manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. SMC Networks
  6. •
  7. Switch
  8. •
  9. SMC Networks ISE70 User manual

SMC Networks ISE70 User manual

Default setting
The default setting is as follows.
If no problem is caused by this setting,
keep these settings.
Outline of Settings [Measurement mode]
Pressure Setting
●[F 0]
Display units, switch output specifications
and diagnostic information selection function
●Other parameter settings
Other Settings
Peak/ ottom value indication
The max. (min.) pressure when the power is supplied is detected and updated.
The value can be displayed on the sub display by pressing the UP or DOWN button in
measurement mode.
Snap shot function
The current pressure value can be stored to the switch output ON/OFF set point.
When the set value and hysteresis are set, press the UP and DOWN buttons for 1 sec. or
longer simultaneously. Then, the set value of the sub display (right) shows [- - -], and then
values corresponding to the current pressure values are automatically displayed.
Zero-clear function
In measurement mode, when the UP and DOWN buttons are pressed for 1 sec. or longer
simultaneously, the main display shows [- - -], and the reset to zero.
The display returns to measurement mode automatically.
Key-lock function
To set each of these functions, refer to the SMC website (UR http://www.smcworld.com)
for more detailed information, or contact SMC.
Maintenance
How to reset the product after a power cut or forci le de-energizing
The setting of the product will be retained as it was before a power cut or de-energizing.
The output condition is also basically recovered to that before a power cut or deenergizing,
but may change depending on the operating environment. Therefore, check the safety of the
whole installation before operating the product. If the installation is using accurate control,
wait until the product has warmed up (approximately 10 to 15 minutes).
Default settings
When the pressure exceeds the set
value, the switch will be turned on.
When the pressure falls below the
set value by the amount of
hysteresis or more, the switch will
be turned off.
The default setting is to turn on
the pressure switch when the
pressure reaches the center of the atmospheric pressure and upper limit of the rated
pressure range. If this condition, shown to the right, is acceptable, then keep these settings.
DOWN buttonUP button
SET button
LCD display
Operation light (OUT1)
Operation light (OUT2)
IO-Link status indicator light
3 Step Setting Mode
Function Selection Mode
Simple Setting Mode
When tightening, do not hold the upper part
of the product (Display) with a spanner.
Trou leshooting
Error indication function
This function is to display error location and content when a problem or error has occurred.
How to use connector
Align the cable connector key groove with the product connector key to insert and
rotate the knurled part of the connector.
Connect the wires of the lead wire with M12 connector as shown below.
M12 connector (Port Class A)
4
13
2
Power is supplied
The product code is displayed for approximately 3 sec. after power is supplied.
Then, measurement mode will be displayed.
∗: Within approximately 0.2 sec. after power-on, the switch starts.
Press the SET
button once.
Press the SET
button between
1 and 3 sec.
[3 step setting
mode]
Set either of set
value or
hysteresis.
[Simple setting
mode]
Select the set
value, hysteresis
and delay time.
[Function
selection mode]
Change the
function settings.
[Other Settings]
•Zero-clear function
•Key-lock function
∗: The outputs will continue to operate during setting.
∗: If a button operation is not performed for 30 sec. during the setting, the display will flash.
(This is to prevent the setting from remaining incomplete if, for instance, an operator were to leave
during setting.)
∗:
3 step setting mode, simple setting mode and function selection mode settings are reflected each other.
Switch ON
At normal output Switch OFF
Set value
P_1 Hysteresis
H_1
Time
Pressure
[3 step setting mode (hysteresis mode)]
In the 3 step setting mode, the set value (P_1 or n_1) and hysteresis (H_1) can be changed.
Set the items on the sub display (set value or hysteresis) with the UP or DOWN button.
When changing the set value, follow the operation below. The hysteresis setting can be
changed in the same way.
(1) Press the SET button once when the item to be
changed is displayed on the sub display. The set
value on the sub display (right) will star flashing.
(2) Press the UP or DOWN button to change the set
value.
The set value can be increased with the UP button
and can be reduced with the DOWN button. When
the UP and DOWN buttons are pressed and held
simultaneously for 1 sec. or longer, the set value is displayed as [- - -], and the set value
will be the same as the current pressure value automatically (snap shot function).
Afterwards, it is possible to adjust the value by pressing the UP or DOWN button.
(3) Press the SET button to complete the setting.
The pressure switch turns on within a set pressure range (from P1 to P1H) during window
comparator mode. Set P1 , the lower limit of the switch operation, and P1H, the upper limit
of the switch operation and WH1 (hysteresis) following the instructions given above.
(When reversed output is selected, the sub display (left) shows [n1 ] and [n1H].)
∗: Set OUT2 in the same way. (ex. P_2, H_2)
∗: Setting of the normal/reverse output switching and hysteresis/window comparator mode switching are
performed with the function selection mode [F 1] OUT1 setting and [F 2] OUT2 setting.
(1) Press and hold the SET button between 1 and 3 sec. in measurement
mode. [SEt] is displayed on the main display. When the button is released
while in the [SEt] display, the current pressure value is displayed on the
main display, [P_1] or [n_1] is displayed on the sub display (left), and the
set value is displayed on the sub display (right) (Flashing).
(2) Change the set value with the UP or DOWN button, and press the SET
button to set the value. Then, the setting moves to hysteresis setting.
(The snap shot function can be used.)
(3) Change the set value with the UP or DOWN button, and press the SET button to set the
value. Then, the setting moves to the delay time of the switch output.
(The snap shot function can be used.)
(4) Press the UP or DOWN button, the delay time of the switch output can be selected.
Delay time setting can prevent the output from chattering.
The delay time can be set in the range 0.00 to 60.00 sec. in 0.01 sec. increments.
(5) Press the SET button for less than 2 seconds to complete the OUT1 setting.
[P_2] or [n_2] is displayed on the sub screen (left). Continue with setting the OUT2.
Press and hold the SET button for 2 seconds or longer to complete the setting.
The product will return to measurement mode.
In the window comparator mode, set P1 , the lower limit of the switch operation, and P1H,
the upper limit of the switch operation, WH1 (hysteresis) and dt1 (delay time) following the
instructions given above.
(When reversed output is selected, the sub display (left) shows [n1 ] and [n1H].)
∗: Set OUT2 in the same way.
Function selection mode
In measurement mode, press the SET
button between 3 and 5 sec., to display
[F 0]. Select to display the function to
be changed [F□□]. Press and hold the
SET button for 2 sec. or longer in
function selection mode to return to
measurement mode.
∗:
Some products do not have all the functions.
If no function is available or selected due
to configuration of other functions, [- - -] is
displayed on the sub display (right).
If the product is to be used with changed settings, refer to the SMC website
(UR http://www.smcworld.com) for more detailed information, or contact SMC.
•The upper part (display) of the product can be
rotated by 336°.
Rotating the display with excessive force will damage
the end stopper.
Item Default setting
Unit specification ["Nil" or M]:
MPa
PNP
A
Display unit
Switch output
specifications
Item Default setting
[F 3] Digital filter setting 0.00 ms
[F 4] Auto-preset function OFF
0.0%[F 6]
Fine adjustment of display value
std (Standard)[F10] Sub display setting
1000-split[F11]
Display resolution setting
0.0[F14] Zero cut-off setting
Item Default setting
[F80] Power saving mode OFF
[F81] Security code OFF
[F90] Setting of all functions OFF
-[F96]
Number of pressurizing errors
N/A (normal output)
[F98] Output check
OFF[F99]
Reset to default settings
Press the SET button
between 3 and 5 seconds.
Measurement mode
Function selection mode
F0 Function
setting F1 Function
setting F99 Function
setting
Item
Output mode
Reversed output
Pressure setting
Default setting
Hysteresis
Hysteresis mode
ISE70(G): 0.500 MPa/ISE71: 0.800 MPa/
ISE75G: 1.000 MPa/ISE76G: 2.50 MPa/
ISE77G: 5.00 MPa
Normal output
ISE70(G): 0.050 MPa/ISE71: 0.080 MPa/
ISE75G: 0.100 MPa/ISE76G: 0.25 MPa/
ISE77G: 0.50 MPa
Delay time 0.00 sec.
●[F 1] Setting of OUT1
Display colour
Output ON: Green Output OFF: Red
( inked to OUT1)
●[F 2] Setting of OUT2
Same setting as [F 1] OUT1.
The switch output load current is
80 mA or more.
Turn the power off and remove
the cause of the over current.
Then supply the power again.
Over
current
error
Error Error displayed Description Measures
Residual
pressure
error
Pressurizing
error
Version
does not
match
During zero clear operation, pressure
greater than ±7%F.S. is present. Note that
the mode is returned to measurement
mode automatically 1 sec. later. The zero
clear range varies by ±1%F.S. due to
variation between individual products.
Release the applied pressure
to atmospheric pressure, and
retry the zero clear operation.
Pressure exceeding the upper limit of
the set pressure range is applied.
Displayed if an internal data error has
occurred.
Reset applied pressure to a
level within the set pressure
range.
Turn the power off and on
again.
If the failure cannot be solved,
contact SMC.
Pressure exceeding the lower limit of
the set pressure range is applied.
,
,
,
System
error
IO- ink Master and product version
are not matched.
Mismatch because master version is 1.0.
Align the master IO- ink
version to the device.
,
No. Name
1 +
2 DO
-3
C/Q4
Function
18 to 30 VDC
Switch output 2
0 V
Communication data (IO- ink)/
Switch output 1 (SIO)
ead wire colour
Brown
White
Blue
Black
[Measurement mode]
Detects the pressure after power is supplied, and indicates the display and switch
operating status. This is the basic mode; other modes should be selected for set-point
changes and other function settings.
Current pressure value (Main display)
Set value or peak/bottom value
(Sub display (right))
Item (Sub display (left))
Sub display
In measurement mode, the sub display can be temporarily changed by pressing the UP
or DOWN buttons.
OUT1
set value
OUT2
set value
OUT2
hysteresis
OUT1
hysteresis
Peak
value
Bottom
value
∗: Arbitrary
display
Mode
display
∗: One arbitrary display mode can be added to the sub display by setting the [F10] sub display.
If the sub display is switched during the arbitrary display setting, the display will be returned to
the arbitrary display 30 sec. later. (The default setting does not include arbitrary display.)
Measurement mode screen
Press the SET
button between
3 and 5 sec.
Current
pressure
value
Diagnostic
information
Before Use
High-precision Digital Pressure Switch
ISE70/ISE71/
ISE70G/ISE75G/ISE76G/ISE77G
Thank you for purchasing an SMC ISE70/ISE71/ISE70G/ISE75G/ISE76G/ISE77G
Series High-precision Digital Pressure Switch.
Please read this manual carefully before operating the product and make sure you
understand its capabilities and limitations. Please keep this manual handy for
future reference.
To obtain the operation manual about this product and
control unit, please refer to the SMC website
(UR http://www.smcworld.com) or contact SMC directly.
Safety Instructions
CAUTION indicates a hazard with a low level of risk which, if
not avoided, could result in minor or moderate injury.
Caution:
Warning:
Danger:
WARNING indicates a hazard with a medium level of risk
which, if not avoided, could result in death or serious injury.
DANGER indicates a hazard with a high level of risk which,
if not avoided, will result in death or serious injury.
These safety instructions are intended to prevent hazardous situations and/or
equipment damage.
These instructions indicate the level of potential hazard with the labels of
"Caution", "Warning" or "Danger". They are all important notes for safety and must
be followed in addition to International standards (ISO/IEC) and other safety
regulations.
Operator
The operation manual is intended for those who have knowledge of machinery
using pneumatic equipment, and have sufficient knowledge of assembly,
operation and maintenance of such equipment. Only those persons are
allowed to perform assembly, operation and maintenance.
Read and understand the operation manual carefully before assembling,
operating or providing maintenance to the product.
Safety Instructions
Do not disassem le, modify (including changing the printed circuit oard) or repair.
An injury or failure can result.
Do not operate in an atmosphere containing flamma le or explosive gases.
Fire or an explosion can result.
This product is not designed to be explosion proof.
Do not use the product in a place where static electricity is a pro lem.
Otherwise it can cause failure or malfunction of the system.
If using the product in an interlocking circuit:
•Provide a dou le interlocking system, for example a mechanical system
•Check the product regularly for proper operation
Otherwise malfunction can result, causing an accident.
The following instructions must e followed during maintenance:
•Turn off the power supply
•Stop the air supply, exhaust the residual pressure and verify that the air is released efore performing
maintenance work
Otherwise an injury can result.
After maintenance is complete, perform appropriate functional inspections and leak tests.
Stop operation if the equipment does not function properly or there is a leakage of fluid.
When leakage occurs from parts other than the piping, the product might be faulty.
Disconnect the power supply and stop the fluid supply.
Do not apply fluid under leaking conditions.
Safety cannot be assured in the case of unexpected malfunction.
Do not touch the terminals and connectors while the power is on.
Otherwise electric shock, malfunction or damage to the product can result.
Do not operate the product outside of the specifications.
Do not use for flammable or harmful fluids.
Fire, malfunction, or damage to the product can result.
Verify the specifications before use.
Warning
Caution
Summary of Product parts
Names of individual parts
Mounting and Installation
Piping
Piping specification: -02 and -N02
After hand tightening, tighten the fitting using
a spanner on the flat surfaces of the fitting.
The tightening torque must be 8 to 12 Nm.
Piping specification: -F02
After hand tightening, tighten the fitting using
a spanner on the flat surfaces of the fitting.
The tightening torque must be 4 to 5 Nm.
336° rotation
Wiring
Connections should be made with the power supply turned off.
Use a separate route for the product wiring and any power or high voltage wiring.
Otherwise, malfunction may result due to noise.
If a commercially available switching power supply is used, be sure to ground the frame
ground (FG) terminal. If the switching power supply is connected, switching noise will
be superimposed and it will not be able to meet the product specifications.
In that case, insert a noise filter such as a line noise filter/ferrite between the switching
power supplies or change the switching power supply to the series power supply.
Note: Specifications are subject to change without prior notice and any obligation on the part of the manufacturer.
© 2017 SMC Corporation All Rights Reserved
Akihabara UDX 15F, 4-14-1, Sotokanda, Chiyoda-ku, Tokyo 101-0021, JAPAN
Phone: +81 3-5207-8249 Fax: +81 3-5298-5362
URL http://www.smcworld.com
PS※※-OMV0002-B
Specifications/Outline with Dimensions (in
mm)
Refer to the product catalog or SMC website (UR http://www.smcworld.com) for more
information about the product specifications and outline dimensions.
If the error cannot be reset after the above measures are taken, or errors other than above
are displayed, please contact SMC.
Refer to the SMC website (UR http://www.smcworld.com) for more information about
troubleshooting.
NOTE
•The direct current power supply to be used should be U approved as follows:
Circuit (of Class 2) which is of maximum 30 Vrms (42.4 V peak), with U 1310
Class 2 power supply unit or U 1585 Class 2 transformer.
•The product is a U approved product only if it has a mark on the body.
Installation
Mounting with racket
Mount the bracket to the body with mounting screws (Cross recessed round head
screw: Nominal size M4 x 6 (2 pcs)), then set the body to the specified position.
∗: Tighten the bracket mounting screws to a torque of 0.76±0.1 Nm.
Cross recessed
round head screw:
Nominal size M4 x 6L
Bracket A
Cross recessed
round head screw:
Nominal size M4 x 6L
Bracket B
•Bracket A (Part No.: ZS-50-A) •Bracket B (Part No.: ZS-50-B)
Mounting with orifice
∗: Tighten the orifice to a torque of 1.5±0.1 Nm.
∗: ISE70G/ISE75G/ISE76G/ISE77G can be mounted.
Orifice
•Orifice (Part No.: ZS-48-A)
Spanner
Upper part
of the product
(Display)
Fitting
(width across flats: 24 mm)

Other manuals for ISE70

3

This manual suits for next models

5

Other SMC Networks Switch manuals

SMC Networks EX600-WD Series User manual

SMC Networks

SMC Networks EX600-WD Series User manual

SMC Networks 6152L2 Programming manual

SMC Networks

SMC Networks 6152L2 Programming manual

SMC Networks D-M9 Series User manual

SMC Networks

SMC Networks D-M9 Series User manual

SMC Networks SMC TigerAccess SMC7824M/ESW User manual

SMC Networks

SMC Networks SMC TigerAccess SMC7824M/ESW User manual

SMC Networks 6724AL2 Programming manual

SMC Networks

SMC Networks 6724AL2 Programming manual

SMC Networks TigerSwitch SMC8124PL2 User manual

SMC Networks

SMC Networks TigerSwitch SMC8124PL2 User manual

SMC Networks ISE70 User manual

SMC Networks

SMC Networks ISE70 User manual

SMC Networks 6624FMSC - annexe 1 User manual

SMC Networks

SMC Networks 6624FMSC - annexe 1 User manual

SMC Networks 6900FSC - annexe 1 Programming manual

SMC Networks

SMC Networks 6900FSC - annexe 1 Programming manual

SMC Networks EZ Connect User manual

SMC Networks

SMC Networks EZ Connect User manual

SMC Networks ZSE40A User manual

SMC Networks

SMC Networks ZSE40A User manual

SMC Networks EZ Connect User manual

SMC Networks

SMC Networks EZ Connect User manual

SMC Networks SMC6724L2GLSC User manual

SMC Networks

SMC Networks SMC6724L2GLSC User manual

SMC Networks SMC8612XL3 F 1.0.1.3 Programming manual

SMC Networks

SMC Networks SMC8612XL3 F 1.0.1.3 Programming manual

SMC Networks EZ Switch SMC8506T User manual

SMC Networks

SMC Networks EZ Switch SMC8506T User manual

SMC Networks Edge-core ES4704BD User manual

SMC Networks

SMC Networks Edge-core ES4704BD User manual

SMC Networks TigerStack II SMC8848M User manual

SMC Networks

SMC Networks TigerStack II SMC8848M User manual

SMC Networks ZSE40A User manual

SMC Networks

SMC Networks ZSE40A User manual

SMC Networks TigerSwitch SMC6128PL2 User manual

SMC Networks

SMC Networks TigerSwitch SMC6128PL2 User manual

SMC Networks EZ Connect User manual

SMC Networks

SMC Networks EZ Connect User manual

SMC Networks 6900FSC - annexe 5 Programming manual

SMC Networks

SMC Networks 6900FSC - annexe 5 Programming manual

SMC Networks SMC6224M User manual

SMC Networks

SMC Networks SMC6224M User manual

SMC Networks ZSE20C Series User manual

SMC Networks

SMC Networks ZSE20C Series User manual

SMC Networks ZSE20CF Series User manual

SMC Networks

SMC Networks ZSE20CF Series User manual

Popular Switch manuals by other brands

Samsung WAM250 user manual

Samsung

Samsung WAM250 user manual

FXC FXC5024 user manual

FXC

FXC FXC5024 user manual

EuroLite LVH-7 user manual

EuroLite

EuroLite LVH-7 user manual

Dungs AA-A1-0 Series installation instructions

Dungs

Dungs AA-A1-0 Series installation instructions

Lightwave Communications ServerSwitch 8200 user manual

Lightwave Communications

Lightwave Communications ServerSwitch 8200 user manual

Lutron Electronics Maestro MS-OPS6-DDV manual

Lutron Electronics

Lutron Electronics Maestro MS-OPS6-DDV manual

Digitus DN-80111 manual

Digitus

Digitus DN-80111 manual

SsecureNetworx StreamBuddy user guide

SsecureNetworx

SsecureNetworx StreamBuddy user guide

Hawking HKS104 Quick installation guide

Hawking

Hawking HKS104 Quick installation guide

Huawei CR-SRUB Specifications

Huawei

Huawei CR-SRUB Specifications

EUCHNER STP AS1 Series operating instructions

EUCHNER

EUCHNER STP AS1 Series operating instructions

D-Link DSS-100E-18P Quick installation guide

D-Link

D-Link DSS-100E-18P Quick installation guide

Clas Ohlson MA-001 quick start guide

Clas Ohlson

Clas Ohlson MA-001 quick start guide

Kathrein EXD 158 Twin manual

Kathrein

Kathrein EXD 158 Twin manual

3Com SuperStack 4 Configuration guide

3Com

3Com SuperStack 4 Configuration guide

Cisco Intelligent Gigabit Ethernet Switch Module manual

Cisco

Cisco Intelligent Gigabit Ethernet Switch Module manual

Farnell SY-UHD-S41A installation guide

Farnell

Farnell SY-UHD-S41A installation guide

DRIESCHER-WEGBERG LDTM Operation and assembly instructions

DRIESCHER-WEGBERG

DRIESCHER-WEGBERG LDTM Operation and assembly instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.