4 Summing amplifier
The summing amplifier (module AMPLIFIER) is located near the power supply
unit. The magnetic field compensation system must not be operated without
amplifier module installed because the amplifier also provides the system’s analog
ground reference signal. The amplifier adds and amplifies the signals of up to three
magnetometer/filter modules. The amplifier’s gain can be adjusted in steps from
0 to 11. The LED bar graph display shows the DC drive level of the amplifier.
The display characteristic is non-linear. The amplifier output signal is fed to the
BNC connector OUT.
The amplifier module contains a window comparator which automatically resets
the magnetometers in case of an overrange condition, see section 5.
5 Magnetometer
The measurement range of the fluxgate magnetometer is ±199.9 µT. The mea-
surement direction is axial, pointing away from the sensor cable. The sensor is
labeled with the serial number and a letter A, B, or C. It must be connected to
the corresponding receptacle SENSOR A, SENSOR B, or SENSOR C on the rear
panel. If two or three magnetometer modules are installed, then they must not be
interchanged.
The digital display shows the DC magnetic field in µT. The display range can
be changed between ±19.99 µT and ±199.9 µT with the switch RANGE. The
magnetometer is zeroed by pressing the RESET button. The actually measured
DC magnetic field component is subtracted from the magnetometer signal. This
procedure is indicated by the RESET LED and takes four seconds. After zeroing
the display shows a small value. The maximum field which can be zeroed is
±500 µT (5 Gauss). The magnetometer is automatically reset when the amplifier
detects an overrange condition, see section 4. The auto-reset function can be
deactivated by replacing a jumper on the magnetometer board.
The analog output signal at the BNC connector has a bandwidth of 1 kHz (−3 dB).
The scale factor is 3 V/100 µT in range 199.9 µT (RANGE switch up) and
3 V/10 µT in range 19.99 µT (RANGE switch down). The scale factor of the
signal internally routed to the filter module does not depend on the range switch
position and is 3 V/100 µT.
5