to the right vertical
unit. Returnthe
TRIG SOURCEswitch
to VERT MODE. Remove calibrator signal from left
verticalunit and connect it to right vertical
unit. Repeat
steps
20 ro 22. The trigger signalwill come from right
vertical.
When the LEFT button is pressed
of the TRIG
SOURCE switch the displav
is not stablebecausethere is
no input signal
connectedto the left vertical.Return the
TRIG SOURCEswitch
to VERT
MODE.
Horizontal
Deflection
System
23. Position the start of the sweepto the left
graticule
linewith the
time-baseunit position
control.
24. Connect a 10X probe to the input of the right
vertical
unit. Set the right vertical unit for a deflection
factor of 10 volts/divisionand
setthe VERT MODE switch
to RIGHT. Set the time-base
unit for a sweeprateof five
milliseconds/div
ision.
25. Connect the probe
tip to a line-voltagesource.The
display should show three complete cycles over the 10
divisionswithin 0.3 division.
A correctdisplay
indicates
that the 7603 Horizontal Deflection System and the
time-base
plug-in
unit are
correctlycalibrated.
lf the display
is outside the given tolerance, either the 7603 or the
time-base
unit needs to be recalibrated.
Refer to the
Calibration
section of this manual. and to the time-base
unit manualfor adjustment
procedure.
NOTE
This step is based on an accurate 60-Hertz line
frequency. For other line frequencies, this procedure
will need to be changed accordingly.
26. Disconnect
the probe
from the line-voltage
source
and the right
vertical
unit
and
connectthe
probe
to the
left
vertical
unit. Set the VERT MODE
switch
to LEFT andset
the time-base
unit for a sweep
rate of 0.5 millisecond/
division.
Beam
Finder
27. Set the deflectionfactor of the left vertical
unit to
0.1 volt/division.
Notice that a square-wave
display
is not
visible.since
the deflection exceedsthe scan areaof the
CRT.
28. Pressand
hold the BEAM FINDER switch.Notice
that the display is returned to the viewing area in
compressedform. Increasethe vertical and horizontal
deflectionfactorsuntil the displav
is
reducedto abouttwo
divisionsverticallyand
horizontally
(when
the horizontal
unit is operated
in the time-base
mode, change
only the
REVB SEP1979
OperatingInstructions-7603/R7603 Service
deflection factor of the vertical
unit).
Adjust
the position
controls
of the displayed
vertical
unit and the time-base
unit to centerthe compresseddisplay
aboutthe centerlines
of the
graticule.
Release
the
BEAM FINDER
switch.Notice
that the displayremainswithin the viewing
area.
Z-Axislnput
29. lf an external signal
is
available
(five
volts
peak-to-
peak
minimum at two megahertz
or less),the function
of
the EXT Z AXIS input can be
demonstrated.Connect
the
external
signalto both the input of the right vertical
unit
and the EXf Z AXIS connectorwith two BNC
cables
anda
BNC T connector.Set the
VERT l\4ODE
switchto RIGHT
and set the vertical unit for a deflection factor of two
volts/division.Set
the
time-base
unit for a
sweeD
ratewhich
displaysseveralcycles
of the signal.Adjust
the
amplitudeof
the signal
generator
until intensitymodulation
isvisible
on
the display.
The positivepeaks
of the
waveformshouldbe
blanked
out and the negative
peaks
intensified.Notice
that
the setting of the INTENSITY control determinesthe
amountof intensitymodulation
that isvisible.
30. Disconnect the signal from the EXT Z AX lS
connector,but leaveit connected
to the right vertical
unit
input.
Check
that peak-to-peak
amplitude
of the displayed
signal
is
four
divisions
maximum.
31. This completes
the Operating
Checkout
procedure
for the 7603. Instrument
operationsnot
explained
here,
or
operationswhich need further explanation
are discussed
under
GeneralOperatingInformation.
SIMPLI
F
I
ED
OPERATINGINSTRUCTIONS
The following information
is
provided
to aid in
quickly
obtaining the correct setting for the 7603 controls to
present
a
display.The operator
shouldbe familiar
with the
complete function and operation of this instrument
as
described
elsewhere
in this section before using this
procedure.
For detailed operating information for the
plug-in
units,see
the
instruction
manuals
for the
applicable
un
its.
Single-TraceDisplay
The following procedure
will provide a display
of a
single-trace
verticalunit against
one time-base
unit. For
simplicity
of explanation,
thevertical
unit isinstalled
in
the
left vertical
compartment.The right vertical
compartment
canbe usedif
the
procedure
is
changed
accordingly.
1. Install
a 7A-series
com
partment.
unit in the left
1-7