
2.2 Device configuration
Item Description
Measurement
Record Mode Select values to be stored or just displayed.
Sync at Full Hour Synchronize time intervals to full hours.
Fixed Duration Measurement ends after a fixed time.
Repeat Mode If a Fixed Duration is specified an automatic repeat can be enabled.
Repeat Interval If a Fixed Duration is specified and the Repeat Mode is enabled, an interval must be configured
in which the measurement is repeated. So can be configured for example to measure every day
for 1 hour at 6 o’clock.
Start Time Predefine a time to start the measurement automatically.
Calibration
Start Calibration Start the process of calibration.
Calibration Level Set the calibration level (94.0 dB, 104.0 dB, 114.0 dB, Other).
Reset Calibration Reset the calibration level to factory value.
Measurement Values (Select a sound level e.g. LAFmax and adjust the following properties.)
Display Location Select on which graph the value shall be shown.
Display Interval Select the rate of display refresh (500 ms, . . . , 60 min, End Result only).
Storage Interval Set the storage interval (No storage, 62.5 ms, . . . , 60 min, End Result only).
Print Result Decide whether the result shall be printed.
Octave Spectrum
Octave Mode Select whether 1/3 or full octaves shall be measured.
Display Range Adjust the scaling of the Y-axis.
Display Weigthing Available weigthings: Z, A, C
Display Interval Select the rate of display refresh (125 ms, . . . , 60 min, End Result only).
Storage Weigthing Available weigthings: Z, A, C
Storage Interval Set the storage interval (No storage, 125 ms, . . . , 60 min, End Result only).
History Graph
Display Range Adjust the scaling of the Y-axis.
History Time Adjust the scaling of the X-axis.
Graph 1 Select which value shall be displayed as yellow.
Graph 2 Select which value shall be displayed as green.
Storage Files
Stored Files View Results, Delete File
Reset Storage File ID Resets File ID to the smallest possible number, so that no gaps are filled with new measurements:
Assume stored measurements are labled with IDs 1 to 6. If for example measurements 3, 5, and
6 are deleted and the File ID is reset, the next measurement will be labled with the number 5.
Erase all Storage Files Clear all Files.
Device
Time Set time and select 12h or 24h format.
Display Set Display properties.
Panels Set, whether Panels should be displayed or not:
Octave Spectrum, History Graph, Results, Measurement Info
Options Additional configurational settings:
Microphone Connection (Direct connection, cable connection, Weather Protector)
Power (automatic: On by USB, On by Reset, Off by Inactivity)
Info Provides information about firmware and hardware (see also figure 1.6).
Reset Configuration Reset to factory settings.
Print Results
Table 2.2: Manual configuration menu of Tango_Plus
SINUS Messtechnik GmbH 13 of 47 Manual Tango_Plus