manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. TransLite
  6. •
  7. Medical Equipment
  8. •
  9. TransLite Veinlite PEDI2 User manual

TransLite Veinlite PEDI2 User manual

Power
Depressing the Power Button [B] turns on the orange
LEDs. Depressing the button again, adds the red LEDs to
the orange LEDs. Depressing the button again turn the
unit off.
Depressing the Brightness Button [C] for one second will
turn the unit off.
The LEDs will automatically shut off after 7 minutes.
Brightness Control
The Veinlite PEDI2 has three levels of brightness that are
selected by pressing the Brightness Button [C] after the
unit is turned on. Brightness control is important when
working with small children.
Disposable Plastic Cover
The single-use disposable plastic covers for
the Veinlite PEDI2 (Product #VLP2-DPC)
MUST be used when the device is placed in
contact with a patient’s skin.
CAUTION: These plastic covers are not sterile. They
should be wiped with a germicidal wipe prior to use.
Before using Veinlite PEDI2, place it inside the disposable
cover with the flap on the same side as the LEDs.
Pull the flap over the device to the other side.
Wipe the LED side of the Veinlite PEDI2 and any part of it
coming in contact with the patient, with a medically
approved germicidal wipe.
Batteries
The Veinlite PEDI2 is powered by 2 AAA batteries. They
provide between 2 and 4 hours of continuous illumination
depending on the number of LEDs turned on and the
brightness level used.
Low Battery Indicator
When the batteries get low, the orange lights will flicker
when turned on. Replace the batteries as soon as the
orange light flickers when turned on.
Installing Or Replacing Batteries
To install or replace the batteries, press firmly on the
Battery Cover [D] and slide it off. Remove any used
batteries. Install 2 new AAA batteries, oriented as
indicated by the markings inside the battery compartment
then slide the battery cover back into place.
Description
The PEDI2 utilizes the patented1transillumination
technique known as side-transillumination and a
patented2orange color LED light. A ring of bright light
from 10 orange and 4 red LEDs is aimed through a clear
plastic LED Window [A] and focused below the skin
surface and towards the ring’s center.
The focused light creates a volume of illunimation, with
the central focus acting as a virtual light source under the
skin. This technique achieves unifrom illumination of a
small region of tissue anywhere on the body.
The patented orange light is selected because of high
absorption in venus blood. The red LEDs have high
penetration in darkly pigmented skins. The combination of
the two provides optimum imaging of veins in humans.
Vein Access
1Place the Veinlite PEDI2 inside the disposable plastic
cover so that the device is totally covered.
2Place the Veinlite PEDI2 on the skin, with the LED
Window [A] facing down. Apply gentle pressure to the
Veinlite PEDI2 to ensure it is in contact with the skin.
3Depress the Power Button [B] to turn on the orange
and then the red LEDs. Use the Brightness Button [C]
to control the brightness level of the device to the
desired level.
4To scan for veins, place
Veinlite PEDI2 against skin
perpendicular to vein
direction. Scan it
forward and
backward. Veins will be
visible as dark lines due to
transillumination.
5Once a suitable vein
has been identified,
rotate device 90˚,
parallel to vein.
6Apply gentle pressure on the
Veinlite PEDI2 to secure the
vein.
7Push back slightly on the
Veinlite PEDI2 to stretch the
skin and provide traction.
8Insert the needle into
the vein, through the
opening of the
device.
NOTE: Discard the
single-use disposable plastic
cover after each patient.
READ INSTRUCTIONS BEFORE USE
TransLite LLC
345 Commerce Green Blvd
Sugar Land TX 77478 USA
+1 281 240 3111
www.veinlite.com
[email protected]
Caution
•Veinlite PEDI2 is designed for external examination
only.
•To avoid contamination of the device, Veinlite PEDI2
must be used with the disposable plastic cover.
•To avoid the spread of infection, always change the
disposable plastic cover between patients and wipe
the new cover with a germicidal wipe.
•These devices are intended for use only by clinicians
trained in performing vascular access, or under the
supervision of a clinician.
•Do not look directly into the LEDs while on.
•Device may be wiped down with a germicidal wipe.
•Do not immerse the device in liquid.
•Do not autoclave the device.
Introduction
The TransLite Veinlite PEDI2 is a hand-held
battery-powered transillumination device for visualizing
veins and superficial blood vessels in pediatric patients.

Veinlite PEDI2
LED Window [A]
Battery Cover [D]
Power Button [B]
Brightness Button [C]
Warranty
The Veinlite PEDI2 has a one year parts and labor
warranty. It is not necessary to register your Veinlite PEDI2
for warranty. The process is automatic and the serial
number of the unit provides all the information needed.
Troubleshooting
Before returning the device to TransLite for repair, please
call or e-mail to obtain an authorization number and
shipping details.
No refunds will be given for devices shipped without
authorization.
Specifications
Number of LEDs: 10 Orange
4 Red
Weight with battery: 29g
Dimensions: 40.8×86.6×21mm
Battery: 2x AAA 1.5V
Intended Use/Indications
•Emergency venous access in infants and children.
•Venous access in infants and pediatric patients with a
history of difficult venous access.
•Easier visualization of veins in dark
pigmented patients.
•Easier visualization of veins in overweight
infants and children.
Contraindications
•When used for internal examination
•When used without the disposable plastic cover
Explanation of Symbols
Model Number
Serial Number
Manufacturer
EC Representative (EU Only)
Do not dispose of as unsorted municipal waste
M. Devices Group
Health Education Centre
The Church, Portland Street
Southport PR8 1HU
United Kingdom
+44 1704 544 944
5
6
7
8
4
Power
Depressing the Power Button [B] turns on the orange
LEDs. Depressing the button again, adds the red LEDs to
the orange LEDs. Depressing the button again turn the
unit off.
Depressing the Brightness Button [C] for one second will
turn the unit off.
The LEDs will automatically shut off after 7 minutes.
Brightness Control
The Veinlite PEDI2 has three levels of brightness that are
selected by pressing the Brightness Button [C] after the
unit is turned on. Brightness control is important when
working with small children.
Disposable Plastic Cover
The single-use disposable plastic covers for
the Veinlite PEDI2 (Product #VLP2-DPC)
MUST be used when the device is placed in
contact with a patient’s skin.
CAUTION: These plastic covers are not sterile. They
should be wiped with a germicidal wipe prior to use.
Before using Veinlite PEDI2, place it inside the disposable
cover with the flap on the same side as the LEDs.
Pull the flap over the device to the other side.
Wipe the LED side of the Veinlite PEDI2 and any part of it
coming in contact with the patient, with a medically
approved germicidal wipe.
Batteries
The Veinlite PEDI2 is powered by 2 AAA batteries. They
provide between 2 and 4 hours of continuous illumination
depending on the number of LEDs turned on and the
brightness level used.
Low Battery Indicator
When the batteries get low, the orange lights will flicker
when turned on. Replace the batteries as soon as the
orange light flickers when turned on.
Installing Or Replacing Batteries
To install or replace the batteries, press firmly on the
Battery Cover [D] and slide it off. Remove any used
batteries. Install 2 new AAA batteries, oriented as
indicated by the markings inside the battery compartment
then slide the battery cover back into place.
Description
The PEDI2 utilizes the patented1transillumination
technique known as side-transillumination and a
patented2orange color LED light. A ring of bright light
from 10 orange and 4 red LEDs is aimed through a clear
plastic LED Window [A] and focused below the skin
surface and towards the ring’s center.
The focused light creates a volume of illunimation, with
the central focus acting as a virtual light source under the
skin. This technique achieves unifrom illumination of a
small region of tissue anywhere on the body.
The patented orange light is selected because of high
absorption in venus blood. The red LEDs have high
penetration in darkly pigmented skins. The combination of
the two provides optimum imaging of veins in humans.
Vein Access
1Place the Veinlite PEDI2 inside the disposable plastic
cover so that the device is totally covered.
2Place the Veinlite PEDI2 on the skin, with the LED
Window [A] facing down. Apply gentle pressure to the
Veinlite PEDI2 to ensure it is in contact with the skin.
3Depress the Power Button [B] to turn on the orange
and then the red LEDs. Use the Brightness Button [C]
to control the brightness level of the device to the
desired level.
4To scan for veins, place
Veinlite PEDI2 against skin
perpendicular to vein
direction. Scan it
forward and
backward. Veins will be
visible as dark lines due to
transillumination.
5Once a suitable vein
has been identified,
rotate device 90˚,
parallel to vein.
6Apply gentle pressure on the
Veinlite PEDI2 to secure the
vein.
7Push back slightly on the
Veinlite PEDI2 to stretch the
skin and provide traction.
8Insert the needle into
the vein, through the
opening of the
device.
NOTE: Discard the
single-use disposable plastic
cover after each patient.
READ INSTRUCTIONS BEFORE USE
TransLite LLC
345 Commerce Green Blvd
Sugar Land TX 77478 USA
+1 281 240 3111
www.veinlite.com
[email protected]
Caution
•Veinlite PEDI2 is designed for external examination
only.
•To avoid contamination of the device, Veinlite PEDI2
must be used with the disposable plastic cover.
•To avoid the spread of infection, always change the
disposable plastic cover between patients and wipe
the new cover with a germicidal wipe.
•These devices are intended for use only by clinicians
trained in performing vascular access, or under the
supervision of a clinician.
•Do not look directly into the LEDs while on.
•Device may be wiped down with a germicidal wipe.
•Do not immerse the device in liquid.
•Do not autoclave the device.
Introduction
The TransLite Veinlite PEDI2 is a hand-held
battery-powered transillumination device for visualizing
veins and superficial blood vessels in pediatric patients.

VP2 DFU Rev A 0220
Made in USA
1 US Patent Number 5,146,923 2 US Patent Number 7,874,698
©2019 TransLite, LLC. All rights reserved. These instructions or any part
thereof may not be reproduced by any means, without prior permission
of TransLite, LLC.
FM 633388
Warranty
The Veinlite PEDI2 has a one year parts and labor
warranty. It is not necessary to register your Veinlite PEDI2
for warranty. The process is automatic and the serial
number of the unit provides all the information needed.
Troubleshooting
Before returning the device to TransLite for repair, please
call or e-mail to obtain an authorization number and
shipping details.
No refunds will be given for devices shipped without
authorization.
Specifications
Number of LEDs: 10 Orange
4 Red
Weight with battery: 29g
Dimensions: 40.8×86.6×21mm
Battery: 2x AAA 1.5V
Intended Use/Indications
•Emergency venous access in infants and children.
•Venous access in infants and pediatric patients with a
history of difficult venous access.
•Easier visualization of veins in dark
pigmented patients.
•Easier visualization of veins in overweight
infants and children.
Contraindications
•When used for internal examination
•When used without the disposable plastic cover
Explanation of Symbols
Model Number
Serial Number
Manufacturer
EC Representative (EU Only)
Do not dispose of as unsorted municipal waste
M. Devices Group
Health Education Centre
The Church, Portland Street
Southport PR8 1HU
United Kingdom
+44 1704 544 944

Other TransLite Medical Equipment manuals

TransLite Veinlite EMS User manual

TransLite

TransLite Veinlite EMS User manual

TransLite Veinlite EMS User manual

TransLite

TransLite Veinlite EMS User manual

TransLite Veinlite PEDI User manual

TransLite

TransLite Veinlite PEDI User manual

TransLite Veinlite LED User manual

TransLite

TransLite Veinlite LED User manual

Popular Medical Equipment manuals by other brands

IASIS TECHNOLOGIES MCN Training manual

IASIS TECHNOLOGIES

IASIS TECHNOLOGIES MCN Training manual

Optomed HALO user manual

Optomed

Optomed HALO user manual

EDAN SE-12 Series user manual

EDAN

EDAN SE-12 Series user manual

cosinuss c-med alpha user manual

cosinuss

cosinuss c-med alpha user manual

Physio Control Lifepak CR2 LIFELINKcentral Getting started guide

Physio Control

Physio Control Lifepak CR2 LIFELINKcentral Getting started guide

LPG HUBER 360 EVOLUTION user guide

LPG

LPG HUBER 360 EVOLUTION user guide

Molift Mover 180 manual

Molift

Molift Mover 180 manual

Cantel SCOPE BUDDY PLUS manual

Cantel

Cantel SCOPE BUDDY PLUS manual

Chison SonoEye P5 user manual

Chison

Chison SonoEye P5 user manual

Microtek Medi-6000 Plus quick start guide

Microtek

Microtek Medi-6000 Plus quick start guide

Blatchford ESK+ 4-Bolt Instructions for use

Blatchford

Blatchford ESK+ 4-Bolt Instructions for use

berlinger BEREG-Kit small user manual

berlinger

berlinger BEREG-Kit small user manual

Chattanooga Neo Vacuum Module installation instructions

Chattanooga

Chattanooga Neo Vacuum Module installation instructions

Ferno SCOOP EXL Stretcher User and maintenance manual

Ferno

Ferno SCOOP EXL Stretcher User and maintenance manual

Stanley Healthcare T2s Tag user guide

Stanley Healthcare

Stanley Healthcare T2s Tag user guide

Planmeca intra installation manual

Planmeca

Planmeca intra installation manual

Alcoa SFBTT20-IT SFBTT20-IT instruction manual

Alcoa

Alcoa SFBTT20-IT SFBTT20-IT instruction manual

WARDRAY PREMISE MR5501 Operator's manual

WARDRAY PREMISE

WARDRAY PREMISE MR5501 Operator's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.