30. CHECK HOLD-OFF
Connec a 10X probe o he ungrounded side of C312. Free run he 321 sweep
and check for sufficien hold-off on all sweep ranges.
RANGE HOLD-OFF RANGE HOLD-OFF
.5 nsec 13 nsec 10 nsec 130 nsec
1 nsec 13 nsec 20 nsec 130 nsec
2 nsec 13 nsec 50 nsec 130 nsec
5 nsec 13 nsec
.1 Millisec .8 Millisec 10 Millisec 150 Millisec
.2 Millisec .8 Millisec 20 Millisec 150 Millisec
.5 Millisec .8 Millisec 50 Millisec 150 Millisec
1 Millisec 8 Millisec .1 Sec 150 Millisec
2 Millisec 8 Millisec .2 Sec 150 Millisec
5 Millisec 8 Millisec •5 Sec• 150 Millisec
MAKE NOTE OF CRT TYPE AND SERIAL NUMBER AND RECORD ON CALIBRATION RECORD.
32. CHECK OPERATION OF BATTERY CHARGER CIRCUIT.
Connec he scope o 117V wi h he power swi ch off. Wi h he ba ery pack
in place, connec he nega ive lead of he vol me er o he -10V regula ed
supply. Connec he + lead of he vol me er o he fron end of he 117Q
10W resis or on he lef side of he scope, (viewed from he fron ) Wi h
he swi ch in he DRY CELLS posi ion, he reading will be 0 vol s. Swi ch
o he 2.5AH posi ion and he me er should read 37-40V. Swi ch o he 4.3AH
posi ion and he me er should read 34-36V.
Now swi ch he scope ON and he me er should read 14-16V in he 4.3AH posi
ion, 16-18V in he 2.5AH posi ion and 0 V in he DRY CELL posi ion if he
charger is working properly.
33. CHECK BATTERY OPERATION
Remove he AC cord. Inser he ba ery pack. Hie scope should func ion
normally as long as he ba ery vol age under load is higher han 11.5 v.
34. CHECK DC OPERATION
Remove he ba ery pack from he scope and inser he DC cord. Connec he
ba ery pack o he black and whi e leads. (Nega ive o he whi e lead.)
The scope should opera e normally. Check be ween he black and green leads
for a resis ance of 13&»
Connec a DC supply o he black and whi e leads. Hie scope should regula e
from 11.5 v o 23 vol s.
Connec he DC supply o he whi e and green leads. The scope should regu
la e from 22 o 35 vol s.