Sensica Sensifirm CEL100EU Series User manual

Table of Contents
Intended use 4
WARNINGS 7
CAUTIONS 7
Possible Adverse Effects 8
User interface description
11
11
11
12
13
14
Device Operation
Preliminary test
15
Recommended Treatment
17Post-treatment care
19
Troubleshooting
20
Customer Service
© All rights reserved to EL Global Trade Ltd.
3
Device Specifications
Who Should NOT Use the Device?
CONTRAINDICATIONS
6
6
Device description 9
10
User Instructions
What to Expect When Using Sensifirm
Preparing the treatment area
English
Treatment areas
18

English
4
Please make sure to
carefully read these
Instructions for Use
before you start your
Sensifirm treatment.
This guide, as well as an
instructional video, can
also be found at
www.sensica.com
of adipocytes' (fat cells); and
improving microcirculation and
lymphatic drainage. Together,
these effects lead to firming
and smoothing of the skin,
improvement in cellulite
appearance and
circumferential reduction, for a
tighter body with the best
possible visible, lasting results.
Introducing
LIPOTHERM RF™
technology- how it works
Radio Frequency (RF) is a form
of electromagnetic energy
that has been used for a long
time in aesthetic medicine. The
spacing between Sensifirm’s
four RF electrodes enables
great treatment volume and
allows multiple RF waves to
reach deeply into the skin
layers, heating the dermis,
hypodermis (inner skin layers)
and subcutaneous fat layers,
without damaging the
epidermis (outer skin layer).
This deep thermal effect in
turn activates several
biological mechanisms,
among them remodelling, and
later producing, collagen and
elastin; decreasing the volume
Intended use
By women who want to
improve cellulite appearance.
By men and women who want
to tighten and reduce the
circumference of their body.
Sensifirm can be used on
buttocks, thighs, abdomen,
waist and arms.
Note: This device is especially
designed for your own personal
treatment and is intended for
a single user and home use only.

English
5
Who should NOT
use the device?
CONTRAINDICATIONS
have an active implant,
such as a pacemaker or
insulin pump, or if you have
any known severe heart
disorder, such as arrhythmias.
are pregnant or
breastfeeding.
suffer from severe or
uncontrolled endocrine or
metabolic conditions, such
as diabetes or uncontrolled
thyroid disease.
have a damaged immune
system or autoimmune disorder,
such as AIDS or Lupus, or if you
use drugs that influence the
immune system.
Do NOT use the device if you:
suffer from blood clotting
disorders, excessive bleeding
or bruising, damaged wound
healing or if you routinely take
blood-thinning medication.
suffer from a history of deep
vein thrombosis.
have a history of collagen
disorders or suffer from keloid
scarring (growth of extra scar
tissue where the skin has healed
after an injury).
have a general extreme illness,
such as hepatitis, liver disease
or epilepsy.
suffer from sensitive or damaged skin,
such as in cases of frequent redness
or rash, inflammations, breakouts
(such as psoriasis), blisters, sores,
infections, acne or extra dry skin.
have diminished or exaggerated
perception of temperature change.
have taken Accutane in the past
6 months.
have a heat-induced disease or
an active skin disease, such as
herpes, eczema, rashes etc.
in the treatment area.
Carefully read the
information below before you
begin a treatment with the
device. If you still have any
doubts or are unsure that the
device is safe for your use,
please consult with your
physician.
have skin cancer, areas at risk of
malignancy, or if you have received
radiation therapy or chemotherapy
within the past 3 months.

English
6
Note: If you have undergone
any other aesthetic procedure
in the designated treatment
area, or if you have treated the
skin with a prescribed drug
within the past 3 months, or if
you have any concerns
regarding potential side
physician before using
Sensifirm.
If you have sunburn, or dark
brown or black spots such as
freckles, birth marks, moles or
blisters in the treatment area.
If you are allergic or sensitive
to chrome or nickel or any
component of the Base Gel.
If you are under the age of 18.
If you have reduced physical,
sensory or mental
capabilities.
Do not use Sensifirm over
breasts, genitals or areas
not recommended in this
Instructions for Use.
Do not use Sensifirm over
hairy areas of the body, as
this may result in the hairs
being burnt.
WARNINGS
If you have any permanent
implants in the designated
treatment area, including
but not limited to: silicon or
injected chemical materials,
metal plates or other metal
materials.
If in the past 3 months you
have had surgery, including
plastic surgery, or you have
undergone any aesthetic
procedure, such as skin
resurfacing, liposuction,
deep peeling, cryolipolysis
(fat freezing), or have had
aesthetic injections of fillers,
such as hyaluronic acid,
collagen, fat or Botox, in the
treatment area.
If you have any metal items,
such as jewelry or piercings,
or tattoos or permanent
make up on the designated
treatment areas.
Do NOT use the device:

English
7
Note: the Base Gel is for
external use only - avoid direct
contact with eyes. In the event
that such contact does occur,
wash the eyes immediately
with water.
CAUTIONS
Never use flammable liquids,
such as alcohol (including
perfumes, sanitizers, or other
applications containing
alcohol) or acetone to clean
the skin in the hour before
using Sensifirm as it may
cause burning of the skin.
Do not open or repair the
device, as it may expose you
to dangerous electrical
components.
Do not place, store or use
Sensifirm near bathtubs,
showers, basins or other
vessels containing water,
as this may cause severe
electrocution.
Do not leave Sensifirm
unattended while plugged
into the electricity supply, as
it may lead to electrocution.
Keep out of reach of children
and pets.
Use the Sensifirm only in
accordance with these
Instructions for Use and for
the intended use.
Do not use the device if any
part of the device is missing,
damaged or loose, if the
device is not working
properly, if you see or smell
smoke, or if the device
becomes damp or wet.
Make sure the device is
clean, dry and intact before
each use. Cleaning
instructions are detailed in
the Cleaning section below
Do not use the Sensifirm
device without first applying
the Base Gel.
Do not use the Base Gel
after its expiration date.

English
8
When using this device
according to the Instructions
for Use, adverse effects are
rare. However, every
aesthetic procedure may
cause some degree of
adverse effects (as listed in
the table below). In case
adverse effects appear, stop
using the device, and consult
with your physician if
necessary.
Mild adverse effects should
be spontaneously relieved
within a few hours.
Possible Adverse Effects
Severity of
Adverse
Effect
Frequency of
Adverse effect Adverse effects
Mild
Moderate
Local pain
Skin irritation
Skin redness
Skin sensitivity
Allergies or sensitivity to the
Base Gel
Mild skin burns
Pigment changes
Excessive redness, rash, swelling
Subcutaneous bleeding
Severe
Common
Rare
Extremely rare Severe burns
Infection
Bruising
Blistering

English
9
Device description
Status
indicator
RF electrodes
RF energy
selection
button
1-3 energy
level lights
RF electrodes
(front pair)
RF electrodes
(rear pair)
Temperature
sensors
Power
Socket

English
Status
Indicator
Sensifirm is in standby mode
and ready to work.
Sensifirm is in treatment mode.
RF is activated, skin contact is
detected and the treatment is
underway*.
Sensifirm is in error mode. The
device has detected a
malfunction and has blocked
the activation of the RF
function.
The green light indicates which
RF energy level has been
activated. Press the RF energy
level selection button
repeatedly until you reach the
desired level: 1, 2, or 3.
User interface description
*Note: RF energy is delivered only when proper contact with your skin
is detected. When treatment is underway, the status indicator will
have a steady green light and you will feel vibration.
* Upon contacting the treatment area, the massage function is
automatically turned on.
RF energy
level
selection
button
Steady green light
Alternating red
and orange lights
1-3 energy levels
lights
10
Transient orange
light (for ~5 sec)
followed by
transient green
light (for ~1 sec)

English
User instructions
Sensifirm is intended for use
on buttocks, thighs, abdomen,
waist and arms.
Preparing the treatment
Important:
area
1.
2.
3.
4.
Make sure your skin is dry,
clean and free of topical
products, such as
moisturizers and lotions.
Caution: Never use
flammable liquids, such as
alcohol (including perfumes,
sanitizers, or other
applications containing
alcohol or acetone) to clean
the skin in the hour before
using Sensifirm, as it may
cause burning of the skin.
Remove all jewelry from the
treatment area and from
your hands.
11
What to expect
when using Sensifirm
Check the expiration date of
the Base Gel.
Before beginning the treatment,
make sure that the electrodes
are clean, dry and intact.
During your treatment,
it is normal to experience:
Sensations of warmth created
by the RF energy.
Some mild red or pink colour
on the skin, during and just
after the treatment session.
Remove long and/or thick
hair from the treatment area
(mostly applicable for male
users).
Perform a preliminary test on
a small patch of skin in the
intended treatment area, 48
hours before the first use of
Sensifirm. In the preliminary
test, you can determine the
energy level settings that
best suit you (see
Preliminary test section
below).

English
Preliminary Test
Perform a preliminary test on
each area you intend to treat
48 hours before the first full
treatment.
Apply a thin, even layer of
Base Gel, included in your kit,
on the treatment area.
Plug in the Sensifirm device
and select RF energy level 3.
Place the Sensifirm device on
the desired treatment area
and move it in a circular
motion, for 3 minutes.
The treatment will create a
warm sensation on your skin.
However, if you feel your skin
is heating up too fast, reduce
the RF energy level, adjusting
it to suit your own comfort.
Stop the test immediately if
you do not feel comfortable,
such as severe redness or
blistering.
Wait 48 hours and examine
the tested area.
Prepare the treatment area
as detailed above.
1.
2.
3.
4.
5.
6.
If the skin in the test area
seems normal (no adverse
a full treatment at the highest
comfortable energy level.
Note: The RF energy level
indicates the rate at which the
temperature rises when in
heating mode. A high RF energy
level will enable you faster and
more even heating.
We recommend performing all
treatments at energy level 3
(highest). However, you can
achieve maximum results at all
RF energy levels: at lower levels,
it will just take longer to reach
7.
8.
12
The preliminary test is
intended to determine your
skin's reaction to treatment
with the device and help you
set the optimal setting for
each treatment area.

English
the skin.
If you do not feel a sensation
of warmth after approximately
30 seconds, try moving the
Sensifirm over a smaller
treatment area, and/or
increasing the RF energy.
At the end of each treatment
pressing on the energy
selection button. An orange
indicator light will light up
while the device is shutting
down. After the light has
from the electrical outlet.
After completing the
remaining Base Gel from the
treated area.
Clean the device with a soft
damp cloth or paper towel,
according to the cleaning
instructions in the Cleaning
section below.
6.
7.
8.
9.
Device Operation
Apply an even layer of
the Base Gel included in your
kit, on the area that is
intended for treatment.
Set the RF energy level to
the level you found most
comfortable when you
carried out the preliminary
test. Press the RF energy
level selection button
repeatedly until you reach
the desired level: 1, 2, or 3.
Place the device on the
intended treatment area,
ensuring full contact of the
electrodes with your skin
throughout the entire
treatment. If the device is not
properly placed, the RF energy
and massage will not be
delivered. When the device is
working, you will feel and hear
a gentle vibration, and the
indicator will be lit up in green.
1.
2.
3.
4.
Note: A preliminary test should be
conducted before proceeding to
a full treatment. (See Preliminary
test section above).
See treatment areas, treatment
time and treatments schedule in
the Recommended treatment
section below.
Prepare the treatment area
as detailed in “preparing the
treatment area” section
above.
5. Move the device over the
treatment area in slow circular
motions, while maintaining
constant, firm contact with
13

English
14
1. 20 minutes for a body area
the size of 15cmX30cm/
6inX12in (this is about the size
of your two palms).
2. 10 minutes for a body area
the size of 8cmX15cm/
3inX6in (this is about the size
of one palm).
Recommended treatment
For long-lasting results, use
Sensifirm once a week on each
treatment area, for a period of
8 weeks.
To sustain optimal results, you
can repeat the Sensifirm
treatment protocol as needed.
You can use the Sensifirm device
on the buttocks, thighs, abdomen,
waist or arms. Treatment session
depends on area size.

English
15
Treatment areas
Buttocks (area 1 )
Place the electrodes on the
treatment area, as indicated
in the diagram.
Move the Sensifirm in
continuous circular motions
around the area, as
indicated by the arrows in
the diagram.
Thighs (area 2)
Place the electrodes on the
treatment area, as indicated
in the diagram.
Move the Sensifirm in
continuous circular motions
around the area, as
indicated by the arrows in
the diagram.
Do not treat full areas with
energy levels you have not
tested before (as detailed in the
Preliminary test section above).
Abdomen (area 3)
Place the electrodes on the
treatment area, as indicated
in the diagram.
Move the Sensifirm in
continuous circular motions
around the area, as
indicated by the arrows in
the diagram.

English
16
Arms (area 5)
Place the electrodes on the
treatment area, as indicated
in the diagram.
Move the Sensifirm in
continuous circular motions
around the area, as
indicated by the arrows in
the diagram. If this is
uncomfortable, switch to
linear motions.
Waist (area 4)
Place the electrodes on the
treatment area, as indicated
in the diagram.
Move the Sensifirm in
continuous circular motions
around the area, as
indicated by the arrows in
the diagram. If this is
uncomfortable, switch to
linear motions.
In this area you can use just
2 electrodes

English
17
Do not rinse the Sensifirm
device under running
water, as this may cause
device malfunction.
Post-treatment care
After completing the
remaining Base Gel from the
treated area.
Do not expose the treatment
area to direct sun light
without applying a
high-factor sunscreen (at
least 30 SPF). Uncontrolled
sun exposure might cause
your skin to be sensitive
For 2 days after treatment,
avoid the use of extremely
hot water and do not use
abrasive or harsh products.
Since skin cellulite formation
is an ongoing process, you
may repeat the Sensifirm
treatment according to your
needs.
Cleaning
unplug the device before
cleaning.
It is recommended to use a
soft damp cloth, paper towel
or pre-moistened wipe to
clean any residual Base Gel
from the Sensifirm. Make sure
the entire underside of the
device, including the
electrodes area, temperature
sensors and plastic cover, is
cleaned thoroughly.
1.
2.

English
18
Troubleshooting
If the device does not turn on
If you experience any problems with your Sensifirm, please refer to
the information below:
In case of general error
If Sensifirm does not deliver
RF signal energy and massage
(i.e. you cannot feel the warm
sensation or feel the gentle
vibration)
Make sure that the power cord is
plugged in properly, both to the
electrical outlet and into the
device socket.
Make sure that the power outlet
has power.
Make sure that there is full contact of
the electrodes with the skin
Gently press the device onto your skin
Make sure that there is enough Base
Gel on the treatment area.
The RF signal will be activated only
when at least 2 electrodes are in full
contact with the skin. (See indicator
details set out in the User Interface
description section above.)
If dirt or debris get stuck on the
electrodes, you should clean them in
accordance with the Cleaning section
above. If you cannot get the area
clean, contact our customer service.
Reset the device – unplug the power
cord from the electrical outlet on the
wall, wait 10 seconds and then
reconnect it.
If you still have problems with your
device, please contact our customer
service.

English
19
For more information about
Sensifirm, if your Sensifirm is
broken or stops working, if
you need assistance or if you
would like to order more
Base Gel (for additional
treatments or if your Base
Gel has expired), please
contact us:
0800 923 4833
Device specificationCustomer Service parameters
The RF handheld device is
powered by multipolar RF at
10±0.5 Watt
Frequency:
1±0.05 MHz
Selectable output power level:
3 levels
Temperature sensors:
RF heating is automatically
deactivated when skin
temperature reaches 42oC
(106.7 F). The heating is
reactivated when
temperature decreases.
Treatment area:
External power supply:
Input 100-240VAC, 50-60Hz,
~10- 20cm2(~3.25-7.5 in²)
0.5A
Output 12VDC, 1.8A
Device size:
146mm X 74mm X 63mm
(LxHxW)
5.75in X 2.91in X 2.48in (LxHxW)
Weight:
192g (6.77oz)

English
20
Storage Instructions
Storage and Transport
Conditions
Temperature
Humidity
Non-condensingPressure
Operational Conditions
Temperature
30% to 70% RHHumidity
3,000 m.
Max
Operating
Altitude
-10 C to 70 C
(14 F to 158 F)
10% to 90% RH
at 55 C/131 F
10 C to 40 C
(50 F to 104 F)
Labeling
Model: CEL100EUxxyyzzz
Triple insulation
Consult Instructions
for use
Waste Electronic and
Electrical Equipment
Please read all
warnings and safety
precautions prior to use
Do not rinse Sensifirm
under running water,
as it may cause device
malfunction. Keep dry
Use Sensifirm only with the original power supply.
For your safety, never use DC adapters which are
not approved by the manufacturer.
MD 04194 A3 / RD 12003 C0

21
Treatment log
Treatment date Treated area Comments

22
Treatment log
Treatment date Treated area Comments