ATTACHMENTS
Two patented* attachments are included with the
Veinlite PEDI:
Veinlite PEDI Light Shield
A snap-on light shield to reduce interference from
bright overhead lighting.
Veinlite PEDI Neonate Adapter
A snap-on neonate adapter to reduce the size of the
viewing area for use with infants.
To scan for veins, place device against skin
perpendicular to vein direction. Scan it
forward and backward. Veins will be visible
as dark lines due to transillumination.
1
2
3a
3b
4
Once a suitable vein has been identified,
rotate device 90˚, parallel to vein.
Apply gentle pressure on the Veinlite
PEDI to secure the vein.
Push back slightly on the Veinlite PEDI to
stretch the skin and provide traction.
Insert the needle into the vein, through the
opening of the device. NOTE: Discard the
single-use disposable plastic cover after
each patient.
OPERATION
• Place the Veinlite PEDI against the patient’s skin
with the LED Window [A] in direct contact with the
skin perpendicular to vein direction. Apply gentle
pressure to maintain optimal skin contact.
• To turn on the LEDs, press the Orange Button [B]
or Red Button [C] located on the left and right
side of the device.
• Press the button a second time to turn off.
• Press the Orange Button [B] if you are examining
small superficial veins on fair skin.
• Press the Red Button [C] if you are examining
deeper veins or if the patient’s skin is darkly
pigmented.
WARNING
DO NOT SHINE THE LED LIGHTS
IN THE PATIENT’S EYES.
WARNING
Veinlite PEDI is designed for external
examination only.
To avoid contamination of the device, Veinlite
PEDI must be used with the disposable plastic
cover.
To avoid the spread of infection, always change
the disposable plastic cover between patients
and wipe the new cover with a germicidal wipe.
These devices are intended for use only by
clinicians trained in performing vascular
access, or under the supervision of a clinician.
Do not look directly into the LEDs while on.
USING THE DISPOSABLE PLASTIC COVER
The single-use disposable plastic covers for the
Veinlite PEDI (Product #VPED-DPC) MUST be used
when the devise is placed in contact with a patient’s
skin.
CAUTION: These plastic covers are not sterile. They
should be wiped with a germicidal wipe prior to use.
• Before using Veinlite PEDI, place it inside the
disposable cover with the flap on the same side
as the LEDs.
• Pull the flap over the device to the other side.
• Wipe the LED side of the Veinlite PEDI and any
part of it coming in contact with the patient, with a
medically approved germicidal wipe.
PRECAUTIONS
• Device may be wiped down with a germicidal
wipe however, avoid using alcohol on the clear
plastic window as this may cause fogging.
• Do not immerse the device in liquid.
• Do not autoclave the device.
Scan Direction
Access Vein
a
b
53 4