Shaw SDHmini User manual

Model SDHmini
Dewpoint Meter
Instrucon Manual
3006 - SDHmini User Manual - Iss 1.1


Contents
1
2
3
4
5
6
Introducon ................................................................................ 5
1.1 Unpacking Your Shaw Moisture Meters SDHmini ................ 5
1.2 General Descripon ............................................................. 5
Important Informaon ................................................................ 6
2.1 Symbols Table ...................................................................... 6
2.2 Lithium-ion Cell - Baery ...................................................... 6
2.3 Pressure Exposure ................................................................ 7
Installaon .................................................................................. 7
3.1 Gases to Avoid ..................................................................... 7
3.2 Installing the Air/Gas Sampling System ............................... 7
3.3 Piping Installaon Schemac ............................................... 8
3.4 Piping Schemac Component Index .................................... 9
3.5 Purging the Sample Connecon .......................................... 9
3.6 Using the SDHmini to Take a Moisture Reading ................ 10
3.7 Desiccant Dry-Down Technology ....................................... 11
3.8 Sample Connecons .......................................................... 11
3.9 Pressure and Flow Control ................................................ 12
3.10 Parculate Contaminaon ............................................... 12
3.11 Liquid Contaminaon ...................................................... 12
Normal Operaons ................................................................... 13
4.1 Powering On/O ............................................................... 13
4.2 Graphical Viewing Opons ................................................ 13
4.3 Display Opons .................................................................. 13
4.4 Choice of Units .................................................................. 14
Calibraon ................................................................................ 15
5.1 Auto Calibraon (AutoCal) ................................................ 15
5.2 Calibraon Setpoint (CalSet) ............................................. 16
5.3 Calibraon Procedure Error Messages .............................. 17
Sensor Related Sengs ............................................................. 17
6.1 Sensor Type ....................................................................... 17
6.2 Selecng Gas Types ........................................................... 17
6.3 Selecng Gas Molecular Weights ...................................... 18
6.4 Pressure ............................................................................. 18

Contents
7
8
9
10
11
12
13
Non Measurement Related Conguraon Funcons ................. 19
7.1 Changing the Date ............................................................. 19
7.2 Changing the Time ............................................................. 19
7.3 Changing Internaonal Sengs ........................................ 19
7.4 Hotkeys .............................................................................. 20
Installing the SDHmini User Tool ............................................... 21
8.1 Extracng the SDHmini User Tool Soware from the USB
Flash Drive (Memory Sck) ................................................ 21
8.2 Extracng the SDHmini User Tool Soware from the
Instrument .......................................................................... 22
Connecng the SDHmini and SDHmini User Tool ...................... 24
Using the SDHmini User Tool .................................................... 25
10.1 Menu items ...................................................................... 25
10.1.1 File - SDHmini Report ............................................... 25
10.1.2 Language .................................................................. 25
10.1.3 Help .......................................................................... 26
10.2 Instrument Live Reading ................................................... 26
10.3 Instrument Stac Informaon .......................................... 26
10.4 Ulies ............................................................................. 27
10.5 Sengs ............................................................................. 28
10.6 Firmware Update ............................................................. 29
10.7 Factory Reset .................................................................... 30
Addional Informaon ............................................................. 31
11.1 Desiccant and Head Seal Replacement ........................... 31
11.2 Temperature .................................................................... 32
11.3 Response Time ................................................................ 32
11.4 Guarantee ........................................................................ 32
11.5 Basic Denions .............................................................. 32
11.6 Problem Guide ................................................................. 33
SDHmini Specicaons .............................................................. 34
SDHmini General Assembly Diagram ......................................... 35

1 Introducon
1.1 Unpacking Your Shaw Moisture Meters SDHmini
Please examine the SDHmini package for any damage or mishandling. If any
damage is evident please nofy the carrier and the Shaw Moisture Meters
representave from where this unit was purchased.
You should have received (if ordered):
1 model SDHmini instrument
1 carry case
1 Allen key
1 two metre length of PTFE tubing
1 USB charging cable
1 Fir Tree ng plus 1 x ngs as
ordered
If anything is missing please contact your distributor immediately.
1.2 General Descripon
The SDHmini is a micro-controller based baery operated dewpoint hygrometer
with state of the art features and enhancements with support of USB for
communicang with PC.
It is a fully, self-contained unit incorporang the moisture sensor, signal
condioning circuitry, memory management, graphics display, membrane
keyboard, real me clock, on-board baery and charger circuit.
The graphical user interface features three main viewing screens, digital, graphical
(historical readings in a scrolling strip chart) and a meter (a tradional moving
coil). There is also a secondary display in the top right which allows display of
readings in alternave units of measurement.
The unit operates with the following 10 language opons: English, German,
French, Spanish, Portuguese, Russian, Italian, Chinese, Japanese and Korean.
A full colour graphics LCD display with 320 x 240 resoluon landscape and a 6 key
input including an on/o key.
The displayed units can be changed on both the primary and secondary displays for
the selectable moisture unit e.g. °C, °F, g/m3, ppm(v), ppm(w) and lb/MMSCF.
5
●
●
●
●
●
●
●
●
●
●
1 AC/DC adaptor
1 pressure dewpoint circular
calculator
1 USB ash drive with SHAW
SDHmini User Tool soware
1 instrucon manual

The SDHmini sengs can be adjusted by a user-friendly soware interface
incorporang mulple menus, hot key combinaons, informaon displays and
control funcons.
Use the supplied USB cable, aach between the PC and the instrument, an
automac connecon over USB is established.
The USB port allows:
● Charging of the baery (Li-ion) when connected to PC or dedicated
power supply
● Downloading of new rmware to the unit
● Downloading of setup data from dedicated PC soware
The view panel allows access to many of the features, sengs and informaon
pages within the instrument.
A micro SD card is ed to the instrument, the contents of which can be explored
by clicking the SD card icon in the boom le of the view panel. The contents may
include user manuals, service notes, calibraon data sheets, rmware backups etc.
2 Important Informaon
2.1 Symbols Table
6
2.2 Lithium-ion Cell - Baery
The SDHmini is powered by a single lithium-ion cell. The cell is not
eld replaceable by the user as it is completely enclosed within the
instrument. However, the cell may be exchanged by an authorised
agent who has received the necessary training, support systems
and accreditaon. It is recommended to only use the USB power
supply supplied with the instrument.
It is recommended that the SDHmini be put on charge overnight
when the unit is rst received.
To aain maximum cell life, the SDHmini should always be charged
at room temperature. The discharging cell life (hours of use from a
fully charged cell) will reduce when used in cold condions. Switch
on the instrument to check if fully charged.
The symbol indicates that the device contains a baery which may
be charged via the USB connector.
Rang of instrument is less than 1 W in operaon. Charging
voltage is 5 V nominal.

3 Installaon
3.1 Gases to Avoid
The SHAW moisture sensors are suitable for many dierent industrial and research
applicaons. Most gases can be checked for their moisture content with no need
for the calibraon to be altered when changing between dierent gases, as the
sensor operates only with reference to the water vapour content.
There are however, some gases that must be avoided as they are not compable
with the material of construcon of the sensor. Ammonia (NH3), Ozone (O3) and
Chlorine (Cl2) must be avoided at all mes, even in small quanes. Hydrogen
Chloride (HCl) also aacks the sensors very quickly. Some, less aggressive acidic
gases such as Sulphur Dioxide (SO2) can be monitored as long as the moisture
content is low, generally less than 100ppm(v). If in doubt, please ask your supplier.
Sulphur Hexauoride (SF6) has no eect on the sensor. If the gas has been exposed
to arcing, however, it is possible that various acidic species will have been formed
that will corrode the sensor. When tesng SF6that may have been arced, therefore,
an acidity test should be carried out rst; if the gas proves to be acidic then the
moisture test should not be carried out.
3.2 Installing the Air/Gas Sampling System
The Piping Installaon Schemac diagram (see secon 3.3) shows all components,
which could be used in a dry gas measurement applicaon. Although not all the
items shown will be required for every installaon.
The main principle when considering the presentaon of the gas sample to the
SDHmini dewpoint meter is to keep the sample system as simple as possible and
of minimum swept volume, so that it has the least possible inuence on the gas
sample.
The maximum pressure to which the telescopic
measuring head is exposed should not be more than
0.3 bar, 4 psi
Exposing the measuring head to higher pressures may
damage the instrument and result in injury to the
operator or other personnel in the area.
7
2.3 Pressure Exposure
!

The SDHmini should be placed on a stable surface prior to connecng to the
sample pipework. Care should be taken to ensure that the sample presented to the
SDHmini is not contaminated with any component that will damage, contaminate
or aect in a way that will impair the unit’s accuracy.
The ow rate, although not crical to the sensor measurement, should be low
enough to avoid abrasion to the sensor surface without being so low as to extend
the SDHmini response me to an unacceptable level. In general, a ow rate of
between 2 and 5 litres/min at normal temperature and pressure will give the right
balance.
The SDHmini is designed to operate at atmospheric pressure with a maximum of
0.3 bar.
3.3 Piping Installaon Schemac
8
MAIN
PROCESS
LINE
1
3
2
4567
8
9
a.
b.
The sample point should be on the upper surface of the horizontal
pipe or from a vercal secon of pipe, wherever possible.
The sample tube should run upwards from the sample point. If
this is not possible, then an inspecon port or drain tap should be
installed at the lowest point in the sample system.
Notes

3.4 Piping Schemac Component Index
Sample Isolaon Valve - This is a recommended item as it allows access to
the sample system without interrupng the main process line.
Filter Unit – A lter unit is recommended when the samples are likely
to contain parculate maer. If the air/gas sample contains heavy
hydrocarbon condensate, the lter must be of the coalescing type with a
drain. The lter unit should be posioned as close to the sample point as
praccal.
Sample Tube – This should be stainless steel for dry air or gas applicaons
but copper or carbon steel can be used where weer gases are to be
measured. If any secon of the sample tube must be exible then PTFE
should be used. In most cases, 3 mm OD (⅛”) is sucient as it provides
good system response me with minimum ow. 6 mm OD (¼”) tube can
be used where pressure drops across the 3 mm tube are too high.
Pressure Reducon Valve or Pressure Regulator – the sample is measured
at atmospheric pressure requiring that valve 4 is ed to the system.
Flow Control Valve.
Sample Connecon.
SDHmini.
Flow Indicator – The recommended sample ow is 2 to 5 L/min.
Sample Exhaust – The exhaust is vented to atmosphere or returned to an
atmospheric pressure line.
3.5 Purging the Sample Connecon
Refer to the Piping Installaon Schemac in secon 3.3.
It is advisable to carry out an inial purge of the sample loop without connecng
the instrument, in order to avoid the possibility of sensor damage on start-up.
Open the sample isolaon valve (1) slowly, unl there is a small ow of air/gas at
atmospheric pressure from the pressure reducon valve (4). Allow the air/gas to
exhaust through the sample connecon (9).
Connue this for two minutes to remove any residual moisture from the sample
pipe work and components. Check that no liquid or parculate contaminaon,
which could damage the sensor, is passing through the sample pipe. Direcng the
sample onto a piece of white paper (such as a lter paper) will make it easier to
see any dust or oil mist in the sample.
1.
2.
3.
4.
5.
6.
7.
8.
9.
9

Connect the SDHmini. The SDHmini is not ow sensive, however, the sample
ow needs to be enough to ll the sensor head with gas eecvely, so anything
less than 1 litre/min would mean that the instrument would take too long to sele
to a stable reading. The ow should not be so high that the gas velocity could
physically damage the sensor or cause back-pressure in the sensor head, so should
not exceed approximately 10 litres/min.
3.6 Using the SDHmini to Take a Moisture Reading
The rst purpose of the SDHmini is to provide a rapid, easy and reliable means of
carrying out moisture content measurement of gases. To ensure rapid response
the sensor is kept dry by surrounding the sensor with desiccant in the head
assembly when it is not in use.
Aer purging the system to be monitored t the SDHmini sample pipe to the
sample connecon (6). Allow the gas to purge for a further two minutes with the
head down. Restrict the outlet of the SDHmini (7) by placing a nger over the
sample exhaust (9), so that the pressure of the sample lis the telescopic head.
As soon as the head is fully extended remove the nger, to prevent pressure build
up. Pull gently on the head to make sure it is fully extended. If there is not enough
pressure to li the head assembly, assist it by pulling it up, do this very slowly so
that ambient air is not drawn in through the outlet.
The sensor will respond to the moisture level present in the sampled gas. The
displayed moisture value will connue to change unl the sensor reaches
equilibrium at which point the displayed value is representave of the moisture
level in the sampled gas. When a reading has been taken and the test is nished,
push down the telescopic head fully, carefully turn o the sample ow and
disconnect the sample pipe. With the head in the down posion the desiccant will
dry the sensor, ready for the next test.
Note: It is important that the head assembly is either fully up or fully down to
keep the desiccant isolated from the ambient air so avoiding weng it. When the
instrument is not in use it should be stored with the head in the down posion.
10

11
3.7 Desiccant Dry-Down Technology
Keeping the sensor dry between tests ensures that the SDHmini is ready to carry
out rapid spot checks. The unique design of the desiccant head achieves this by
surrounding the sensor with desiccant, keeping the sensor dry. The head is then
lied for sampling, exposing the sensor to the sample gas.
Between tests the sensor is not allowed to come into contact with ambient air. The
chamber is designed so that the void space and chamber wall surfaces are purged
with sample gas before exposure of the sensor, giving faster, more accurate and
reliable results.
3.8 Sample Connecons
The SDHmini is supplied with two connecons: a Fir Tree on one side and a Swagelok
compression ng for ¼”, ⅛”, 6 mm OD pipe or another Fir Tree ng (as specied
on your order) on the other. Either can be used as the inlet connecon (6).
PTFE and stainless steel are recommended materials for sample pipes. Some specialist
exible tubing may be suitable as sample pipe but common exible tubing such as
PVC, nylon, polythene or rubber, should not be used, as they are permeable and
water vapour from the atmosphere will diuse through them into the sample gas.
A 2 m PTFE sample pipe is supplied for use with the Fir Tree connecon. The rst
me this is used it is advisable to warm the end that will go over the Fir Tree for a few
seconds with a heat source to soen it slightly, which helps to get a gas ght seal.
Dry-Down Posion Read Posion
The unique design
allows purging of the
gas in the “Dry-Down
Posion” which will
bring all surfaces and
voids to equilibrium.
Sample exhaust
Sample gas
inlet pipe Sensor
The desiccant and
sensor are always
isolated from
atmospheric air to
ensure an accurate
reading.
Sensor exposed
to gas sample.
Sensor surrounded by desiccant
ensuring it is dry, ready for the
next measurement.

3.9 Pressure and Flow Control
In most cases the gas to be sampled will be at a posive pressure, which will need
to be reduced to atmospheric pressure. At low pressures, up to around 8 barg, this
can be done by including a simple needle valve (5) to control the ow and thereby
the pressure dropping across it. Ideally this would be the isolaon valve (1) on the
sample point, thus keeping the number of joints and glands, which are points of
potenal moisture ingress to a minimum.
At higher pressures it is advisable to reduce the pressure across a pressure
regulator (4). Any regulator used should be stainless steel and of a low swept
volume. Typical laboratory or welding type regulators are not suitable because
they have elastomeric diaphragms which are permeable to water vapour and the
high swept volume will be slow to purge.
3.10 Parculate Contaminaon
If the sample gas is likely to contain parculate material this should be kept away
from the sensor by including a small inline parcle lter (2) in the sample system.
If the surface of the sensor becomes partly coated with parcles it will not aect
the calibraon of the sensor but will restrict the movement of water molecules
across the surface so increasing the me it takes for the sensor to come to
equilibrium with the water vapour concentraon around it. If the solid material is
hygroscopic, such as desiccant dust, whenever there is a change in the moisture
concentraon in the sample, this material must come to equilibrium with the
moisture content before the sensor can respond to the change, which will cause
delays in obtaining accurate results.
3.11 Liquid Contaminaon
If the sample gas may contain liquid contaminaon, such as oil mist, this should
be kept away from the sensor by including a coalescing lter or an appropriate
membrane lter (2) in the sample system.
If the surface of the sensor becomes partly coated with liquid, again, it will
not aect the calibraon of the sensor. It will restrict the movement of water
molecules across the surface so increasing the me it takes for the sensor to
come to equilibrium with the water vapour concentraon around it. If the liquid is
hygroscopic, such as glycol, the buering eect described for desiccant dust may
be so great that the sensor becomes too slow to use.
12

13
4 Normal Operaons
4.1 Powering On/O
To switch on (and o), press , the SDHmini
briey displays a splash screen showing instrument
status, error informaon and rmware version number.
4.2 Graphical Viewing Opons
The start up screen is soon replaced by the normal measurement mode screen
(see below) displaying the current moisture measurement in a variety of views.
The last used or last selected mode will be ulised automacally.
4.3 Display Opons
The display can be changed by pressing the following keys to cycle
through all of the graphical views and will take you to the home screen.
Here is a view starng as a digital number followed by simulated strip chart and
moving coil instruments.
The digital display mode has the benet of showing the moisture measurement
in main choice of engineering units but also a secondary display with perhaps
alternave engineering units. This mode also allows applicaon of pressure
dewpoint to evaluate the measurement in other samples pressures.
Top Bar
Secondary Display or
Measurement Value
Buon Icon Area
These are so keys which
are always visible to assist
users in navigang through
the graphical screens.
Top Bar
Baery Status - Time
Viewing Portal
Main display resides here
but can also show alternave
Measurement Data,
Informaon Screens and
Conguraons in a variety of
formats.

14
4.4 Choice of Units
In normal measurement mode. The secondary display is the measurement
displayed in the top bar on the right hand side (A). It allows alternave
engineering units or concentraons to be displayed
alongside those chosen in the main display.
To choose the engineering units to be displayed, select the
units key from the main digital view.
You will be presented with the following screen to select engineering units for the
main and secondary displays. You may toggle between adjusng the main and
secondary displays by pressing the key 3.
Use the key to move to the desired
measurement type and press to
select it. Pressing will exit the
current screen without making any
further changes.
Here the secondary display is being
changed to ppm(v) engineering units
(B).
Please note: when you change the
display to either the moving coil meter or
the strip chart the secondary engineering
units display is replaced with the primary
units (C).
This connues to allow high precision measurement whilst having the benet of
the graphical display.
5 Calibraon
Field calibraons of the SDHmini may be performed periodically, every two to three
months, to ensure accuracy between laboratory calibraons by checking the span
of the instrument and correcng for any deviaon. These funcons can be used to
verify operaon of the SDHmini if conrmaon of an unexpected result is required.
This may be accomplished by two similar in built calibraon funcons, namely
AutoCal and CalSet.
A
B
C

15
5.1 Auto Calibraon (AutoCal)
The AutoCal facility uses the fact that each sensor is designed and manufactured
to give no further increase in response when it reaches its maximum moisture
level. For example a SDHmini with a full scale value of 0 °C dewpoint will read
0 °C dewpoint when exposed to a gas with this moisture concentraon and will
connue to read 0 °C dewpoint as the moisture concentraon rises above 0 °C
dewpoint.
The SDHmini can therefore be adjusted to read 0 °C dewpoint on any gas with a
moisture concentraon weer than 0 °C dewpoint, ambient air for instance, thus
seng the span to match the original calibraon.
NOTE: For instruments with a dewpoint range above 0 °C, e.g. the Blue range, -80 to +20 °C,
please refer to secon 5.2 Calibraon Setpoint (CalSet).
Select AutoCal from the Conguraon Menu.
Please ensure that the inlet and outlet valves are open to
atmosphere.
Extend the desiccant head on the instrument (as illustrated on
the instrument screen) nocing that the value of the full scale
moisture level has already been automacally preselected as a
target of 0 °C. You may need to li the head two to three mes
to allow ambient air through to the chamber, ensuring inlet
and outlet are unrestricted. Press .
The SDHmini will display a moving bar unl it assesses that
the sensor has stabilised (please allow up to ten minutes for
this to occur).
The SDHmini will now display the full scale value of the unit,
either -20 °C or 0 °C depending on the sensor range.
Close the desiccant head when instructed to do so and press
.

16
Pressing during the AutoCal aborts the process.
Press to accept.
The value present at the start of the aborted AutoCal process
is reinserted.
The SDHmini will return the instrument to normal measurement mode.
Note: Avoid carrying out the AutoCal procedure on a SDHmini with a full scale value of 0 °C
dewpoint on days when the ambient may be below this value, for example a dry day when
the ambient temperature is below approximately 12 °C.
The entered AutoCal value must not be pressure compensated.
5.2 Calibraon Setpoint (CalSet)
The funcon CalSet is suitable for sensors with an upper
range limit of +20 °C dewpoint, for example the Blue range
-80 to +20 °C dewpoint, where this upper range limit will
exceed typical ambient condions.
The unit may be checked against a known moisture
level, for example from a moisture generator or other
reference standard, such as a chilled mirror hygrometer or a
psychrometer using the funcon CalSet.
Select CalSet from the Conguraon Menu.
Extend the desiccant head and press .
Whether the instrument is being checked against the
measured ambient level or against a known calibraon
standard from a moisture generator, and are used
to set the value accordingly.
Selecng +10 °C for instance involves changing the sign and
number by use of both keys.
The entered CalSet value must not be a pressure
compensated one and must be within the range of the
sensor.
Press when done.

17
The SDHmini will display a moving bar unl it assesses that
the sensor has stabilised (up to 10 minutes).
The full scale value, 10 °C dewpoint, is now displayed.
Press when complete.
Close the desiccant head when instructed and press .
Pressing during CalSet aborts the process and the value
present at the start of the aborted CalSet process is reinserted.
Press to connue/accept changes.
The SDHmini will return to the normal
operaonal mode.
5.3 Calibraon Procedure Error Messages
If the instrument fails calibraon, a message will be displayed on the screen unl
acknowledged by the user, e.g.
AutoCal Failed
Sensor Open Circuit
or
AutoCal Failed
Sensor Short Circuit
This will mean replacing the sensor or its interconnecng cable in some cases.
6 Sensor Related Sengs
On the Conguraon screens there exists a collecon of measurement sengs
associated with the sensors and its environment.
6.1 Sensor Type
Here you may view the type of sensor
installed in the SDHmini and its range.
6.2 Selecng Gas Types
The SDHmini default seng is for ideal gas behaviour, however, there is an opon
to change the seng
to natural gas which
uses the appropriate
calculaon.

18
6.3 Selecng Gas Molecular Weights
The SDHmini default seng is for air (molecular weight 28.99), however, this
funcon allows the molecular weight to be set appropriately for other sample
gases, for example 18.99 .
6.4 Pressure
This sub-menu has, for both the primary and secondary displays, ve opons
allowing the user to select either atmospheric pressure or one of four pressure
engineering units and applied pressure.
When this funcon is selected, the main and secondary display screens will display
the pressure at which the dewpoint reading is calculated.
The SDHmini requires that a numerical pressure value be entered in the selected
units at the boom of the units selecon secon, for example selecng pressure
dewpoint of 2,000 psig on secondary display.
Warning: This funcon is purely an internal calculaon and DOES NOT allow an
elevated pressure to be applied to the measuring head.

19
7 Non Measurement Related Conguraon Funcons
7.1 Changing the Date
In order to change the present date press the arrow key
to select the Date funcon from the Conguraon menu.
Press .
Use and to change the date.
Here for example the date is changing from 06/03/15 to 06/04/15.
Press to enter new date or to abort changes.
7.2 Changing the Time
In order to change the present me
press to select the Time funcon
from the Conguraon menu.
To change the me use and .
Here for example the me is changing from 10:48 to 11:48.
Press to enter new me or to abort changes.
7.3 Changing Internaonal Sengs
Use the selecon cursor to select opons to change the
sengs for Language, Date Format and Numerical Delimiter.
Select Internaonal Sengs from the Conguraon Menu.
To change the Language opon select .
The language choice area is highlighted. A ag of a country
most associated with one of the pre-installed languages
indicates which language is currently selected. The

20
SDHmini operates with the following 10 language opons:
English, German, French, Spanish, Portuguese, Russian,
Italian, Chinese, Japanese and Korean.
Use the to select the ag required.
And to accept the choice.
The Date Format i.e. DD/MM/YY or MM/DD/YY is also chosen
from within the Internaonal Sengs menu.
And nally, numbers can be presented 123.45 or 123,45.
Note: These sengs are carried through to any data les and
printer outputs where appropriate.
7.4 Hotkeys
To access the hotkeys from the main screen, press key 5 on the keypad.
The four so keys are replaced by a new set of hotkeys for a short while. These are
available for rapid access to certain areas of the menu structures or funcons.
Turns ON and OFF the inbuilt sounder
Accesses the alternave moving coil graphical viewing screen
Takes the user to the Informaon screen
If the SDHmini is connected to a PC then the second hotkey is replaced
by an SD card symbol
Table of contents
Other Shaw Measuring Instrument manuals