Operation Manual Shodex OHpak LB-800 series Version 210215E
6
6.3. Eluents
(1) Degas the eluent fully to prevent the formation of air bubbles.
(2) Presence of small debris or insoluble substances may result in deterioration of the column and/or
they appear as noise on the chromatograms. Filter the eluent with a 0.45 μm disposable filter to
prevent the problems.
(3) When replacing the eluent with miscible solvents, first use 1:1 mixture of the current solvent and
the new solvent, and then replace it with 100 % of the new eluent.
e.g. When replacing highly concentrated buffer solution (or aqueous salt solution) to methanol, first
run 1:1 mixture of buffer (or aqueous salt solution) and methanol, then replace it with methanol.
(4) Check miscibility/solubility of the desired new eluent and eluent previously used in the system. If
the new eluent has low miscibility/solubility to the eluent previously used, first use the eluent that is
miscible/soluble to both eluents, and then replace it with the desired eluent.
e.g. When replacing highly concentrated buffer solution (or aqueous salt solution) to
water/acetonitrile, first run water, and then replace it with water/acetonitrile.
(5) When replacing the solvent, start the system at less than 0.3 mL/min.
Attention! • Whenever water is required, use ultra-pure water freshly generated by a water
purification system or water from a newly opened HPLC grade distilled water bottle.
Solvents left in an opened bottle for a long time should not be used. The content may
have been changed or has been contaminated.
• Always use freshly prepared solvents. Solvents stored for a long time may have
changed their compositions and may influence elution patterns and/or damage the
column.
• Frequent solvent replacement deteriorates the column, and thus not recommended.
• When using highly corrosive salts such as sodium chloride, wash out the salts at the
end of analysis. The metal parts of the devices and/or the columns may rust.
Note • Use of on-line degasser is recommended.
•It is recommended to set the pump limiter to avoid exceeding the maximum
pressure.
6.4. Column Cleaning
Problems in peak shapes and elution time changes or elevated column pressure etc. are often caused
by insoluble or adsorbing components present in the eluent and reagents being deposited inside the
column. These problems may be resolved by cleaning the column.
If a guard column is used with the analytical column, first remove the guard column, and check the
performance of the analytical column alone. If the problem is solved, most likely the cause is from the
guard column. In this case, clean the guard column.
If the problem is not solved by removing the guard column, clean both guard and analytical columns.
Make sure to clean the guard and the analytical columns separately. When washing the column, let
the washing solution flow from the column outlet go directly into the waste container and not let the
solution go through the detector.
<Cleaning Method>
(1) Insoluble components that block the column inlet may be removed by reversing the flow direction,
i.e., introducing the eluent from the column outlet, with flow rate at less than half of the recommended
flow rate.
(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow
direction and use the flow rate at less than 0.5 mL/min. In general, the solvent volume introduced
should be 5 to 10 times the column volume.
Method 1 Adsorption of hydrophobic compounds (when using aqueous eluent)
Introduce the eluent with a high polar organic solvent (acetonitrile or methanol) added.
Method 2: Adsorption of ionic compounds
Introduce the eluent with a higher salt concentration.