manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Thermo Scientific
  6. •
  7. Measuring Instrument
  8. •
  9. Thermo Scientific Orion Star User manual

Thermo Scientific Orion Star User manual

A
B
C
D
E
F
North America
166 Cummings Center
Beverly, MA 01915 USA
Toll Free: 1-800-225-1480
Tel: 1-978-232-6000
Dom. Fax: 1-978-232-6015
Int’l Fax: 978-232-6031
Europe
P.O. Box 254, 3860 AG Nijkerk
Wallerstraat 125K, 3862 BN
Nijkerk, The Netherlands
Tel: (31) 033-2463887
Fax: (31) 033-2460832
Asia Pacific
Blk 55, Ayer Rajah Crescent
#04-16/24, Singapore 139949
Tel: 65-6778-6876
Fax: 65-6773-0836
www.thermo.com/water
Environmental Instruments
Water Analysis Instruments
©2008 Thermo Fisher Scientific Inc. All rights reserved. RDO is a registered trademark of In-Situ Inc.,
Fort Collins, CO USA. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
RDO
Important Note – Read Before Assembling or
Operating the Equipment
The RDO optical DO sensor has an internal clock that counts down the 365 day
lifespan of a new optical cap. The countdown begins when the RDO cap is
installed on the RDO sensor, the RDO sensor is connected to the meter and the first
measurement is taken. The 365 day countdown cannot be reset or changed once
the first measurement is taken.
To View the Remaining Lifespan of the RDO Cap
In the measurement mode, press the setup key. Press the up arrow key until dO
is displayed on the top line. Press the line select key to move the icon to the
middle line. Press the up arrow key until LIFE is displayed on the middle line.
The remaining lifespan of the RDO cap, in days, will be displayed on the bottom
line. Press the measure key to return to the measurement mode.
Preparation
1. Power adapter (benchtop meters only) – Select the appropriate wall outlet plug
and slide the plug plate into the groove on the back of the adapter. See A.
Batteries – Select four AA alkaline batteries. Confirm that the meter is off and
remove the battery cover. Orient and insert the batteries as depicted in the
battery compartment housing. Replace the cover. See B and C.
2. Prepare the RDO optical DO sensor according to the directions in the RDO
sensor user guide. Install the RDO cap on the RDO sensor before connecting
the RDO sensor to the meter.
Note: An Err 880 message on the meter indicates that the RDO sensor is
not fully connected to the meter. Reconnect the sensor by firmly pushing the
sensor connector into the 9 pin miniDIN meter input.
Note: An Err 882 message on the meter indicates that the RDO cap is missing
or not properly installed on the RDO sensor.
3. Calibration sleeve – To prepare the calibration sleeve, remove the cap from the
sleeve and remove the sponge from the cap. Saturate the sponge with distilled
water and squeeze all of the excess water out of the sponge. Reassemble the
calibration sleeve and insert the RDO sensor.
4. Meter connections – Connect the power adapter to the meter and to the wall
outlet (benchtop meters only). See D. Connect the RDO sensor to the 9 pin
miniDIN input on the meter. See E. Connect the stirrer sensor to the stir jack
input on the meter (benchtop meters only). See F.
Note: All unused inputs on the meter should be covered with the black caps.
Key Description
power up arrow line select stir
(benchtop
meters only)
calibrate down arrow setup
measure digit view log
Meter Connections
All connectors on the 5-Star Plus benchtop meter are depicted above. All connectors on the 5-Star
Plus portable meter are highlighted in gray.
The 3-Star Plus and 4-Star Plus meters will have fewer connectors. For example, the 3-Star Plus
portable RDO meter will only have 9 pin miniDIN, 8 pin miniDIN (for ATC probe only) and
RS232 connectors.
For the complete Orion Star™and Star Plus meter user guide, visit www.thermo.com/water.
Thermo Scientific Orion Star™
RDO®Optical Dissolved Oxygen Meter
Quick Start Guide
English
257251-001 Rev. A 09-08
Stir Jack –
Stirrer Sensor
Pin Tip –
Reference
BNC – pH,
ORP or ISE
8 Pin MiniDIN –
ATC or Cond
9 Pin MiniDIN
– RDO
RS232
Ground
Power
1
3
5
2
4
6
RDO
Meter Overview
1. To power on the meter, press the power key.
2. Press the line select key to choose the top, middle
or bottom display line. The icon will point to the
selected line.
3. In the measurement mode, press the up arrow or
down arrow key to change the measurement units
on the bottom display line to dissolved oxygen (%
saturation), dissolved oxygen (mg/L), barometric
pressure (no icon), temperature (no icon) or a
blank line.
4. To escape out of any meter function, press and hold
the measure key until the meter returns to the
measurement mode.
5. The 3-Star, 4-Star and 5-Star RDO optical dissolved
oxygen meters can perform an air calibration using
water saturated air, water calibration using air
saturated water, manual calibration (Winkler titration)
or zero point calibration.
Air Calibration
1. Press the power key to turn on the meter. See 1.
2. In the measurement mode, press the line select key
until the icon points to the bottom line and press
the up arrow key until the %sat or mg/L icon is
shown. See 2.
3. Make sure that the RDO sensor is connected to the
meter. See 3.
4. Prepare the calibration sleeve, insert the RDO sensor
into the prepared calibration sleeve and wait five
minutes for the RDO sensor to equilibrate. See 4.
To prepare the calibration sleeve, remove the cap
from the sleeve and remove the sponge from the cap.
Saturate the sponge with distilled water, squeeze all
the excess water out of the sponge and reassemble
the calibration sleeve.
5. Press the calibrate key. See 5.
6. When the reading stabilizes, the meter will display
100.0% saturation and proceed to the measurement
mode. See 6.
RDO Calibration Type Selection
1. In the measurement mode, press the setup key.
2. Press the up arrow key until dO is displayed on the
top line.
3. Press the line select key to move the icon to the
middle line. Press the up arrow key until CALt is
displayed on the middle line.
4. Press the line select key to move the icon to
the bottom line. Press the up arrow key to select
the calibration type as AIr for a water saturated
air calibration, H20 for an air saturated water
calibration, mAn for a manual calibration (Winkler
titration) or SEt0 for a zero point calibration. AIr is
the default setting.
5. Press the line select key to move the icon to the
top line and press the measure key.
Measurement Mode Selection
1. In the measurement mode, press the setup key.
2. Press the up arrow key until rEAd is displayed on
the top line.
3. Press the line select key to move the icon to the
middle line. Press the up arrow key until tYPE is
displayed on the middle line.
4. Press the line select key to move the icon to the
bottom line. Press the up arrow key to select AUt0
for the AUTO-READ™ mode, tImE for the timed
mode or C0nt for the continuous mode. AUt0 is the
default setting.
5. Press the line select key to move the icon to the
top line and press the measure key.
Automatic Datalog Feature
1. In the measurement mode, press the setup key.
2. Press the up arrow key until LOg is displayed on
the top line.
3. Press the line select key to move the icon to the
middle line. Press the up arrow key until AUt0 is
displayed on the middle line.
4. Press the line select key to move the icon to the
bottom line. Press the up arrow key to select OFF
or On. OFF is the default setting.
5. Press the line select key to move the icon to the
top line and press the measure key.
Thermo Scientific Orion Star™
RDO®Optical DO Meter Quick Start Guide
RDO Optical DO Measurements
Turn on the automatic datalog feature to send
measurements to the meter datalog or connect the
meter to a printer or computer.
1. Rinse the RDO sensor with distilled water, blot it
dry with a lint-free tissue and insert the sensor
into the sample.
2. If the meter is in the AUTO-READ measurement
mode, press the measure key to start a reading.
Once the reading is stable, the AR icon will stop
flashing and the display will freeze. Press the
measure key to take a new reading.
If the meter is in the timed measurement mode,
the meter will take a dissolved oxygen reading
at the frequency specified by the operator. If
parameters other than dissolved oxygen are
measured, the meter will immediately take
the other readings and continuously update
the display.
If the meter is in the continuous measurement
mode, the meter will immediately start taking
readings and continuously update the display.
Once the reading is stable, the %sat or mg/L
icon will stop flashing. This mode is not
recommended for everyday use, as it consumes a
significant amount of the meter’s battery power.
3. Remove the RDO sensor from the sample, rinse
it with distilled water, blot it dry with a lint-free
tissue, insert the sensor into the next sample and
repeat step 2 or see the RDO sensor user guide
for recommended storage.
Printing Data
If the automatic datalog feature is on, the meter will
log a measurement every time the measure key is
pressed (AUTO-READ and continuous modes) or at
the preset time interval (timed mode).
1. Connect the meter to a printer or computer and
verify the baud rate and output settings in the
meter setup menu.
2. In the measurement mode, press the log
view key.
3. Press the up arrow key to select SEnd to print
the datalog or CALS to print the calibration log.
4. Press the log view key to send the selected data
to the printer or computer.
English

Other manuals for Orion Star

6

Other Thermo Scientific Measuring Instrument manuals

Thermo Scientific TSQ 8000 Evo User manual

Thermo Scientific

Thermo Scientific TSQ 8000 Evo User manual

Thermo Scientific Nicolet iS50 AEM User manual

Thermo Scientific

Thermo Scientific Nicolet iS50 AEM User manual

Thermo Scientific 43iQ User manual

Thermo Scientific

Thermo Scientific 43iQ User manual

Thermo Scientific Orion Star LogR Meter User manual

Thermo Scientific

Thermo Scientific Orion Star LogR Meter User manual

Thermo Scientific Star Series User manual

Thermo Scientific

Thermo Scientific Star Series User manual

Thermo Scientific Orion Versa Star Pro User manual

Thermo Scientific

Thermo Scientific Orion Versa Star Pro User manual

Thermo Scientific SOLA II User manual

Thermo Scientific

Thermo Scientific SOLA II User manual

Thermo Scientific Orion Versa Star User manual

Thermo Scientific

Thermo Scientific Orion Versa Star User manual

Thermo Scientific Evolution 300 User manual

Thermo Scientific

Thermo Scientific Evolution 300 User manual

Thermo Scientific Orion 720Aplus User manual

Thermo Scientific

Thermo Scientific Orion 720Aplus User manual

Thermo Scientific LevelPRO Series User manual

Thermo Scientific

Thermo Scientific LevelPRO Series User manual

Thermo Scientific Orion Star User manual

Thermo Scientific

Thermo Scientific Orion Star User manual

Thermo Scientific Orion Star Plus User manual

Thermo Scientific

Thermo Scientific Orion Star Plus User manual

Thermo Scientific AQUAfast AQ3700 User manual

Thermo Scientific

Thermo Scientific AQUAfast AQ3700 User manual

Thermo Scientific Orion Star A112 User manual

Thermo Scientific

Thermo Scientific Orion Star A112 User manual

Thermo Scientific Partisol 2000i User manual

Thermo Scientific

Thermo Scientific Partisol 2000i User manual

Thermo Scientific AQUAfast AQ3170 User manual

Thermo Scientific

Thermo Scientific AQUAfast AQ3170 User manual

Thermo Scientific 48i User manual

Thermo Scientific

Thermo Scientific 48i User manual

Thermo Scientific ERGOSTAR HM 200 User manual

Thermo Scientific

Thermo Scientific ERGOSTAR HM 200 User manual

Thermo Scientific RadEye GX User manual

Thermo Scientific

Thermo Scientific RadEye GX User manual

Thermo Scientific Orion Versa Star User manual

Thermo Scientific

Thermo Scientific Orion Versa Star User manual

Thermo Scientific Nicolet Summit PRO User manual

Thermo Scientific

Thermo Scientific Nicolet Summit PRO User manual

Thermo Scientific 49iQ User manual

Thermo Scientific

Thermo Scientific 49iQ User manual

Thermo Scientific EPD Mk2 User manual

Thermo Scientific

Thermo Scientific EPD Mk2 User manual

Popular Measuring Instrument manuals by other brands

Weston BV01 user manual

Weston

Weston BV01 user manual

Agilent Technologies ESA-E E4402B user guide

Agilent Technologies

Agilent Technologies ESA-E E4402B user guide

Helmholz TB20 manual

Helmholz

Helmholz TB20 manual

Endress+Hauser Micropilot S FMR531 operating instructions

Endress+Hauser

Endress+Hauser Micropilot S FMR531 operating instructions

DVA MINI instruction manual

DVA

DVA MINI instruction manual

Iwatsu SS-540 instruction manual

Iwatsu

Iwatsu SS-540 instruction manual

Vision Engineering Hawk QC-200 user guide

Vision Engineering

Vision Engineering Hawk QC-200 user guide

Nohken GW100NR instruction manual

Nohken

Nohken GW100NR instruction manual

Bartec SYSCOM ROCK user manual

Bartec

Bartec SYSCOM ROCK user manual

Calscan Falcon user manual

Calscan

Calscan Falcon user manual

PCB Piezotronics A357B14 Installation and operating manual

PCB Piezotronics

PCB Piezotronics A357B14 Installation and operating manual

HYTEM 08-02-48 Product specifications and user's guide

HYTEM

HYTEM 08-02-48 Product specifications and user's guide

RFL 904 instructions

RFL

RFL 904 instructions

Siemens SITRANS 7MF1570 operating instructions

Siemens

Siemens SITRANS 7MF1570 operating instructions

PRECISION DIGITAL PD6720 Series instruction manual

PRECISION DIGITAL

PRECISION DIGITAL PD6720 Series instruction manual

Underground Magnetics MAG 3S manual

Underground Magnetics

Underground Magnetics MAG 3S manual

Benning Sun 2 operating manual

Benning

Benning Sun 2 operating manual

Wanf WF-D02A user manual

Wanf

Wanf WF-D02A user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.