Section 1 - Control and Indicating Devices
1.1 Installation
Remove all packaging materials and tape. Open the doors
to ventilate the unit. If the outside of the unit is dirty, clean with
a neutral detergent and rinse
with clear water.
Adjust the leveling feet by
rotating counterclockwise
(Figure 1-1) until all feet are in
full contact with the floor.
Ensure the unit is level.
Secure the unit to the wall
using the two fixtures (hooks)
attached to the back frame
(Figure 1-2). Attach a rope or
chain between the wall and the
fixtures. If holes can be drilled
in the wall, use the 10.5mm
bolt/nut supplied. This can be
used on a concrete wall only.
Locate the Model 3678 Refrigerator on a level, sturdy sur-
face capable of supporting the unit weight [277.8 lbs. (126kg)].
The refrigerator needs to be out of direct sunlight and away
from heat-generating sources. Four inches (10cm) clearance is
required on all sides for ventilation.
Make sure the power outlet is grounded. Do not use an out-
let without a ground.
1.2 Start-Up
1. Connect the unit to dedicated power supply and turn on
the power switch. Do not put any product in the unit at
this time.
2. On start-up, the alarm buzzer sometimes
operates. In this case, stop the buzzer by
pressing the alarm buzzer stop key
(BUZZER).
3. Set the desired temperature.
4. Confirm that the chamber temperature is at
the desired temperature.
5. When the unit is working properly, begin
slowly placing product into the chamber to
minimize temperature rise.
Model 3678 Refrigerator/Freezer ________________________________________________Control and Indicating Devices
1 - 1
Figure 1-1
Figure 1-2
Figure 1-3
1.3 Control Panel (Figure 1-3)
1. Alarm buzzer stop key (BZ): Press this key to silence
the buzzer in the event that the alarm activates and the
buzzer sounds. See Section 3 for details.
2. Display select key (REF./FREEZ.): By pressing this
key, the chamber temperature display of refrigerator or
freezer and each setting is selected.
3. Set key (SET): Pressing this key activates temperature
set mode and the digit which can be set flashes. By
pressing the key again after setting, the set value is
saved.
4. Digit shift key ( ): Pressing this key in set mode caus-
es the choice of digits to move. Key lock is activated by
pressing this key for more than 5 seconds in the temper-
ature display mode. See "Key lock operation" in
Section 2.2.
5. Numeric value shift key ( ): Pressing this key in set
mode causes the numeric value to change. "ON-OFF"
of key lock can be selected by pressing this key in the
key lock mode.
6. Temperature display: This screen shows the chamber
temperature, set temperature, or error code.
7. Refrigerator indicator (REFRIGERATOR): This indica-
tor is lit when the refrigerator is selected.
8. Freezer indicator (FREEZER): This indicator is lit when
the freezer is selected.
9. Alarm indicator (ALARM): This indicator flashes dur-
ing an alarm condition. Refer to Section 2.4 "Alarm
Temperature Setting".
10. Door check indicator (DOOR): This indicator lights
when the door is open.