SMC Networks HRW Series User manual

Fig. 1
HRW***-* -Y
DI control kit (optional)
Operation Manual
Applicable model: Series HRW
(Thermo Chiller)
- Understand the meaning of the following signs before reading the body of this manual, and keep the instruction.
1
Accessory installation
2
Procedure 1
Confirm that a nut, a ferrule, and an insert are installed at the end of the DI tube for IN (Fig. 1). Insert the DI tube
for IN to the end of the DI circuitSupply, and tighten the nut by hand. At that time, use two spanners, (1) and (2),
not to rotate the self-align fitting. Turn the hand-tightened nut by 1/4 rotation.
3
HRX-OM-K018-B
Circulating fluid Supply
Circulating fluid Return
- It is possible to control the electric resistance of circulating fluid as you like by using an ion exchange resin filter
(hereafter DI filter) and electric resistance meter (hereafter DI sensor).
- DI filter is not attached. Purchase our HRZ-DF001 separately if you need. If the DI filter is used at temperature outof
the range, 20 to 40degC, also purchase a thermal insulator, HRZ-DF002, to prevent frostbite and a burn.
Things to prepare
Spanner (width across flats: (1) 9/16 inch, (2) 1/2 inch, (3) 5/8 inch, (4) 11/16 inch)
DI circuit Supply
(DI tube connecting port for IN)
DI circuit Return
(DI tube connecting port for OUT)
Procedure 4
Supply circulating fluid from IN side of the DI filter to exhaust air. Continue to supply until OUT side is filled
with circulating fluid.
Fig. 2
Procedure 3
Mount a therml insulator for DI filter as shown at Fig. 2, and set it in the drain pan (only for the customers who purchased
it).
Procedure 5
Connect the DI tube for OUT and DI tube for IN to OUT side and IN side of the DI filter respectively. As theDI
tube has a strainer and a packing at the DI filter connection part, be careful not to drop them (Refer to Fig. 3).
Procedure 6
Perform trial operation in accordance with Thermo Chiller operation manual, HRX-OM-K003. Initially, theDI
filter has water. If ethylene glycol solvent is used, the concentration may be decreased. Confirm the concentration
after the trial operation. The proper concentration of ethylene glycol is 60%. If no leakage is found, installa
insulator to a ball valve connected to the DI tube, and fix it with a band(Refer to Fig. 4).
Fig. 3 Fig. 4
Introduction
Thank you very much for purchasing DI control kit for our Thermo Chiller, Series HRW.
This manual is for operators who have enough knowledge for general industrial equipment and devices, and also
thoroughly understand assembling, handling, and maintenance of them. Before assembling, handling, and
maintenance, read and understand this manual and the one for Thermo Chiller, Series HRW carefully.
Installing of accessories has to be performed by a customer according to this manual.
Refer to the operation manual for Thermo Chiller for warranty.
This manual is subject to change without prior notice.
Summary of option
Indication Meaning
Operator error could result in serious injury or loss of life.
!Warning
Operator error could result in injury or equipment damage.
Caution
Warning
Safety instruction
Caution
!
!
- Before using, understand the specification range thoroughly.
This product is designed as an option for Thermo Chiller, Series HRW. Do not use this product for the other
purpose, or outside of the specification range.
- Perform installation work without any circulating fluid in Thermo Chiller.
Perform the work before supplying circulating fluid in or after exhausting all the circulating fluid from
Thermo Chiller. The circulating fluid shoud be exhausted after the fluid gets to be ambient temperature.
Otherwise, an operator may get a burn or frostbite.
- Do not performinstallation work while power is on.
Perform lock-out and tag-out of the power securely. Otherwise, Thermo Chiller may operate unexpectedly.
- Understand the contents of thismanual and the working procedure thoroughly.
Understand this manual and the one for Thermo Chiller (Document no: HRX-OM-K003) thoroughly. Keep
this operation manual so that you can refer whenever necessary.
- Confirm there is no leakage or condensation after installation.
Confirm there is no leakage or condensation with Thermo Chiller operated. If leakage is found, stop Thermo
Chiller immediately.
- Do not disassemble or modify.
Otherwise it may cause leakage and operation failure.
Introduction
Refer to the accessory list on the back, and confirm the number of the accessory. In addition, confirm the
model no. plate on the body has "-Y" indication, which is symbol for this option.
DI tube forIN
DI tube for OUT
Nut
Ferrule
Insert
Procedure 2
Confirm that a nut, a ferrule, and a insert are installed at the end of the DI tube for OUT (Fig. 1). Insert the DI tube
for OUT to the end of the DI circuit RETURN, and tighten the nut by hand. At that time, use two spanners, (3) and
(4), not to rotate the self-align fitting. Turn the hand-tightened nut by 1/4 rotation.
Indication on the operation display panel
Status 1
Status 2
Model indication Menu 1
Maintenance
DI ACC reset
Initial setting1
Operation panel sequence
Status 3
Setting
(1) Setting of DI SP
Plate a cursor on "DI SP" on the screen "Setting", and press [ENT] key. Set the value to the one you desire
with [ ], [ ], and [ ] keys. After setting the value, press [ENT] key.
Description
DI PV
DI SP
DI SV
Indication
DI level at the circulating fluid Supply, measured by the DI sensor
Setting for DI level
Accumulated ON time of the solenoid valve (accumulated fluid-running time through the DI filter)
Status of the solenoid valve (ON=open, OFF=closed)
DI ACC
Operation with operation display panel
(2) Setting of DI HYS
Set a cursor on "DI HYS" on the screen "Initial setting 4", and press [ENT] key. Set the value to the one you
desire with[ ], [ ], and [ ] keys. After setting the value, press [ENT] key.
DI HYS Refer to "DI HYS" below.
(3) Setting of DI level lowering alarm (LOW DI)
Set a cursor on "LOW DI" on the screen "Initial setting 4", and press [ENT] key. Set the value to the one you
desire with [ ], [ ], and [ ]keys. After setting the value, press [ENT] key.
LOW DI When the DI level gets lower than the setting, an indication "DI LOW
LEVEL WRN" will be shown, and a buzzer will be given off.
If the setting has "0.0", no alarm will be given off.
(4) Reset of DI ACC
Press [ ] and [ ] keys while the screen "DI ACC reset" is shown, and select [YES]. Then, press [ENT] key.
Set range
-
0 to 20Mohm
Yes/No
ON/OFF
0 to 0.9Mohm
0 to 20Mohm
When shipped
from factory
-
0.5Mohm
No
-
0.3Mohm
0.0Mohm
(5) Status of DI circuit
You can confirm it on the screen "Status 3".
External communincation
Only DI PV can be output by external communication (serial RS-485). Refer to the separate manual,
"Communication Specification/ HRX-PS-K005", for detail. If analog communication (option) is used
refer to "Communication Specification/HRX-PS-K018".
DI HYS
Solenoid
valve
Operation
condition
DI LEVEL
DI SP
OFF(closed)
ON(open)
Stop
Operation
DI HYS
!
Menu 2
Initial setting2
Initial setting3
Initial setting4
© 2006 SMC CORPORATION All Rights Reserved
1
2
Thermo Chiller
HRW□□□-□□-Y
Rev.□□□
SMC Co.
TEMP PV 23.8℃
▲
▼
SEL
▲
▼
SEL
▲
▼
SEL
FLOW PV 20.0LPM
PRESS 0.85MPa
R.TEMP 23.8℃
TEMP PV 23.8℃
TEMP SV 25.0℃
OFFSET 0.0℃
OFFSET: OFF
DI PV 1.0MΩ
DI SP 0.5MΩ
DI ACC 0h
DI SV OFF
<MENU>
1.SETTING
2.REMOTE/LOCAL
3.INITIAL SET
<MENU>
4.MAINTENANCE
5.OPTION
<MAINTENANCE>
1.DI ACC RESET
2.VALUE OPEN
<DI ACC RESET>
DI ACC 0h
DATA RESET? No
▲
▼
▲
▼
▲
▼
SEL ENT
▲
▼
SEL ENT
<SETTING>
TEMP SP 25.0℃
DI SP 0.5MΩ
<INITIAL SET>
OFFSET 0.0℃
OFFSET: OFF
HIGH TEMP 93℃
<INITIAL SET>
LOW FLOW 0LPM
FLOW UNIT LPM
PRESS UNIT MPa
<INITIAL SET>
SLAVE ADRS. 1CH
OUT: N/A
E OUT: N/A
<INITIAL SET>
DI HYS 0.3MΩ
LOW DI 0.0MΩ
F.LOW TEMP 10℃

HRW***-* -Y
Material for DI control kit (optional)
1
2
Circulating fluid return
(Rc3/4)
Circulating fluidsupply
(Rc3/4)
DI circuit return
(3/8 self-align fitting/with plug)
DI circuit supply
(3/8 self-align fitting/with plug)
Applicable model
Available circulating fluid for DI circuit
DI level indication range
DI level setting range
Control solenoid valve hysteresis
-
MΩ・cm
MΩ・cm
MΩ・cm
-
0~20*2
0~20*1
HRW***-H1-Y/HRW***-H1S-Y HRW***-H2-Y/HRW***-H2S-Y
60% Ethylene glycol solvent DI water
Thermo Chiller DI circuit supply
:Option
Solenoid valve
DI sensor
Circulating fluid tank in Thermo Chiller
Piping forcustomer's system
Circurating
fluid supply
(Rc3/4)
Circulating fluid
return
(Rc3/4)
Strainer
Ball valve
(IN/OUT)
DI filter
(not attached)
DI circuit
Thermo Chiller DI circuit Return
*This option does not have DI filter.
For details, please refer to specification table and notes.
Facility water
Note
1.This product is designed as an option for Thermo chiller, Series HRW. Do not use this product for the other purpose, or outsi de
of the specification range.
2.Installation of accessories has to be performed by a customer according to the operation manual.
3.Supply circulating fluid from IN side of the DI filter to exhaust air. Continue to supply unitl OUT side is filled with circulati ng
fluid.
4.The DI filter has water inside. If ethylene glycol solvent is used, the concentration may be decreased.
Confirm the concentration after the trial operation. The proper concentration of ethylene glycol is 60 %.
Dimensional tolerance for each: 10mm
Accessories
Specification Table
DI tube for IN/OUT
(with a insulator) x 1pc for each Operation manual (this document)
[document no:HRX-OM-K018]Jpn x 1 copy, Eng x 1 copy
0~0.9
*1: DI filter (Part no.: HRZ-DF001) is necessary to control DI level. This optional product does not have DI filter,so
purchase it separately. If this product is used out of the range of 20 to 40degC, also purchase a thermal insulat or
for DI filter (our part no.: HRX-DF002) to avoid frostbite and burn.
*2: Temperature correction is not applied to the DI level indication value.
Note: Temperature stability at supply of circulating fluid in the Thermo Chiller could be out of +/-0.3degC depending
on DI filter replacement and operating condition s.
Flowchart
DI filter/ Thermal insulator for DI filter
(This option does not have DI filter and Thermal insulator.)
DI filter
Part No.:HRZ-DF001
Thermal insulator for DI filter
Part No.:HRZ-DF002
DI filter/ insulator for DI filter
mounted condition
(380)(860)
(665)
(242)
(294)
DI level lowering alarm setting range 0~20
(120)
(143)
MΩ・cm
1
2
Applicable model
HRW***-* -Y
HRW***-* S-Y
1
2
Other SMC Networks Control System manuals
Popular Control System manuals by other brands

Crestron
Crestron 3 Series Operation guide

Klimor
Klimor EVO -T Operation and maintenance manual

Instron
Instron Torsion Add-on Version 3 Operator's guide

Siemens
Siemens SINUMERIK 840DE Operation & programming manual

Cardin Elettronica
Cardin Elettronica ACD-148 manual

Mitsubishi Electric
Mitsubishi Electric M700V Series manual

Polyaire
Polyaire ZoneMaster ZoneSwitch V2 Operation and installation instructions

Tigo
Tigo TS4-R Installation & quick start guide

Weil-McLain
Weil-McLain BMC Install and operation instructions

Whelen Engineering Company
Whelen Engineering Company PCCS9R installation guide

CAME
CAME RBM21 - HW installation manual

Woodbridge
Woodbridge B0940S installation manual