When all changes are completed, use the START/STOP key to move to SET, then use the
UP or DOWN key to set the new date and time (this needs to be done even if the date
and time are correct).
• The machine will reboot, when saving is completed, the main screen will appear
11. WARNING: The Elara11 is equipped with a pre-heat mode for the chamber. Upon initial set
up after turning on the power switch and having set the date and time, it is critical to open
the door, select the “Vacuum Test” cycle and empty the chamber of all packing materials.
To do so:
• Open the door immediately after System Ready has flashed across the screen followed
by the fill reservoir icon and the jacket temperature icon and the generator
pressure icon . The door can be opened by following these steps:
• Place your thumb on the plastic door cover and the other fingers in the handle pocket.
• Pull the handle until the latch of the door is released.
• Now use the arrow keys on the EZPad to advance to the “Vacuum Test” cycle.
• This will turn off the pre heating mode
12. Remove all accessories and plastic material from the chamber.
13. Fill the reservoir, using one of the two options below, with steam distilled water having
the characteristics described in sec 1.18 of the Operator’s Manual.
NOTE: The reservoir has a built in overflow. Over filling the reservoir will result in water
emptying on to the counter top under unit
Fill from the front funnel as follows:
• Open the door. (The autoclave needs to be on for the door to open).
• Pour steam distilled water, gently, into the front funnel until it reaches the top of the blue
area on the level gauge.
• If water is filled above the blue area into the red area then use the drain hose to drain off
the over fill (see sec 10.4 in the Operator’s Manual).
Fill from the opening at the top as follows:
• Remove the water reservoir cover.
• Pour steam distilled water into the reservoir through the opening on top of the autoclave
until it reaches the base of the safety valve holder or reaches the top of the blue area of
the front level gauge.
• Under no circumstance should water be filled above the safety valve holder (located in the
reservoir).
• In case more water is accidentally filled above the blue area, decrease the water level by
draining the reservoir before starting a cycle (see sec. 10.4 in Operator’s Manual).
Using the tray rack reinsert the trays back into the chamber. In the Elara 11 it is normal and
necessary that the bottom tray protrude more than the other trays. DO NOT push this tray all
the way in, it will interfere with the sensors at the back of the chamber.
The unit is now ready to run sterilization cycles.
This Quick Installation Guide is an excerpt from the Operator’s Manual. The Operator’s Manual should be read in full to gain an
understanding of proper use and maintenance of this autoclave. The Operator’s Manual should also be saved for reference.
Elara11 Autoclave Quick Installation Guide
Water reservoir cover
Safety valve holder
14a.
14b.
15.
16.