Thermo Scientific Elite CTS User manual

Elite CTS Tester Operating Instructions
68X001308 Rev.0 June, 2017

To Insert battery plate,
align the small taps into
the guide ribs on the
housing
To lock the battery plate,
press down and slide in To Release the battery plate,
Push away this tap.
Battery Plate Unlock position Battery Plate locking position
To Insert battery plate,
align the small taps into
the guide ribs on the
housing
To lock the battery plate,
press down and slide in To Release the battery plate,
Push away this tap.
Battery Plate Unlock position Battery Plate locking position
To Insert battery plate,
align the small taps into
the guide ribs on the
housing
To lock the battery plate,
press down and slide in To Release the battery plate,
Push away this tap.
Battery Plate Unlock position Battery Plate locking position
Figure 1: Removing battery
cover
Figure 2: Push to unlock Figure 3: Align tabs Figure 4: Push down to unlock
Figure 5: Battery plate unlock
position
Figure 6: Battery plate lock
position
To Insert battery plate,
align the small taps into
the guide ribs on the
housing
To lock the battery plate,
press down and slide in To Release the battery plate,
Push away this tap.
Battery Plate Unlock position Battery Plate locking position
To Insert battery plate,
align the small taps into
the guide ribs on the
housing
To lock the battery plate,
press down and slide in To Release the battery plate,
Push away this tap.
Battery Plate Unlock position Battery Plate locking position
2 Elite CTS Tester Operating Instructions
This instruction manual is applicable for Elite CTS Tester.
Replacing the Batteries
The Elite CTS tester uses four AAA 1.5 V batteries.
1. To remove the battery cover, See Figure 1. Clear the front catch and then the back
catch, before sliding the cover off.
2. To remove the battery plate, push the center tab towards the front of the tester as
show in Figure 2. Once unlocked, remove the plate to access the batteries.
3. Invert the tester upside down to remove the batteries. Each side uses two
AAA batteries. Orient each battery with positive terminal facing downward.
4. To lock the battery plate, align the small taps into the guide ribs
on the housing and then press down. See Figure 4.
Getting Started
The Elite CTS tester has been factory calibrated and usually works well out of the box. However, after extended
periods of non-use, it is best to remove the sensor cap and soak the sensor in warm tap water for 10 minutes or so.
Prior to taking the measurements, periodic callibaration with certied standards is recommended for best accuracy.
Measurement Parameter Setting
1. Press to power on the tester.
2. Press to enter setup window. Press to select Measure. The display shows Cond, TDS and Salinity.
3. Scroll down by pressing to toggle between Cond, TDS and Salinity. Press to select Cond.
4. The display shows the selected measure with a .
Settings
Temp Set
Reset
Key Info
Scroll
Accept
EXIT
TDS
Salinity
Cond
Measure Cond

3Elite CTS Tester Operating Instructions
TDS Factor Setting
1. Press to enter setup window. Scroll down by pressing to select Settings.
2. Press to select settings. The display shows TDS factor and Backlight.
3. Press to select the default TDS factory setting or to adjust the setting.
4. Press to conrm the selection of the setting. The display shows the set TDS factor with a .
Backlight Settings
1. Press to enter setup window. Scroll down by pressing to select Settings.
2. Press to select Settings. The display shows TDS Factor and Backlight.
3. Scroll down by pressing to toggle between TDS Factor and Backlight. Press to select Backlight.
4. The display shows ON and OFF. Scroll down by pressing to toggle between ON and OFF. Backlight ON
increases readability in low light conditions.
5. Press to select the desired backlight option. The display shows the selected backlight option with a .
Temperature Settings
1. Press to enter setup window. Scroll down by pressing to select Temp Set. Press to select
Temp set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
2. Press to select Set °C/°F. Scroll down by pressing to toggle between °C and °F.
3. Press to select tempareture unit. The display shows the selected temperature setting with a .
Temperature Calibration
1. Press to enter setup window. Scroll down by pressing to select Temp Set.
2. Press to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
3. Scroll down by pressing to toggle between Set °C/°F, Temp Cal and Temp Coeff. Press to select
Temp Cal.
4. The lower display shows the current measured temperature reading based on the last set offset and the upper
display shows the current measured temperature reading based on factory default calibration.
Measure
Settings
Temp Set
Reset
Key Info
Scroll
Accept
EXIT
TDS Factor Backlight
ON OFF
Backlight OFF
Measure
Settings
Temp Set
Reset
Key Info
Scroll
Accept
EXIT
Backlight
Adjust
TDSFactor
0.71
0.66
TDS Factor
Measure
Settings
Reset
Key Info
Scroll
Accept
EXIT
Temp Cal
Temp Coeff
°C
°F
°C
Temp Set
Set °C/°F

4 Elite CTS Tester Operating Instructions
5. Dip the tester into a solution of known temperature and allow time for the in built temperature sensor to stabilize.
6. Press to adjust the temperature value or press the to conrm the calibrated value as the new temperature
value of the solution.
Note: To exit this program without conrming the calibration, press .
Temp Coefficient
1. Press to enter setup window. Scroll down by pressing to select Temp Set.
2. Press to select Temp Set. The display shows Set °C/°F, Temp Cal and Temp Coeff.
3. Scroll down by pressing to toggle between Set °C/°F, Temp Cal and Temp Coeff.
4. Press to select Temp Coeff or to adjust the Temp Coeff.
5. Press to conrm the Temp Coeff value. the new value is automatically conrmed with a .
Calibration for Conductivity, TDS, or Salinity
For best results, periodic calibration with an accurate standard is recommended prior to measurement. Use the
calibration standard value that is close to your intended sample value.The tester will retain one calibration value in
each mode (conductivity, TDS, salinity) when the instrument is powered off. The conductivity value can be calibrated
automatically or manually, while the TDS & salinity values require manual calibration. The tester will begin in the
measurement mode that was used when it was powered off. See “Elite CTS Measurement Parameter Setting” to
change the desired parameter.
Automatic Calibration for Conductivity
1. Remove the cap and press the to power on.
2. Dip the sensor in at least 30 mm of calibration standard.
3. Stir gently and press to begin the calibration.
4. The display will show CAL followed by the default value. CAL is indicated on the display during calibration mode.
5. If the reading is within the calibration range of the automatically recognized standards; 80 (84 µS/cm),
1410 (1413 µS/cm), or 12.90 (12.88 mS/cm), the icon is displayed when the automatic calibration standard
value has been detected.
6. Press to accept the auto conductivity standard and nish the calibration.
7. Display returns to Measurement window.
Measure
Settings
Reset
Key Info
Scroll
Accept
EXIT
Set °C/°F
Temp Cal Adjust
Temp Coeff
2.00
2.00
Temp Set Temp Coeff
Measure
Settings
Reset
Key Info
Scroll
Accept
EXIT
Set °C/°F
Temp Coeff
Temp Cal
23.4 °C
23.4 °C
23.4°C
Temp Set Temp Cal
CAL
µS/cm
844
1413
Place the tester
in the calibration
STD to start
or press to
escape.
COND
µS/cm
844
23.5 °C
CAL

5Elite CTS Tester Operating Instructions
Manual Calibration
When the conductivity reading is outside calibration range of the automatic conductivity standards or when TDS or
salinity is used, the tester will require manual adjustment.
1. Repeat steps 1 to 4 from “Automatic Calibration for Conductivity”.
2. Press to manually adjust the value to the desired reading.
Note: The adjustment will decrease only, however the adjustment will eventually cycle to the highest available value
after decreasing by 40% of the initial value.
3. Press to accept and nish the calibration when the desired value is selected.
4. To abort calibration, press to escape.
5. Once the calibration is nished and user has accepted the changes, measurement window will now show the
calibrated reading.
Note: The auto conductivity standards are 84 µS/cm, 1413 µS/cm & 12.88 mS/cm.
CTS Measurement
1. Press to power the tester on if needed.
2. Dip the electrode about 2 cm to 3 cm into the test solution. Stir and let the reading stabilize. The timer icon will
blink during this time. Once the reading is stabilized, the timer stops blinking and will appear to indicate the
stability of the reading.
CAUTION: Testing dry samples is not accurate and can lead to sensor damage or
breakage. Soils must be wet and free of particulates that may scratch the
glass sensor. Excessive force into dry samples can cause glass breakage.
3. Note the CTS value or press to freeze the reading. To release the reading, press again.
4. Press to turn off tester. If key is not pressed for 8.5 minutes, the tester will automatically shut off to conserve
batteries.
User Reset
Reset the CTS calibration to the user’s default settings by using the user reset function. Temperature user calibration
is not affected by the user reset function.
1. Press to enter setup window. Scroll down by pressing to select Reset. Press to select Reset. The
display shows User Reset and Fact Reset.
2. Press to select User Reset.
3. The display automatically shows No and Yes. Scroll down by pressing to toggle between No and Yes.
4. Press to conrm either No or Yes. The display shows the User Reset option with a .
COND
µs/cm
820
24.8 °C
Measure
Settings
Temp Set
Key Info
Scroll
Accept
EXIT
Fact Reset
No
Yes
User Reset
Reset
CAL
µS/cm
844
1413
CAL
Place the tester
in the calibration
STD to start
or press to
escape.
CAL
µS/cm
844
820
COND
µS/cm
844
24.8 °C

6 Elite CTS Tester Operating Instructions
Factory Reset
Reset the CTS tester to the factory default by using the factory reset function.
1. Press to enter setup window. Scroll down by pressing the to select Reset. Press to select Reset. The
display shows User Reset and Fact Reset.
2. Scroll down by pressing to toggle between the resets. Press to select Fact Reset.
3. The display automatically shows No and Yes. Scroll down by pressing to toggle between No and Yes.
4. Press to conrm either No or Yes. The display shows the Factory Reset option with a .
HOLD Function
This feature helps to freeze the display for a delayed observation.
1. Press button to freeze the measurement. 2. Press again to release the measurement.
Sensor Maintenance
1. Always keep the sensors electrodes clean. Rinse the electrodes with de-ionized water and wipe them dry with
clean cloth before storing with its protective cap. For cup type electrodes, remove the white plastic cup and insert
to thoroughly clean viscous solutions. Never scratch electrodes with a hard substance.
2. For better performance, soak the electrode in alcohol for 10 to 15 minutes and rinse with de-ionized water before
starting any measurement process. This is to remove dirt and oil stains on the electrode which may affect the
accuracy of the measurements.
Applications
The Elite CTS tester can be used for the following:
• Agriculture • Drinking water Hydroponics • Verication of reverse osmosis system operation
• Aquaculture • Electroplating rinse tanks • Water and wastewater treatment
• Aquariums and sh farms • Printing industry • Others
• Boiler blow-down • Swimming pools
Warranty
This instrument is supplied with a warranty against manufacturing defects for a period of one year from the date of
purchase.
Measure
Settings
Temp Set
Key Info
Scroll
Accept
EXIT
User Reset No
Yes
Reset
Fact Reset
COND
µs/cm
820
23.5 °C
µs/cm
820
23.5 °C
COND
Press to
release
COND
µs/cm
820
23.5 °C
COND
µs/cm
820
23.5 °C
Press to
release

Oring ID22XCS3
Ring
Oring ID21XCS2
Sensor
Small Tab
Insert Sensor Module
Turn anti-clockwise to release
7Elite CTS Tester Operating Instructions
Return of Items
Authorization must be obtained from your distributor before returning items for any reason. When applying for
authorization, please include information regarding the reason the item(s) are to be returned.
We reserve the right to make improvements in design, construction and appearance of products without notice.
Prices are subject to change without notice.
Sensor Replacement
Replace the sensor module at the fraction of the cost of a new tester. When the tester fails to calibrate or gives
uctuating readings in calibration standards, you need to change the electrode.
1. With dry hands, grip the ring with sensor facing you. Twist the ring clockwise. Save the ring for later use.
2. Pull the old sensor module away from the tester.
3. Align the four tabs on the new module so that they match the four slots on the tester.
4. Gently push the module onto the slots to sit it in position. Push the smaller O-ring fully onto the new sensor
module. Push the other O-ring over the module and thread it into place by rmly twisting counter clockwise.
Note: It is necessary that you recalibrate your tester prior to measurement after a sensor replacement.
Self Diagnostic Messages
Batteries are weak and need replacement soon.
stable error Appears when calibration is attempted but the reading is not yet stable. Wait for the
reading to stabilize or manually conrm the calibration by pressing enter.
over range The reading is above the measuring range of tester.

8 Elite CTS Tester Operating Instructions
Specication Elite CTS
Conductivity
Conductivity range 0.0 to 200.0 µS, 200 to 2000 µS, 2.00 to 20.00 mS
Resolution 0.1 µS, 1 µS, 0.01 mS
Relative accuracy ±1% full scale
Normalization temperature 25.0ºC (77ºF)
Temperature Co-efcient 0.0% to 10.0%
Calibration Points Up to 3 points
TDS
TDS Range 0 .0 to 100.0 ppm 100 to 1000 ppm 0.10 to 10.00 ppt
(TDS Factor 0.5)
Resolution 0.1 ppm, 1 ppm, 0.01 ppt
Relative accuracy ±1% full scale
Calibration points Up to 3 points
TDS Factor 0.40 to 1.00 ( Selectable)
Salinity
Salinity range 0.00 to 10.00 ppt
Resolution 0.10 ppt
Relative accuracy ±1% full scale
Calibration points One
Temperature
Temperature range 0°C to 60°C (32.0°F to 140.0°F)
Temperature resolution 0.1°C / 0.1°F
Temperature accuracy from 0°C to 50ºC (±0.5ºC / ±0.9ºF + 1 LSD);
from 50ºC to 60ºC (±1.0ºC / ±1.8 ºF + 1 LSD)
Temperature compensation Yes (Automatic Temperature Compensation)

9Elite CTS Tester Operating Instructions
Specication Elite CTS
General
Display Graphics, Dot Matrix 80X100 Pixel
Auto off 8.5 minutes ( From Last Key press)
Reset User / Factory
Power requirement Four AAA 1.5 V batteries
Battey life >150 hours
Water proong IP67
Environmental operating conditions
Ambient operating temperature 5°C to 45°C
Relative humidity 5% to 85% Non-Condensing
Storage temperature -20°C to 60°C
Storage humidity 5% to 85% Non-Condensing
Accessories
Ordering Code Product Description
CTS Tester
ELITECTSPIN Conductivity/TDS/salinity pocket tester with batteries
ELITECTSCUP Conductivity/TDS/salinity pocket tester with batteries
Conductivity Standard Solutions and Sachets
ECCON100BT 100 µS/cm KCl, 480 mL
ECCON1413BT 1413 µS/cm KCl, 480 mL
ECCON1288BT 12.88 mS/cm KCl, 480 mL
ECCON1413BS 1413 µS/cm KCl, box of 20 x 20 mL sachets
TDS 442* Standard Solutions
EC442300BT 300 ppm 442, 480 mL
EC4421000BT 1000 ppm 442, 480 mL
Salinity (NaCl) Solutions
ECNACL5PPT 5 ppt NaCl, 480 mL
Other Accessories
ELITECAP Replacement sensor cap
TFLANYARD Tester lanyard
*The 442 scale simulates fresh water dissolved solids as a mixture of 4 parts sodium sulfate, 4 parts sodium
bicarbonate, and 2 parts sodium chloride (40% Na2SO4,40% NaHCO3, 20% NaCl).
Table of contents
Other Thermo Scientific Test Equipment manuals

Thermo Scientific
Thermo Scientific Elite ORP User manual

Thermo Scientific
Thermo Scientific PCTestr 35 User manual

Thermo Scientific
Thermo Scientific EUTECH INSTRUMENTS PCSTestr35 User manual

Thermo Scientific
Thermo Scientific HAAKE Viscotester iQ User manual

Thermo Scientific
Thermo Scientific Thermolyne NCAT F85930-33 User manual

Thermo Scientific
Thermo Scientific CM-3CDEFT-32 User manual

Thermo Scientific
Thermo Scientific HAAKE Viscotester iQ User manual

Thermo Scientific
Thermo Scientific Elite PCTS User manual