Sterlitech 1812 Supplement

1812 Spiral Wound
Membrane Housing
Assembly & Operation Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com

www.sterlitech.com 1-877-544-4420info@sterlitech.com
Table of Contents
2
3
5
8
4
10
10
10
11
11
10
13
13
13
14
12
Page
1. Introduction
3. 1812 Housing Assembly
6. Supplementary Operating Information
8. Return Material Authorization
2. 1812 Housing Components
2.1. Additional Equipment
6.1. Membrane Performance Results
6.3. Temperature Limits
6.2. Membrane Performance
6.4. Membrane Cleaning
5. Operation of the 1812 Housing Assembly
4. 1812 Housing Assembly Connections and Parts
7. Accessory and Spare Part Ordering Information
9. Warranty
10. Technical Assistance
Appendix 1: Bench Scale Filtration Products

2
1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
The 1812 Spiral Wound Membrane Housing is a bench-scale crossflow filtration system that is designed
to evaluate 1812 spiral wound membrane elements in a variety of applications. It simulates the flow
dynamics of commercially available spiral wound membraneelements. Users can vary the operating
conditions and fluid dynamics over broad ranges.
Table 1 below outlines the features and technical specifications of the 1812 Spiral Wound Membrane
Housing (hereafter referred to as 1812 Housing).
Prior to operating or servicing this device, this manual must be read and understood.
1. Introduction
Parameter Description
Membrane Active Area
Maximum Pressure
Cross Flow Velocity
Inlet/Outlet/Permeate Tube ID
Variable (1 to 5 ft2)
SS316
41.3 bar (600 psig)
Variable
O-rings
Length
Buna-N (Other materials) available)
374.6 mm (14.75 inches)
9.3 mm (0.365 inches)
48 mm (1.88 inches)
12.7 mm (1/2 inches)
1812 Housing
Maximum Temperature
Dimensions
Inlet/Outlet/Permeate Tube OD
177 °C (350 °F)
pH Range
Diameter
Membrane Dependent
Table 1: 1812 Housing Features and Technical Specification

1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
3
The 1812 housing consists of the following components. Figure 1 illustrates a typical
assembly of the 1812 Housing.
Figure 1: 1812 Housing Assembly
Legend:
1. Stainless Steel 316 housing
2. 2inch gaskets for top and bottom end caps
3. Permeate line adapter for top, hollow
4. Permeate line adapter for bottom, solid
5. Top endcap
6. Bottom end cap
7. 2 inch Tri clamp fittings
8. O-rings for permeate line adapter
3
5
7
7
4
8
8
6
12
2
2. 1812 Housing Compnents

4
1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
Operation of the 1812 housing cell requires additional equipment that are sold
separately as listed below. To inquire about a complete system contact
sales@sterlitech.com.
• Feed Pump
• Feed Tank
• 1812 Spiralwound Membrane Elements
• Concentrate Control Valve
2.1. Additional Equipment

1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
5
After verifying that all of the necessary components were shipped and present, you can begin
the assembly of the system.
To assemble the 1812 Element into the housing
1. If you are using a female 1812 element, Insert the adaptors into the element (Figure 2,
#1,2)
2. Remove clamp (Figure 3, #1) and open up the Bottom End Cap (keep the Top End Cap
intact, Figure 4, #1).
3. Insert element into the housing in the correct orientation, i.e. inserted from the side of
the housing where the bottom end cap is, so that O-rings on the permeate line adaptor fits
snugly in the Top End Cap.
4. Push Bottom End Cap until element is fully inserted into the housing. This may take a bit
oorce.
Note: Applying Glycerin to the two O-rings installed on the permeate tube is
recommended to ensure that there will be no leaks.
5. Put high-pressure clamp back on the Bottom End Cap and tighten.
WARNING: Do not exceed 600 psi (41 bar) in the 1812 Housing.
To remove the 1812 Element from the housing:
1. Remove the clamp (Figure 4, #1) and open up Top End Cap
2. Remove the element by pulling the element through the Top End Cap (Figure 4, #1).
3. Separate the element from the Top End Cap by pulling it out.
4. If you are not using this element any more, Remove the adaptors and save them for next
element.
3. 1812 Housing Assembly

6
1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
Figure 2: Permeate adaptors insertion
Figure 3: 1812 Element insertion into the housing and Sanitary Fittings on
Bottom End Cap
3
21
Legend:
1. Permeate line adaptor for bottom, solid
2. Permeate line adaptor for top, hollow
3. 1812 element
Legend:
1. Bottom End Cap
2. Permeate line adaptor for bottom, solid
3. 1812 element
12
3

1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
7
The 1812 Housing is now assembled and ready to be connected to a feed pump for operation.
Figure 4: 1812 element insertion and Sanitary Fittings on Top End Cap
Legend:
1. Sanitary Fittings on the Top EndCap
2. Permeate line adaptor for top, hollow
3. Feed inlet
4. Retentate outlet
5. Top end cap connected to permeate line adaptor
6. Top end cap disconnected permeate line adaptor
12
4
5
6

8
1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
Figure 5: 1812 Housing Assembly Connections and Parts
4. 1812 Housing Assembly Connections and Parts
1
2
3
5
6
4
7
11
A
A
B
C
D
9
10
8
1812 FLOW DIAGRAM
ITEM NO.
PART DESCRIPTION
1
FEED TANK
2
PRESSURE RELIEF VALVE
3
FEED PUMP
4
FEED BY-PASS VALVE
5
FEED BY-PASS FLOW METER
6
FEED PRESSURE GAUGE
7
1812 HOUSING
8
CONCENTRATE CONTROL VALVE ASSEMBLY, (0-1000PSI)
9
CONCENTRATE FLOW METER
10
PERMEATE FLOW METER
11
PERMEATE TANK
A
3/8" HIGH PRESSURE TUBING, FEED LINE
B
3/8" LOW PRESSURE TUBING, BY-PASS
C
3/8" LOW PRESSURE TUBING, RETURN LINE
D
3/8" LOW PRESSURE TUBING, PERMEATE OUTPUT LINE
1812 Flow Diagram
Item No.
4
8
Part Description
Feed By-Pass Valve
Concentrate Control Valve Assembly, (0-1000 PSI)
2
6
10
B
Pressure Relief Valve
Feed Pressure Gauge
Premeate Flow Meter
3/8” Low Pressure Tubing, By-Pass
1
5
9
Feed Tank
Feed By-Pass Flow Meter
Concentrate Flow Meter
3
7
11
C
A
D
Feed Pump
1812 Housing
Permeate Tank
3/8” Low Pressure Tubing, Return Line
3/8” High Pressure Tubing, Feed Line
3/8” Low Pressure Tubing, Permeate Output Line

1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
9
Figure 5 shows typical plumbing connections that need to be made for the operation of
the 1812 Housing, the configuration can change slightly depending on the objectives of the
user. In general, the connections to be made are:
1. Feed vessel to the pump inlet
2. Pump outlet to the feed inlet of the 1812 housing
3. Concentrate outlet of the 1812 housing to the feed vessel
4. Permeate collection tube to the permeate collectionvessel
Figure 5 also lists the additional components that are necessary to operate the 1812
Housing. These components are sold separately and can be found in the Membrane/
Process Development section of the Sterlitech website (http://www.sterlitech.com/
membrane-process-development.html). The final assembly of the system will vary with the
feed vessels, pumps, and tubing used.

10
1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
5. Operation of the 1812 Housing Assembly
6. Supplementary Operating Information
6.1. Membrane Performance Results
6.2 Membrane Performance
Once the 1812 Housing has been assembled and connected to a feed pump, it can be used
in reverse osmosis, ultrafiltration, nanofiltration, and microfiltration separation processes
for applications in a variety of industries, including food and beverage, pharmaceutical, or
industrial.
To operate the 1812 Housing Assembly:
1. Open all of the valves in the system to relieve any pressure build-up before turning on
the feedpump.
2. Turn the feed pumpon.
3. Set the feed pressure to be less than 600 psig by gradually closing the concentrate
control valve (Figure 5,8).
4. Verify the feed pressure reading on the concentrate pressure gauge (sold separately)
(Figure 5, 8).
5. Adjust the bypass valve (Figure 5, 4) and concentrate control valve (Figure 5, 8) to
obtain the desired pressure and flow. Experimentation enables you to determine the
optimum settings for pressure, and flow rate to use on the chosen membrane and the fluid
being processed.
6. Once the experiment is complete, shut down the system by turning the feed pump o and
opening all of the valves to relieve any pressure buildup.
Spiral wound membranes containing mesh spacers are usually operated with a fluid velocity
across the membrane surface of 0.1–0.5 m/sec (0.3–1.6 ft/sec). Higher velocities in membranes
may lead to excessive pressure dierential across the membrane and possible damage.
Other parameters such as viscosity, pressure, and suspended solids may also aect
performance/operation. Experimentation with the 1812 Housing can help predict the best
operating parameters. If your pump is delivering too much flow, a portion of the flow can
be diverted back to the feed container before entering the feed inlet of the cell body. This
requires installation of an optional bypass valve and fitting on the pump outlet (Figure 4, 4),
which is not supplied with the housing.

1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
11
6.3. Temperature Limits
6.4. Membrane Cleaning
The membranes, feed spacer, permeate, O-rings, and cell body materials construction dictates
maximum operating temperatures.
The 1812 housing can be cleaned easily after the membrane is removed. However, you may
wish to simulate the actual cleaning conditions of Cleaning-In-Place (CIP) in larger systems
with spiral-wound or tubular membranes. CIP chemically cleans without removing the
membrane to mechanically scrub the unit.
During CIP, cleaning solutions are re-circulated and, in some cases, allowed to sit for a period
of time within the cell body. In some cases, the feed pump can be used to re- circulate the
cleaning solutions.
Components Maximum Temperature
1812 Housing
Buna-N
177 °C (350 °F)
Variable
Viton 200 °C (400 °F)
121 °C (250 °F)
150 °C (300 °F)
Membrane Elements
O-rings
EPDM
Table 2: Upper Temperature Limits

12
1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
7. Accessory and Spare Part Ordering Information
Accessories and spare parts for the 1812 housing can be ordered by calling Sterlitech Corporation at
1-877-544-4420 or by visiting www.sterlitech.com.
Table 3: Accessory and Spare Part Ordering Information
Product
Square Parts
Shipping Weight Ordering Number
1812 Housing, for operation to 41.3 bar (600 psig)
Buna-N O-rings for adapters
1812 Membrane Elements
4.67 kg (10.3 lbs)
Contact Sterlitech Contact Sterlitech
1180147
1180146
Contact Sterlitech
1180145
1149419
1180140
0.05 kg (0.028 kg)
Contact Sterlitech
Permeate line adaptor for bottom, solid 0.5 lbs (0.22 kg)
Feed Flow Pump
Permeate line adapter for top, hollow 0.5 lbs (0.22 kg)
Concentrate Control Valve 2 lbs (0.9 kg)

1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
13
8. Return Material Authorization
9. Warranty
10. Technical Assistance
If materials are to be returned to Sterlitech for repair, evaluation, or warranty consideration, a Return
Material Authorization (RMA) number and form must be obtained from Sterlitech prior to the return.
Contact Sterlitech’s Customer Service Department for these forms.
The form must be completed and returned with the material. Be sure to include a complete, detailed
written reason for the return. Also, include serial numbers, installation and removal dates, and any other
pertinent information that is available. 1812 Housings have a serial number imprinted on the cell bottom.
Indicate the proposed disposition of the material, and reference the RMA number on all packages or
cartons. All material must be shipped to Sterlitech with freight prepared by the customer.
The following is made in lieu of all other warranties expressed or implied. Sterlitech Corporation
guarantees equipment to be free from defects in material and workmanship when operated in
accordance with written instructions for a period of one year from receipt. Parts not manufactured by
Sterlitech are covered by their manufacturer’s warranties, which are normally for one year.
Manufacturers and Seller’s only obligation shall be to issue credit against the purchase or replacement of
equipment proved to be defective in material or workmanship. Neither manufacturer nor seller shall be
liable for any injury, loss or damage, direct or indirect, special or consequential, arising out of the use of,
misuse, or the inability to use such product.
The information contained herein is based on technical data and tests, which we believe to be reliable,
and is intended for use by persons having technical skill at their discretion and risk. Since conditions
of use are outside Sterlitech’s control, we can assume noliability whatsoever for results obtained or
damages incurred through the application of the data presented.
This information is not intended as a license to operate under, or a recommendation to infringe upon,
any patent of Sterlitech or others covering any material or use.
The foregoing may not be altered except by a written agreement signed by ocers of the manufacturer.
Please contact us if you have any questions or technical inquiries about our products by calling
Sterlitech Corporation at 1-877-544-4420 or by visiting www.sterlitech.com.

14
1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
Appendix 1: Sterlitech Bench Scale Filtration Products
Filter Holder
Membrane
Active Area
Typical
Permeate Flux
Typical
Permeate Flow
Rate (per Cell)
System
Capacity
Min. Batch
Volume (per
Cell)
Filtration
Mode(s)
Cell Material
Max Operating
Pressure
HP4750(X)
14.6 cm2(2.26in2)
30-300 LMH
(17.6-176 GFD)
1.5-15 mL/min
0.7-7L/day
(0.2-2 GPD)
300mL
300mL
Dead-End Stirred
Cell
SS316,HastelloyTM
HP4750: 69bar
(1,000 psi)
HP4759X: 172bar
(2,500 psi)
20.6 cm2(3.2in2)
30-300 LMH
(17.6-176 GFD)
1-10mL/min
1.5-15L/day
(0.4-4 GPD)
3.7 L(1gal)
3.7 L(1gal)
Cross Flow, Forward
Osmosis
PTFE, Delrin,Acrylic,
SS316, HastelloyTM
PTFE, Delrin,Acrylic,
SS316, HastelloyTM
PTFE/Acrylic: 27.6bar
(400psi)
Delrin/SS316/
HastelloyTM :69bar
(1,000 psi)
PTFE/Acrylic: 27.6bar
(400psi)
Delrin/SS316/
HastelloyTM :69bar
(1,000 psi)
Acrylic: 15.2bar
(200psi)
SS316/HastelloyTM
:69bar (1,000 psi)
Acrylic, SS316,
HastelloyTM
SS316
41.4bar (600 psi)
42 cm2(6.5in2)
30-300 LMH
(17.6-176 GFD)
2-20 mL/min
3-30L/day
(0.5-5GPD)
3.7L(1gal)
3.7L(1gal)
Cross Flow, Forward
Osmosis
140 cm2(6.5in2)
30-300 LMH
(17.6-176 GFD)
7-70mL/min
10-100L/day
(0.5-5GPD)
3.7L(1gal)
3.7L(1gal)
Cross Flow, Forward
Osmosis
0.27-0.46 m
(3-5ft2)
30-300 LMH
(17.6-176 GFD)
350-2,300mL/min
194-1,940 L/gal
(51-510 GPD)
15L(4gal)
15L(4gal)
Cross Flow
Discoverer
CF016
Innovator
CF042
Explorer
Sepa
Developer
1812
Investigator

1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
15
Notes:

16
1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
Notes:

1812 Housing Manual
www.sterlitech.com 1-877-544-4420info@sterlitech.com
17

18
www.sterlitech.com 1-877-544-4420info@sterlitech.com
Founded in 2001 in Kent, WA, Sterlitech Corporation manufactures and markets filtrationfocused
laboratory products to a broad spectrum of scientific and industrial sectors. Its line of
flat sheet membranes and tangential flow cells deliver industry-leading performance and reliable
results. Configured for reverse osmosis, nanofiltration, ultrafiltration, and microfiltration
applications, Sterlitech’s bench scale test equipment provides the versatility required to
innovate.
Sterlitech’s comprehensive line of products is supported by the expertise of its technical
specialists who can assist with application-specific product selection, and provide customized
solutions where necessary. Unique problem-solving approaches, flexibility, and consistent
quality have made Sterlitech Corporation a renowned global provider of filtration products and
equipment.
General Corporate Information
Sterlitech Corporation
22027 70th Avenue S
Kent, WA 98032-1911 USA
Tel: 877-544-4420 or 1-253-437-0844
Fax: 1-253-437-0845
info@sterlitech.com
Sales Inquiries
sales@sterlitech.com
Accounts Payable
AP@sterlitech.com
Accounts Receiveable
AR@sterlitech.com
Press Inquiries
PR@sterlitech.com

22027 70th Ave. S
Kent, WA 98032-1911 USA
Phone: (253) 437-0844
Fax: (253) 437-0845
Email: sales@sterlitech.com
© 2019 Sterlitech Corporation
2019.V.1
Other manuals for 1812
1
Table of contents
Other Sterlitech Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Julabo
Julabo CORIO CP Original operating instructions

Shodex
Shodex IC YK-421 Operation manual

HAVER & BOECKER
HAVER & BOECKER HAVER EML 200 Premium operating instructions

Cole Parmer
Cole Parmer ShatterBox SM-300 Operation manual

ABS
ABS PH-ABT-UCFS-0220M owner's guide

CLIMET INSTRUMENTS
CLIMET INSTRUMENTS CI-154 user manual