STORAGE & CLEANING
It should be stored at room temperature. The sensor lens is the most delicate part of the thermometer. The lens should be kept
clean at all times, care should be taken when cleaning the lens using only a soft cloth or cotton swab with water or medical
alcohol, allowing the lens to fully dry before using the thermometer. Do not submerge any part of the thermometer.
LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the settings of HAL and LAL.
‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. ‘Er3’ is displayed
when the ambient temperature exceeds 0℃ (32°F) or +50℃ (122°F). The thermometer should be allowed plenty of
time (minimum 30 minutes) to stabilize to the working/room temperature.
Error 5~9, for all other error messages it is necessary to reset the thermometer. To reset it, turn the instrument off,
remove the battery and wait for a minimum of one minute, reinsert the battery and turn on. If the error message
remains, please contact the Service Department for further assistance
BATTERIES
The thermometer incorporates visual low battery indication as follows:
BATTERY REPLACEMENT
1. Pull out the rubber gasket (10)
2. Loosen the screw and pull out the battery cover
3. Replace and reinstall with new batteries
4. Place the battery cover back and fasten the screw
5. Plug the rubber gasket (10) to avoid failure of water proof
When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V batteries.
Please note: It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction
Dispose of used battery promptly and keep away from children.
SPECIFICATION
Measurement Range -60 to 550 ℃ (-76 to +1022 ºF)
Operating Range 0 ~ 50℃ (32~122 ºF)
Accuracy Measured value ±2% or ±2℃
Emissivity 0.1 ~ 1.0 (default value: 0.95)
Response Time 1 sec
D:S (Distance:Spot) 12:1
Battery Life min 14 hours continuous use
Dimensions 143.9 x 116.73 x 42.78mm
Weight 180 grams(6.35 oz) including batteries (AAA*2pcs)
Note: Under the electromagnetic field of 3V/m from 200 to 700 MHz, the maximum error is 5℃ (9ºF). The thermometer will
automatically shut off if left idle for more than 15sec.
EMC/RFI: Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of
approximately 3 volts per meter, but the performance of the instrument will not be permanently affected.
‘Battery Exhausted’:
measurements are not possible
‘Battery Low’: battery needs to be replaced,
measurements are still possible
‘Battery OK’: measurements
are possible