Tempo Communications OFL100 User manual

OFL100 OTDR
55501342 REV 0 © 2022 Tempo Communications Inc. 03/2022
INSTRUCTION MANUAL
Read and understand all of the instructions
and safety information in this manual
before operating or servicing this tool.
Register this product at www.TempoCom.com
SET ESC
ON
CHARGE
www.lambdaphoto.co.uk
T: +44 (0)1582 764334

OFL100
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
2
Table of Contents
Preface 3
Description ............................................................................................................................................................ 3
Safety....................................................................................................................................................................... 3
Purpose of This Manual ................................................................................................................................. 3
Warranty................................................................................................................................................................. 3
Important Safety Information 4
Section 1. Introduction 7
Section 2. Basic OTDR Information 8
Principle of OTDRs ............................................................................................................................................ 8
%DVLF'HƬQLWLRQDQG&ODVVLƬFDWLRQRI(YHQWV....................................................................................... 8
Section 3. OFL100 Operation 8
Host .......................................................................................................................................................................... 8
Function Keys ...................................................................................................................................................... 9
Main Interface ...................................................................................................................................................... 9
Auto OTDR ..........................................................................................................................................................10
Expert OTDR ......................................................................................................................................................10
Parameter Settings..........................................................................................................................................11
OFL100 Pulsewidth Selection .....................................................................................................................11
Threshold/Criterion ........................................................................................................................................12
Threshold Settings................................................................................................................................ 12
Eligibility Criteria ..................................................................................................................................... 12
OFL Trace ............................................................................................................................................................. 12
OTDR File Saving.............................................................................................................................................. 13
File Operation.....................................................................................................................................................13
Event Map............................................................................................................................................................14
OPM ........................................................................................................................................................................14
VFL ..........................................................................................................................................................................14
Laser Source....................................................................................................................................................... 15
Optical Loss Test.............................................................................................................................................. 15
RJ45 Sequence Testing................................................................................................................................. 15
System Settings ............................................................................................................................................... 16
Section 4. OFL100 Troubleshooting 17
Section 5. Trace Viewer 17

OFL100
3
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
Preface
Description
7KH7HPSR&RPPXQLFDWLRQV2)/LVDQ2SWLFDO7LPH'RPDLQ5HƮHFWRPHWHU27'5WKDWLVXVHGWRPHDVXUH
WKHOHQJWKDQGLQGLYLGXDOORVVHYHQWVRQDVLQJOHPRGHƬEHURSWLFOLQN
Safety
Safety is essential in the use and maintenance of Tempo Communications tools and equipment.
7KLVLQVWUXFWLRQPDQXDODQGDQ\PDUNLQJVRQWKHWRROSURYLGHLQIRUPDWLRQIRUDYRLGLQJKD]DUGVDQGXQVDIH
practices related to the use of this tool. Observe all of the safety information provided.
Purpose of This Manual
7KLVLQVWUXFWLRQPDQXDOLVLQWHQGHGWRIDPLOLDUL]HDOOSHUVRQQHOZLWKWKHVDIHRSHUDWLRQDQGPDLQWHQDQFHSUR-
cedures for the Tempo Communications OFL100 handheld OTDR.
This instruction manual is free to download from www.tempocom.com
Warranty
Tempo Communications Inc. warrants to the original purchaser of these goods for use that these products will
EHIUHHIURPGHIHFWVLQZRUNPDQVKLSDQGPDWHULDOIRUWZR\HDUV7KLVZDUUDQW\LVVXEMHFWWRWKHVDPHWHUPV
and conditions contained in Tempo Communication’s standard one-year limited warranty.
)RUDOO7HVW,QVWUXPHQWUHSDLUVFRQWDFW&XVWRPHU6HUYLFHDWWRUHTXHVWD5HWXUQ$XWKRUL]DWLRQ
)RULWHPVQRWFRYHUHGXQGHUZDUUDQW\VXFKDVLWHPVGURSSHGDEXVHGHWFDUHSDLUFRVWTXRWHLVDYDLODEOH
upon request.
1RWH3ULRUWRUHWXUQLQJDQ\WHVWLQVWUXPHQWSOHDVHFKHFNWKHEDWWHU\LVDWIXOOFKDUJH

OFL100
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
4
Important Safety Information
SAFETY
ALERT
SYMBOL
7KLVV\PEROLVXVHGWRFDOO\RXUDWWHQWLRQWRKD]DUGVRUXQVDIHSUDFWLFHVZKLFKFRXOG
UHVXOWLQDQLQMXU\RUSURSHUW\GDPDJH7KHVLJQDOZRUGGHƬQHGEHORZLQGLFDWHVWKH
VHYHULW\RIWKHKD]DUG7KHPHVVDJHDƱHUWKHVLJQDOZRUGSURYLGHVLQIRUPDWLRQIRU
SUHYHQWLQJRUDYRLGLQJWKHKD]DUG
,PPHGLDWHKD]DUGVZKLFKLIQRWDYRLGHG:,//UHVXOWLQVHYHUHLQMXU\RUGHDWK
+D]DUGVZKLFKLIQRWDYRLGHG&28/'UHVXOWLQVHYHUHLQMXU\RUGHDWK
+D]DUGVRUXQVDIHSUDFWLFHVZKLFKLIQRWDYRLGHG0$<UHVXOWLQLQMXU\RUSURSHUW\
damage.
Read and understand all of the instructions and safety information in
this manual before operating or servicing this tool.
)DLOXUHWRREVHUYHWKLVZDUQLQJFRXOGUHVXOWLQVHYHUHLQMXU\RUGHDWK
(OHFWULFVKRFNKD]DUG
• 'RQRWRSHUDWHLQDQH[SORVLYHDWPRVSKHUHVXFKDVLQWKHSUHVHQFHRIƮDPPDEOH
gases or fumes.
• Before applying power, verify that the unit is set to match the available line
voltage.
)DLOXUHWRREVHUYHWKHVHZDUQLQJVFRXOGUHVXOWLQVHYHUHLQMXU\RUGHDWK

OFL100
5
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
Important Safety Information FRQWpG
&RQQHFWLQJWKH2)/WRDOLYHƬEHUPD\FDXVHSHUPDQHQWGDPDJH
'RQRWFRQQHFWWROLYHƬEHUV
(OHFWULFVKRFNKD]DUG
• )XVHUHSODFHPHQWVKRXOGEHSHUIRUPHGE\7HPSRTXDOLƬHGSHUVRQQHO
• Do not use repaired fuses or short-circuited fuse holders.
)DLOXUHWRREVHUYHWKHVHZDUQLQJVFRXOGUHVXOWLQVHYHUHLQMXU\RUGHDWK
While there is no potential for eye damage due to
unaided direct exposure, users should always avoid
ORRNLQJGLUHFWO\LQWRWKHRXWSXWSRUW7KHXVHRI
optical viewing instruments, such as microscopes,
PDJQLƬHUVHWFVKRXOGDOZD\VEHDYRLGHG7KHXVH
RIVXFKGHYLFHVDURXQGDFWLYHƬEHUVFDQIRFXVDQ
intense beam of light energy onto the retina of the
eye, which can result in permanent damage.
/DVHUKD]DUG
• :KHQSHUIRUPLQJPHDVXUHPHQWVRQƬEHURSWLFV\VWHPVDYRLGH\HH[SRVXUHWR
DQ\RSHQHQGHGƬEHUVRSWLFDOFRQQHFWRUVRSWLFDOLQWHUIDFHVRURWKHUVRXUFHV
because they may be connected to active laser transmitters.
• 'RQRWORRNLQWRWKHRSWLFDOSRUWZKHQDVRXUFHLVWXUQHGRQ
• $YRLGORRNLQJDWWKHIUHHHQGRIDWHVWƬEHULHWKHHQGQRWFRQQHFWHGWRWKH
LQVWUXPHQW,ISRVVLEOHGLUHFWWKHIUHHHQGWRZDUGDQRQUHƮHFWLYHVXUIDFH
)DLOXUHWRREVHUYHWKHVHSUHFDXWLRQVPD\UHVXOWLQLQMXU\
VISIBLE LASER RADIATION
AVOID DIRECT EYE EXPOSURE
CLASS 2 LASER PRODUCT
MAX OUTPUT < 1mW
WAVELENGTH 650±20nm
PULSE DURATION CW
INVISIBLE LASER RADIATION DO NOT VIEW
DIRECTLY WITH OPTICAL INSTRUMENTS
CLASS 1M LASER PRODUCT
IEC 60825-1:2014
IEC 60825-1:2014
WARNING

OFL100
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
6
Important Safety Information FRQWpG
(OHFWULFVKRFNKD]DUG
• 'RQRWH[SRVHEDWWHULHVWRƬUHRULQWHQVHKHDW'RQRWRSHQRUPXWLODWHEDWWHULHV
Avoid touching the electrolyte in the batteries, which is corrosive and may cause
GDPDJHWRH\HVRUVNLQ
• Do not open the case of the unit for any reason. It contains no user-serviceable
parts.
• 8VHWKLVXQLWIRUWKHPDQXIDFWXUHUpVLQWHQGHGSXUSRVHRQO\DVGHVFULEHGLQWKLV
manual. Any other use can impair the protection provided by the unit.
)DLOXUHWRREVHUYHWKHVHSUHFDXWLRQVPD\UHVXOWLQLQMXU\DQGPD\GDPDJHWKHXQLW
,QVWUXPHQWGDPDJHKD]DUG
• 0DNHVXUHWKDWWKHRSWLFDOƬEHURUFDEOHLVQRWLQXVHDQGWKHUHLVQRODVHUEHDPLQ
WKHƬEHUEHIRUHWHVWLQJZLWKWKLVXQLW
• Do not leave the unit in direct sunlight or near direct sources of heat.
• 3URWHFWWKHXQLWIURPVWURQJLPSDFWVRUVKRFN
• Do not immerse the unit in water or store in areas with high humidity.
• When necessary, clean the case, front panel, and rubber cover with a damp cloth.
Do not use abrasives, harsh chemicals, or solvents.
• Replace the protective dust cap when the unit is not in use.
• Store the unit and interface adapters in a cool, dry, and clean place.
)DLOXUHWRREVHUYHWKHVHSUHFDXWLRQVPD\UHVXOWLQLQMXU\DQGPD\GDPDJHWKHXQLW
(OHFWULFVKRFNKD]DUG
• Do not expose the unit to extremes in temperature or high humidity. Refer to
q6SHFLƬFDWLRQVr
• 7XUQWKHXQLWRƪGLVFRQQHFWIURPSRZHUDQGPDNHVXUHWKHODVHUVRXUFHLVRƪ
before cleaning.
)DLOXUHWRREVHUYHWKHVHSUHFDXWLRQVPD\UHVXOWLQLQMXU\DQGPD\GDPDJHWKHXQLW
The OFL100 uses a polymer lithium battery and is to be charged with the supplied
charger at 5V/2A at a temperature range of 0° to 50°. The OFL100 will not charge.
Do not use any other charger to charge the OFL100.

OFL100
7
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
6HFWLRQ ,QWURGXFWLRQ
7HPSRpV2)/KDQGKHOG27'5FDQPHDVXUHDVLQJOHRSWLFDOƬEHURUDQHQWLUHƬEHURSWLFOLQN
7KH2)/FDQFKHFNWKHWUDQVPLVVLRQTXDOLW\RIRSWLFDOƬEHUWKURXJKWKHPHDVXUHPHQWRIEDFNVFDWWHU
6WDQGDUGRUJDQL]DWLRQVOLNHWKH,QWHUQDWLRQDO7HOHFRP8QLRQ,78GHƬQHEDFNVFDWWHUDVDQHƪHFWLYHDQDO\VLV
PHDQVWRPHDVXUHRSWLFDOƬEHUORVV%DFNVFDWWHULQJLVDOVRWKHRQO\HƪHFWLYHZD\RIFRQQHFWRULQVSHFWLRQ
ZKLFKFDQEHDSSOLHGWRPHDVXUHWKHOHQJWKRIRSWLFDOƬEHU
7KH2)/27'5RSHUDWHVE\UHYLHZLQJHYHQWVLQRSWLFDOƬEHUIRUH[DPSOHLUUHJXODULWLHVDQGFRQQHFWRUV
7KHVHWRROVFDQKHOSLGHQWLI\LUUHJXODULWLHVLQRSWLFDOƬEHUORFDWHWKHPDQGPHDVXUHWKHLUDWWHQXDWLRQUHOHYDQW
loss, and homogeneity.
The OFL100 is easy to use, small, and compact with large touchscreen LCD display with graphical interfaces.
7KH\FDQVDYHDQGWUDQVIHUWKHPHDVXUHPHQWFXUYHVGDWDWRD3&E\XVLQJWKHSURYLGHG7UDFH9LHZHUVRƱ-
ZDUHIRUIXUWKHUDQDO\]LQJUHSRUWLQJDQGSULQWLQJ
Basic applications:
s 0HDVXUHWKHOHQJWKRIRSWLFDOƬEHUDQGFDEOH
s 0HDVXUHWKHGLVWDQFHEHWZHHQWZRSRLQWVRQRSWLFDOƬEHUDQGFDEOH
Ǵ 0HDVXUHORVVEHWZHHQWZRSRLQWVRQRSWLFDOƬEHUDQGFDEOH
• Measure fusion splice losses.
s0HDVXUHUHƮHFWLRQRIUHƮHFWLYHHYHQWVRIƬEHURSWLFOLQNV
s /RFDWHIDXOWVRIRSWLFDOƬEHURSWLFFDEOHV
s 'LVSOD\GLVWULEXWLRQFXUYHRIRSWLFDOƬEHUDQGFDEOHORVV

OFL100
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
8
6HFWLRQBasic OTDR Information
Principle of OTDRs
$Q27'52SWLFDO7LPH'RPDLQ5HƮHFWRPHWHULVDPHDVXUHPHQWLQVWUXPHQWIRULGHQWLI\LQJRSWLFDOƬEHU
WUDQVPLVVLRQIHDWXUHV7KH27'5LVXVHGWRPHDVXUHWKHRYHUDOODWWHQXDWLRQRIDƬEHURSWLFOLQNDQGWRSURYLGH
GHWDLOVUHODWLQJWRWKHSRVLWLRQRIHDFKHYHQWLQWKDWOLQN(YHQWVLQFOXGHVSOLFHVFRQQHFWRUVEHQGVDQGRSWLFDO
components. Its non-destructive, single-ended connection and rapid measurement have made the OTDR an
LQGLVSHQVDEOHWRROIRUPDQXIDFWXULQJFRQVWUXFWLRQDQGPDLQWHQDQFHRIRSWLFDOƬEHUOLQNV
7KHIDXOWVDQGKHWHURJHQHLW\RIRSWLFDOƬEHULWVHOIFDXVH5D\OHLJKVFDWWHULQJRIOLJKWWUDQVPLWWHGLQRSWLFDOƬEHU
3DUWRIWKHOLJKWSXOVHLVVFDWWHUHGLQWKHUHYHUVHGLUHFWLRQDQGWKLVLVFDOOHG5D\OHLJKEDFNVFDWWHULQJZKLFK
DFWXDOO\SURYLGHVDWWHQXDWLRQGHWDLOVUHODWLQJWRƬEHUOHQJWK
,QIRUPDWLRQUHODWLQJWRGLVWDQFHLVREWDLQHGWKURXJKWLPHLQIRUPDWLRQWKXVqWLPHGRPDLQrLQWKHQDPH27'5
)UHVQHOUHƮHFWLRQRFFXUVDWWKHERXQGDU\EHWZHHQWZRPHGLDRIGLƪHUHQW,25IRUH[DPSOHFRQQHFWLRQVRI
IDXOWVFRQQHFWRUVRURSWLFDOƬEHUHQG7KLVUHƮHFWLRQLVXVHGWRORFDWHWKHGLVFRQWLQXRXVSRLQWVRQRSWLFDOƬEHU
7KHPDJQLWXGHRIUHƮHFWLRQGHSHQGVRQWKHGLƪHUHQFH
between IOR and the smoothness of boundary.
$Q27'5VHQGVRXWDOLJKWSXOVHLQWRWKHRSWLFDOƬEHUDQGUHFHLYHVUHƮHFWLRQVRIHYHQWVDQGEDFNVFDWWHULQJ
SRZHURIWKHSXOVHLQWLPH/RFDWLRQVZLOOEHGLVSOD\HGRQWKH/&'7KH\D[LVLVWKHG%YDOXHRIEDFNVFDWWHULQJ
power, and the x-axis is the distance.
%DVLF'HƬQLWLRQDQG&ODVVLƬFDWLRQRI(YHQWV
Events refer to any abnormal points causing attenuation or sudden change of scattering power besides the
QRUPDOVFDWWHULQJRIRSWLFDOƬEHUZKLFKLQFOXGHORVVHVVXFKDVEHQGLQJFRQQHFWLRQVDQGUXSWXUHV
Event points displayed on the LCD are abnormal points that cause traces to deviate from a straight line.
(YHQWVFDQEHFODVVLƬHGDVUHƮHFWLRQHYHQWVRUQRQUHƮHFWLRQHYHQWV
6HFWLRQOFL100 Operation
1 OTDR/LS port
2 OPM port
3 VFL port
4 Flashlight
5 LAN 1
6 LAN 2
7 SD Card
8 Type C USB
1 Dust cover
2 4.3 inch color LCD
3 Function keys
4 Charging indicator
5 Power on status indicator
1 RJ45 Sequence Test
Remote
Main ViewTop
Host
Bottom

OFL100
9
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
Function Keys
Main Interface
Press less than two seconds to turn on the OFL100 on.
3UHVVIRUORQJHUWKDQWZRVHFRQGVWRREWDLQWKHSRSXSPHVVDJHWRWXUQWKH2)/Rƪ
:KLOHWKH2)/LVRQDVKRUWSUHVVZLOOWRJJOHWKHƮDVKOLJKWRQDQGRƪ
Direction key
SET key
Press to access basic OFL100 test settings.
ESC key
Exit current function.
Zoom control / AB cursor switch key
Test / Stop key
Press to start or stop test.
Press to PRYHVHOHFWLRQXSGRZQOHƱRUULJKW
Toggles between cursor movement and
zooming in on the currently displcayed trace.
Power ON/OFF key:
SET ESC
ON
CHARGE
8VHWKHFXUVRUNH\VWRQDYLJDWHWRWKHGHVLUHG
IXQFWLRQDQGSUHVV2.NH\WRVHOHFWWKDW
function or directly select on the screen.
3UHVVWKHq6KRUWFXWPHQXrLFRQWRHQWHUWKH
TXLFNRSHUDWLRQPHQXDQGSUHVVVHOHFWWKH
desired function.
3UHVVWKH3ULQW6FUHHQFDPHUDLFRQWRPDNHD
screen shot. The screen shot is automatically
VDYHGZLWKDƬOHQDPHEHLQJWKHFXUUHQWWLPH
Date & Time 86%
SD Card Shortcut menu
Battery
Flashlight

OFL100
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
10
Auto OTDR
All parameters are automatically determined and the
OFL100 will measure the length and attenuation of a
ƬEHURSWLFOLQN
([SHUW27'5
7KHXVHUFDQVHWHDFKWHVWSDUDPHWHUWRRSWLPL]HIRU
the most accurate test result.
AutoTest
Perform an automatic test with OFL100
determined settings.
Real Test
Perform an updated real time test at a rate of
DSSUR[LPDWHO\+]IRUUHDOWLPHLGHQWLƬFDWLRQRI
events.
Averaging Test
Perform a test with user determined settings with a
static display upon completing the measurement.
The trace and event list are
displayed at the same time.
/LQNUHVXOWVDUH
VXPPDUL]HGLQDOLVW
Switch to event icon
display mode.
6DYHWKHFXUUHQWWUDFHƬOH
TXLFNO\
Enter parameter setting
interface.

OFL100
11
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
Parameter Settings
Wavelength: The wavelength can be set to
1310nm or 1550nm.
Range: The range can be set to capture the entire
ƬEHUOLQNOHQJWK
3XOVHZLGWKRefers to the time width of the optical
pulse signal emitted during test. The larger the pulse
ZLGWKWKHVWURQJHUWKHRSWLFDOSRZHULQMHFWHGLQWRWKH
RSWLFDOƬEHUWKHVWURQJHUWKHEDFNVFDWWHULQJVLJQDO
RIWKHRSWLFDOƬEHULVDQGWKHIDUWKHUWKHHƪHFWLYH
detection distance of the OTDR can be. However, the
large pulse width will cause saturation of the initial
UHƮHFWLRQVLJQDODQGDORQJHUGHDG]RQH7KHFKRLFHRI
SXOVHZLGWKLVUHODWHGWRWKHOHQJWKRIWKHRSWLFDOƬEHU
The longer the length, the larger the pulse width,
ZKLFKFDQRQO\EHPRGLƬHGLQUHDOWLPHDYHUDJH
measurement mode.
IOR: The Index of Refraction is provided by optical
FDEOHƬEHUPDQXIDFWXUHUDQGFDQEHIRXQGRQWKH
LQGLYLGXDOVSRRORIƬEHU7KHWHFKQLFLDQFDQHQWHUWKH
IOR for the most accurate measurements.
Test time:8VHGLQWKHDYHUDJHWHVWPRGHWRVHWWKH
total averaging time. A longer averaging time will
result in the best signal to noise ratio.
Unit: 8VHGWRVHOHFWWKHGHVLUHGXQLWVWRNPNIHHW
or miles.
2)/3XOVHZLGWK6HOHFWLRQ

OFL100
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
12
OFL100 Trace
Select correct parameter, the test results such as trace
DQGHYHQWOLVWZLOOEHGLVSOD\HGDƱHUWHVWFRPSOHWHG
Curve Zoom:
3UHVVWKH>]RRP@PHQXWRHQWHUWKH]RRPLQDQG]RRP
out mode.
(YHQW/LVW
Total length: 7KHWRWDOOHQJWKRIWKHOLQNXQGHUWHVW
Total loss:7KHWRWDOORVVRIWKHOLQNXQGHUWHVW
Average Loss:7KHORVVSHUNLORPHWHURIWKHOLQN
under test.
Total events: The total number of events, the number
of passed events and the number of failed events of
WKHOLQNXQGHUWHVW
12: The order of the current event.
Type: The type of the current event.
Distance: The location of the current event.
Segment Section: the distance between the previous
event and the current event.
Loss: the loss of the current event.
5HƮHFWLYLW\ the return loss of the current event.
/
/
Zoom in or out in X direction
Zoom in or out in Y direction
7KUHVKROG&ULWHULRQ
7KUHVKROG6HWWLQJV
(YHQWORVVWKUHVKROG Set the loss threshold for
connectors and fusion splices between 0.2 and 30dB.
The default is 0.2dB. Events larger than the threshold
will be listed in the event table.
5HIHOFWLRQWKUHVKROG Set the return loss threshold
IRUUHƮHFWLYHHYHQWVUDQJLQJEHWZHHQG%DQGG%
7KHGHIDXOWLVG%5HƮHFWLYHHYHQWVPRUHLQWHQVH
than 40dB will be recorded in the event table.
(QGWKUHVKROG7KHHQGRIWKHƬEHUOLQNZLOOEH
determined by the end threshold setting. The range
FDQEHDGMXVWHGIURPWRG%ZLWKWKHGHIDXOWVHWWR
G%7KHƬUVWORVVHYHQW!G%ZLOOLQGLFDWHWKHHQG
RIWKHƬEHU
(OLJLELOLW\&ULWHULD
Setting the threshold determines the pass or fail
criteria for each event.
Fresnel events:5HƮHFWLYHHYHQWVW\SLFDORI
connectors.
Raleigh events:1RQUHƮHFWLYHHYHQWVW\SLFDORI
fusion splices.

OFL100
13
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
OTDR File Saving
$ƱHUWKHPHDVXUHPHQWSUHVV>VDYH@WRVDYHWKHƬOH
HQWHUWKHƬOHQDPHDQGSUHVV>HQWHU@WRVDYHWKHƬOH
<RXFDQDOVRSUHVV>)DVWVDYH@WRVDYHWKHƬOH7KHƬOHLV
saved in a folder named the same day’s date.
$XWRVDYHRSHQWKHDXWRVDYHIXQFWLRQWKHƬOHQDPH
will be automatically generated according to the rules;
)LOHQDPLQJPHWKRGRQO\YDOLGIRUqDXWRVDYHrDQG
qRQHFOLFNVDYHr
2SWLFDO)LEHU&RGH,'7KHRSWLFDOƬEHUQXPEHUDQG
code set when the line is installed:
Location A: /LQNVWDUWSRLQWORFDWLRQ
Location B: /LQNWHUPLQDWLRQSRLQWORFDWLRQ
Direction: 2SWLFDOƬEHUWHVWGLUHFWLRQIURP$WR%
from B to A;
Operator: Enter the name of the technician.
ƬOHQDPHƬEHUQXPEHUQDPLQJƬEHU
number
ƬOHQDPHZDYHOHQJWKƬEHUQXPEHU
QDPLQJƬEHUQXPEHULQFUHDVLQJLQRUGHU
ƬOHQDPHZDYHOHQJWKSXOVHZLGWK
ƬEHUQXPEHUDQGWKHƬEHUQXPEHU
increases in order.
)LOHQDPHHQWHUWKHƬOHQDPHPDQXDOO\
1
4
+
1
2 4
++
3
1
2 4
+++
File Operation
File Operation
All the test curves are saved in the standard SD
FDUGRIWKHLQVWUXPHQW3UHVV>)LOH@WRHQWHUWKH
ƬOHRSHUDWLRQLQWHUIDFH<RXFDQRSHQGHOHWHDQG
UHQDPHƬOHV

OFL100
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
14
VFL
7KHYLVXDOIDXOWORFDWRULVXVHGWRLQMHFWUHGODVHUOLJKW
LQWRWKHƬEHUXQGHUWHVWWRDOORZWKHWHFKQLFLDQWR
YLVXDOO\ORFDWHEUHDNVRU0DFUREHQGVLQƬEHURSWLF
cables. The VFL can also locate contaminated and
damaged connectors.
Open: Turns the visual fault locator on in continuous
wave mode.
1Hz: 9LVXDOIDXOWORFDWRUƮDVKHVDWDUDWHRI+]
2Hz: 9LVXDOIDXOWORFDWRUƮDVKHVDWDUDWHRI+]
Close: 7XUQRƪYLVXDOIDXOWORFDWRU
WARNING
Avoid looking directly at the laser output port;
permanent eye damage may occur
OPM
The OPM can be used to measure absolute and
relative power. The OPM can identify and measure
WRQHZDYHOHQJWKVRI+]+]N+]DQGN+]
emitted from a compatible laser.
Wavelength: Set the desired wavelength.
Reference: Set the current power as the reference.
The Relative power will be set to 0dB where loss
measurements can be made.
Calibration: Enter the calibration mode.
7KUHVKROGSet the threshold value of power
measurement. If it exceeds the threshold value, it will
be displayed in red; if it is lower than the threshold
value, it will be displayed in green.
(YHQW0DS
6WDUWRIƬEHU
Raleigh loss event such as a fusion
Connector
Splitter
End of link
5LVLQJHYHQWFDXVHGE\WKH
inconsistency of refractive index of two
VHFWLRQVRIƬEHU

OFL100
15
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
Optical Loss Test
The optical loss test allows the technician to measure
the insertion loss of passive optical components.
To measure insertion loss:
6HWWKH230DQG6/6WRWKHGHVLUHGZDYHOHQJWK
&RQQHFWWKH6/627'5SRUWWRWKH230SRUW
3UHVV5HIHUHQFH7KH5HO3RZVKRXOGEHG%
,QVHUWGHYLFHWREHPHDVXUHG
5HDGWKHORVVRIWKHGHYLFHXQGHUWHVWXQGHU5HO3RZ
Laser Source
7KH2)/SURYLGHVD&:VWDELOL]HGOLJKWVRXUFH
that is 1310nm or 1550nm. These can be used with a
FRPSDWLEOH230WRPHDVXUHLQVHUWLRQORVVRIƬEHURSWLF
OLQNV$WRQHFDQEHLQVHUWHGDW+]+]N+]DQG
N+]WRDLGLQƬEHULGHQWLƬFDWLRQ
Open: Turns the laser source on.
Wavelength: Allows the technician to set the
wavelength of laser source.
0RGH6HWWKHODVHUIUHTXHQFH\EHWZHHQ&:+]
+]N+]DQGN+]
Power + : Increase the output power.
Power - : Reduce the output power.
The slider bar at the bottom of the screen can be used
WRDGMXVWWKHRXWSXWSRZHU
WARNING
Avoid looking directly at the laser output port;
permanent eye damage may occur
RJ45 Sequence Testing
Connect the cable under test to LAN2 port and the
RWKHUHQGWRWKHUHPRWHVRFNHWRQWKHERWWRPRIWKH
instrument. Straight connect and Interleaved type
cables can be tested.
Direct Cable Test: The indicators of the host and
UHPRWHZLOOƮDVKRQHDWDWLPHIURPWR
,QWHUOHDYHG&DEOH7HVWThe indicators of the host and
WKHUHPRWHZLOOƮDVKRQHDWDWLPHLQWKLVRUGHU
4,5,2,7,8.
Both T568A and T568B cables can be tested.
Do not connect live cables to the OFL100. Doing so may
result in permanent damage.

OFL100
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
16
System Settings
$XWRPDWLFVKXWGRZQThe technician can select the
GHVLUHGWLPHWRDXWRPDWLFDOO\WXUQWKH2)/RƪDƱHU
a period of inactivity equal to 5/15/30/45/60 minutes or
to never shut down.
Backlight brightness: 7KHWHFKQLFLDQFDQDGMXVWWKH
brightness to 20%/40%/60%/80% or 100%.
Beeper: The technician can enable or disable the beeper
IRUWRXFKFRQƬUPDWLRQ
Language: Allows the technician to select the desired
language.
USB connection: Allows the technician to connect and
WUDQVIHUGDWDYLDWKH86%SRUW
7LPHGDWHAllows the technician to set the time
and date.
Restore factory settings: Allows the technician to
UHVWRUHWKH2)/EDFNWRIDFWRU\GHIDXOW
8SJUDGH Provides the technician the utility to update
WKHƬUPZDUHDSSOLFDWLRQ

OFL100
17
Tempo Communications Inc. 1390 Aspen Way • Vista, CA 92081 USA • 800-642-2155
Fault Description Cause of Failure Solution
OTDR will not turn on. The battery is dead. Charge the battery and observe the charging
indicator. If the red light is displayed, continue
charging.
OTDR cannot be charged normally. Ambient temperature not within acceptable
levels.
Replace charger. Try charging the OFL100 with the
USB port.
Check that the charger is working. 7IXVERKIXSKVIEXIVXLERXLIPIRKXLSJXLIƼFIV
OTDR does not measure and display
test results.
OTDR parameters are not set correctly. 7IXVERKIXSKVIEXIVXLERXLIPIRKXLSJXLIƼFIV
OTDR bulkhead is contaminated. 'PIER38(6FYPOLIEHERHƼIPHƼFIVJIVVYPI
OTDR bulkhead is damaged. Send to Vista for bulkhead repair.
Optical output connector mismatch. 9WIGSVVIGXƼFIVWM^IERHGSRRIGXSVXSQEXGL
bulkhead.
The noise of test curve is excessive,
and the waveform is not smooth.
8LIƼIPHƼFIVMWRSXGSRRIGXIHTVSTIVP]XS
the bulkhead.
'PIERERHVIGSRRIGXXLIƼIPHƼFIVXSXLIFYPO-
head.
The pulse width setting is too small. Increase the test pulse width.
Averaging time is set too low. Increase the averaging time.
Events are saturated. The pulse width is too long. Decrease test pulse width.
8LIPEWXGSRRIGXSVMWSTIRERHPMOIP]EƽEX
polished connector.
Terminate the last connector or place a Mandrel
[VETSRXLIƼFIVFIJSVIXLIPEWXGSRRIGXSV
%JXIVEVIƽIGXMZIIZIRXXLIVIMWEPSRK
tail.
OTDR bulkhead is contaminated. 'PIER38(6FYPOLIEHERHƼIPHƼFIVJIVVYPI
OTDR bulkhead is damaged. Send to Vista for bulkhead repair.
8LIVIƽIGXMSRTIEOEXXLIIRHSJXLI
ƼFIVGERRSXFIQIEWYVIH
OTDR parameters are not set correctly. 7IXVERKIXSKVIEXIVXLERXLIPIRKXLSJXLIƼFIV
The pulse width is too short. Increase the test pulse width.
False positive in curve analysis. Event threshold setting is too low. Increase the threshold analysis setting.
6HFWLRQOFL100 Troubleshooting
7UDFH9LHZHUVRƱZDUHLVDQDSSOLFDWLRQGHYHORSHGIRUWKH2)/27'5,WDOORZVWKHSUHYLRXVO\VWRUHG625
PHDVXUHPHQWUHFRUGVLQWKHLQVWUXPHQWWREHXSORDGHGWRD3&YLDWKH86%SRUW
The measurements can then be displayed, saved, or printed as a PDF for close out reporting.
The Trace Viewer Documents:
- Fiber trace
- Test conditions/settings
- Fiber information
- File name/date
All events are listed with test results documented in tabular form.
)HDWXUHVRIDWUDFHFDQEH]RRPHGLQRQDQGFXUVRUVPDQLSXODWHGIRUFORVHUH[DPLQDWLRQ
The Tempo Communications Trace Viewer is available for download at www.tempocom.com.
6HFWLRQTrace Viewer

Distribution in the UK & Ireland
Lambda Photometrics Limited
Lambda House Batford Mill
Harpenden Herts AL5 5BZ
United Kingdom
W: www.lambdaphoto.co.uk
T: +44 (0)1582 764334
F: +44 (0)1582 712084
Other manuals for OFL100
2
Table of contents
Other Tempo Communications Measuring Instrument manuals

Tempo Communications
Tempo Communications CableScout CS90B User manual

Tempo Communications
Tempo Communications TV220E Reference guide

Tempo Communications
Tempo Communications OPM510 User manual

Tempo Communications
Tempo Communications OFL100 User manual

Tempo Communications
Tempo Communications CmA-360 TRMS User manual

Tempo Communications
Tempo Communications OPM5 Series Reference guide

Tempo Communications
Tempo Communications OFL100 User manual

Tempo Communications
Tempo Communications BLL-200 User manual

Tempo Communications
Tempo Communications Marker-Mate EML100 User manual

Tempo Communications
Tempo Communications OPM510 Reference guide