Wheelchair & Bed Alarm
Continued on reverse side
A
C
B
Mounting Bracket
LED
Reorder# Description Unit
909510 MultiPro Alarm Unit Each
909515 MultiPro Alarm Unit 10/pk
Condition Status
When battery is installed, no sensor On Steady - Green Flash
Sensor pad plugged in Green Slow Flash
Monitoring Green/Red Flash
Alarm ring Red Flash
LED Conditions
Features:
• Low battery alert - alarm sounds when battery is low
• Sounds - two tone buzzer, high (90db), low (80db)
• Battery Test
• Auto-Reset
• Two Interfaces (connections) - Modular (telephone type)
foratsensorsandmini-plugforalternativesensors
• Cable strain relief
• Shock resistance bumpers
• Nurse call capable
1. RemovebatterycompartmentcoveronbackofalarmunitandinstalltwoAAbatteries,LEDwilllight,steady.
2. Press“BatteryTest”button.Ifalarmdoesn’tsoundcheckbatteryconnectionandretest.
Ifalarmfailstosoundreplacebattery.
a.Thisalarmhasalowbatteryalertandanalarmwillsoundduringa“lowbatterycondition”.
3. Inserteithermodularconnectorormini-phonepluginthealarm(IllustrationA).Threadthewirethrough
the“CableStainRelief”cavityasshownin(IllustrationB)LEDwillash.Makesurethatthereissome
slacksothatifthecordistuggeditdoesn’tputpressureontheconnector.Alarmwillsound.
Press“Auto-Reset”buttontosilencealarm.
4. Beforeusing,testunitbypressingthepalmofyourhandrmlyagainstthesensorpad.
Whiledoingthis,checkalarmtomakecertainthatLEDisashing.Removeyourhand.
Alarmshouldsound.Press“Auto-Reset”buttontosilencealarm.Ifalarmfailstosound,
makecertainthetelephonetypeplugissecure.
5. SelectHighorLowalarmvolumeusingtheswitchonrightsideofalarm,andselectTone1or2
6. Whenalarmisactivatedbytheresidentleavingthewheelchair,itmaybesilencedbypressingthe
“Auto-Reset”button.Thealarmisnowinstand-bymode.Thesystemwillautomaticallygointothe
monitoringmodewhentheresidentisreseated.There is no on/off switch.
Safety Reminders
1. Checkbatterybeforeeachuse.Ifalarmis
weak or fails to sound, replace the
twoAAbatteries.
2. Whenunitisinuse,checkLEDregularly.
Ifitisnotglowing,replacethetwoAAbatteries.
You may choose one of the three methods to attach the alarm unit to the wheelchair
Mounting Bracket
1. Removethesecond-from-topscrewfromeithertheleftorrightsideofthe
wheelchairbackrest(otherlocationsmaybechosenbythecaregiver).
2. Removegrommetfromscrew,placethescrewthroughholeonalarmunit
mountingbracket,andfastenscrew(withthegrommet)towheelchairbackrest(IllustrationC).
OPERATION
WHEELCHAIR INSTALLATION
Condition
Alarm sounds when resident stands
Alarm soundswhenmodularconnectorormini-phoneplugareremoved,
whileresident’ssittingonsensor
Alarm sounds when sensor wire (grey) is cut, while resident is sitting on sen-
sor.Alarmdoesnotsoundwhenauxiliarywire(black)iscut
Alarm sounds when batteries are low
Alarm Conditions
Note: The alarm will not sound when the sensor pad wire (grey) is cut or accidentally pulled out
unless it is activated. Also, when the sensor is activated by a resident sitting on it and the plug is
accidentally pulled out it will ring but when re-inserted, the alarm will stop ringing. All of the
conditions above apply either to the telephone type plug or mini-plug. The alarm sounds only when
the mini-plug (black) is removed while sensor is activated. It will not ring when this wire is cut.