Q 3
CALORIMETER OPERATING PARAMETERS
Calorimeter 1: Calorimeter 2:
Model No:
Serial No:
Model No:
Serial No:
Calibration Wavelength nm Calibration Wavelength: nm
Output Sensitivity
(S): Time Constant
(1/e): Calibration
Temp:
V/W Output Sensitivity (S): V/W
sec. Time Constant (1/e): sec.
oC Calibration Temp: °C
Sub. Heater Resistance (Rc): ohms Sub. Heater Resistance
(Rc): ohms
Sub. Heater Voltage (Vh): volts Sub. Heater Voltage
(Vh): volts
Sub. Heater Wattage (Wh) watts Sub. Heater Wattage
(Wh): watts

PN9458Q 4
VECTOR™CALORIMETER SPECIFICATIONS
The words "calorimeter" and "detector" are synonymous as used in this manual.
Table 1. Vector™ 8mm and 25mm Aperture Calorimeters
Model
AC2500
AC25HD
ACX25HD
AC2501
ACX2501
AC25UV
AC2504
Type of Absorber
Surface
Surface
Surface
Volume
Volume
Volume
Volume
Aperture Diameter
25.4 mm
25.4 mm
8 mm
25.4 mm
8 mm
25.4 mm
25.4 mm
Spectral Response
.25-35 µm
.193-12 µm
.4-2 µm
.266-1.2 µm
.4-1.2 µm
.193-.36 µm
.85-4.2 µm
Average Power Maximum
10 W
Average Power Minimum
1 mW when installed in an Isoperibol Enclosure
Noise Level
10 µW or µJ
Maximum Power Density
200 W/cm²
1.5 kW/cm²
12 kW/cm²
Note 1
Note 2
Note 3
Note 4
Maximum Peak Power Density
1 MW/cm²
100 MW/cm²
800 MW/cm²
Note 5
8.5 GW/cm²
Note 6
Note 7
Maximum Single Pulse Energy
10 J
Maximum Energy Density
Note 8
Note 9
Note 10
Note 11
Note 12
Note 13
Note 14
Precision
< 1 %
Accuracy
±
3 %
Response Time
3 sec when connected to a Scientech power meter in Watts Mode
Dimensions DxL - inches
cm
3.75 x 2.2
9.53 x 5.6
3.75 x 2.2
9.53 x 5.6
3.75 x 3.82
9.53 x 9.7
3.75 x 2.2
9.53 x 5.6
3.75 x 3.82
9.53 x 9.7
3.75 x 2.2
9.53 x 5.6
3.75 x 2.2
9.53 x 5.6
Weight - pounds
kg
1.5
0.68
1.5
0.68
1.7
0.77
1.5
0.68
1.7
0.77
1.5
0.68
1.5
0.68
Power Meter Compatibility
H410, H410D, S310, S310D
Table 2. Vector™ 16mm and 50mm Aperture Calorimeters
Model
AC5000
AC50HD
ACX50HD
AC5001
ACX5001
AC50UV
AC5004
Type Absorber
Surface
Surface
Surface
Volume
Volume
Volume
Volume
Aperture Diameter
50.8 mm
50.8 mm
16 mm
50.8 mm
16 mm
50.8 mm
50.8 mm
Spectral Response
.25-35 µm
.193-12 µm
.4-2 µm
.266-1.2 µm
.4-1.2 µm
.193-.36 µm
.85-4.2 µm
Average Power Maximum
30 W
Average Power Minimum
40 mW
Noise Level
400
µ
W or
µ
J
Maximum Power Density
200 W/cm²
1.5 kW/cm²
12 kW/cm²
Note 1
Note 2
Note 3
Note 4
Maximum Peak Power Density
1 MW/cm²
100 MW/cm²
800 MW/cm²
Note 5
8.5 GW/cm²
Note 6
Note 7
Maximum Single Pulse Energy
30 J
Maximum Energy Density
Note 8
Note 9
Note 10
Note 11
Note 12
Note 13
Note 14
Precision
< 1 %
Accuracy
±
3 %
Response Time
3 sec when connected to a Scientech power meter in Watts Mode
Dimensions DxL - inches
cm
4.75 x 2.3
12.07 x 5.8
4.75 x 2.3
12.07 x 5.8
4.75 x 3.92
12.07 x 9.96
4.75 x 2.3
12.07 x 5.8
4.75 x 3.92
12.07 x 9.96
4.75 x 2.3
12.07 x 5.8
4.75 x 2.3
12.07 x 5.8
Weight - pounds
kgs
2.9
1.3
2.9
1.3
3.1
1.4
2.9
1.3
3.1
1.4
2.9
1.3
2.9
1.3
Power Meter Compatibility
H410, H410D, S310, S310D

PN9458Q 5
Note 1: AC2501, AC5001 30W/cm
2
@ 1064nm, 23W/cm
2
@ 532nm, 8.5W/cm
2
@ 355nm,
175mW/cm
2
@ 266nm
Note 2: ACX2501, ACX5001 Note 1 specs x 8 for 400nm to1.2µm
Note 3: AC25UV, AC50UV 50W/cm
2
@ 355nm Note 4: AC2504, AC5004 35W/cm
2
@1064nm
Note 5: AC2501, AC5001 100GW/cm
2
@ 1064nm, 78GW/cm
2
@532nm, 29GW/cm
2
@ 355nm,
580MW/cm
2
@266nm
Note 6: AC25UV, AC50UV Repetitive pulses: 101MW/cm
2
@355nm
Single pulses: 3.5GW/cm2 @ 355nm
Note 7: AC2504, AC5004 125GW/cm
2
@1064nm
Note 8: AC2500, AC5000 Max J/cm
2
= 1,000 x (pulse width)
1/2
to a
maximum of 200J/cm
2
.
Note 9: AC25HD, AC50HD Max J/cm
2
= 4,500 x (pulse width)
1/2
to a
maximum of 14J/cm
2
.
Note 10:ACX25HD, ACX50HD Max J/cm
2
= 36,000 x (pulse width)
1/2
to a
maximum of 42.5J/cm
2
.
Note 11: AC2501, AC5001 Repetitive pulses: 4.1J/cm
2
@1064nm, 3.2J/cm
2
@532nm, 1.2J/cm
2
@355nm,
24mJ/cm
2
@266nm Single pulses: 8J/cm
2
@1064nm, 6.2J/cm
2
@532nm,
2.3J/cm
2
@355nm, 46mJ/cm
2
@266nm
Note 12: ACX2501, ACX5001 Note 11 specs x 8
for 400nm to 1.2µm Note 13: AC25UV, AC50UV
Repetitive
pulses: 1.1J/cm
2
@ 355nm
Single pulses: 40J/cm2 @ 355nm
Note 14: AC2504, AC5004 Repetitive pulses: 4.8J/cm
2
@1064nm
Single pulses: 10J/cm2 @ 1064nm

PN9458Q 6
Absorption of HD Material
Warning: You must exercise caution when using HD detectors. They exhibit spectral reflection of
between 7% and 18%, of the input power, back out of the aperture. Please refer to Figure 1 to determine
the reflectance for the wavelength you are measuring. These detectors should be treated as a partial mirror
or any other type of reflective optic and the appropriate caution level observed, especially at the CO2
wavelength.
A B S O R P T I O N O F H D A B S O R B E R V s . W A V E L E N G T H
1 0 0
9 5
9 0
8 5
8 0
7 5
7 0
6 5
6 0
5 5
5 0
0 1 1 0 1 00
W A V E L E N G T H [ M I C R O M E T E R S ]
Figure 1. HD Absorption vs Wavelength
HD calorimeters might show “beam” marks on the absorbing surface. These marks are characteristic of
the material and do not affect the performance of the calorimeter. Do not rub off or remove the marks.
Polishing or cleaning the absorbing surface might change the performance of the calorimeter.
Please see the table at the end of this manual for specific absorption vs. wavelength information.
ENVIRONMENTAL SPECIFICATIONS
This product is intended for indoor use at altitudes up to 2000 meters, Pollution Degree 2 in accordance
with IEC 664 and transient over voltages according to Installation Categories (Overvoltage Categories)
II. Note that each of the above detectors will not pass the IEC 801 Publication, Part 3, Radiated
Electromagnetic Field Requirements. The system, meter and detector, is designed to measure radiation
within the test's radiation band. The detectors were held outside the radiated electromagnetic field during
ABSORPTION [%]

PN9458Q 7
this test. It is up to the user to be aware of RF fields present during measurements and their effects if any
on those measurements.
CE Certification
All of the calorimeters listed in this manual have been certified for the European CE mark.
UNPACKING AND SETUP
The calorimeter and accessories are shipped in custom packing materials. All packing materials should be
saved for future damage free shipments.
Each detector has a 1/2" diameter mounting hole for installing the detector onto a mounting base assembly.
Screw the mounting base post into one of the mounting holes in the body of the calorimeter. Optional
mounting base assemblies are available for purchase from Scientech for mounting the detector/mounting
base assembly to an optical bench.
A 3 meter mini DIN interconnect cable is also included. If you are using a Scientech indicator, connect
the calorimeter to the indicator with the interconnect cable. Follow the detailed set up instructions that are
in the indicator’s instruction manual. If you are not using a Scientech indicator operating requirements are
contained in this manual.
Note: Vector™ calorimeters are sensitive to all types of thermal input. Due to the handling of the
calorimeter during setup and possible environmental temperature differences, thermal gradients may exist
in the calorimeter. Allow the calorimeter to sit undisturbed for several minutes to reach thermal
equilibrium before using.
Note: When using a 25mm Vector calorimeter for measuring average power levels below 30mW and
single pulse energy levels below 30mJ, a Scientech Model 36-0203A, Isoperibol™ Enclosure, is highly
recommended. The Isoperibol enclosure should not be used at average power levels above 30mW, and
single pulse energy levels above 100mJ because heat build-up will occur.

PN9458Q 8
CORRECTING VECTOR™ HD CALORIMETER OPERATING
PARAMETERS FOR USE AT DIFFERENT WAVELENGTHS
Scientech Vector calorimeters in general have a flat response to all wavelengths within their specified
spectral response. HD and HDX calorimeters are an exception to that rule and are calibrated at a specific
wavelength by adjusting the calorimeter’s gain circuitry for that wavelength. The calibration wavelength
is recorded in the Operating Parameters section at the front of the manual and on the detector’s serial tag.
When a HD or HDX calorimeter is used at a wavelength other than the calibration wavelength, the
calorimeter’s output sensitivity can be adjusted to compensate for the absorption rate at the new
wavelength as follows:
1. Find the absorption rate from the chart at the end of this manual for the calibration wavelength of
your calorimeter.
2. Find the absorption rate for the wavelength where you will be working.
3. Determine the new output sensitivity using the following formula:
absorption rate of new wavelength x output sensitivity (V/J) = output sensitivity for new
absorption rate of calibration wavelength wavelength from serial tag
NOTE: Due to variability in the manufacturing process the absorption characteristics of the HD model
calorimeters can vary in the UV region (190 to 400nm). Scientech recommends optical calibration at
266nm if the detector is to be used in the UV region instead of relying on an absorption rate in the chart
at the end of this manual.
VECTOR™CALORIMETER OPERATION WITHOUT POWER
METER
Cable Requirements
Vector calorimeters are powered up by the indicators. To use a Vector calorimeter without a Scientech

PN9458Q 9
power meter, but with a volt meter, you must apply +/-8Vdc to the mini DIN connector as shown in Figure
2. The voltage output of the calorimeter, from pin 8, should be connected to the positive side of the DVM
or chart recorder. All 3 of the grounds should be tied together at the negative side. Pins 2 and 3 are not
used.
Figure 2. Calorimeter mini DIN Connector Pin Out
Operation of VectorCalorimeters with a Digital Volt Meter
The calorimeters may be used with any digital volt meter (DVM) capable of reading 5 volts full scale.
1. Connect the output of the calorimeter to the DVM.
2. Select the DC volts mode.
3. Direct the laser beam on to the absorbing surface of the calorimeter.
4. When the display of the DVM has stabilized (about 2 minutes), calculate the laser power using the
formula:
W = V/S
where:
W = Laser power in watts
V = Voltage reading of the DVM in volts
S = Sensitivity of the calorimeter from page 2.
CALIBRATION OF VECTOR™CALORIMETERS USING

PN9458Q 10
ELECTRIC SUBSTITUTION HEATING
For Vector calorimeters the electric substitution heating option must be ordered and installed at the factory
when the calorimeter is purchased. It cannot be retrofitted to a calorimeter at a later time. To calibrate
using electric substitution heating proceed as follows.
A. Remove the screws holding the calorimeter's ID tag and remove the plate to expose the circuit
board. Figure 3 shows the pin out of the calorimeter circuit board.
Figure 3. Vector Calorimeter Circuit Board
B. Connect a DVM to the test points labeled SUB and HTR on the calorimeter circuit board.
C. Measure the resistance of the substitution heater making sure to subtract the resistance of
the patch cables from the total resistance measurement. Compare this resistance to Rc in
the calibration data in the front of the manual. The two should agree within 2%. If not
contact Scientech.

PN9458Q 11
D. Remove the DVM. Connect a power supply to the SUB and HTR test points and connect
the DVM to monitor the power supply.
E. Connect a second DVM to the output and ground of the Vector calorimeter.
F. Apply Vh volts, stated in the calibration data you received with the calorimeter, to the
substitution heater.
G. Verify that the output voltage is one of the following values depending on the type of
calorimeter being tested:
0.750 volts for 25 mm calorimeters
0.333 volts for 50 mm calorimeters
H. If needed, adjust the calibration trim pot, R4 on the calorimeter circuit board, until the
voltage measured in step G is from the calibration data, is displayed by the indicator.

PN9458Q 12
VECTOR™LARGE APERTURE CALORIMETER SPECIFICATIONS
Table 3. Vector™100mm Aperture Calorimeter Specifications
Model
360401
380401
380402
384UV5
Type of Absorber
Surface
Volume
Volume
Volume
Aperture Diameter
100mm
Minimum Beam Diameter
5cm
Spectral Response
.25 - 35μm
.266 - 1.2µm
9 -11µm
.193 - .36μm
Max Average Power
50W with full illumination of absorbing surface
Min Average Power
150mW
Noise Level
1.5mJ - mW
Max Power Density
200W/cm²
See Note 1
4W/cm²
Note 2
Max Peak Power Density
1MW/cm²
See Note 4
100MW/cm²
See Note 5
Max Single Pulse Energy
300J
Max Energy Density
Note 7
Note 8
4J/cm2
Note 9
Precision
< 1%
Accuracy
5%
Dimensions DxL
inches
cm
6.00 x 8.00
15.24 x 20.32
Weight
pounds
kgs
6
2.72
Table 4. Vector™200mm Aperture Calorimeter Specifications
Model
360801
380801
380802
384UV5
Type of Absorber
Surface
Volume*
Volume
Volume
Aperture Diameter
200mm
Minimum Beam Diameter
7.5cm
Spectral Response
.25 - 35μm
.266 - 1.2µm
9 -11µm
.193 - .36μm
Max Average Power
100W with full illumination of absorbing surface
Min Average Power
700mW
Noise Level
7mJ or 7 mW
Max Power Density
200W/cm²
See Note 3
4W/cm²
Note 2
Max Peak Power Density
1MW/cm²
See Note 6
100MW/cm²
See Note 5
Max Single Pulse Energy
1000J
500J
1000J
Max Energy Density
Note 7
Note 10
4J/cm2
Note 9
Precision
< 1%
Accuracy
5%
Dimensions DxL
inches
cm
9.00 x 10.00
22.86 x 25.40
Weight
pounds
kgs
16.27
7.26

PN9458Q 13
Notes for 100 & 200mm Calorimeters:
Note 1:
380401S
27W/cm2@ 1064 nm
21W/cm2 @ 532 nm
7.7W/cm2@ 355 nm
158mW/cm2@ 266nm
Note 2:
384UV5S, 388UV5S
50W/cm2@ 355nm
Note 3:
380801S
13.5W/cm2@ 1064 nm
10.5W/cm2@ 532 nm
3.85W/cm2@ 355 nm
79mW/cm2@ 266nm
Note 4:
380401S
90GW/cm2@ 1064 nm
71GW/cm2@ 532 nm
27GW/cm2@ 355 nm
530MW/cm2@ 266nm
Note 5:
384UV5S, 388UV5S
Repetitive pulses:
101MW/cm2@ 355nm
Single pulses:
3.5GW/cm
2
@ 355nm
Note 6:
380801S
45GW/cm2@ 1064 nm
35.5GW/cm2@ 532 nm
13.5GW/cm2@ 355 nm
265MW/cm2@ 266nm
Note 7:
360401S, 360801S
Max J/cm2 = 1000 x (pulse width)1/2 to a maximum of 200J/cm2
Note 8:
380401S
Repetitive pulses:
3.7J/cm2@ 1064nm
2.9J/cm2@ 532nm
1J/cm2@ 355nm
20mJ/cm2@ 266nm
Single pulses:
7J/cm2@ 1064nm
5.6J/cm2@ 532nm
2.1J/cm2@ 355nm
41mJ/cm
2
@ 266nm
Note 9:
384UV5S, 388UV5S
Repetitive pulses:
1.1J/cm2@ 355nm
Single pulses:
40J/cm2 @ 355nm
Note 10:
38-0801S
Repetitive pulses:
1.85J/cm2@ 1064nm
1.45J/cm2@ 532nm
0.5J/cm2@ 355nm
10mJ/cm2@ 266nm

PN9458Q 14
UNPACKING AND SET UP
The calorimeter and accessories are shipped in custom packing materials. All packing materials should
be saved for future damage free shipments for repair or calibration.
A ¾” diameter mounting post and mounting base are available for the larger calorimeters. The post screws
into the ¾” mounting hole in the body of the large calorimeter. Mount the post to your optical bench or
working surface.
If you are using a Scientech power meter, connect the calorimeter to the interface module and the interface
module to the indicator. Follow the detailed set up instructions that are in the power meter’s instruction
manual. If you are not using a Scientech power meter operating requirements are contained in this manual.
Note: Large aperture calorimeters are sensitive to all types of thermal input. Due to the handling of the
calorimeter during setup and possible environmental temperature differences, thermal gradients may exist
in the calorimeter. Allow the calorimeter to sit undisturbed for several minutes to, reach thermal
equilibrium, before using.
Cable Requirements for Large Calorimeters
Refer to Figure 1. The output of the calorimeter is connected directly to the DVM. Large aperture
calorimeters do not require any power. The voltage output is on pin 1 of the DIN connector and should
be connected to the positive side of the DVM. Ground is on pin 3 and should be connected to the negative
side. Pin 2 is not used.

PN9458Q 15
Figure 4. 100 & 200mm Calibration Heater Connectors
LARGE CALORIMETER OPERATION WITHOUT POWRE METER
Note: Whenever a large aperture calorimeter is used without an indicator the interface module, which is
required to connect the calorimeter to the indicator, is not used.
Operation of Large Aperture Calorimeters with a Digital Volt Meter
The calorimeters may be used with any digital volt meter (DVM) capable of reading 5 volts full scale.
A. Refer to Figure 1. Connect the output of the calorimeter to the DVM.
B. Select the DC volts mode.
C. Direct the laser beam on to the absorbing surface of the calorimeter.
D. When the display of the DVM has stabilized (about 2 minutes), calculate the laser power using the
formula:
W = V/S
where:
W = Laser power in watts

PN9458Q 16
V = Voltage reading of the DVM in volts
S = Sensitivity of the calorimeter from page 2.
Calibration of Large Aperture Calorimeters
A. Refer to Figure 1. Connect a DVM to the white jacks of the calorimeter.
B. Measure the resistance of the substitution heater making sure to subtract the resistance of the patch
cables from the total resistance measurement.
When measuring the substitution heater resistance of a 200 mm calorimeter, R1 and R2 must be connected
together in series.
Compare this resistance to Rcin the calibration data in the front of the manual. The two should agree
within 2%. If not, contact Scientech.
C. Calculate the voltage equivalent to laser power using the following formula:
V = (Rcx C x W)1/2
where:
V = voltage applied to the heater coil
Rc= substitution heater resistance from step B
C = Cal coefficient 360401 = 1.018 360801 = 1.000
380401 = 0.974 380801 = 1.008
380402 = 1.024 380802 = 1.008
384UV5 = 1.021 388UV5 = 1.002
W = desired laser power in watts
D. Connect the DVM to the calorimeter’s DIN connector.
E. Apply the calculated voltage (V) to the electrical substitution heater.
F. Record the voltage reading of the DVM (Vc).
G. Calculate the calorimeter’s output sensitivity (S) as follows:
S = Vc/W

PN9458Q 17
where:
S = calorimeter’s output sensitivity
Vc= voltage output from the calorimeter in mV
W = desired laser power output.
The measured sensitivity should be ± 3 % of the calorimeter’s original sensitivity value.
SCIENTECH CALIBRATION SERVICE
Scientech recommends that a complete calibration be performed annually to verify system accuracy.
Please visit Scientech’s website at www.scientechinc.com to obtain an RMA (returned material
authorization) number and complete the form noting that you are requesting calibration of your detector.
You may also contact our Product Service Department at (800)525-0522 or (303)444-1361 or Fax
(303)444-9229 or email [email protected]. Be prepared to provide model number, serial number
along with contact information. to arrange for a NIST traceable, factory calibration. Scientech calibrates
the detector with its power meter for the same price.
LIMITED WARRANTY
Scientech warrants and represents that the laser power measurement product will be free from defects in
design, materials and workmanship and conform with applicable Scientech product specifications for a
period of three (3) years. The product warranty period begins on the date of shipment from Scientech.
Scientech warrants that its products shall conform to applicable Scientech specifications and drawings and
will meet all the functional and performance requirements when properly installed, operated, and
maintained in accordance with Scientech’s operating manual. Warranty does not extend to any Scientech
products that have been subjected to misuse, abuse, or accidents, or improper installation, maintenance or
applications, repaired by unauthorized personnel, or Products in which the tamper proof sticker has been
removed or broken.
During the warranty period, Scientech will repair, or at its option replace at no charge, components that
prove to be defective. The product must be returned, shipping prepaid, to Scientech's authorized repair

PN9458Q 18
facility. Products repaired by Scientech’s authorized repair personnel/facilities will be warranted against
defects in the repaired component and workmanship for a period of 365 days from the date of shipment
of the repaired Product.
RETURN MATERIAL PROCEDURE
Should it become necessary to return any product to Scientech for any reason including calibration, please
visit Scientech’s website at www.scientechinc.com to obtain an RMA (returned material authorization)
number and complete the form. You may also contact our Product Service Department at (800)525-0522
number, serial number, and a description of the problem along with contact information. Frequently we
can provide self-help information which will eliminate the need for returning the product.
If product return is required, please pack the items in the original box and packing material. As an alternate,
place the equipment in a snug-fitting box, and then pack that box in a larger box with at least four inches
of packing material. Scientech does not assume responsibility for products damaged during shipping and
shipping damage will not be treated as a warranty repair. Please include a point of contact, email address,
and phone number of the person we should contact regarding repair questions.
Normally, products are repaired and shipped within five (5) business days following receipt of the product
at the authorized service facility. The repair turn-around time could vary depending on the workload.
Shipping Address: Scientech, Inc.
Product Service Department
5649 Arapahoe Ave.
Boulder, Colorado 80303 U.S.A.
DISPOSAL OF ELECTRONIC EQUIPMENT
Scientech recommends the following for disposal of electrical and electronic equipment:
1. The best option is to reuse the equipment in its entirety.
2. Where the equipment cannot be reused in its entirety, priority should be given to reuse of
its subassemblies and components.

PN9458Q 19
3. Where reuse is not appropriate, electrical and electronic equipment, including batteries,
should be recycled according to local ordinances.
4. Waste electrical and electronic equipment should never be mixed with municipal waste.

PN9458Q 20
HD ABSORPTION VS. WAVELENGTH
Use this table for all standard and slim profile HD model pyroelectric detectors.
Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorption
μm %μm %μm %μm %
0.200 93.66 0.480 94.71 0.735 93.99 0.990 93.34
0.210 93.78 0.485 94.75 0.740 94.21 0.995 93.35
0.220 92.18 0.490 94.80 0.745 93.98 1.000 93.34
0.230 90.44 0.495 94.76 0.750 93.98 1.005 93.33
0.240 88.79 0.500 94.73 0.755 94.08 1.010 93.33
0.250 90.07 0.505 94.70 0.760 94.04 1.015 93.33
0.255 90.42 0.510 94.67 0.765 93.89 1.020 93.29
0.260 90.41 0.515 94.68 0.770 93.94 1.025 93.28
0.265 90.56 0.520 94.70 0.775 94.01 1.030 93.28
0.270 90.71 0.525 94.66 0.780 93.91 1.035 93.25
0.275 91.13 0.530 94.66 0.785 94.03 1.040 93.23
0.280 91.63 0.535 94.70 0.790 93.96 1.045 93.17
0.285 91.85 0.540 94.75 0.795 93.89 1.050 93.14
0.290 92.01 0.545 94.73 0.800 93.64 1.055 93.12
0.295 92.73 0.550 94.66 0.805 93.65 1.060 93.12
0.300 93.02 0.555 94.63 0.810 93.67 1.065 93.14
0.305 93.04 0.560 94.64 0.815 93.53 1.070 93.17
0.310 93.15 0.565 94.64 0.820 94.21 1.075 93.16
0.315 93.53 0.570 94.62 0.825 93.61 1.080 93.13
0.320 93.45 0.575 94.56 0.830 93.81 1.085 93.12
0.325 93.82 0.580 94.59 0.835 93.83 1.090 93.09
0.330 93.83 0.585 94.56 0.840 93.83 1.095 93.06
0.335 93.96 0.590 94.56 0.845 93.93 1.100 93.03
0.340 94.07 0.595 94.61 0.850 93.84 1.105 93.01
0.345 94.14 0.600 94.55 0.855 93.31 1.110 93.00
0.350 94.15 0.605 94.51 0.860 93.65 1.115 92.83
0.355 94.19 0.610 94.50 0.865 93.90 1.120 92.76
0.360 94.31 0.615 94.45 0.870 93.84 1.125 92.67
0.365 94.30 0.620 94.38 0.875 93.92 1.130 92.60
0.370 94.44 0.625 94.44 0.880 93.81 1.135 92.57
0.375 94.48 0.630 94.42 0.885 93.85 1.140 92.56
0.380 94.46 0.635 94.44 0.890 93.54 1.145 92.51
0.385 94.53 0.640 94.48 0.895 93.88 1.150 92.47
0.390 94.53 0.645 94.50 0.900 93.60 1.155 92.43
0.395 94.57 0.650 94.46 0.905 93.70 1.160 92.42
0.400 94.62 0.655 94.49 0.910 93.62 1.165 92.37
0.405 94.63 0.660 94.39 0.915 93.68 1.170 92.32
0.410 94.65 0.665 94.43 0.920 93.66 1.175 92.30
0.415 94.72 0.670 94.41 0.925 93.64 1.180 92.24
0.420 94.70 0.675 94.44 0.930 93.60 1.185 92.23
0.425 94.78 0.680 94.36 0.935 93.67 1.190 92.21
0.430 94.74 0.685 94.37 0.940 93.59 1.195 92.16
0.435 94.80 0.690 94.36 0.945 93.60 1.200 92.08
0.440 94.76 0.695 94.19 0.950 93.50 1.205 92.02
0.445 94.80 0.700 94.37 0.955 93.49 1.210 91.94
0.450 94.79 0.705 94.25 0.960 93.51 1.215 91.91
0.455 94.79 0.710 94.20 0.965 93.45 1.220 91.92
0.460 94.79 0.715 94.16 0.970 93.42 1.225 91.88
0.465 94.76 0.720 94.08 0.975 93.40 1.230 91.87
0.470 94.78 0.725 94.11 0.980 93.35 1.235 91.82
0.475 94.75 0.730 94.21 0.985 93.37 1.240 91.81
Other manuals for AC2500
1
This manual suits for next models
14
Table of contents
Other ScienTECH Security Sensor manuals
Popular Security Sensor manuals by other brands

Herz
Herz clever&smart Room Controller installation manual

Aritech
Aritech Mini 6920P Installation sheet

DATACOM TEXTRON
DATACOM TEXTRON FiberLive R Series user manual

Gardner Denver
Gardner Denver THOMAS ASF VAKUMATIK IVF R34 Installation and operation manual

SPX
SPX RADIODETECTION RD5100 H2O+ Operation manual

Goldak
Goldak 777-A quick guide