Operation Manual Shodex PROTEIN LS-G 4J Version 190422E
4
7. Eluents
(1) The following solvents are usable:
Various buffer solutions such as phosphate buffer solution, tris-HCI buffer solution, acetate buffer
solution with salts are generally used. Salts to be added include sodium chloride, sodium sulfate and
ammonium sulfate. The suitable salt concentration is 0.1 to 0.3 M.
(2) Acetonitrile and methanol can be used up to 100%.
(3) Protein denaturation reagents such as urea or guanidine hydrochloride, and surfactants such as
SDS or Brij can also be used. Since the addition of urea or guanidine hydrochloride in the eluent
increases the viscosity, so set the flow rate at less than 0.17 mL/min.
(4) Degas the eluent fully to prevent the formation of air bubbles.
(5) Presence of small debris or insoluble substances may result in deterioration of the column and/or
they appear as noise on the chromatograms. Filter the eluent with a 0.45 μm disposable filter to
prevent the problems.
(6) Start the system at less than 0.17 mL/min when replacing the solvent.
(7) Use of viscous solvents will increase the column pressure. Make sure to adjust the flow rate and/or
column temperature so that thecolumn pressure remains below the maximum column pressure stated
in section "6. Usable Conditions".
Attention! When using a mixture of organic solvent and buffer solution, make sure that there is no
precipitation of salt in the eluent before use.
Attention! Frequent solvent replacement can deteriorate the column, and thus not recommended.
Attention! When using highly corrosive salts such as sodium chloride, wash out the salts after the
analysis. The metal parts of the devices and/or the columns may rust.
Attention! Since protein denaturants may remain on the column, it is recommended to use the same
eluent conditions without replacing other eluent conditions.
Note! Use of on-line degasser is recommended.
Note! It is recommended to set the pump limiter to avoid use exceeding the maximum pressure.
8. Sample Preparation
(1) If possible, use the eluent for analysis to dissolve or dilute the samples. If this is difficult, use a
solvent which has a composition that is as close as possible to the eluent's composition, but fully
dissolves or dilutes the sample.
(2) Filter the sample solution using disposable 0.45 µm filter to prevent the column from clogging or
deteriorating.
9. Column Cleaning
Problems of peak shapes and elution timing changes or elevated column pressure etc. are often caused
by insoluble or adsorbing components in the eluent and reagents being deposited inside the column.
These problems may be resolved by cleaning the column. If a guard column is used with the analytical
column, first remove the guard column and check the performance of the analytical column alone. If
the problem is solved, most likely the cause is from the guard column. In this case, clean the guard
column. If the problem is not resolved after the cleaning, replace the guard column with a new one.
Insoluble components that block the column inlet may be removed by reversing the direction and
introducing the eluent from the column outlet with the flow rate at less than half of the recommended
flow rate. Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse
the direction and waste the washing solution directly from the column outlet into the waste container
without the detector and reduce the flow rate at less than 0.17 mL/min. A suggested cleaning solvent
volume is 5 - 10 times the column volume. Make sure to clean only the guard without the analytical
columns.