
of 1 cycle or the time etween 2 or more events is measu
red (see Frequency Measurements). A portion of 1 cycle,
or a full cycle is usually displayed when measuring or o
serving small portions of a wave train (such as risetime,
pulse duration, and a errations). When part of a wave
train or waveform that occurs later than the sweep trigger
are to e examined, the X10 position of the SWEEP MAG
switch can e used to expand the center portion of the dis
play. The effective sweep rate for the expanded portion is
decreased y a factor of 10; therefore, timing measurements
must e divided y 10. Portions of the display to e expand
ed should e centered with the POSITION control efore
switching to the X10 position.
SIGNAL APPLICATION TO VERTICAL INPUT
For most applications, the P6049B, or similar pro e, should
e used as an interface etween the signal source and the
oscilloscope vertical input. The pro e increases the input
resistance of the oscilloscope system to a out 10 Mf2 and
decreases the effective capacitive loading to a out 13.5 pF.
The higher the impedance ratio etween the input to the
oscilloscope system and the signal source, the less the cir
cuit under test is loaded, and the more accurate the measure
ment. Because of circuit loading considerations, all
measurements should e made at low impedance points in
the circuit eing tested.
NOTE
To insure that measurements are accurate al
ways compensate the probe before using. To
compensate the probe touch the probe tip to
the .5 V CAL OUT connector on the 314 front
panel and adjust the compensation for the
best front-corner response to the square wave
signal (set VOL TS/Di V switch to 10 m and
TIME/DIV switch to 1 ms).
For applications in which the source, such as the output of
a signal generator, is a coaxial connector, use a coaxial ca le
etween the oscilloscope and the signal source. In this
application, loading is not a pro lem; however, to preserve
signal fidelity, the ca le impedance should equal the source
impedance and the ca le should e terminated in its charac
teristic impedance at the oscilloscope input.
Use a common-ground return etween the signal source and
the oscilloscope. When a pro e is used, connect the shortest
possi le ground lead near the signal source to avoid a long
signal path to ground.
WARNINGH
If the 314 is powered from an external dc source
(not connected to an ac power source) provide
@314 Operators 13