Model 7709 6 × 8 Matrix Module Instructions for use with DAQ6510
077147500 / June 2018
Installation
Before operating an instrument with a switching module, verify that the switching module is
properly installed and the mounting screws are tightly fastened. If the mounting screws are
not properly connected, an electrical shock hazard may be present.
If you are installing two switching modules, it is easier to install one switching module into slot 2 first, then install
the second switching module into slot 1.
If you have a Keithley Instruments Model 2700, 2701, or 2750 instrument, you can use your existing
switching module in the DAQ6510. Follow the instructions in your original equipment documentation
to remove the module from the instrument, then use the following instructions to install it in the
DAQ6510. You do not need to remove wiring to the module.
For inexperienced users, it is recommended that you do not connect a device under test (DUT) and
external circuitry to the switching module. This allows you to exercise close and open operations
without the dangers associated with live test circuits. You can also set up pseudocards to experiment
with switching. Refer to "Pseudocards" in the Model DAQ6510 Data Acquisition and Multimeter
System Reference Manual for information on setting up pseudocards.
To prevent electric shock that could result in injury or death, never handle a switching
module that has power applied to it. Before installing or removing a switching module, make
sure the DAQ6510 is turned off and disconnected from line power. If the switching module is
connected to a DUT, make sure power is removed from all external circuitry.
If a card slot is unused, you must install slot covers to prevent personal contact with high
voltage circuits. Failure to install slot covers could result in personal exposure to hazardous
voltages, which could cause personal injury or death if contacted.
Before installing or removing a switching module, make sure the DAQ6510 power is turned
off and disconnected from line power. Failure to comply may result in incorrect operation and
loss of data in the memory.
Required equipment:
▪Medium flat-blade screwdriver
▪Medium Phillips screwdriver