PIN protection: This is for both the advanced and standard features.
Accumulated total: Accessible and resettable from both the front panel and
through the software.
Minimum flow cut-off: This is to ensure that inaccurate low ow measurements are
not recorded or returned. This is normally set to the minimum
ow for the meter.
Median filter: This is a lter that will remove short term erroneous ow
results.
Signal quality: The bar graph should be green, orange is a low signal quality
and red is very poor. This usually signies gas break-out in the
uid or a system not fully bled.
PIN protection.
There are 2 possibilities for the PIN protection:-
To use these features it is rst necessary to create a PIN. This is a four digit number in
the range 0001 to 9999. Setting a value of 0000 tells the software there is no PIN and
this is how to remove the PIN. After allocating a PIN these check boxes can lock all
of the settings or just the advanced user, including the accumulated total reset. If you
forget your PIN please contact you supplier for the default value. Once entered, all the
setting parameters remain visible. In the ‘Help’ menu, the user can ‘Log On using PIN...’
but the user will be prompted for the PIN prior to making any changes. If the display
version of the meter is being used and the display is set to view totals or rate and total
the left hand button will cycle through the various options.
Rate The display shows units and time base e.g. 1.0 Ltr/min.
Total The display is static with just the volume e.g. 10.3Ltr.
Accumulated total The display cycles “Ac tot” and the volume e.g. 120.3Ltr.
If the accumulated total is displayed and “right button resets totals” is selected in the
conguration screen it can be reset by pressing both buttons simultaneously for 4
seconds. If PIN protection is selected the display will change to a ashing curser awaiting
the pass word. The left button increments the value and the right button advances to
the next digit. After all 4 numbers are entered pressing and holding the right key for 4
seconds resets the Accumulated Total.
If a computer is being used for the reset simply press the “reset accumulated total” box.
Accumulated total.
Accumulated total DP position.
The chevron in the decimal point box calls a
drop down selection showing the possible
decimal point positions. Please ensure that the
value selected has sufcient resolution for your
anticipated accumulation period.
This resets the accumulated total. If PIN
protection is selected a PIN will be required.
Measurement control.
This sets the minimum ow for the meter. The
default is the normal minimum under ideal
conditions for the size of meter. This value can
be adjusted to suit the operating conditions.
e.g. If the automatic gain control is working
hard with an attenuating (ultrasound absorbing)
uid the threshold of operation is automatically
raised assuming that the background noise will
also be higher. This may not be the case and
the meter may perform reliably at a lower ow.
One method of establishing a safe minimum
ow is to set a low minimum cut off value with
no ow and a full pipe. Check the totaliser or
rate meter to see if any ow is being recorded if
yes increase the minimum ow value and repeat
until no ow is seen with zero ow in the pipe.
Median filter.
This lter removes short term anomalies in the recorded e.g. an air bubble passing
through. It should be used with caution as it is theoretically possible that over use of this
feature could lead to incorrect results in certain circumstances. Under normal operating
conditions the meter returns around 25 results per second and a regular occurrence at
this frequency could be completely ignored thus causing problematic readings.
The lter is a moving window taking the middle number from an odd number of results
selectable between 1 and 21. It is not an average or mean and it is designed to totally
ignore one or more results. A simple example is shown on page 18 for a median value
of “3”.