J2010-6
V1924-ADD1 REV 0
Pulse Volume
Use1. Up and Down keys (no) to raise or lower pulse beep volume.
Continue to press the2. Down key to silence the pulse tone completely.
Entering Setup Menu
Power oximeter on by pressing the1. On/O key ( x) located on the device’s upper right side.
When the patient number (Example:2. P1) is displayed, press and hold the Exit key ( ) and then
press the Menu key ( ).
The Averaging Menu will be the rst option.3. AVG will be displayed. Release the Exit key ( ).
Setting Averaging
While1. AVG is being displayed, use the Up and Down keys (no) to select the desired SpO2averaging (1-4).
Press the2. Menu key ( ) to advance to the next menu.
To exit menus, press the3. Exit key ( ) or wait 20 seconds and menu will be exited automatically.
Setting Sensitivity
Advance from previous menu or see1. Entering Setup Menu. KS (for High Sensitivity) will be displayed.
Use the2. Up and Down keys (no) to turn the High Sensitivity setting 0n or 0ff.
Display Brightness
Advance from previous menu or see1. Entering Setup Menu. The display will show db.
Use the2. Up and Down keys (no) to select the desired brightness, ranging from 1-10. The default setting is 10
(brightest).
Setting the Date
Advance from previous menu or see1. Entering Setup Menu. To skip initial menus, if setting only date, press Menu key
() three (3) times until dt1 (Date Setting 1) is displayed.
Use the2. Up and Down keys (no) to select the year. Range is 00 (2000) to 99 (2099).
Press the3. Menu key ( ) to advance to the Month Menu. The display will show dt2 (Date Setting 2).
Use the4. Up and Down keys (no) to select the month. Range is 01 (January) to 12 (December).
Press the5. Menu key ( ) to advance to the Day Menu. The display will show dt3 (Date Setting 3).
Use the6. Up and Down keys (no) to select the day. Range is from 01 to 31.
Setting the Time
Advance from previous menu or see1. Entering Setup Menu. To skip initial menus, if setting only time, press Menu key
() six (6) times until t1 (Time Setting 1) is displayed.
The display will show2. t1. Use the Up and Down keys (no) to select the hour. Range is 00 (Midnight)
to 23 (11PM).
Press the3. Menu key ( ) to advance to Minute Menu. The display will show t2. Range is 0-59.
Press4. Exit key ( ) to exit Setup Menus.
f Attention, see operation manual for Warnings, Cautions, Notes and additional instructions for use.
Page 2 of 2
Quick Reference Guide
V1030
Manufactured By: Smiths Medical PM, Inc.
N7W22025 Johnson Drive
Waukesha, Wisconsin 53186-1856 USA
Phone: 1-262-513-8500
Toll-Free: (U.S.A. only) 1-888-745-6562
Fax: 1-262-542-0718
www.surgivet.com
SurgiVet and the Smiths design mark are trademarks of the Smiths Medical family of companies. The
symbol ®indicates the trademark is registered in the U.S. Patent and Trademark Oce and certain
other countries.
©2010 Smiths Medical Family of companies. All rights reserved.
H