
Superdew Operating Instructions
Page 6
omitted from the system. If measurements are to be taken at full line pressure
and the sample is to be returned to a part of the main line or vent, which is at a
pressure higher than atmospheric, and the input to that line needs a controlled
pressure, then both 4A and 4B will be required.
5) Sample Pressure Gauge – This is not a critical part of the moisture
measurement but may be required if Dew/Frost point measurements are to be
made at higher than atmospheric pressure.
6) Measuring Sensor.
7) Sensor Holder.
8) Desiccant Chamber – This item is required when the sampling is to be
intermittent. When installed, it prevents the ingress of wet air to the sample
system while the sample is not flowing, improving the response time.
9) Flow Control Valve – This can be a separate item or combined with the flow
indicator.
10) Flow Indicator – The recommended sample flow is 2 to 3 SL/M.
11) Sample Exhaust – The exhaust can be vented to atmosphere or returned to the
process line as discussed above.
5
Installing and commissioning the sensor
It is advisable to carry out an initial purge routine of the sample loop before installing the sensor.
This is to remove the possibility of sensor damage on start-up.
Refer to the sample system schematic in section 4.1. Open the inlet isolation valve slowly until a
small flow of air/gas (at atmospheric pressure) flows through the inlet pipe work to the sensor
holder, exhausting through the sensor entry port of the sensor holder.
Allow this purge to continue for about 15 to 20 minutes to remove any residual moisture from
the sample pipe work and components.
Close the inlet isolation valve and install the sensor into the sensor holder. Locate and connect
coaxial cable in position on the sensor.
Open the inlet valve slowly, by opening all valves after the sensor holder, allow a low-pressure
purge through the whole sample system. (Note: If a closed by-pass loop is installed, this section
of the procedure is not possible).
Set the required flows within the sample loop.
This completes the installation and commissioning, but on initial start-up, it could take several
hours for the system to reach equilibrium. The Superdew 3 will now indicate the dewpoint of the
air/gas surrounding the sensor, and the analogue output would be giving mA signals
proportional to the indicated dewpoint.