Tektronix 284 User manual

INSTRUCTION
MIAN
UAL
serial
Number
226
2
BI?
Tektronix,
Inc.
S.W.
Millikan
Way
@
P.
©.
Box
500
@
Beaverton,
Oregon
97005
@
Phone
644-0161
@
Cables:
Tektronix
070-0754-00
1067

K4XL's
“@
BAMA
This
manual
is
provided
FREE
OF
CHARGE
from
the
“BoatAnchor
Manual
Archive”
as
a
service
to
the
Boatanchor
community.
It
was
uploaded
by
someone
who
wanted
to
help
you
repair
and
maintain
your
equipment.
If
you
paid
anyone
other
than
BAMA
for
this
manual,
you
paid
someone
who
is
making
a
profit
from
the
free
labor
of
others
without
asking
their
permission.
You
may
pass
on
copies
of
this
manual
to
anyone
who
needs
it.
But
do
it
without
charge.
Thousands
of
files
are
available
without
charge
from
BAMA.
Visit
us
at
http://bama.sbc.edu

or
service.
-
a
oe
ec
cba
.
sean
nin
ee
.
oe
-
Specifications
and
price
change
privi-
_leges
‘feserved,
Type
284
.
®

cia
ptueons
pee
Type
284

TYPE
284
PULSE
GENERATOR
SOUSARE
WAVE
AMPLITUDE
PERIOD
BATE:
SOS)
SQUARE
Wave
SINE
WAVE
100
av
WOns
TOMS
ode
Ce
elllaldiesd
INFO
808”
Satta
[ie
wrioes
OR
QUTPUT
eae
~
SINE
WAVE
OUTPUT
TeS
POpey
EF
BOOMV
a
Py
BOR
=
|
TERTRONIX,
INC,
PORTLAND,
OREGON,
U.S.
A
TEKTRONIX
«
Fig.
1-1.
Type
284
Pulse
Generator.
Type
284
@i

Type
284
SECTION
|!
SPECIFICATION
Change
information,
if
any,
affecting
this
section
will
be
found
at
the
rear
of
the
manual.
Introduction
The
Tektronix
Type
284
Pulse
Generator
provides
special-
ized
signals
that
are
defined
sufficiently
fo
allow
them
to
be
used
to
verify
the
performance
of
fast
general
purpose
and
sampling
oscilloscopes.
Specifically,
the
signals
allow
a
check
of
the
vertical
amplifier
risetime,
deflection
factor
accuracy,
transient
response
and
pulse
flatness
deviation
(aberrations);
plus
checking
upon
the
accuracy
of
horizontal
sweep
rates.
The
fast
pulse
may
be
used
as
a
signal
source
for
testing
the
risetime
of
transistors
and
amplifiers,
or
for
Time
Domain
Re-
flectometry.
Each
signal
provides
a
trigger
signal
for
externally
trig-
gering
an
oscilloscope
time
base
generator.
The
fast
pulser
also
generates
a
trigger
in
advance
of
the
fast
step,
provid-
ing
the
pretrigger
signal
necessary
for
some
sampling
oscil-
loscopes.
Regulated
DC
power
supplies
assure
stable
operation
over
a
wide
range
of
line
voltage
and
frequency.
Circuits
are
temperature
compensated
to
maintain
both
amplitude
and
frequency
accuracies
over
a
wide
temperature
range.
The
electrical
characteristics
that
follow
are
divided
into
two
categories.
Information
listed
in
the
Performance
Requirement
column
applies
directly
to
the
instrument
per-
formance
and
is
a
commitment
by
Tektronix,
Inc.,
to
the
customer.
Information
listed
in
the
Operational
Information
column
is
for
clarification
and
operator
convenience
and
is
not
to
be
used
to
specify
any
performance
limits
for
the
in-
strument.
Characteristics
listed
in
the
Performance
Require-
ment
column
are
checked
by
the
Performance
Check
pro-
cedure,
Section
5,
of
this
manual.
Any
instrument
not
meeting
these
performance
limits
can
be
brought
within
specifications
by
performing
the
Calibration
Procedure,
Section
6
of
this
manual.
The
performance
requirements
apply
over
an
ambient
temperature
range
of
0°
C
to
+50°C,
at
altitudes
up
to
5,000
feet,
with
upper
temperature
limit
decreased
by
1°
C
per
1000
feet
from
5000
feet
to
15,000
feet,
after
a
warm
up
period
of
five
minutes,
providing
the
convection
cooling
holes
(bottom
and
both
sides)
are
not
obstructed.
The
unit
may
be
stored
in
temperatures
from
—40°
C
to
+65°
C.
ELECTRICAL
CHARACTERISTICS
Characteristic
|
Operational
Information
Performance
Requirement
|
SINE
WAVE
OR
SQUARE
WAVE
OUTPUT
Calibrated
PERIOD
Signals
Square
Wave
10
ws,
1
ws
and
100
ns
For
checking
sweep
rates.
Sine
Wave
10
ns
and
J
ns
Signal
Amplitude
Into
500
Square
Wave
1V,
100
mV
and
10
mV
For
checking
deflection
factors.
Sine
Wave
100
mV
For
checking
horizontal
timing
only.
Accuracies
SIGNAL
SIGNAL
Square
Wave
Sine
Wave
Square
Wave
PERIOD
10
ss
1
us
100
ns
10ns
|
Ins
10us
|
lps
|
100ns
Timing
+0.5%
+0.5%
+0.05%
|
1%
|
1%
|
Amplitude
accuracies
in
this
col-
Amplitude
1V
+0.5%
|
+05%
|
42%
umn
are
based
upon
DC
testing
of
attenuators.
100
mV
+20%
|
20%
|
+1% +1%
+2.5%)
10
mV
+1.5%
+15%
+3%7
Square
Wave
Duty
Factor
48%
to
52%
Trigger
Output
Signa!
Amplitude,
into
500
Square
Wave
+200
mV,
20%
Sine
Wave
200
mV
peak
to
peak,
+40%
1Measured
20ns
after
pulse
reaches
100%
on
Tektronix
Type
383
or
453
Sampling
Probe
Unit.
100%
on
Tektronix
Type
451
Dual
Trace
Sampling
Unit.
®
Measured
30
ns
after
pulse
reaches
1-1

Specifications—Type
284
ELECTRICAL
CHARACTERISTICS
(cont)
PULSE
GENERATOR
Characteristic
Performance
Requirement
Operational
Information
Repetition
Rate
50
kHz,
+10%
20
ws
period
Risetime,
10%
to
90%
|
<70
ps
Amplitude
into
500
>
+200
mV
Step
beginning
adjusted
to
zero
volts
during
calibration.
Pulse
Flatness
Deviation
During
first
2s
after
pulse
reaches
100%
amplitude:
<3%
p-p
total
deviation
over
a
range
from
+3%
to
—3%.
After
2ns:
<2%
p-p
total
deviation
over
a
range
from
+2%
to
—2%.
Aberrations
Internal
Pulser
Assembly
Signal
Reflections
due
to
changes
of
im-
pedance
<+5%,
measured
by
Tektronix
Type
182
TDR
unit.
L
BS,
+50%,
—0%
Pulse
Duration,
50%
to
50%
TRIGGER
OUTPUT
SIGNAL
WHEN
USING
PULSE
GENERATOR
Amplitude
+200
mV,
+20%
Risetime,
10%
to
90%
<3
ns
Pulse
Duration,
50%
to
50%
>10
ns
Trigger
Pulse
Occurrence
In
Ad-
vance
of
PULSE
OUTPUT
Signal
5
ns
or
75
ns,
5
ns?
Selectable
by
LEAD
TIME
switch
POWER
LINE
REQUIREMENTS
Line
Voltage
Ranges
115
Volt
Line
90
VAC
to
136
VAC
230
Volt
Line
180
VAC
to
272
VAC
Frequency
Range
48
Hz
to
440
Hz
Maximum
Power
Consumption
Approximately
6.5
Watts
at
115
VAC
60
Hz
Fuse
Data
Use
Fuses
listed
in
Parts
List
MECHANICAL
CHARACTERISTICS
Dimensions
Height
63/,
inches
Width
41/,
inches
Length
15
inches
Approximate
dimensions
include
knobs
and
connectors.
Construction—Aluminum
alloy
chassis
with
epoxy
laminated
circuit
boards.
Front
panel
is
anodized
aluminum.
Accessories—An
illustrated
list
of
the
accessories
supplied
with
the
Type
284
is
at
the
end
of
the
Mechanical
Parts
List
pullout
pages.
Serial
numbers
BO10100-B020235;
5
ns
or
50ns,
+5
ns.
1

Type
284
SECTION
2
OPERATING
INSTRUCTIONS
Change
information,
if
any,
affecting
this
section
will
be
found
at
the
rear
of
the
manual.
General
This
section
of
the
manual
provides
the
basic
information
required
for
operation
of
the
Type
284.
Instructions
include
function
of
the
front
panel
controls
and
connectors,
first
time
operation,
general
operating
infromation,
and
basic
appli-
cations.
AC
Power
Considerations
The
Type
284
can
be
operated
from
either
a
115
or
a
230-volt
line,
at
frequencies
from
48
to
440Hz.
The
AC
line
voltage
selector
panel
and
fuse
holder
is
located
on
the
rear
panel
of
the
Type
284.
CAUTION
The
Type
284
should
not
be
operated
with
the
volt-
age
selector
in
an
incorrect
position
for
the
nominal
line
voltage
applied.
Operation
of
the
instrument
in
a
wrong
voltage
range
will
either
provide
incor-
rect
operation
or
damage
the
instrument.
The
AC
line
voltage
selector
panel
on
the
rear
of
the
Type
284
indicates
nominal
line
voltage
selected.
See
Fig.
2-2.
Check
to
see
that
the
plastic
indicating
tab
of
the
line
se-
lector
is
protruding
through
the
selector
panel,
indicating
that
its
position
is
proper
for
applied
nominal
line
voltage
(115
or
230
AC
Volts).
The
selector
assembly
cover
can
be
removed
to
change
the
nominal
line
from
115
Volts
to
230
Volts
or
vice
versa,
or
to
replace
the
fuses.
Use
the
following
procedure:
1.
Disconnect
the
instrument
from
the
power
source.
2.
Loosen
the
two
captive
screws
which
hold
the
cover
onto
the
voltage
selector
assembly,
then
pull
to
remove
the
cover.
3.
To
convert
from
115-volt
to
230-volt
nominal
line
volt-
age
or
vice
versa,
pull
out
the
Voltage
line
selector
switch
bar;
turn
it
180°
and
plug
it
back
into
the
remaining
holes.
Change
the
line
cord
to
fit
the
power
source
receptacle
or
use
a
line
cord
adapter.
4,
It
is
not
necessary
to
change
fuses
if
the
range
is
changed;
however,
in
the
230
volt
range
both
fuses
are
in
use.
Also
in
checking
the
fuses,
the
upper
fuse
(1/10
A)
for
the
115
volt
range
has
a
nominal
resistance
of
approximate-
ly
80
ohms,
and
the
lower
fuse
(1/16
A)
for
the
230
volt
range
has
a
nominal
resistance
of
approximately
150
ohms.
5.
Re-install
the
selector
assembly
cover
and
tighten
the
two
captive
screws.
6.
Before
applying
power
to
the
instrument,
check
that
the
indicating
tab
of
the
line
selector
is
protruding
through
@
the
selector
panel
indicating
that
its
position
is
proper
for
the
applied
nominal
line
voltage
range
(115
or
230).
In
the
115
volt
nominal
range,
the
regulating
range
is
from
90
to
136
volts.
In
the
230
volt
nominal
range
the
regulating
range
is
from
180
to
272
volts.
Handle
and
Stand
The
bail-type
handle
of
the
Type
284
can
be
pulled
out
for
convenient
carrying
of
the
instrument.
When
not
in
use,
the
handle
folds
out
of
the
way
into
the
trim
of
the
instru-
ment
cabinet.
See
Fig.
2-1.
The
bail-type
stands
are
mounted
beneath
the
cabinet.
The
stands permit
the
Type
284
to
be
tilted
for
convenient
operation
as
shown
in
Fig.
2-1.
The
instrument
may
also
be
set
on
the
rear
feet
either
for
operation
or
storage.
Fig.
2-1.
Type
284
showing
handle
and
stand.
CONTROLS
AND
CONNECTORS
A
brief
description
of
the
function
and
operation
of
the
front
and
rear
panel
controls
and
connectors
follows.
Fig.
2-2
shows
the
front
and
rear
panels
of
the
instrument.
More
detailed
operation
is
given
in
this
section
under
General
Operating
Information.
2-1

Operating
Instructions—Type
284
(A)
Front
Panel
TYPE
284
PULSE
GENERATOR
SQUARE
WAVE
AMPLITUDE
PERIOD
ARTO
Sa
SQUARE
WAVE
SINE
WAVE
WOO
nv
2ANNMV
WDOns
100s
—
NTO
SON
ns
>
SQUARE
WAVE.
_
PULSE
i
SINE“
WAVE
-OUTPUT
font
500:
oo:
OUTPUT
TB
Wipe
=
+AODAN
(B})
Rear
Panel
bes
FREQ:
48-TO
440Hz
LING
FUSE
>
-
-
(LRea
SOW
}
80
re
138
{BU-to
272
30
_
PABA
SLOW
AMP:
(MAK)
0:05
“oe
AAT.
TBV-60H2)
Nominal
115
AC
Line
V
position
Nominal
230
AC
Line
V
position
Fig.
2-2.
Front
and
rear-panel
controls
and
connectors.
@2

Front
Panel
SQUARE
WAVE
AMPLITUDE
PERIOD
MODE
PULSE
OUTPUT
TD
BIAS
LEAD
TIME
TRIGGER
OUTPUT
SQUARE
WAVE
OR
SINE
WAVE
OUTPUT
POWER
Rear
Panel
Line
voltage
selector
Power
Connector
Selects
three
amplitudes
of
square
wave
output
signal
voltages:
10mV,
100
mV,
and
1
V (into
a
50Q
load).
Selects
the
time
periods
of
the
SQUARE
WAVE
OR
SINE
WAVE
outputs.
The
SQUARE
WAVE
output
time
periods
are
10
ys,
1
ys
and
100ns
with
amplitude
selected
by
the
SQUARE
WAVE
AMPLI-
TUDE
switch.
The
SINE
WAVE
time
peri-
ods
are
10ns
and
Ins
with
approximate
amplitudes
of
100
mV
(into
a
50
Q
load).
Selects
either
the
PULSE
OUTPUT
or
the
SQUARE
WAVE
OR
SINE
WAVE
OUT-
PUT
to
their
respective
connectors.
Output
connector
for
the
PULSE
OUTPUT
(rise
time
<
70
ps,
amplitude
>
200
mV).
Adjusts
the
bias
of
the
tunnel
diode
af-
fecting
the
PULSE
OUTPUT.
Selects
either
5ns
or
75ns
of
LEAD
TIME
of
the
TRIGGER
OUTPUT
signal
with
ref-
erence
to
the
PULSE
OUTPUT
signal.
Output
connector
for
the
trigger
signal
pickoff,
a
portion
of
the
signal
from
either
the
PULSE
OUTPUT
or
the
SQUARE
WAVE
OR
SINE
WAVE
OUTPUT
depending
upon
the
position
of
the
MODE
switch.
Output
connector
for
the
SQUARE
WAVE
OR
SINE
WAVE
OUTPUT
as
selected
by
the
SQUARE
WAVE
AMPLITUDE
and
PERI-
OD
switches.
Light:
Indicates
when
POWER
switch
is
ON
and
the
instrument
is
connected
to
a
power
source.
Switch:
Applies
power
to
the
instrument.
Switching
assembly
to
select
the
nominal
operating
voltage
(115V
or
230V).
This
assembly
also
includes
the
line
fuses.
Input
connector
for
line
power.
FIRST-TIME
OPERATION
The
following
steps
will
demonstrate
the
use
of
the
con-
trols
and
connectors
of
the
Type
284.
It
is
recommended
that
this
procedure
be
followed
for
familiarization
with
this
instrument.
In
this
procedure
a
Tektronix
Type
561A
Oscil-
loscope
is
used
with
a
Type
3S]
Dual-Trace
Sampling
Unit
and
a
Type
3T2
Sampling
Sweep
Unit.
If
other
equipment
is
used
the
control
settings
and
the
procedure
will
be
dif-
ferent.
If
you
are
not
familiar
with
the
operation
of
the
oscilloscope
with
its
sampling
plug-in
units,
read
the
First
Time
Operation
portions
of
the
manuals
for
these
instruments
before
proceeding.
@1
Operating
Instructions—Type
284
Setup
information
1.
Set
controls
as
follows:
Type
284
SQUARE
WAVE
AMPLITUDE
PERIOD
MODE
LEAD
TIME
Type
3581
Display
Mode
switch
Smooth-Normal
switch
A
Position
B
Position
DC
Offset
+1
V,
Channel
A
DC
Offset
-E1
V,
Channel
B
mVolts/Div,
Channel
A
mVolts/Div,
Channel
B
Variable
mVolts/Div
(both
channels)
Invert-Norm
switch,
Channel
A
Invert-Norm
switch,
Channel
B
Internal
Trigger
Sampling
Mode
Type
372
Time
Position
Time
Position
Fine
Horiz
Position
Samples/Div
Range
Start
Point
Display
Mag
Time
Magnifier
Time
Magnifier
Variable
Display
Mode
Trigger
Sensitivity
Recovery
Time
Polarity
Source
Samples/Div
(an
Internal
Switch)
1.0
V
10
ps
SQUARE
WAVE
OR
SINE
WAVE
OUTPUT
(to
right)
5ns
Chan
A
Normal
Midrange
Midrange
Midrange
(5
turns
from
one
end)
Midrange
(5
turns
from
one
end)
200
200
Cal
Norm
Norm
Off
Triggered
Fully
Clockwise
Fully
Clockwise
Midrange
Midrange
10 us
With
Trigger
X1
x1
Cal
Normal
Fully
Clockwise
Midrange
Ext
100
(up
position)
2.
Connect
the
Type
284
to
a
power
source
that
meets
the
voltage
and
frequency
requirements
of
the
instrument.
Check
to
see
that
the
plastic
indicating
tab
on
the
line
se-
lector
panel
on
the
rear
of
the
instrument
is
protruding
through
the
selector
panel
indicating
that
its
position
is
2-3

Operating
Instructions—Type
284
proper
for
the
applied
nominal
line
voltage
(115
V
or
230
V).
See
AC
Power
Considerations
in
this
section
for
instructions
to
change
the
nominal
line
voltage
if
necessary.
3.
Set
the
Power
switch
to
ON
on
the
Type
284
and
also
on
the
Test
Oscilloscope,
and
allow
five
minutes
warm
up
before
proceeding.
Square
wave
output
4.
Connect
the
Type
284
SQUARE
WAVE
OUTPUT
signal
through
a
50
coaxial
cable
with
GR
type
connectors
to
the
A
input
Connector
on
the
Type
3S1.
Connect
the
TRIG-
GER
OUTPUT
signal
through
a
509
BNC
type
coaxial
cable
to
the
50Q
Trigger
Input
connector
on
the
Type
3712.
5.
Turn
the
Trigger
Sensitivity
fully
counterclockwise,
then
slowly
clockwise
until
a
stable
display
is
obtained.
See
Fig.
2-3A.
Normal
slight
changes
of
the
Samples/Div
con-
trol
on
the
Type
312,
the
A
Position
control
on
the
Type
3S1
may
be
necessary
to
display
this
waveform.
Fig.
2-3.
Typical
SQUARE
WAVE
OUTPUT
display
(1
Volt
into
502
load)
for
two
time
periods.
2-4
6.
Change
the
Type
284
PERIOD
switch
to
1
us
and
change
the
Type
3T2
Range
switch
to
1
us
(Time/Div
100
ns).
Ob-
serve
a
similar
display
to
that
in
preceeding
step.
7.
Change
the
Type
284
PERIOD
switch
to
100ns
and
change
the
Type
312
Range
switch
to
100ns
(Time/Div
10
ns).
Observe
a
display
similar
to
Fig.
2-3B.
8.
Change
the
Type
284
SQUARE
WAVE
AMPLITUDE
switch
to
100mV
and
change
the
Type
3S1
Channel
A
mVolts/Div
switch
to
50.
Observe
a 2
division
display
sim-
ilar
to
the
5
division
display
in
Fig.
2-3B.
9.
Change
the
Type
284
SQUARE
WAVE
AMPLITUDE
switch
to
10
mV
and
change
the
Type
3S1
Channel
A
mVolts/
Div
switch
to
5.
Observe
2
divisions
of
display
with
the
nor-
mal
noise
of
the
Type
381
showing
up
at
this
sensitivity.
Sine
Wave
Output
10.
Move
the
trigger
signal
coaxial cable
on
the
Type
312
from
the
50Q
input
connector
to
the
1
MQ/UHF
Sync
connector.
Set
the
Type
284
PERIOD
switch
to
10
ns
SINE
WAVE
and
set
the
Channel
A
mVolts/Div
switch
on
the
Type
3S1
to
50.
Slight
change
in
the
Trig.
Sensitivity
control
on
the
Type
312,
and
in
the
A
Position
contro]
on
the
Type
3S]
may
be
necessary
to
obtain
a
display
similar
to
Fig.
2-4A.
11.
Set
the
PERIOD
switch
to
1
ns
SINE
WAVE
on
the
Type
284
and
set
the
Time
Magnifier
switch
on
the
Type
312
to
X10
(Time/Div
1
ns).
Observe
the
1
ns
period
sine
wave
similar
to
Fig.
2-4B.
Slight
change
in
the
Trig.
Sensitivity
control,
and
the
Recovery
Time
control
on
the
Type
312
may
be
necessary
to
obtain
a
stable
display.
Pulse
Output
12.
Move
the
signal
coaxial cable
on
the
Type
284
to
the
PULSE
OUTPUT
connector,
and
set
the
MODE
switch
to
the
PULSE
OUTPUT
position
(to
the
left).
Set the
Time
Mag-
nifier
switch
on
the
Type
312
to
X1
(Time/Div
10
ns),
and
turn
the
Trig.
Sensitivity
switch
fully
counterclockwise
then
clockwise
for
a
stable
trace.
Change
the
LEAD
TIME
switch
to
75ns
position
and
note
the
change
in
delay
time
of
ap-
proximately
70
ns
as
shown
in
Fig.
2-5B.
NOTE
Since
the
tunnel
diode
pulse
generator
is
an
inte-
gral
part
of
the
PULSE
OUTPUT
connector,
mechan-
ical
strain
on
the
PULSE
OUTPUT
connector
will
cause
slight
mechanical
changes
in
the
tunnel
di-
ode
assembly.
Slight
adjustment
of
the
TD
BIAS
control,
a
screwdriver
adjustment
on
the
front
panel,
will
correct
the
display.
For
the
TD
BIAS
adjustment
procedure
see
the
General
Operating
Instructions.
CONTROL
SETUP
CHART
Fig.
2-6
is
a
control
setup
chart
for
the
front
and
rear
panels
of
the
Type
284.
This
figure
may
be
reproduced
and
used
as
a
test
setup
record
for
special
applications
or
procedures.
It
may
also
serve
as
a
training
aid
to
facilitate
control
operation.

Fig.
2-4.
Typical
SINE
WAVE
OUTPUT
display
showing
(A)
10
ns
and
(B)
1ns
time
period.
GENERAL
OPERATING
INFORMATION
Mode
Switch
The
MODE
switch
selects
one
output
with
its
trigger
signal.
The
MODE
switch
always
turns
the
power
off
on
one
output
which
prevents
both
trigger
signals
from
appearing
at
the
TRIGGER
OUTPUT
at
the
same
time.
Square
Wave
Output
Tne
SQUARE
WAVE
OUTPUT
amplitude
is
selected
by
the
SQUARE
WAVE
AMPLITUDE
switch,
and
the
time
period
is
selected
by
the
PERIOD
switch
on
three
of
its
positions.
Note
that
the
other
two
positions
of
the
PERIOD
switch
select
the
SINE
WAVE
time
period.
In
order
to
maintain
the
output
selected
by
the
SQUARE
WAVE
AMPLITUDE
switch,
an
output
load
of
509
is
required.
For
further
loads
use
the
following
formula
to
calculate
the
output
voltage:
Rioad
ut
Rioad
+
50
x
(
Operating
Instructions—Type
284
Pare
a
a
Fig.
2-5.
Typical
PULSE
OUTPUT
display,
showing
a
double
ex-
posure
of
two
positions
of
LEAD
TIME
switch.
(E
is
the
Voltage
selected
by
the
SQUARE
WAVE
AMPLI-
TUDE
switch)
Fig.
2-7
shows
the
time
period
and
amplitude
measurement
points.
In
this
display
the
output
of
the
Type
284
is
properly
terminated
by
being
coupled
to
the
50Q
input
of
a
sam-
pling
oscilloscope.
Disregard
the
overshoot
at
the
top
corner
of
the
square
wave
in
making
amplitude
measure-
ments.
The
time
period
measurement
(here
a
period
of
100
ns)
is
the
time
of
one
cycle
of
the
square
wave.
The
square
wave
output
desired
can
be
selected
for
am-
plitude
from
1V,
100mV
or
10
mV
and
for
the
time
period
from
10
ps,
1
us
or
100
ns.
Sine
Wave
Output
The
SINE
WAVE
OUTPUT
is
selected
by
the
PERIOD
switch.
The
period
switch
selects
either
a
10ns
or
a
Ins
sine
wave
time
period.
The
time
period
is
the
time
of
one
cycle
of
the
sine
wave.
The
amplitude
is
approximately
100
mV
when
it
is
connected
into
a
50Q
load.
2-5

Operating
Instructions—Type
284
(—
)
TYPE
284
PULSE
GENERATOR
SQUARE
WAVE
AMPLITUDE
PERIOD
UNTO
S081
SQUARE
WAVE
SINE
WAVE
100mV
100ns
10ns
=
100m
INTO
son
10mV
Py
10us
MoprE
SQUARE
WAVE
PULSE
cr
sarees
ans
OR
OUTPUT
ae
SINE
WAVE
OUTPUT
7,
S
70ps,
=
+200mV
Bias
(cc
=)
2)
©
(“=r
050
TRIGGER
OUTPUT
LEAD
TIME
Sas
(_L
js.
SERIAL
=
500
TEKTRONIX,
INC.
SY
PORTLAND,
OREGON,
U.S.
A.
q
®
(A)
FRONT
PANEL
eb)
@
LINE
FUSE
[ex
stow
|p
230V
FUSE
[izza
stow
|p
FREQ
45
TO
440Hz
90
To
136
()
\}
180
10
272
230
G
NS
Ya
WATTS
(Max)
6.5
AMP
=
(MAX)
0.05
(AT
T15V-60Hz)
©
386-128
())
(B)
REAR
PANEL
2-6
Fig.
2-6.
Control
setup
chart.
@2

Mm
200
mV/Div
Fig.
2-7.
Typical
display
showing
amplitude
and
period
measure-
ments.
A
1
Volt
amplitude,
100
ns
time
period
pulse
from
the
Type
284
into
502
input
of
Sampling
oscilloscope.
Pulse
Output
The
Pulse
Output
power
is
turned
on
by
setting
the
Mode
switch
to
its
left
position.
The
pulse
is
generated
by
a
tun-
nel
diode
in
a
coaxial
environment
and
connected
directly
to
the
PULSE
OUTPUT
connector.
A
screwdriver
adjustment
marked
TD
BIAS
(tunnel
diode
bias)
on
the
front
panel
is
used
to
adjust
the
optimum
voltage
of
the
tunnel
diode.
Mechanical
strain
on
the
front
panel
connector
may
change
the
tunnel
diode
environment
and
a
slight
adjustment
of
the
TD
BIAS
may
be
necessary.
To
obtain
the
power
adjustment
of
the
TD
BIAS,
it
is
necessary
to
observe
the
display
on
a
sampling
oscilloscope.
Following
is
a
setup
procedure
us-
ing
a
Tektronix
Type
561A
with
a
Type
381
Dual-Trace
Sampling
Unit
and
a
Type
3T2
Sampling
Sweep
Unit.
If
other
equipment
is
used,
the
control
settings
and
the
pro-
cedure
will
be
different.
NOTE
An
internal
TD
Bias
adjustment
(R133)
is
provided
for
coarse
adjustment
of
tunnel
diode
bias.
If
the
tunnel
diode
is
replaced
or
if
the
front
panel
con-
trol
doesn't
provide
a
satisfactory
setting
adjust
the
TD
Bias
(internal)
in
the
manner
given
in
the
Cali-
bration
section
of
this
manual.
1.
Set
the
controls
as
follows:
Type
284
MODE
switch
Pulse
Output
(to
(left)
LEAD
TIME
5
ns
POWER
ON
Type
351
Display
Mode
switch
Chan
A
Smooth-Normal
switch
Normal
A
Position
Midrange
DC
Offset
+1
V,
Midrange
(5
turns
from
Channel
A
one
end)
Operating
Instructions—Type
284
mVolts/Div,
Channel
A
50
Variable
mVolts/Div
Cal
(Channel
A)
Invert-Norm
switch,
Norm
Channel
A
Sampling
Mode
switch
Triggered
Type
312
Time
Position
Time
Position
Fine
fully
clockwise
fully
clockwise
Horiz
Position
Midrange
Samples/Div
Midrange
Range
100
ns
Start
Point
with
Trigger
Display
Mag
Xx
Time
Magnifier
x1
Time
Magnifier
Variable
Cal
Display
Mode
Normal
Trigger
Sensitivity
Fully
clockwise
Polarity
+
Source
Ext
Samples/Div
(an
internal
100
(up
position)
switch}
2.
Connect
the
Type
284
PULSE
OUTPUT
signal
through
a
50
coaxial
cable
with
GR
type
connectors
to
the
A
input
connector
on
the
Type
3S1.
Connect
the
TRIGGER
OUTPUT
signal
through
a
500 BNC
type
coaxial
cable
to
the
50
Trigger
Input
connector
on
the
Type
372.
3.
If
the
TD
BIAS
is
at
optimum
adjustment
you
will
ob-
serve
an
untriggered
display
of
about
4
divisions
or
about
200
mV
peak
to
peak
as
in
Fig.
2-8A,
and
upon
turning
the
Trigger
Sensitivity
control
almost
fully
counterclockwise,
observe
a
clean
triggered
step
display
similar
to
Fig.
2-8B.
Normal
slight
changes
to
the’
Type
3T2
Samples/Div
control
and
the
Type
3S1
A
Position
control
may
be
necessary
to
display
this
waveform.
4.
If
upon
turning
the
Trigger
Sensitivity
control
almost
fully
counterclockwise
a
trace
display
appears
showing
the
position
of
the
upper
portion
of
the
pulse,
this
indicates
that
the
TD
BIAS
control
is
clockwise
from
the
optimum
adjust-
ment.
See
Fig.
2-8C
(upper
trace).
Slight
counterclockwise
rotation
of
the
TD
BIAS
control
will
be
necessary
to
obtain
a
display
similar
to
Fig.
2-8B.
5.
Fig.
2-8C
(lower
trace)
display
indicates
that
the
TD
BIAS
control
is
counterclockwise
from
the
optimum
adjust-
ment.
If
the
display
as
in
Fig.
2-8B
cannot
be
attained
with
clockwise
rotation
of
TD
BIAS
control,
refer
to
the
calibration
section
of
this
manual.
The
PULSE
OUTPUT
of
the
Type
284
has
an
amplitude
of
approximately
+200
mV
into
a
50
load.
The
pulse
width
is
approximately
1
ys
out
of
a
pulse
period
of
about
20
us.
The
rise
time
of
the
pulse
is
equal
to
or
less
than
70
ps.
With
these
specifications,
several
precautions
should
be
considered
when
connecting
the
Type
284
PULSE
OUTPUT
signal
to
a
test
device
or
to
a
display
oscilloscope.
Cable
Considerations
The
cables
that
conduct
the
output
to
the
device
under
test
should
be
low-loss
509
coaxial
cables
to
assure
that
all
information
contained
in
the
pulse
will
be
delivered
to
the
test
point
without
distortion.
The
physical
and
electrical
2-7

Operating
Instructions—Type
284
Fig.
2-8.
Typical
display
showing
adjustment
of
TD
BIAS
control.
characteristics
of
the
cable
determine
the
characteristic
im-
pedance,
velocity
of
propagation
and
nature
of
signal
loss.
Since
the
signal
losses
caused
by
energy
dissipation
in
the
dielectric
are
proportional
to
the
signal
frequency,
any
very
high
frequency
information
in
a
fast-rise
pulse
will
be
lost
in
a
very
few
feet
of
cable.
Therefore
it
is
important
to
use
cables
or
airlines
that
are
as
short
as
possible.
2-8
Impedance
Matching
To
provide
a
smooth
transition
between
devices
of
differ-
ent
characteristic
impedance,
each
device
must
encounter
a
total
impedance
that
is
equal
to
its
own
characteristic
im-
pedance.
Thus,
when
the
OUTPUT
PULSE
signal
of
the
Type
284
is
applied
to
a
load
other
than
509,
a
suitable
im-
pedance
matching
device
must
be
provided.
If
the
imped-
ances
are
not
matched,
reflections
and
standing
waves
in
the
cables
will
result
in
distortion
of
the
signal
at
the
load.
In
many
cases
the
load
required
will
be
509,
then
with
the
use
of
available
attenuators,
2X,
5X
and
10X
(10X
Attenuator,
Tektronix
Part
No.
017-0078-00)
will
attenuate
the
signal
and
provide
the
impedance
match.
Fig.
2-9
illustrates
a
simple
resistive
impedance-matching
network
that
provides
minimum
attenuation.
To
match
im-
pedances
with
the
network,
the
following
conditions
must
exist:
(Ri
+
Za)
Re
must
equal
Z,;
and
R,
+
ZiRe
must
(Ri
+
Zs)
+
Re
Zi
+
Re
equal
Zo».
Therefore:
RiRo
=
Z,Z.
and
RyZ;
=
Re
(Z2
—
Z;,)
or
Ry
=
Zp
(Zp
—
Zy);
and
R,
=
Z,
3
As an
example,
to
match
a
50-ohm
system
to
a
125-ohm
system:
Z,
=
50
ohms;
and
Z,
=
125
ohms.
Therefore:
R,
=
125
(125
—
50)
=
96.8
ohms
_
125
_
Ro
—
50
125
—
50.
=
64.6
ohms
Z,
E,
R,
E.
Z5
—>
e—
°
w—9)
<~—.
_
Ro
=
~
Z,
<Z
To
match
impedances:
or:
R,
=y
Zo
(Z,
—
Z,)
and
R,
=
Z,
2,
2
Vz,
—2,
Fig.
2-9.
Simple
resistive
impedance-matching
network
providing
minimum
attenuation.
(1

Though
the
network
in
Fig.
2-9
provides
minimum
attenu-
ation
for
a
purely
resistive
impedance-matching
device,
the
attenuation
as
seen
from
one
end
does
not
equal
that
seen
from
the
other
end.
A
signal
applied
from
the
lower
im-
pedance
source
(Z,)
encounters
a
voltage
attenuation
{(A;)
that
may
be
determined
as
follows:
EF,
—
E
Ep
Since:
Ip,
=
Iza;
————
=
‘
Ry
Zo
Therefore:
A,
=
Es
=
Ry
E.
1;
(1
<<
Ar
<
2)
A
signal
applied
from
the
higher
impedance
source
(Z.)
will
encounter
a
greater
voltage
attenuation
(Az)
that
may
be
determined
similarly:
E.
_
Ey __
E,
4
E,
Since
Ir1
=
Ire
+
!21;
R,
=~
RR,
Z,
Therefore:
_
&
| R
Ry
.
22,
Ag
—
E,
—
Ro
+
1;
(1
<
Az
<
Z,
)
In
the
example
of
matching
50
ohms
to
125
ohms,
96.8
Ai
=
125
+
]
=
1.77;
968
96.8
_
and
A,
=
oA6
+
50
+1=
444
Note
that
if
the
50-ohm
source
were
used
for
pulsing
a
high-impedance
load,
R;
would
approximately
equal
the
impedance
of
the
load
(high
R}
and
R,
would
approximately
equal
the
50
ohms
of
the
pulse
source.
In
this
situation,
voltage
attenuation
would
be
about
2.
If
a
low-impedance
load
(<Q
50
ohms}
were
to
be
en-
countered,
the
50-ohm
pulse
source
would
be
the
Z,
source.
If
the
load
impedance
were
to
approach
0
ohms,
the
value
of
R,
would
then
approach
the
load
impedance
(low
R).
Voltage
attenuation
in
this
case
would
become
quite
signif-
icant;
2Z,
=
Z.
The
illustrated
network
can
“he
modified
to
provide
dif-
ferent
attenuation
ratios
by
adding
another
resistor
(<Q
Rj)
in
series
between
Z,
and
the
junction
of
Ry
and
Ro.
Attenuation
=
(very
high)
Risetime
Considerations
If
the
PULSE
OUTPUT
signal
from
the
Type
284
is
to
be
used
for
determining
the
risetime
of
a
test
device
or
a
dis-
play
oscilloscope,
the
risetime
of
the
Type
284
may
have
to
be
taken
into
consideration.
Generally
the
risetime
of
the
applied
pulse
(Type
284
PULSE
OUTPUT)
should
be
four
times
or
more
faster
than
the
risetime
to
be
measured.
If
the
risetime
of
the
applied
pulse
approaches
the
risetime
of
the
system
under
test
the
actual
risetime
may
be
approximated
from
the
following
equation:
T,
measured
~
(T,
pulse}?
+
T,
(system)?
(2)
Calculations
are
not
necessary
if
the
oscilloscope
system
or
device
system
under
test
or
a
combination
of
the
device
system
under
test
and
the
oscilloscope
system
has
a
rise-
time
of
more
than
four
times
the
applied
pulse
(70
ps—Type
284)
or
280
ps.
The
above
system
includes
a
short
coaxial
Operating
Instructions—Type
284
connecting
cable
and
matching
attenuators.
For
long
coaxial
cables,
a
display
measurement
is
recommended
because
of
the
losses
introduced
by
different
types
of
cables.
Trigger
Output
The
TRIGGER
OUTPUT
signal
is
selected
with
the
MODE
switch
at
the
same
time
and
synchronized
with
either
the
PULSE
OUTPUT
or
the
SQUARE
WAVE
OR
SINE
WAVE
OUTPUT.
The
amplitude
of
the
TRIGGER
OUTPUT
signal
is
about
200
mV
into
a
50Q
load
in
all
positions.
With
the
SQUARE
WAVE
OUTPUT,
the
200
mV
square
wave
trigger
signal
facilitates
easy
triggering,
especially
for
the
10
mV
SQUARE
WAVE
AMPLITUDE.
The
TRIGGER
OUTPUT
signal
is
a 1
or
10ns
sine
wave,
when
the
PERIOD
and
the
MODE
switch
are
selected
for
1
or
10ns
SINE
WAVE
OUTPUT.
With
the
MODE
switch
in
the
PULSE
OUTPUT
position,
the
TRIGGER
OUTPUT
signal
leads
the
PULSE
OUTPUT
signal
by
5ns
or
75ns
as
selected
by
the
LEAD
TIME
switch.
See
Fig.
2-10.
|)
ft
|
|
TRIGGER
OUTPUT
PULSE
OUTPUT|
OUTPUT
LEAD
TIME
10
ns/Div
|i)
)i
[|
==
TRIGGER
OUTPUT
B)
SN
BO30236
-
up.
Fig.
2-10.
Typical
display
showing
TRIGGER
OUTPUT
signal,
with
PULSE
OUTPUT
signal
in
two
positions
of
the
LEAD
TIME
switch.
2-9

Operating
Instructions—Type
284
End
Line
|
Termination
|
Fig.
2-11.
Setup
for
Amplitude
checks
for
(A)
502
system,
(B)
Sampling
Probe
and
(C)}
Typical
waveform.
2-10
BASIC
APPLICATIONS
General
The
Type
284
is
a
source
of
Square
Wave,
Sine
Wave
and
Pulse
output
signals
useful
to
verify
the
deflection
factor
and
timing
accuracies
of
Sampling
Oscilloscopes.
It
is
also
useful
as
a
signal
source
for
other
applications.
This
part
of
the
operating
instructions
will
show
basic
uses
of
the
Type
284.
Checking
Amplitude
The
Type
284
SQUARE
WAVE
OUTPUT
signal
amplitudes
are
1
V,
100
mV, and
10
mV
into
50
Q
with
a
selection
of
the
time
PERIOD
of
10
ys,
1
us,
and
100
ns.
Fig.
2-11A
shows
Type
284
connected
to
the
Channel
A
input
of
the
Type
381
Dual
Trace
Sampling
Unit,
used
in
the
Type
561A
Oscilloscope
with
the
Type
3T2
Sampling
Sweep
Unit.
The
vertical
deflection
factor
of
other
5092
input
oscilloscope
can
be
checked
with
this
simple
connection.
Since
the
Type
284
uses
the
input
of
the
Sampling
unit
as
a
load,
the
accuracy
of
the
voltage
presented
to
the
input
depends
upon
the
accuracy
of
the
input
resistance.
The
DC
input
resistance
can
be
measured
with
an
accurate
bridge
with
the
oscilloscope
power
off.
If
external
attenuators
are
used,
their
accuracies
must
be
taken
into
account.
Exter-
nally
triggering
the
sampling
oscilloscope
from
the
Type
284
permits
higher
trigger
amplitude
(200
mV)
with
100
mV
and
10
mV
square
wave
output.
Fig.
2-11B
shows
a
method
of
connecting
into
sampling
oscillosopes
with
a
high
impedance
input
or
sampling
probe.
The
Type
VP-2
voltage
pickoff
adapter
permits
use
of
the
high
impedance
sampling
probe
to
look
at
signals
within
a
closed
50
system,
with
a
minimum
effect
upon
the
signal.
Then
by
using
an
accurate
end
line
termination
with
the
VP-2,
accurate
amplitudes
are
conveyed
to
the
Sampling
Probe.
Other
high
impedance
input
oscilloscopes
can
be
checked
with
the
Type
284
if
an
accurate
50Q
termination
is
pro-
vided
and
the
proper
adapter
is
connected
to
the
oscillo-
scope
input.
Fig.
2-11C
shows
a
typical
display
of
a
1
Volt
Square
Wave
Output
100ns
signal
used
in
checking
the
vertical
deflection
of
the
sampling
units.
Checking
Equivalent
Time
Bases
Signals
available
from
the
Type
284
for
checking
time
bases
include
square
waves
with
time
periods
of
10
us,
1
ps,
and
100
ns,
and
sine
waves
with
time
periods
of
10
ns
and
Ins.
The
amplitude
of
the
signal
is
not
important
in
check-
ing
equivalent
time
bases-as
long
as
the
signal
is
adequate
to
obtain
a
stable
display;
however,
with
the
SQUARE
WAVE
output,
amplitude
as
well
as
time
verification
can
be
made.
An
example
of
fast
verification
of
the
amplitude
and
time
checks
is
shown
with
the
Type
182
Sampling
Plug-In
Fig.
2-12B.
A
display
is
shown
of
the
Type
284
SQUARE
WAVE
OUTPUT
of
1
volt
amplitude
at
a
time
period
of
100
ns.
The
SQUARE
WAVE
signal
is
connected
through
a
502
coaxial
cable
to
one
of
the
Thru
Signal
Channel
50
connectors
on
the
Type
182.
The
other
Thru
Channel
509
connector
is
terminated
with
the
End
Line
termination
(Tektronix
Part
No.
017-0083-00).
The
Trigger
Output
signal
from
the
Type
284
is
connected
through
a
BNC
coaxial
cable
and
a
BNC
to
GR
adapter
to
the
Ext.
Trig
Input
connector
on
the
Type
152.
See
Fig.
2-12A.
Fig.
2-12C
is
a
display
of
the
10ns
PERIOD
SINE
WAVE
on
the
Type
182.
The
Type
182
is
set
for
10
ns/Div
(Horizon-
tal
Units)
with
.05
Volts/Div
(Vertical
Unit).
(B)1

502
SOQ
Ceaxial
Cable
=
BNC
Coaxial
Cable
:
i
Sar
ieee
4
}
:
t
oe
|
|
110
ns/DIV.
|
a
(C)
Display
of
10
ns
SINE
WAVE.
Fig.
2-12.
Setup
for
display
showing
timing
and
Amplitude
meas-
urements
for
Type
152.
Checking
Risetime
|
The
Pulse
output
from
the
Type
284
has
a
risetime
of
<70
ps,
a
width
of
about
1
us,
and
a
repetition
rate
of
about
50
kHz.
Risetime
verification
of
Vertical
Sampling
units
with
risetimes
equal
to
or
longer
than
four
times
the
Type
284
applied
pulse
can
be
made
without
calculations.
At
four
times
greater
risetime,
the
error
(by
using
equation
2
in
this
®1
Operating
Instructions—Type
284
on
|
|
’
Fig,
2-13.
Typical
displays
showing
risetime
checks.
section)
is
+3%
of
the
displayed
signal.
Four
times
70
ps
(Type
284
risetime)
or
about
280
ps,
or
slower
risetime
ver-
ticals
units
can
be
verified
directly.
Fig.
2-13A
shows
a
typical
display
with
the
Type
284
Pulse
Output
signal
connected
through
a
5ns
coaxial
cable
to
one
input
connector
of
the
Type
3S1.
A
Type
561A
Oscilloscope
with
a
Type
31T77A
Sampling
Sweep
Unit
was
used
with
the
2-11

Operating
Instructions—Type
284
Type
3S1.
The
risetime
is
verified
from
10%
to
90%
of
the
full
amplitude
to
be
equal
to
or
less
than
350
ps.
Fig.
2-13B
shows
a
display
of
the
Pulse
Output
signal
fed
through
a
60
ns
509
coaxial
cable.
Here
the
measurement
of
the
display
risetime
is
taken
from
the
10%
to
90%
points
of
a
lower
amplitude.
The
Pulse
does
not
reach
full
amplitude
in
this
display
of
100 ps/div.
By
changing
the
equivalent
time
to
5times
slower
or
.5
ns/div,
note
the
long,
gradual
slope
(Fig.
2-13C)
after
the
initial
rise
in
the
waveform.
This
distortion
is
referred
to
as
“dribble
up”
characteristic
of
the
coaxial
cable.
The
risetime
is
measured
from
the
10%
to
the
90%
points
of
the
initial
ramp
portion
of
the
displayed
signal
not
including
the
dribble
up
portion.
The
risetime,
dribble
up
and
other
character-
istics
of
coaxial
cables
can
be
evaluated
with
a
known
rise-
time
generator
and
sampling
oscilloscope
system.
Risetime
verification
of
a
system
including
the
display
oscilloscope
whose
risetime
approaches
the
risetime
of
the
Type
284
applied
pulse,
can
be
approximated
by
equation
(2)
in
this
section.
Checking
Fast
Oscilloscope
Vertical
Response
to
a
Fast
Step
Pulse
A
fast
step
pulse
will
show
up
various
responses
in
ampli-
fiers
due
to
terminations,
mechanical
layout,
and
characteris-
tics
of
input
circuits.
The
following
displays
(Fig.
2-14
through
2-21)
were
taken
showing
the
characteristics
of
some
ver-
tical
amplifiers
when
subjected
to
the
Type
284
Pulse.
The
Pulse
Output
signal
of
the
Type
284
is
connected
through
a
20Cm,
500
Airline
(Tektronix
Part
No.
017-0084-00}
to
the
50Q
input
sampling
oscilloscopes.
External
triggering
was
used
with
the
trigger
signal
from
the
Type
284
Trigger
output
connector
through
a
BNC
coaxial
cable
to
the
Ex-
ternal
Input
of
the
sampling
sweep
unit.
With
the
sampling
probe
type
of
sampling
oscilloscopes,
the
output
of
the
Type
284
is
connected
to
a
Type
VP-2,
and
an
End
Line
Termination
(Tektronix
Type
017-0081-00).
The
Sampling
Probe
was
in-
serted
in
the
Type
VP-2.
Other
high
impedance
oscilloscopes
required
an
accurate
50
Q
termination
and
an
adapter
to
connect
to
the
input.
Fig.
2-14.
Typical
Type
284
PULSE
OUTPUT
signal
display
using
Type
451
Vertical
Sampling
Unit.
tp
pe
Fig.
2-15.
Typical
Type
284
PULSE
OUTPUT
signal
display
using
Type
4$2A
Vertical
Sampling
Unit.
2-12
©!

Operating
Instructions—Type
284
Fig.
2-18.
Typical
Type
284
PULSE
OUTPUT
signal
display
using
Type
353
Vertical
Sampling
Unit.
®
2-13
Table of contents
Other Tektronix Pulse Generator manuals

Tektronix
Tektronix PG 509 User manual

Tektronix
Tektronix TYPE R Use and care manual

Tektronix
Tektronix PG 501 User manual

Tektronix
Tektronix TEK-DPG User manual

Tektronix
Tektronix SPG600 User manual

Tektronix
Tektronix PG 508 User manual

Tektronix
Tektronix PG 509 User manual

Tektronix
Tektronix PG 509 User manual

Tektronix
Tektronix PSPL2600C User manual

Tektronix
Tektronix SPG600 User manual