Teledyne Lecroy PP022-1 User manual

User Manual
PP02
2
Passive Probe

700 Chestnut Ridge Road
Chestnut Ridge, NY, 10977-6499
Tel: (845) 425-2000, Fax: (845) 578 5985
teledynelecroy.com
PP022 Passive Probe Instruction Manual
© 2021 Teledyne LeCroy, Inc. All rights reserved.
Users are permitted to duplicate and distribute Teledyne LeCroy documentation
for internal training purposes only. Resale or unauthorized duplication of
Teledyne LeCroy documentation materials is strictly prohibited.
Teledyne LeCroy is a registered trademark of Teledyne LeCroy, Inc. Other product
or brand names are trademarks or requested trademarks of their respective
holders. Information in this publication supersedes all earlier versions.
Specifications are subject to change without notice.
933091-00 Rev A
pp022-user-manual.pdf
February, 2021

User Manual
1
Terms and Symbols
The following symbols appear on the product or in its documentation:
WARNING
High Voltage. Risk of electric shock or burn.
WARNING or
CAUTION
Condition or practice could result in injury or loss
of life, or damage to equipment.
DOUBLE INSULATION
PROTECTIVE (EARTH) TERMINAL
Introduction
Teledyne LeCroy PP022 is a passive high impedance voltage oscilloscope
probe designed and calibrated for use with general purpose oscilloscopes
having an input impedance of 1 MΩ and can be compensated for use with
instruments having an input capacitance between 10 and 20 pF. It is
compatible with oscilloscopes that automatically detect probe attenuation
and adjust their readout accordingly.
To avoid electric shock, keep fingers behind the probe’s finger
guard during use.
Rated for indoor use only. Do not operate where conductive
pollutants may be present.

PP022 Passive Probe
2
Specifications
Model Attn.
Input
Imped. B/W
Rise
Time Length
Compensation
Range
R (MΩ) C (pF) (MHz) (ps) (m) (pF)
PP022
10x 10 10 500 700 1.3 10 - 20
Attenuation Ratio ......... 10:1 ± 1%
Max. Input Voltage....... Circuits not directly connected to Mains: 500 Vrms,
.........................................1500 V transient overvoltage
.........................................Measurement Category II*: 400 Vrms
To reduce risk of fire or electric shock, do not exceed the voltage or
category ratings of the probe. See voltage derating curve on p.3.
Pollution Degree ........... Pollution Degree 2*
Operating Altitude ................. Up to 3000 meters
Max. Operating Temp.. 0° C to +50° C
Humidity......................... 5% to 95% RH (10 to 30 °C)
.........................................5% to 75% RH (30 to 40 °C)
.........................................5% to 50% RH (above 40 °C)
.........................................RH not controlled below 10 °C
* As defined in IEC/EN 61010-031. See definitions below.
Definitions
Measurement Category II (CAT II) refers to measurements performed on circuits directly connected to
utilization points (socket outlets and similar) of the low-voltage mains installation.
Pollution Degree 2 refers to an operation environment where normally only dry non-conductive
pollution occurs. Temporary conductivity caused by condensation should be expected.

User Manual
3
PP022 Input Impedance Profile
Voltage vs. Frequency Derating Curve

PP022 Passive Probe
4
EC Declaration of Conformity
The product conforms to the applicable European Union
requirements per Low Voltage Directive (LVD) 2014/35/EU.
Compliance was demonstrated to the following specification as listed in
the Official Journal of the European Communities:
IEC/EN 61010-031:2015 Safety requirements for electrical equipment for
measurement, control and laboratory use – Part 031: Safety requirements for
handheld probe assemblies for electrical measurement and test.
EU RoHS2 Compliance
Unless otherwise specified, all materials and processes are compliant with
RoHS Directive 2011/65/EU in its entirety, inclusive of any further
amendments or modifications of said Directive.
Disposal
The probe complies with the applicable European Union
requirements to Directives 2012/19/EU on Waste Electrical and
Electronic Equipment (WEEE).

User Manual
5
General Safety Information
Observe generally accepted safety procedures in addition to those
listed here to avoid personal injury or damage to equipment. The
overall safety of any system incorporating this accessory is the
responsibility of the assembler of the system.
Connect only to grounded instruments. Use only with compatible Teledyne
LeCroy oscilloscopes that have their BNC input connected to an earth
ground. Do not connect the probe reference lead to any point which is at a
potential other than earth ground.
Connect and disconnect properly. Connect probe to the oscilloscope before
connecting it to the test circuit. Disconnect the probe input and reference
lead from the test circuit before disconnecting from the oscilloscope. Do
not connect/disconnect probes while connected to a voltage source.
Do not overload. Do not apply any potential to the probe leads that exceeds
the maximum rating of the probe. Observe all terminal ratings on the
oscilloscope before connecting. Consult the oscilloscope product manual
for further ratings information.
Always comply with the Voltage vs. Frequency Derating Curve.
Do not remove probe casing. Touching exposed connections may result in
electric shock.
Use indoors only within operational environment listed. Do not use in wet or
explosive atmospheres. Keep product surfaces clean and dry.
Handle with care. Probe tips are extremely sharp and may puncture skin or
cause other bodily injury if not handled properly.
Keep fingers behind the finger guard of probe body and accessories.

PP022 Passive Probe
6
Do not operate with suspected failures. Before each use, inspect the probe
and accessories for any potential damage such as tears or other defects in
the probe body, cable jacket, accessories, etc. If any part is damaged, cease
operation immediately and sequester the probe from inadvertent use.
Cleaning
•Clean only the exterior of the probe, cables, and accessories. Use a
soft cotton cloth moistened with a mild detergent and water solution.
Do not allow any portion of the probe to be submerged at any time.
•Dry the probe and accessory thoroughly before attempting to make
any voltage measurement.
•Do not subject the probe to solvents or solvent fumes as these can
cause deterioration of the probe body, cables, and accessories.
Service
Refer all repairs to qualified service personnel. Contact Teledyne LeCroy at
the following address to arrange to return the probe for service:
Teledyne LeCroy, Inc.
700 Chestnut Ridge Road
Chestnut Ridge, NY, 10977-6499
Ph: 800-553-2769 or 845-425-2000
Fax: 845-578-5985
contact.cor[email protected]
For the most complete and up-to-date list of sales and service centers by
country, visit teledynelecroy.com/support/contact.

User Manual
7
Probe Compensation
Proper compensation of the probe is required to assure amplitude accuracy
of the waveform being measured by matching the probe to the
oscilloscope’s input capacitance. Compensation should be adjusted
whenever the probe is connected to or transferred between oscilloscopes.
Low Frequency
1. Apply a 1 kHz square wave to the probe or connect to the
oscilloscope’s calibrator output.
2. Adjust the single LF trimmer located on the PP022 BNC
Compensation Box until you achieve a flat-topped square wave.
High Frequency
1. Apply a 1 MHz square wave to the probe (< 0.7 ns rise-time).
2. Remove the plastic caps from the two, HF trimmers on the PP022
BNC Compensation Box.
3. Adjust HF2 then HF1 until you achieve a flat-topped square wave.

PP022 Passive Probe
8
Over-compensated, correct, and under-compensated waveforms
Probe Accessories
Probe Tips
Both the rigid probe tip and the spring tip can be screwed on to the probe
body by hand.
Be sure not to overtighten the tips to avoid damaging the threads.
Ground Accessories
The ground lead and the short collar are attached to the probe at the time
of shipping. To replace the ground lead with the swivel lead adapter,
unscrew the short collar from the probe and remove the ground lead.
Replace the short collar to hold the swivel lead adapter in place.
Alternatively, the long collar can be pushed onto the probe body and used in
conjunction with the ground spring, BNC adapter or PCB adapter.

User Manual
9
Replaceable Parts
Item
Part Number
Description
QTY
1
PK5-2.5MM-101
Ground Lead with Alligator Clip
1
2
PK5-2.5MM-102
Rigid Probe Tip (1 installed standard)
2
3
PK5-2.5MM-103
Short Collar (installed standard)
1
4
PK5-2.5MM-104
Sprung Hook
1
5
PK5-PK6-001
Trimmer Tool
1
6
PK5-2.5MM-105
Long Collar
1
7
PK5-2.5MM-106
IC Tip Insulators (Black, Brown, Blue, Green, Gray)
5
8
PK5-2.5MM-107
Tip Insulator 2.5mm
1
9
PK5-2.5MM-108
Probe Tip Cover
1
10
PK5-2.5MM-109
BNC Adapter 2.5mm
1
11
PK5-2.5MM-110
Swivel Ground Lead
1
12
PK5-2.5MM-111
Swivel Ground Lead Adapter
1
13
PK5-2.5MM-112
Ground Spring
1
14
PK5-2.5MM-113
PCB Adapter
1
15
PK5-2.5MM-114
Spring Tip
1
16
PK5-PK6-002
Color Coding Rings (2 ea. yellow, red, green, blue)
8
Standard Accessory Replacement Kit part number is PKIT5-2.5MM-101.

PP022 Passive Probe
10
Warranty
NOTE: THE WARRANTY BELOW REPLACES ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED
WARRANTY OF MERCHANTABILITY, FITNESS, OR ADEQUACY FOR ANY
PARTICULAR PURPOSE OR USE. TELEDYNE LECROY SHALL NOT BE LIABLE FOR
ANY SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER IN
CONTRACT OR OTHERWISE. THE CUSTOMER IS RESPONSIBLE FOR THE
TRANSPORTATION AND INSURANCE CHARGES FOR THE RETURN OF PRODUCTS
TO THE SERVICE FACILITY. TELEDYNE LECROY WILL RETURN ALL PRODUCTS
UNDER WARRANTY WITH TRANSPORT PREPAID.
Teledyne LeCroy warrants this oscilloscope accessory for normal use and
operation within specification for a period of one year from the date of shipment.
Spare parts, replacement parts and repairs are warranted for 90 days.
In exercising its warranty, Teledyne LeCroy, at its option, will either repair or replace
any assembly returned within the warranty period to the Customer Service
Department of an authorized service center. However, this will be done only if the
product is determined by Teledyne LeCroy’s examination to be defective because of
workmanship or materials, and the defect is not caused by misuse, neglect,
accident, abnormal conditions of operation; or, damaged by attempted repair or
modifications by a non-authorized service facility.
The customer will be responsible for the transportation and insurance charges for
the return of products to the service facility. Teledyne LeCroy will return all products
under warranty with transportation charges prepaid.
This warranty replaces all other warranties, expressed or implied, including, but not
limited to, any implied warranty of merchantability, fitness, or adequacy for any
particular purposes or use. Teledyne LeCroy shall not be liable for any special,
incidental, or consequential damages, whether in contract or otherwise.
933091-00 Rev A
February, 2021
This manual suits for next models
1
Table of contents
Other Teledyne Lecroy Diagnostic Equipment manuals
Popular Diagnostic Equipment manuals by other brands

iCarsoft
iCarsoft CR IMMO user manual

Mitsubishi MOTORS
Mitsubishi MOTORS M.U.T.-III owner's manual

Truma
Truma Diagnose Tool manual

Continental Refrigerator
Continental Refrigerator AUTODIAGNOS TPMS SE quick start guide

MSW Motor Technics
MSW Motor Technics MSW-DENTPULLER.L01 user manual

Xtool
Xtool D8 user manual