
capacitive
and
res1st1ve
loading due to the oscilloscope
can
be
reduced to a negligible value by using
an
attenu-
ator
probe.
In
addition to providing isolation of the oscilloscope
from the signal source,
an
att
enuator probe also decreases
the amplitude of the displayed waveform by the attenua-
tion factor
of
the probe. Use of a probe a llows you to
increase the vertical deflection factors
of
the oscilloscope
to
look
at
large
amp
litu
de
signals which
are
beyond the
normal
limits
of the oscilloscope
and
plug-
in
combination.
Signal amplitudes, however,
must
be
limited to the maxi-
mum
allowable value of the probe used.
Before using a probe, you must check (and adjust
if
necessary) the compensation
of
the probe to prevent dis-
tortion of the
appl
ied waveform. To adjust the probe
compensation, place the HORIZONTAL
DISPLAY
switch
at
NORMAL
(Xl), the
TIME
BASE
TRIGGERING MODE
switch
at
AUTO,
and
the
TIME
BASE
TRIGGER
SLOPE
switch
at
+INT. Turn up the intensity
until
the trace
is
visible
and
connect the probe tip to the
CAL
OUT con-
nector. Set the
AMPLITUDE
CALIBRATOR
switch for 2
centimeters of displayed signal. Set the
TIME
BASE
TIME/
CM
switch
to
display approximately 3 or 4 cycles
of
the
Calibrator waveform
and
adjust the probe compensation
control to obtain flat tops on the displayed Calibrator square
waves
as
shown
in
Fig.
2-7(6).
The method of setting the compensation control depends
on the probe
in
use.
If
your oscilloscope
is
equipped with
a Type
PSlOA
probe, compensation
is
by means
of
a
screwdriver adjustment through the hole near the nose end
of the probe body.
If
a
P6000
type probe
is
used,
it
is
necessary to first unlock the
Lo
cking Sleeve by turning
it
counterclockwise. The probe
is
th
en compensated by rotat-
ing the probe body while watching the oscilloscope dis-
play for the desired waveform. When compensation
is
INCORRECT
{a)
Operating Instructions
-Type
533/533A
HORIZONTAL
SWEEP
OF
CRT
BEAM
CRT
SCREEN
TRIGGERING WAVEFORM
Fig.
2-8.
In
triggered
sweep
operation
the
horizontal
sweep
is
started
by
the
triggering
waveforms.
completed, carefully turn the Locking Sleeve clockwise to
lock
it
without disturbing the adjustment of the probe.
HORIZONTAL
DEFLECTION
SYSTEM
Trigger
Operation
For
most uses of the oscilloscope a stable display of
some waveforms
is
required.
To
accomplish this the
Fig.
2-7.
When
compen
s
ating
the
pr
obe,
it
is
adjusted
to
obtain
an
undistorted
presentation
of
the
cal
ibrator
square-wave.
FREE
SCAN
IN THE PUBLIC DOMAIN 2-s