Testomat 2000 THCL User manual

Testomat 2000
®
THCL
Combined process photometer for
total chlorine 0 – 2.5 ppm
and process titrator for water hard-
ness 4.47 – 44.7 ppm CaCO
3
Operating Instructions

Contents
2
Contents
Contents...............................................................................................2
Important safetyinformation ..............................................................4
Intended use .........................................................................................4
Qualification of the staff ........................................................................4
Warning notices in these instructions ...................................................5
Further documents................................................................................5
Pay particular attention to .....................................................................5
General instructions ..............................................................................5
Installation .............................................................................................6
Operation ..............................................................................................6
Cleaning ................................................................................................6
De-installation .......................................................................................6
Disposal ................................................................................................6
Scope of delivery ................................................................................7
Performance specifications ...............................................................7
Indicators/Reagents for Testomat 2000
®
-THCL ...................................8
Application instructions.....................................................................9
Installation .........................................................................................10
Operating Testomat 2000
®
THCL in the pressure range 0.3 to 1 bar 10
Installing Testomat 2000
®
THCL .........................................................10
Connecting the water inlet and outlet .................................................11
Water inlet ...........................................................................................11
Water outlet.........................................................................................11
Connecting the power supply and devices .........................................12
Block diagram Testomat 2000
®
THCL ................................................12
Internal design Testomat 2000
®
THCL ...............................................13
Connecting the mains voltage.............................................................14
Connecting the plant components ......................................................15
Connecting the inputs and outputs .....................................................16
Commissioning .................................................................................17
Inserting indicator bottles ....................................................................17
Extracting indicators............................................................................17
Opening the water inlet .......................................................................18
Instrument settings and data input......................................................18
Functions of the operating and display elements.........................18
Switching Testomat 2000
®
THCL on/off..............................................18
Display functions .................................................................................19
Operating elements and function keys ...............................................20
Operating system................................................................................21
Password protection and basic program .......................................22
Entering basic program data...............................................................22
Selecting the display units ..................................................................22
Selecting the operating mode .............................................................23
Entering further basic program data ..............................................25
Internal flushing...................................................................................25
External flushing .................................................................................25
Interval pause .....................................................................................26
Limit value monitoring .........................................................................26
Hysteresis ...........................................................................................27

Contents
3
Switch functions of the limit value outputs LV1 and LV2 ....................28
Switch function 0, duration..................................................................28
Switch function 1, impulse ..................................................................28
Switch function 2, interval ...................................................................28
Function IN1........................................................................................28
Water meter ........................................................................................29
Alarm / Message .................................................................................29
Function AUX ......................................................................................30
Service II .............................................................................................30
Reset operating time...........................................................................30
Maintenance interval...........................................................................31
Description of the signal inputs/outputs........................................32
Interfaces (optional) ............................................................................34
Monitoring the measuring points.........................................................34
Calculating the output currents ...........................................................35
Serial interface RS232 ........................................................................35
Description of the relay outputs......................................................36
Flushing (external flush valve) ............................................................36
LV1 and LV2 limit value outputs .........................................................36
Measuring points 1 or 2 (measuring point switch-over)......................37
AUX (programmable function output) .................................................37
Alarm (fault message output)..............................................................38
Maintenance (output for maintenance message) ...............................38
Information menu "i" ........................................................................39
Request options: Customer service, operating values, program
values, error history, maintenance......................................................39
Program menu "M" ...........................................................................40
Programming of: Indicator, manual mode, flushing, flush chamber,
drain chamber, fill chamber, self-test, confirm maintenance,
diagnosis, date, time, basic program with password ..........................40
Structure of the basic program............................................................42
Error messages / Troubleshooting .................................................43
Further information..............................................................................44
Maintenance ......................................................................................45
Description of maintenance work........................................................45
Service instructions .............................................................................46
Testomat 2000
®
spare parts and accessories................................47
Accessories.........................................................................................48
Technical data ...................................................................................49
Product overview Testomat 2000
®
instruments ............................50

Important safetyinformation
4
Important safetyinformation
Please read these operating instructions carefully and completely
prior to working with the instrument.
Ensure that these operating instructions are always available for all
users.
These operating instructions must always be passed on to the new
owner should Testomat 2000
®
THCL change hands.
Always adhere to hazard warnings and safety tips when using
reagents, chemicals and cleaning agents. Please adhere to the re-
spective safety data sheet! Download the safety data sheets for
the supplied reagents at http://www.heyl.de .
Intended use
Testomat 2000
®
THCL is used for the automatic determination and
monitoring of residual total hardness (water hardness from 0.25 to
2.5 °dH) and the determination of total chlorine content from 0 to 2.5
mg/ltr (ppm) in water. The feed water must be clear, colourless and
free of undissolved particles.
Always adhere to the performance limits stated in the section enti-
tled “Technical data”.
Always observe the application areas/application limits of the indi-
cators and the requirements of the medium being measured.
To ensure correct and intended usage, always read and understand
these instructions, especially the section entitled “Important safety
information”, prior to use.
The instrument is not used as intended if
• it is used in areas not specified in these instructions.
• it is used in areas which do not correspond to the ones described
in these instructions.
Qualification of the staff
Assembly and commissioning require fundamental electrical and
process engineering knowledge as well as knowledge of the respec-
tive technical terms. Assembly and commissioning should therefore
only be carried out by a specialist or by an authorised individual su-
pervised by a specialist.
A specialist is someone who due to his/her technical training, know-
how and experience as well as knowledge of relevant regulations can
assess assigned tasks, recognise potential hazards and ensure ap-
propriate safety measures. A specialist should always adhere to the
relevant technical regulations.

Important safetyinformation
5
Warning notices in these instructions
The warning notices in these instructions warn the user about poten-
tial dangers to individuals and property resulting from incorrect han-
dling of the instrument. The warning notices are structured as follows:
Description of the type or source of danger
Description of the consequences resulting from non-observance
Preventive measures. Always adhere to these preventive meas-
ures.
“DANGER” indicates an immediate hazardous situation which, if not
avoided, will result in death or serious injury.
“WARNING” indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
“CAUTION” indicates a potentially hazardous situation which, if not
avoided, could result in minor or moderate injuries or property dam-
age.
“NOTE” indicates important information. If this information is not ob-
served, it may result in an undesirable result or state.
Further documents
Testomat 2000
®
THCL is a plant component. Therefore, always ob-
serve the maintenance manual of Testomat 2000
®
/ ECO
®
and the
documentation of the plant manufacturer.
Pay particular attention to
General instructions
• Please adhere to health and safety regulations, electrical equip-
ment safety regulations, and environmental protection regulations
valid in the country of use and at the installation site.
• Adhere to national and local regulations during installation and
commissioning.
• Always protect the instrument against moisture and humidity. It
should never come into contact with condensation or splash water.
• Do not carry out any changes or actions at the instrument which
are not described in these instructions
;
failure to adhere to these
instructions will negatively affect any warranty claims that you
make thereafter.
SIGNAL WORD
WARNING
!

Important safetyinformation
6
Installation
• Always completely disconnect the relevant plant part before install-
ing the instrument or connecting/disconnecting it to/from the power
supply. Secure the plant against reconnection.
• Only connect the instrument to the mains voltage specified on the
rating plate.
• Always observe technical data and ambient parameters.
• Testomat 2000
®
THCL requires an interference free and stable
power supply. If necessary, use a mains filter to protect Testomat
2000
®
THCL against interference voltages caused, e.g., by sole-
noid valves or large motors. Never lay connecting cables parallel to
power cables.
Operation
• Ensure that the maximum electrical load capacity of the relay out-
puts is never exceeded.
• Immediately switch off Testomat 2000
®
THCL and contact service
staff if malfunctioning occurs. The warranty will be void if you tam-
per with or attempt to repair Testomat 2000
®
THCL. Repairs must
be carried out by authorised service staff.
Cleaning
• Only use a dry, lint-free cloth for cleaning.
De-installation
• Prior to de-installing a defective instrument, always write down a
description of the error (failure effect). It is only possible to repair a
defective instrument (irrespective of the warranty period) if it has
been de-installed and returned to us with a description of the error.
Disposal
• Dispose of the instrument in accordance with national regulations.
WARNING
!
HINWEIS

Scope of delivery
7
Scope of delivery
1 Testomat 2000
®
THCL
1 plastic bag with screw caps with a hole and an insert for the
screw caps of the indicator bottles
1 operating instructions
Performance specifications
Testomat 2000
®
THCL is used for the automatic determination and
monitoring of residual total hardness (water hardness from 0.25 to
2.5 °dH) via titration and the determination of total chlorine content
from 0 to 2.5 mg/ltr (ppm) in water. The analysis of total chlorine is
carried out by adding three reagents; the analysis result is displayed
after a reaction time of approx. 1 minute (without flushing times).
• Simple, menu-driven operating and programming via a plain
text display
• Measurement via titration of residual hardness and
photometric determination of the total chlorine content
• Freely selectable hardness units in °dH, °f, ppm CaCO3 or
mmol/l and chlorine units in mg/l or ppm
• High measuring accuracy provided by a precise piston-
dosing pump
• Analysis initiation:
- Automatic interval operation
(interval pause can be set from 0-99 minutes)
- External control
- Quantity dependent via impulse water meter
• One limit value with hysteresis (1, 2 or 3 bad analyses) for
residual hardness (LV1) and total chlorine (LV2) and settable
switch functions
• Monitoring of both measuring points (change-over via exter-
nal solenoid valves)
• Internal error documentation
• Programmable service address
• Programmable maintenance interval for a maintenance re-
quest
• Extended operating periods due to 500 ml indicator storage
bottle
• optional
Interface card (0/4-20 mA or 0/2-10 V)
Interface RS 232 (for a protocol printer)

Performance specifications
8
Indicators/Reagents for Testomat 2000
®
-THCL
Water hardness
Indicator type TH 2025
°dH
(Resolution)
0.25 – 2.50
(0.05)
°f
(Resolution)
0.45 – 4.48
(0.1)
ppm
CaCO
3
(Resolution)
4.47 – 44.7
(0.9)
Unit
mmol/l
(Resolution)
0.04 – 0.45
(0.01)
Total chlorine
Reagent CL2250A, CL2250B, CL2250C
Unit
mg/l (ppm)
(Resolution)
0.0 – 2.5
(0.1)

Application instructions
9
Application instructions
• Wait at least 5 seconds before switching the instrument on and
then off again at the main switch.
• In order for Testomat 2000
®
THCL to operate reliably, use Heyl
Testomat 2000
®
indicators in the pH-range 4 – 10.5!
• With Testomat
®
instruments for water hardness monitoring, larger
quantities of heavy metal ions in the softened water might influ-
ence the colour reaction, especially iron above 0.5 mg/I, copper
above 0.1 mg/I and aluminium above 0.1 mg/l (brownish-red col-
our display).
• If the measuring water contains more than 20 mg/I CO
2
(carbonic
acid), incorrect evaluations cannot be excluded.
• The concentration of influencing contents can be determined by
using our colourimetric TESTOVAL
®
test kit.
• Careful handling of the instrument increases both its operational
reliability and service life! Therefore, carry out a visual inspection
at regular intervals as described below.
- Has the use-by date of the indicator expired?
- Are the hose connections of the dosing pump free of leaks?
- Is there air inside the dosing hoses?
- Are all the water connections free of leaks?
- Are the doors of the instrument closed properly?
- Is the instrument heavily soiled?
- Are the measuring chamber and the drain duct/drain hose
clean?
• Trouble-free operation is only possible when maintenance is car-
ried out on a regular basis (For maintenance and service instruc-
tions, please refer to the section entitled “Maintenance” and the
"Maintenance manual of Testomat 2000
®
/ ECO
®
")
• If problems occur, please refer to the section entitled "Error mes-
sages/Troubleshooting.

Installation
10
Installation
Risks resulting from incorrect installation!
Install Testomat 2000
®
THCL at a location where it is protected
against dripping or splash water, dust and aggressive substances
– e. g. in a switch cabinet or on a suitable wall.
Information for trouble-free operation
Install Testomat 2000
®
THCL vertically and without mechanical
stress.
Install Testomat 2000
®
THCL at a vibration-free site.
Operating Testomat 2000
®
THCL in the pres-
sure range 0.3 to 1 bar
Prior to installation, please check whether lower operating pressure is
required. The instrument is factory set for the operating range 1 to 8
bar. Remove the flow controller valve body j
jj
jto operate the instru-
ment in the operating range 0.3 to 1 bar (e.g. when using an aerator
type R). This involves removing the retaining pin i
ii
ifrom the
controller / filter receiver g
gg
g. Subsequently use the metal bracket to
remove the controller plug k
kk
kfrom the borehole. Subsequently
remove the flow controller valve body j
jj
jand reinsert the controller
plug and the retaining pin.
Installing Testomat 2000
®
THCL
Select an installation site where the water inlet hose can be kept as
short as possible (max. 5 m).
Please leave sufficient space on the left-hand side of the instru-
ment to open the door.
Drill the mounting holes as shown in the drawing on the left.
Use three screws to attach the instrument at a suitable position in
the switch cabinet or on a wall.
Remove for
pressure
range 0.3 to
1 bar
g
gg
g
i
ii
i
j
jj
j
k
kk
k
WARNUNG
!
NOTE

Installation
11
Connecting the water inlet and outlet
Information for trouble-free operation
The water pressure must be between 0.3 bar and 8 bar
Avoid strong pressure fluctuations
The measuring water temperature must be between 10 °C and 40
°C
For temperatures above 40°C, the KCN type cooler should be in-
stalled in the branch line of Testomat 2000
®
THCL.
Water inlet
The measuring water is taken from the main water line of the water
treatment plant and fed to the inlet connection of Testomat 2000
®
THCL. The instrument is equipped with a plug connector for plastic
hoses 6/4 x 1 (external diameter 6 mm/ internal diameter 4 mm, wall
thickness 1 mm) as standard.
Install the connection for the branch line of Testomat 2000
®
THCL
directly at the main water line
①
directly after the water treatment
plant
It is important that the branch line connection is laid vertically up-
wards in order to prevent dirt particles from entering the instru-
ment from the main water line
.
Install a manually operated shut-off valve
②
in the branch to Tes-
tomat 2000
®
THCL.
Use an opaque plastic hose 6/4 x 1 (max. length 5 m) for the water
inlet
③.
Flush the inlet to remove any dirt particles
When operating within a pressure range of 0.3 to 1 bar, please re-
move the valve body from the controller and the filter housing.
When using a cooler
The hot water can cause burns and damage wetted parts of
Testomat 2000
®
THCL.
Water outlet
The feed water flows through the measuring chamber to the drain via
the outlet hose.
Connect the outlet connection of Testomat 2000
®
THCL to an
opaque outlet hose
④
(internal diameter 14 mm).
Lay this hose without backwater development and any syphoning
effect, e.g. via an open funnel, to the drain.
⑤
①
②
③
⑤
④
⑤
NO !!
"Sagging"
causes back-
water!
NOTE
CAUTION
!

Installation
12
Connecting the power supply and devices
Risk of electric shocks during installation!
If the power supply is not disconnected prior to installation, it may
result in personal injuries, destruction of the product or damage to
plant parts.
Always disconnect the relevant plant parts before installing Tes-
tomat 2000
®
THCL.
Only use tested cables with sufficient cross-sections for the con-
nections.
Risk of damages caused by electromagnetic fields!
If Testomat 2000
®
THCL or the connecting cables are installed
parallel to power cables or in close proximity to electromagnetic
fields, the instrument may be damaged or measurements incor-
rect.
Ensure that connecting cables are as short as possible.
Always install connecting cables and power cables separately.
Connect the instrument to the protective earth conductor (for
230/115 VAC).
Protect Testomat 2000
®
THCL against interference voltages – e.g.
via a mains filter.
Shield the instrument against strong electromagnetic fields.
Block diagram Testomat 2000
®
THCL
Drawn relay positions: Instrument de-energised
WARNING
!
NOTE

Installation
13
Internal design Testomat 2000
®
THCL
①
Terminal block for inputs Start, Stop, IN1, IN2, and output OUT
②
Mains switch
③
Terminal block for mains inputs and mains outputs
④
Terminal block relay outputs
⑤
Dosing pumps (DosiClip1=left, DosiClip4=right))
⑥
Water connections, inlet and outlet
⑦
Controller / Filter receiver
⑧
Measuring chamber
①
①①
①
②
②②
②
③
③③
③
④
④④
④
⑤
⑤⑤
⑤
⑧
⑧⑧
⑧
⑦
⑦⑦
⑦
⑥
⑥⑥
⑥

Installation
14
Connecting the mains voltage
Only connect the instrument to the specified mains voltage. Refer to
the rating plate for the appropriate mains voltage. Connect the cables
as follows:
Loosen both fastening screws
①
and open the upper door. The
terminal box is now accessible.
Pierce the required rubber cable glands
②
with a screwdriver and
insert the cable through the bush into the terminal box (1)
Subsequently pull back the cable until the bush has been turned
over (2).
Connect the power supply to terminals PE,N,L or for 24 V instru-
ments to terminals U, V.
Connect the conductor to the terminal block as shown on the left
➂
.
Ensure that the leads are held securely in the terminals.
Proceed as shown in figure
➃
to loosen the connection.
Terminal
description
Typ
e Function Comment
PE IN Mains – protective earth
(5x)
Only with mains
115/230 V !
N
(U)
L
(V)
IN Mains, N=neutral (U=24 V)
Mains, L=live (V=24 V)
Mains input
24 V / 115 V / 230 V
n
lOUT
Neutral, switched (8x)
Live, switched (8x)
Mains for consumers,
max. 4 A
Insert the conductor with
ferrule or the solid conductor
into the round input.
➂
①
②
①
1. Insert a screwdriver into the
square opening without force in
order to open the terminal.
2. Once the terminal has been
opened, remove the conductor.
➃

Installation
15
Connecting the plant components
Connect the plant components to the output terminals of relays 1
to 19 (e.g. valves).
If the plant components require mains voltage, connect the
switched mains voltage (l) to the common contact
①
of the re-
spective relay (see the connection example for 230 VAC on the
left).
Connect the neutral conductor of the plant component to one of the
terminals (n).
For components with a protective earth conductor connection,
connect it to the PE connection.
Ensure that the leads are held securely in the terminals.
No.
Terminal
description
Type
Function Comment
1
2 flushing OUT
External flush valve Volt-free relay output,
max. 240 VAC, 4 A
3
4
5
LV1 OUT
Limit value output 1 – Normally closed
Limit value output 1 – Normally open
Limit value output 1 – Common
Volt-free relay output,
max. 240 VAC, 4 A
6
7
8
LV2 OUT
Limit value output 2 – Normally closed
Limit value output 2 – Normally open
Limit value output 2 – Common
Volt-free relay output,
max. 240 VAC, 4 A
9
10
11
Measur-
ing points
1/2
OUT
Measuring point 1 – Normally closed
Measuring point 2 – Normally open
Measuring point switch-over - Common
Volt-free relay output,
max. 240 VAC, 4 A
12
13 AUX OUT
Universal output Volt-free relay output,
max. 240 VAC, 4 A
14
15
16
Alarm OUT
Fault message output – Normally closed
Fault message output – Normally open
Fault message output – Common
Volt-free relay output,
max. 240 VAC, 4 A
17
18
19
Mainte-
nance OUT
Maintenance message – Normally
closed
Maintenance message – Normally open
Maintenance message – Common
Volt-free relay output,
max. 240 VAC, 4 A
Connection example:
Limit value contact LV 1
switches mains voltage
n
l
3
4
5
e.g.
solenoid
valve
PE
①

Installation
16
Connecting the inputs and outputs
Testomat 2000
®
has the following connections for control and moni-
toring functions.
Do not connect external voltage to these connections!
Ensure that the leads are held securely in the terminals.
Use the two fastening screws to close the upper door once installa-
tion has been completed.
For more information, please refer to the section entitled "Description
of the signal inputs/outputs".
No.
Terminal
description
Typ
e Function Comment
20
21 Start IN External analysis start
Common earth for inputs
Only for volt-free normally
open
22
23 Stop IN External analysis stop
Common earth for inputs
Only for volt-free normally
closed/normally open
24
25 IN1 IN
Universal input
1
Common earth for inputs
Only for volt-free normally
closed/normally open
26
27 IN2 IN Universal input 2 (water meter)
Common earth for inputs
Only for volt-free normally
open
┴
28
29
OUT OUT
Earth
0/4 - 20 mA galvanically separated
or serial interface RS232
Earth = ┴
28 = (+) or (TxD)
29 = (-) or (RxD)

Commissioning
17
Commissioning
Handling of reagents/indicators
Please adhere to the respective safety data sheets!
Trouble-free operation of Testomat 2000
®
THCL is only possible
when using Heyl Testomat 2000
®
indicators/reagents!
Inserting indicator bottles
Open the lower housing door by pulling on the right-hand side.
Remove the caps from the indicator bottles.
Remove the plastic bag from inside the housing. The plastic bag
contains the screw caps with hole
①
and the inserts
②
for the
screw caps.
Connect the parts as shown on the left.
Obeserve correction allocation of the indicators to the pumps:
TH2025=DosiClip1 (left), CL2025A=DosiClip2, CL2025B= Dosi-
Clip3 and CL2025C=DosiClip4 (right)
Screw the hose connectors
③
of the intake hoses
④
hand-tight
into the inserts
②.
Place the inserts with the screwed-in intake hoses into the indica-
tor bottles.
Now screw the screw caps with hole
①
hand-tight onto the indica-
tor bottles.
⑤
Switch the instrument on and press the "STANDBY" key
.
Enter the filling levels of all the indicator bottles in "%":
In the menu, select
>SERVICE I
=>
INPUT REAGENTS
=>
INDICATOR TH FILLING (100%)
(New/Full bottle = 100% filling level)
Press "ENTER" to confirm the entry
Enter further filling levels accordingly
Extracting indicators
Switch the instrument on and press the "STANDBY" key
During operation, the pumps (DOSIClip1 - 4)
⑥
automatically ex-
tracts the indicator.
To ensure that indicator is available for the initial analyses, the
intake hose
④
and the transport hose
⑦
must be filled with indi-
cator from the pump up to the measuring chamber.
Press the "manual"
⑧
key several times until the intake hose
④
and the transport hose
⑦
are filled with indicator up to the meas-
uring chamber
If necessary, manually tighten the hose connectors of the intake
and transport hose slightly in case of bubble formation
①
②
③
④
⑤
⑥
⑦
⑧
CAUTION
!

Functions of the operating and display elements
18
①
②
Opening the water inlet
Open the lower housing cover.
Slowly open the manually operated shut-off valve to prevent the
measuring chamber overflowing. The flow regulator requires a few
seconds to function correctly.
Make sure that the water conducting parts are not leaky.
If water sprays from the vent hole
①
of the measuring chamber
②
, reduce the amount of inlet water via the manually operated
shut-off valve. It should take 2 to 6 seconds to fill the measuring
chamber!
Instrument settings and data input
Please read the following information before carrying out settings
and entering data for operating the instrument.
Functions of the operating and dis-
play elements
The Testomat 2000
®
THCL display shows operating statuses and
measured values. The input keys for programming (cursor block) and
the function keys are located underneath the display.
Switching Testomat 2000
®
THCL on/off
(1) Mains switch
Use this switch to switch the instrument on or off.
(2) Instrument fuse (inside the instrument)
This fuse protects Testomat 2000
®
THCL and the outputs
against overloads and short circuits.
(1)
(2)

Functions of the operating and display elements
19
Display functions
1 Status of limit value displays (red/green)
The display 1 illuminates red if limit value 1 (water hardness) has
been reached or exceeded. The display 1 illuminates green if the
value falls below the limit value. The same principle applies to limit
value 2 (total chlorine) and display 2.
2 Text display (4 lines)
Displays the current analysis result as well as all important
statuses and programming data.
2a = The current measured value for measuring point 1 (water
hardness) is displayed in Line 1.
Value falls below the measuring range = "<" e.g. < 4.47 ppm
CaCO
3
Value exceeds the measuring range = ">" e.g. > 44.7 ppm
CaCO
3
The current measured value for measuring point 2 (total chlorine)
is displayed in line 2.
Value exceeds the measuring range = ">" e.g. > 2.5 ppm
2b = The set limit values GW1 (water hardness) and GW2
(total chlorine) are displayed in line 3 and 4
3Alarm (red)
Indicates malfunctioning/error message or warning message
4 Maintenance message (yellow)
Indicates current maintenance requests
5 Status display of the active instrument components (line)
Six displays indicate the current instrument and analysis status
Cancelling error messages/warning messages
Press to acknowledge the message and, if necessary,
eliminate the cause of the fault.
Inlet
valve
Anal
y-
sis
Outlet
valve
Dosing
message
Without
function!
Analysis
stop
2
a
2b
All error and
warning messa
ges
are alternately
shown in line 1 of
the standard
display!
NOTE

Functions of the operating and display elements
20
Operating elements and function keys
If you wish to carry out settings or enter data, or if alterations are
necessary, press the "M" key to open the programming mode. Use
this key when in the menu to go to the higher order menu items or to
exit the programming mode. Please note: The basic program is
password protected!
Programming keys (cursor block)
Use the programming keys (cursor block) to navigate in the menu, to
select the desired functions and to enter necessary instrument and
plant specific data. Press the "ENTER" key to select the submenu
item and to confirm and accept the selection or data input.
Function keys
Manually start an analysis via the "Manual"
①
key
Set the instrument to standby mode via the "STANDBY"
②
key (automatic analyses are not carried out: analysis
stop)
Acknowledge error and warning messages via the "Horn"
③
key
Open the program menu for user-specific and instrument-
specific settings via the "M"
④
key
Retrieve all instrument information and settings via the "i"
key
⑤
(M)neu key
①
①①
①
②
②②
②
③
③③
③
④
④④
④
⑤
⑤⑤
⑤
Programming
keys
(cursor block)
THCL
Cursor block
Table of contents
Other Testomat Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Cole Parmer
Cole Parmer MB-200-40 instruction manual

Megger
Megger SVERKER 900 Application note

Endress+Hauser
Endress+Hauser Cerabar M PMC51? Cerabar M PMP51 Brief operating instructions

HollySys
HollySys HT5 Series Instructions for use

YEO-KAL
YEO-KAL 801C user manual

Bellingham and Stanley
Bellingham and Stanley ADP440 user guide

PCE Instruments
PCE Instruments PCE-TDS 100 H user manual

GRAPHTEC
GRAPHTEC midi LOGGER GL800 quick start guide

VOLTCRAFT
VOLTCRAFT VC-533 AC/DC operating instructions

NSK
NSK Bruel & Kjaer Vibro VCM-3 instructions

Kurth Electronic
Kurth Electronic KE401 IT quick start guide

Benning
Benning CM 9-2 operating manual