While in the “TEST”
mode, if “RETURN
TO SET UP” light
illuminates, go back
to “SET UP” to
ensure the “VACUUM
READY” light is not
flashing. If “VACUUM
READY” light is flashing, go to the
*SECTION TEST section noted to the
left. If the “VACUUM READY”
light illuminates, press the “TEST”
button.
Troubleshooting
If system fails to power
up, check to ensure that
both ends of the ATEC
®
power cord are connected
first to the console and
then into the wall outlet.
Next, check the circuit
breaker.
In “SET UP,” if the
“VACUUM READY”
light is flashing indicating
an air leak is present,
proceed to the *SECTION
TEST section noted below
to isolate the leak
source/point. 2
1
3
4
LIT339:RevB
Remove tissue specimen from the tissue
filter and prepare for pathology.
Finally, remove the ATEC®handpiece
from the patient/breast.
Once the procedure is
complete, power off
the console and
disconnect the red and
black handpiece lines,
remove the silicone
tubing section of the
handpiece saline line
and the vacuum tubing
from the canister lid.
Remove the saline bag
from the hook.
Disconnect the ATEC®Vacuum Line
Assembly from the canister lid.
Dispose of all consumables per your
facility’s protocol. Consumables include:
v ATEC®handpiece and tubing
v Saline bag
v ATEC® suction canister and lid
v Stereotactic needle guide
Choose “TEST”
and system will
run through one
diagnostic cycle.
Once the “TEST”
cycle is complete,
console will automatically
return to “BIOPSY”
and is ready to begin
acquiring tissue samples.
(Biopsy is initiated by
depressing the ATEC®
Footswitch.)
Once the biopsy is
complete, select
“LAVAGE”
and saline will
be introduced and
aspirated to lavage
the biopsy cavity.
151413
181716
321
ATEC® Set Up and Use (continued)
Disposing of the ATEC® Handpiece and Clean Up
These steps are not a substitute for the ATEC®system Instructions for Use, but are intended to serve
only as a quick reference in the set up and use of the ATEC®Breast Biopsy and Excision System.
The system also has a “MANUAL
ASPIRATION” option that evacuates/vacuums
(w/o saline) the biopsy cavity when
depressing the footswitch.
When powering on or
while in the “TEST”
mode, if “RETEST HP”
light is flashing,
press “TEST” again.
If this does not correct
the problem, replace the
ATEC
®
handpiece and
go back through the set up procedure. Do
not dispose of the failed ATEC
®
handpiece.
You must keep the ATEC
®
handpiece that
failed and return it to Suros for evaluation.
Suros Customer Support can assist you with
product return procedures. If the second
ATEC
®
handpiece fails, return both ATEC
®
handpieces for evaluation and call Suros
Customer Support immediately.
*SECTION TEST: The “SECTION TEST” refers to the exercise of
starting at the vacuum source and closing off openings systematically
from the console to the ATEC
®
handpiece to find the source of a
vacuum leak. Your Suros Clinical Education Specialist will train
you on this procedure at the time of in-service.
WIPE TOP OF CONSOLE WITH DISINFECTANT
Questions or Comments: If you are a Suros customer in the U.S.
or Canada, please call Suros Customer Support at 1-877-887-8767.
If you are an international customer, please contact your distributor
directly. Thank you for putting your trust and confidence in Suros
Surgical Systems, Inc.