AG3000
Mechanical Quick Start Guide
Metal Pipe Plastic Pipe
Choose a position that will ensure a full pipe. Choose a position that will minimize ow distortion.
PositioningPositioning
5X 1X
2X 1X
5X 1X
Acceptable Position: Recommended Position:
If air or sediment are
present. Air or sediment
on electrodes can aect
reading.
Electrode moved from
top by rotating meter
Intermittent air
bubbles miss electrode
Electrodes free from
sediment build-up
GroundingGrounding
Refer to the Electrical Quick Start Guide or the complete instruction manual for details on wiring.
WarningsWarnings
• Do not weld or ame cut within 10’ of meter. Heat will
damage electronics and liner. (Heat will void warranty.)
• Install provided full face gaskets. Failure to install gaskets
will void warranty.
• Tube must be completely aligned with adjoining anges.
• Ensure proper grounding.
• Bolts must be tightened evenly. Do not over tighten one side
to ll a gap.
• Install security seal during installation if regulations require.
• Backup batteries are not intended as a primary power source.
Refer to instruction manual for further details.
Equalization Lug
Ground Clamp
(Exothermically weld when
corrosion is a concern)
8’ Ground Rod
Earth
Grounding Ring
Gaskets
Grounding Ring
Gaskets
#6, #8, or #12 AWG Stranded
Copper Ground Wire < 5’
Plastic
Pipe
Plastic
Pipe
Equalization Lug
Meter
Flange Pipe
Flange
Gasket
Pipe Flange
Flange Plate
Lockwasher
Metal Pipe
Metal Pipe
Meter Flange
Gaskets