Safety Warnings
WARNING: INSTALLATION AND MAINTENANCE
MUST BE CARRIED OUT BY SUITABLY
QUALIFIED AND COMPETENT PERSONEL ONLY.
HAZARDOUS VOLTAGES MAY BE PRESENT ON
THE CONNECTION TERMINALS.
INSTALLATION
•Install this product in accordance with local regulations, codes
and instructions.
•All fuses must be 0.5A / 250V Type F with a breaking capacity
of 35A or greater.
•All conductors carrying hazardous voltage must have external
switching or disconnect mechanisms fitted that provide at least
3 mm of contact separation in all poles.
•Signal cables connected to this device must not exceed 30
metres long.
•If signal cables are routed outside the building, install extra
surge-protection devices.
•Power supply, current input, USB and all outputs: Observe
maximum allowable voltages. All circuits connected to these
connectors must be limited-energy and insulated by double/
reinforced insulation from mains voltages according to IEC
61010-1:2010
Failure to install or operate the unit in
accordance with the above requirements may
impair the electrical safety of the unit. Voltage
measurements: An external UL recognized or
listed overcurrent protection device (fuse or
circuit breaker) must be fitted in-line with the
voltage lead. Recommended fuse: 0.5A Type F
with a breaking capacity of 35A or greater. Fuse
voltage rating must be greater than the maximum
voltage that will be applied to the meter.
MAINTENANCE
•Before cleaning, inspection or maintenance, isolate all power
sources to the unit.
•There are no user-serviceable parts inside this unit. Never
open the case.
•Inspect all external wiring connections at regular intervals.
Replace any damaged wiring and tighten any loose
connections.
•To clean the unit, use a dry cloth to wipe the casing.
•Take great care connecting the supply. If you connect power
to the wrong terminals, it may destroy the unit.
030618-01 Issue 1.1 10/20