SurgiVet V3395 User manual

SurgiVet V3395 - Pulse Oximeter,
Respiration & Temperature
Operation Manual
85 98
101.2 15F
SURGIVET, INC.
N7 W22025 JOHNSON ROAD, SUITE A
WAUKESHA, WI 53186
CATALOG V1907
VERSION 2, JUNE 2002
©2002, SURGIVET, INC.


Table of Contents
Operation Manual SurgiVet V3395 i
SurgiVet V3395--Pulse Oximeter, Respirations, & Temperature
Operation Manual
Warranty & Service Information ............................................................................................................................iii
Proprietary Notice...................................................................................................................................................iii
Limited Warranty....................................................................................................................................................iii
Loaner Device (Domestic Sales Only)....................................................................................................................iii
Service Support.......................................................................................................................................................iii
Chapter 1: Introduction 1-1
Intended Use ........................................................................................................................................................ 1-1
Symbol Definitions .............................................................................................................................................. 1-1
2Warnings, Cautions, and Notes....................................................................................................................... 1-1
General Warnings ................................................................................................................................................ 1-2
Oximetry....................................................................................................................................................... 1-2
Cautions ............................................................................................................................................................... 1-3
General Notes ...................................................................................................................................................... 1-3
Oximetry....................................................................................................................................................... 1-4
Respiration & Temperature .......................................................................................................................... 1-4
Chapter 2: Description of Measurements, Controls and Features 2-1
General Description ............................................................................................................................................. 2-1
Front Panel........................................................................................................................................................... 2-1
LCD Display........................................................................................................................................................ 2-3
Keys ..................................................................................................................................................................... 2-4
Top Panel ............................................................................................................................................................. 2-5
Side Panel ............................................................................................................................................................ 2-5
Chapter 3: Setting Up the Monitor 3-1
Unpacking the Monitor and Checking the Shipment ........................................................................................... 3-1
Turning Alarm and Alert Tones On and Off........................................................................................................ 3-1
Working With Menus .......................................................................................................................................... 3-1
Working With System-Wide Settings.................................................................................................................. 3-3
Setup/Volume Menu ............................................................................................................................................ 3-3
Turning Parameter Monitoring On and Off ......................................................................................................... 3-3
Chapter 4: Alarms 4-1
High Priority Alarms ........................................................................................................................................... 4-1
Low Priority Alarms ............................................................................................................................................ 4-1
Working With the Alarms Menu ......................................................................................................................... 4-2
Adjusting or Viewing Alarm Limits ............................................................................................................. 4-3
Alarm Tones ........................................................................................................................................................ 4-3
Turning Alarm and Alert Tones On and Off................................................................................................. 4-3
System Information Signal: Low Battery ..................................................................................................... 4-3
Chapter 5: Oximetry 5-1
General Description ............................................................................................................................................. 5-1
Pulse Oximetry Theory of Operation................................................................................................................... 5-1
Oximeter Display................................................................................................................................................. 5-2
Oximeter Menu .................................................................................................................................................... 5-3
Adjusting the Pulse Beep Volume ....................................................................................................................... 5-3
Averaging Settings............................................................................................................................................... 5-3
Oximeter Messages.............................................................................................................................................. 5-4
Alert Messages..................................................................................................................................................... 5-4
Messages....................................................................................................................................................... 5-4
Attaching the Patient - Oximetry ......................................................................................................................... 5-5
Choose the Sensor......................................................................................................................................... 5-5
Check Sensor Integrity.................................................................................................................................. 5-5
Clean or Disinfect the Reusable Sensor............................................................................................................... 5-6

Table of Contents
ii
Attach the Sensor to the Patient............................................................................................................................5-6
Reflectance R/E Sensor ........................................................................................................................................5-7
Chapter 6: Respiration 6-1
Respiration Theory of Operation ..........................................................................................................................6-1
Respiration Menu .................................................................................................................................................6-1
Chapter 7: Temperature 7-1
Temperature Theory of Operation........................................................................................................................7-1
Temperature Menu ........................................................................................................................................7-1
Units ..............................................................................................................................................................7-1
Chapter 8: Trends 8-1
Trends Menu.........................................................................................................................................................8-2
Chapter 9: Printer 9-1
Printer Menu.........................................................................................................................................................9-1
Output Examples: .................................................................................................................................................9-2
Patient Data ...................................................................................................................................................9-2
Chapter 10: Routine Maintenance 10-1
Operator’s Maintenance .....................................................................................................................................10-1
Charging the Battery...........................................................................................................................................10-1
Cleaning and Disinfecting the Monitor ..............................................................................................................10-2
Long Term Storage.............................................................................................................................................10-2
Chapter 11: Troubleshooting 11-1
Appendices Appendix-1
Appendix A: Supplies and Accessories ................................................................................................................... 1
Appendix B: Specifications ..................................................................................................................................... 2
Parameters Monitored.............................................................................................................................................. 2
Graphics Display ..................................................................................................................................................... 2

Warranty & Service Information
Operation Manual SurgiVet V3395 iii
Warranty & Service Information
Proprietary Notice
Information contained in this document is copyrighted by SurgiVet, Inc. and may not be duplicated in
full or part by any person without prior written approval of SurgiVet. Its purpose is to provide the user
with adequately detailed documentation to efficiently install, operate, maintain and order spare parts
for the device supplied. Every effort has been made to keep the information contained in this
document current and accurate as of the date of publication or revision. However, no guarantee is
given or implied that the document is error free or that it is accurate with regard to any specification.
Limited Warranty
SurgiVet, Inc. warrants each new device to be free from defects in workmanship and materials under
normal use and service for a period of two (2) years from the date of shipment, and reusable Oximetry
sensors for a period of one (1) year from shipment (domestic sales only). SurgiVet’s sole obligation
under this warranty will be to repair or replace, at its option, products that prove to be defective during
the warranty period. The foregoing shall be the sole warranty remedy. Except as set forth herein, seller
makes no warranties, either expressed or implied, including the implied warranties of merchantability
and fitness for a particular purpose. No warranty is provided if the products are modified without the
express written consent of SurgiVet and seller. Neither SurgiVet nor seller shall be liable in any event
for incidental or consequential damage. This warranty is not assignable.
Loaner Device (Domestic Sales Only)
SurgiVet, Inc. will for the period of warranty make available at no charge, loaner devices if, in
SurgiVet's opinion, the repair of the customer's device would require an unreasonable period of time.
Service Support
Repairs of SurgiVet's devices under warranty must be made at authorized repair centers. If the device
needs repair, contact SurgiVet's service department or your local distributor to request a customer
service report number (CSR). When calling, have the device’s model and serial number ready.
NOTE: Shipments received without a CSR number will be returned to sender.
If you need to ship the device, pack the device and its accessories carefully to prevent shipping
damage. All accessories should accompany the device.
SurgiVet, Inc.
Attn: Service
Department
N7 W22025 Johnson
Road, Ste A
Waukesha, WI 53186
Phone: (262) 513-8500
Toll-Free: (888) 745-6562
Fax: (262) 513-9069


Chapter 1: Introduction
Operation Manual SurgiVet V3395 1-1
Chapter 1: Introduction
Intended Use
The SurgiVet V3395 Plus is a low cost portable Pulse Oximetry, respiration, and temperature monitor.
It non-invasively and continuously monitors and displays arterial blood oxygen saturation of arterial
hemoglobin (SpO2), pulse rate, plethysmogram, temperature and respiration. It may be used in the
veterinary hospital or clinical environment. The Oximetry feature works with all SurgiVet Oximetry
sensors, providing SpO2and pulse rate on all patients. The SurgiVet V3395 Plus permits continuous
patient monitoring with adjustable alarm limits for Oximetry, heart rate, and respiration, as well as
visible and audible alarm signals.
Symbol Definitions
SYMBOL DEFINITION
qDefibrillator-proof type CF equipment
Type CF equipment.
2 Attention, consult accompanying documents.
T Class II device
7Refer servicing to qualified service personnel.
G Direct Current
D Output voltage
E Input voltage
IPX1 Drip Proof
F Printer output
B Alarm Silence
0Loudspeaker
2Warnings, Cautions, and Notes
KEYWORD DEFINITION
WARNING Tells you about something that could hurt the patient or hurt the operator.
CAUTION Tells you about something that could damage the monitor.
NOTE Tells you other important information.

Chapter 1: Introduction
1-2 Operation Manual SurgiVet V3395
General Warnings
WARNING: Federal law (USA) restricts the use or sale of this device by, or on the order of, a
licensed veterinarian.
WARNING: Do not use this device in the presence of flammable anesthetics.
WARNING: Do not autoclave, ethylene oxide sterilize, or immerse in liquid. Unplug before
cleaning or disinfecting.
WARNING: ELECTRICAL SHOCK HAZARD when cover is removed. Do not remove covers.
Refer servicing to qualified personnel.
WARNING: Do not use this device in the presence of magnetic resonance imaging (MR or MRI)
equipment.
WARNING: Do not plug the monitor into an outlet controlled by a wall switch.
WARNING: This device must be used in conjunction with clinical signs and symptoms. This
device is only intended to be an adjunct in patient assessment.
WARNING: In the event that earth ground integrity is lost, the performance of this device
and/or other devices nearby may be affected due to excessive RF emissions.
WARNING: When connecting this monitor to any instrument, verify proper operation before
clinical use. Refer to the instrument’s user manual for full instructions. Accessory equipment
connected to the monitor’s data interface must be certified according to the respective IEC
standards, i.e., IEC 950 for data-processing equipment or IEC 601-1 for electromedical
equipment. All combinations of equipment must be in compliance with IEC 601-1-1 systems
requirements. Anyone connecting additional equipment to the signal input port or signal
output port configures a medical system, and, therefore, is responsible that the system
complies with the requirements of the system standard IEC 601-1-1.
WARNING: Where HF (diathermy) is used there is no danger of burning to the patient provided
recommended components are used. Rate meters may be temporarily affected.
WARNING: Equipment is protected against defibrillator discharge. Rate meters and displays
may be temporally affected during defibrillation, but will rapidly recover.
WARNING: Any monitor that has been dropped or damaged should be checked by a qualified
service personnel to insure proper operation prior to use.
WARNING: This monitor will not operate effectively on patients who are experiencing
convulsions or tremors.
Oximetry
WARNING: Use only SpO2sensors supplied with, or specifically intended for use with, this
device.
WARNING:Prolonged use or the patient’s condition may require changing the sensor site
periodically. Change sensor site and check skin integrity, circulatory status, and correct
alignment at least every 4 hours.
WARNING: When attaching SpO2sensors with Microfoam®1tape, do not stretch the tape or
attach the tape too tightly. Tape applied too tightly may cause inaccurate readings and blisters
on the patient's skin (lack of skin respiration, not heat, causes the blisters).
1Microfoam® is a registered trademark of the 3M Company.

Chapter 1: Introduction
Operation Manual SurgiVet V3395 1-3
Cautions
CAUTION: Ensure the device’s AC rating is correct for the AC voltage at your installation site
before using the monitor. The monitor’s AC rating is shown on the external power supply. If the
rating is not correct, do not use the monitor; contact SurgiVet’s service department for help.
CAUTION: Should the device become wet, wipe off all moisture and allow sufficient time for
drying before operating.
CAUTION: This device is intended for use by persons trained in professional health care. The
operator must be thoroughly familiar with the information in this manual before using the
monitor.
CAUTION: Do not allow water or any other liquid to spill onto the monitor. Unplug the external
power supply from the monitor before cleaning or disinfecting the monitor.
CAUTION: Pressing front panel keys with sharp or pointed instruments may permanently
damage the keypad. Press front panel keys only with your finger.
CAUTION: Do not disassemble unit. This unit is not user serviceable. Refer to qualified
service personnel.
CAUTION: Ensure conductive parts including electrodes of the patient cable do not come into
contact with any conductive surfaces or earth parts.
CAUTION: It is the operator’s responsibility to set alarm limits appropriately for each
individual patient.
CAUTION: If the accuracy of any measurement is in question, check the patient’s vital sign(s)
by an alternative method and then check the monitor for proper functioning.
CAUTION: The monitor should be operated from its internal power source (if fitted) if the
integrity of the protective earth conductor is in doubt.
CAUTION: Follow local governing ordinances and recycling instructions regarding disposal
and recycling of device components.
CAUTION: The monitor contains a 1.2 hour lead-acid battery. If the battery fails to hold a
charge or otherwise become inoperable, the battery should be replaced and the old battery
should be disposed of properly. Lead-acid batteries should not be disposed of in normal trash
containers. They should be sent to the proper facilities so that the metals in them may be
reclaimed and/or recycled.
Lead-acid battery: In the US, 1-800-822-8837 will provide information about the proper disposal
of the lead-acid battery. Regulations in Europe vary from country to country. Consult local
authorities for information about proper disposal. SurgiVet, Inc. cannot dispose of monitor
batteries.
General Notes
NOTE: Operation of this device may be adversely affected in the presence of strong EM or RF
sources, such as electrosurgery equipment.
NOTE: Operation of this device may be adversely affected in the presence of computed
tomograph (CT) equipment.
NOTE: The monitor is protected against damage from defibrillator, diathermy, and
electrocautery discharge.
NOTE: All user and patient accessible materials are non-toxic.

Chapter 1: Introduction
1-4 Operation Manual SurgiVet V3395
NOTE: Hazards arising from software errors have been minimized. Hazard analysis was
performed to meet prEN1441: 1994.
NOTE: Each input and output connection of the monitor is electrically isolated. Connection of
this monitor to other equipment will not increase leakage current.
Oximetry
NOTE: Dyes introduced into the bloodstream, such as methylene blue, indocyanine green,
indigo carmine, and fluorescein, may cause an inability to determine accurate SpO2readings.
NOTE: Any condition that restricts blood flow, such as use of a blood pressure cuff or
extremes in systemic vascular resistance, may cause an inability to determine accurate pulse
and SpO2readings.
NOTE: Significant levels of dysfunctional hemoglobins, such as carboxyhemoglobin or
methemoglobin, will affect the accuracy of the SpO2measurement.
NOTE: SpO2measurements may be adversely affected in the presence of high ambient light. If
necessary, shield the sensor area (with a surgical towel, for example).
Respiration & Temperature
NOTE: Use only sensor supplied with, specifically intended for use with this device.

Chapter 2: Description of Measurements, Controls and Features
Operation Manual SurgiVet V3395 2-1
Chapter 2: Description of Measurements, Controls and Features
85 98
101.2 15F
1
2
3
4
5
6
6
FIG. 2-1 FRONT PANEL
General Description
Parameters: The SurgiVet V3395 is Oximetry, temperature, and respiration monitor. Alarm
limits can be set on all monitored parameters.
Oximeter: The monitor continuously measures and displays arterial blood oxygen saturation
(SpO2) and Pulse Rate (HR). Oximetry includes the display of a plethysmogram and pulse
strength bar. The monitor beeps with each pulse beat. The volume of the pulse beep is
adjustable. The pitch of the pulse beep varies with the SpO2value. A variety of veterinary
reusable sensors are available for monitoring patients.
Audio:The monitor uses a multi-frequency speaker for beeps and alarm/alert sounds.
Volumes are adjustable.
Temperature:The monitor continuously measures and displays the temperature. Either
degrees in Celsius or Fahrenheit can be displayed.
Respiration: The monitor continuously measures and displays the respiration rate.
Power: The SurgiVet V3395 operates on power from an external power supply. In addition,
the monitor contains an internal NiMH battery that will allow operation for approximately 2
hours. However, the monitor should remain on the external power supply whenever possible.
Front Panel
LCD Panel
The LCD display provides continuous, real-time updates of waveform or measurement trends,
alarm or alert messages, and temperature and/or respiration values. The display also shows
menus and a pulse strength bar.

Chapter 2: Description of Measurements, Controls and Features
2-2 Operation Manual SurgiVet V3395
Alarm Silence LED
Flashes at ½ Hz during 2 minute alarm silence and remains on steadily for permanent alarm
silence.
Charge LED (Green)
Is on steady while external power is applied. If there is no external power, then this LED is
off.
ALARM LED (Red)
This ALARM indicator flashes during patient alarms.
ALERT LED (Yellow)
This ALERT indicator flashes during a system alarm, but remains on steady if there is no
system alarm and a low battery condition exists.
Keys The front panel keys control the monitor's functions. Dedicated keys are
provided for turning the monitor on and off, silencing alarm and alert tones, selecting
waveforms or trends, and selecting menus.

Chapter 2: Description of Measurements, Controls and Features
Operation Manual SurgiVet V3395 2-3
LCD Display
85 98
101.2 15F
1
2
3
FIGURE 2.2: LCD DISPLAY
Pulse Bar graph The pulse signal strength is displayed here on a 8 segment bar graph.
Message Area Messages for alarms, alerts, and system information are displayed
here on two lines. If more than one message must be displayed on the same line, then they
alternate once per second.
Waveforms, Trends, Menus The waveform is displayed here. A trend graph or a
menu can be displayed in place of the waveform.

Chapter 2: Description of Measurements, Controls and Features
2-4 Operation Manual SurgiVet V3395
Keys
FIGURE 2.3: KEYS
LSTANDBY/ON Pressing Lswitches the monitor between ON (monitoring a
patient) and STANDBY (monitor off, but power is applied if the green indicator is lit.).
aWAVE/TRENDThis key controls the waveform area display. Press this key
to display the plethysmogram or a trend.
Press awhile menus are displayed for a quick menu exit. The waveform or trend
previously displayed will replace the menu.
BALARM SILENCE Pressing Bdisables the audible alarm tone for two
minutes. (TheAlarm Silence LED on the keypad flashes.) Pressing and holding this key for
about three seconds disables the alarm tone indefinitely. The Alarm Silence LED is lit and not
flashing). Pressing this key momentarily cancels either alarm silence condition.
The monitor defaults to two minute alarm silence at power up.
bMENU/ENTER Press this key to display the list of menus. While menus are
displayed, press bto select a menu item or to accept a value that has been adjusted.
ARROWS If a Menu is displayed, press the UP/DOWN ARROWS
()to move among menu items or to adjust the value of a selected item.
If a Trend is displayed, the ARROW ( ) keys cycle between trended parameters. If a
plethysmogram is displayed, the ARROW ( ) keys adjust the display’s contrast.
85 98
101.2 15F
1
23
4
5

Chapter 2: Description of Measurements, Controls and Features
Operation Manual SurgiVet V3395 2-5
Top Panel
12
3
FIGURE 2.4: TOP PANEL
SpO2Connector The oximetry sensor patient cable is connected here.
Respiration Connector The respiration cable is connected here.
Temperature Connector The temperature cable is connected here.
CAUTION: Ensure the device’s AC rating is correct for the AC voltage at your
installation site before using the monitor. The monitor’s AC rating is shown on the
external power supply. If the rating is not correct, do not use the monitor; contact
SurgiVet’s service department for help.
Side Panel
4
FIGURE 2.5: SIDE PANEL
Power Input The external power supply (V3396, 110V, 50/60 Hz AC Charger)
attaches to this connector.


Chapter 3: Setting Up the Monitor
Operation Manual SurgiVet V3395 3-1
Chapter 3: Setting Up the Monitor
Unpacking the Monitor and Checking the Shipment
Carefully remove the monitor and accessories from the shipping carton. Save the packing materials in
case the monitor or accessories must be shipped or stored. Compare the packing list with the
accessories received to make sure the shipment is complete.
Turning Alarm and Alert Tones On and Off
When the monitor is turned on, the alarm and alert tones are silenced for two minutes. The Alarm
Silence LED, on the LCD keypad, flashes during the two minute time-out.
• To silence the alarm and alert tones indefinitely: Press and hold Bfor about three
seconds; the Alarm Silence LED lights steady.
NOTE: To comply with government requirements for patient monitoring, the indefinite
alarm and alert tone silence feature may not be available in monitors shipped to your
country.
• To silence the alarm and alert tones for two minutes: Momentarily press B; the
Alarm Silence indicator flashes. If tones are already silenced, press Btwice (the
first press cancels alarm silenced; the second press silences the alarms for two
minutes).
• To cancel either two minute or indefinite alarm silence and enable alarm and alert
tones: Momentarily press B; the Alarm Silence LED turns off.
Working With Menus
ALARMS
VOLUME
OXIMETER
RESPIRATION
TEMPERATURE
TRENDS
PRINTER
[EXIT]
SPO2 SENSOR!
85 98
101.2 F 15
FIGURE 3.1: MENU

Chapter 3: Setting Up the Monitor
3-2 Operation Manual SurgiVet V3395
ALARMS
Set SpO2High and Low Alarms
Set Pulse Rate High and Low Alarms
Set Respiration Rate High and Low Alarms
VOLUME
Adjust Alarm Volume
Adjust Pulse Volume
OXIMETER
Select Averaging Parameters
Disable/Enable Oximeter Function
RESPIRATION
Sensitivity Setting
Enable/Disable Respiration Function
TEMPERATURE
Units
Enable/Disable Temperature Function
TRENDS
Select Display Time
Select Trend Display Scales for Each Parameter
Clear Trend Memory
PRINTER
Start/Stop Serial Output
Select Data Format
Select Printer Type
FIGURE 3.2: MENU STRUCTURE
The operator chooses various monitor settings through menus that appear in the waveform
area. The bkey and ARROW keys ( ) are used to select menu items and change
settings.
1. To show the Main menu: Press the bkey. Menus replace any waveforms displayed.
2. To select a menu item: Press the ARROW keys ( ) until the selector arrow points
to the desired item, then press b. This will either take you to a new menu or allow the
selected item to be adjusted.
3. To change a setting for a selected menu item: After pressing bas in step 2, the item,
which can be changed, will be highlighted. Use the ARROW keys ( ) to select the
desired setting, then press bto accept that setting.
4. To exit the current menu: Select the [EXIT] menu item
5. To exit all menus quickly and return to the previous monitoring screen: Press a.
NOTE: If menus are shown and you do not press menu keys for 20 seconds, the
waveform display will return, and any current menu selections will be accepted.

Chapter 3: Setting Up the Monitor
Operation Manual SurgiVet V3395 3-3
Working With System-Wide Settings
This section describes working with system-wide settings using the VOLUME menu.
Setup/Volume Menu
The following system-wide settings are viewed and/or adjusted from the Volume menu:
VOLUME
[EXIT]
SPO2 SENSOR!
85 98
101.2 F 15
(volume menu)
_ _ _
PULSE OFF
FIGURE 3.3: SETUP MENU
VOLUME Indicates the Alarm and Pulse volume. Allows the volume to be adjusted in the
range of 1 to 15. Pulse volume can be set to OFF. Alarm volume can not be shut off.
Turning Parameter Monitoring On and Off
Monitoring for the following parameters can be turned on or off:
• Respiration (RR/M)
• Oximeter (Plethysmogram, SpO2%, pulse rate, pulse strength)
• Temperature (°C / °F)
When turned off, displays, indicators, and alarms related to the parameter are disabled. The
parameter occupies a space on the display, but its value is shown as one dash (—).
To turn a parameter’s monitoring on or off, do the following:
1. From the Main menu, select the desired parameter’s menu item: OXIMETER,RESPIRATION,
or TEMPERATURE.
2. In the parameter’s menu, select the monitor on/off item.
3. Press
band use the ARROWS ( ) to adjust the setting.
4. Press
b to accept the selection.
5. Select [EXIT] or press ato exit menus.

Table of contents
Other SurgiVet Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Cianna Medical
Cianna Medical SCOUT Check quick start guide

bort medical
bort medical 103 360 quick guide

Global Halotherapy Solutions
Global Halotherapy Solutions HaloFX Mobile manual

Lowenstein Medical
Lowenstein Medical JOYCEone Instructions for use

Dermlite
Dermlite GL instructions

Aquolab
Aquolab Aquolab Professional user manual