Tyco Kendall Genius 2 Operation manual

kENDALL
GENIUS™
2
Checker / Calibrator
Checker / Calibrator
Operation and Service Manual
Appareil de vérification/Calibrateur
Manuel d’utilisation et d’entretien
Prüfgerät/Kalibrator
Benutzer- und Service-Handbuch
Sistema di controllo della
calibrazione/calibratore
Manuale operativo e di assistenza
Verificador de calibración
Manual de operación y servicio
Kalibreringsenhet
Användar- och servicemanual
Validator
Gebruiks- en onderhoudshandleiding
Verificador/Calibrador
Manual de utilização e assistência
Kalibrointilaite
Käyttö- ja huolto-opas
Kontrol- og kalibreringsapparat
Betjenings- og servicehåndbog
™
™
GENIUS
F/C˚
F/C˚
Checker / Calibrator
™
GENIUS
™
º
F
C
º
Ελεγκτής βαθμονόμησης
Εγχειρίδιο λειτουργίας και σέρβις
Ověřovač/Kalibrátor
Uživatelská a servisní příručka
Ellenőrző/kalibráló készülék
Működtetési és karbantartási kézikönyv
Контрольно-измерительное
устройство/Калибратор
Руководство по эксплуатации и
техническому обслуживанию
Kalibrator
Instrukcja obsługi i serwisowania
Kontrolör/Kalibratör
Kullanım ve Servis Kılavuzu
Kontroll- og kalibreringsapparat
Bruker- og servicehåndbok
™ Trademark ofTyco Healthcare Group LP or its affiliates.
TYCO HEALTHCARE GROUP LP • 15 HAMPSHIRE ST. • MANSFIELD, MA 02048 • MADE IN U.S.A. • ©2008 TYCO HEALTHCARE GROUP LP • ALL RIGHTS
RESERVED • 090909 • PRODUCT INFO IN U.S. • 1-800-962-9888 • www.tycohealthcare.com
TYCO HEALTHCARE U.K. LTD GOSPORT, PO13 0AS, U.K.
182563-01
CE MARK SYMBOLS
NON-CE MARK SYMBOLS
ELECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS
Corrugated
Recycles
STERILE
Keep Away
From Sunlight Keep Dry
Single Use
LOT
0123
FRAGILE
THIS SIDE UP
DEHP–FREE
0°
40°
Avoid Temperature
Extremes
Do not use if package is
opened or damaged.
STERILE R
A
STERILE
EOSTERILE
STERILE
STERILE R
A
STERILE
EOSTERILE
Humidity Limitation
Atmospheric pressure limitation
Use By
C US
LATEX–FREE
Caution, consult
accompanying
documents
Consult
Instructions
For Use
MR
MR UNSAFE NON-PYROGENIC
PROTECT FROM HEAT AND
RADIOACTIVE SOURCES
NON
STERILE
2
STERILIZE
DO NOT RESTERILIZE
POWDER–FREE
POWDER
CE MARK SYMBOLS
NON-CE MARK SYMBOLS
ELECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS
Corrugated
Recycles
STERILE
Keep Away
From Sunlight Keep Dry
Single Use
LOT
0123
FRAGILE
THIS SIDE UP
DEHP–FREE
0°
40°
Avoid Temperature
Extremes
Do not use if package is
opened or damaged.
STERILE R
A
STERILE
EOSTERILE
STERILE
STERILE R
A
STERILE
EOSTERILE
Humidity Limitation
Atmospheric pressure limitation
Use By
C US
LATEX–FREE
Caution, consult
accompanying
documents
Consult
Instructions
For Use
MR
MR UNSAFE NON-PYROGENIC
PROTECT FROM HEAT AND
RADIOACTIVE SOURCES
NON
STERILE
2
STERILIZE
DO NOT RESTERILIZE
POWDER–FREE
POWDER
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

Table of Contents
Page
Section I - Functional Description................................................................................................................................................... 1
Section II - Precautions and Warnings........................................................................................................................................... 2
Section III - Icon Identification........................................................................................................................................................ 3
Section IV - Required Equipment.................................................................................................................................................... 4
Section V - System Setup.................................................................................................................................................................. 4
Section VI - Operational Procedure................................................................................................................................................. 5
Apply Power.................................................................................................................................................................................. 5
Start-up Screen............................................................................................................................................................................. 5
Reminder Screen........................................................................................................................................................................... 5
Main Menu.................................................................................................................................................................................... 5
Preferences.................................................................................................................................................................................... 5
Language Selection....................................................................................................................................................................... 5
Set User/Site Information.............................................................................................................................................................. 5
Set Date and Time ......................................................................................................................................................................... 6
Check Calibration........................................................................................................................................................................... 6
Section VII - Error Messages............................................................................................................................................................. 8
Section VIII - Checker/Calibrator Specifications............................................................................................................................ 9
Section IX - Cleaning Instructions................................................................................................................................................... 9
Section X - Troubleshooting........................................................................................................................................................... 10
Section XI - Customer Service ........................................................................................................................................................ 11
Section XII - Warranty..................................................................................................................................................................... 11
Section XIII - Electromagnetic Conformity Declaration............................................................................................................. 12
This product contains software solely owned by Tyco Healthcare Group LP. Tyco Healthcare Group LP grants the user a non-exclusive,
limited license to use the software according to the operating instructions. A copy of the license can be obtained from Tyco Healthcare
Group LP.
English
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

1
GENIUS™2Checker/Calibrator Operation and Service Manual
Section I - Functional Description
This manual describes the operation of the GENIUS 2 Checker/Calibrator. This device has been developed to check the accuracy
of the GENIUS 2 Tympanic Thermometer and automatically recalibrate the thermometer, if necessary. In addition, the GENIUS 2
Checker/Calibrator has the capability to generate a test report for each thermometer tested and save it to a USB ash drive. The
GENIUS 2 Checker/Calibrator will work with all GENIUS 2 Thermometer software revisions.
The GENIUS 2 Checker/Calibrator contains two independently controlled infrared calibration targets that are similar to factory
calibration targets. These “blackbody” targets are designed to have ecient radiative heat transfer and produce infrared radiation
that approximates the human ear at temperatures of 32.2° C (90.0° F ) and 40.6° C (105.0° F ).
™
GENIUS
F/C˚
F/C˚
1 2 3 4
8910
5
6
7
11
Checker /Calibrator
F 3.15A 250V
™
Figure 1 - Front Panel Layout
Legend
1 Power Connector 7 High Temperature Target
2 Power Switch 8 High Temperature Target LED
3 Fuse Holder 9 LCD Display
4 Low Temperature Target LED 10 Encoder Knob
5 Low Temperature Target 11 USB port
6 GENIUS 2 Thermometer Connector
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

2
GENIUS™2Checker/Calibrator Operation and Service Manual
Section II - Precautions and Warnings
• The GENIUS 2 Checker/Calibrator is designed to be operated as specified in this manual. Protection provided by the safety
devices in the GENIUS 2 Checker/Calibrator may be impaired if the device is not used in accordance with the directions contained
in this manual.
• The GENIUS 2 Checker/Calibrator enclosure is not waterproof. Avoid spilling liquid of any kind onto the front panel of the device.
Likewise, the GENIUS 2 Checker/Calibrator is powered from an AC power source. Avoid the potential for contact between liquid
and any external surface of the device while connected to the mains source.
• The surface quality of the interior of the reference blackbody targets is critical to their operation as efficient infrared radiators.
Care should be exercised when working near the entrance to the blackbody targets to ensure that no contact is made with the
surface inside the rubber flaps.
• In the event of a blown fuse, replace the fuse with a fuse that exactly matches the rated voltage, current, and form factor as
specified in the GENIUS 2 Checker/Calibrator Specifications section of this manual. Make sure the device is disconnected from
the AC power source before changing a blown fuse.
• This device should not be used in the presence of flammable anesthetics. There is risk of an explosion in the presence of these
anesthetics.
• There are no user-serviceable components in this device. Do not remove the front panel. Refer servicing to qualified service
personnel.
• Use only the supplied power module with this device.
• When prompted to insert the GENIUS 2 Thermometer into one of the blackbody targets, there is a 60 second window of time to
accomplish the insertion. If the time delay exceeds this period, the procedure will be terminated.
• Prior to checking/calibrating a GENIUS 2 Thermometer, ensure that the thermometer lens is clean. Effective cleaning will ensure
that the calibration tolerances are correctly verified. Refer to the cleaning instructions in the Genius 2 Thermometer operating
manual.
• When installing a GENIUS 2 probe cover on to the tip of the GENIUS 2 Thermometer, use only covers that are in a cassette and
not released. Try to avoid all hand contact with the probe covers in order to prevent inaccurate readings. Do not reuse probe
covers. Use only GENIUS 2 probe covers.
• Let the GENIUS 2 Checker /Calibrator warm up after power-up for at least 15 minutes before use. Make sure that the GENIUS 2
Thermometers, probe covers and the GENIUS 2 Checker/Calibrator have had enough time to equilibrate to room temperature
before proceeding. Allow a longer warm up period if the device or thermometers were at the extreme limits of allowed
ambient temperature. The portability of the GENIUS 2 Checker/Calibrator will allow for relocation of the device to areas that
meet the ambient temperature requirements.
• To maintain Electromagnetic Compatibility (EMC) conformance, use a USB flash drive that is CE marked and meets EN55022
standard.
• Do not put or rest the GENIUS 2 Thermometer on the surface of the GENIUS 2 Checker/Calibrator at any time. The thermal
sensors inside the thermometer probe tip will detect the heat transferred from the blackbody targets through the front panel
surface. This will cause instability of the thermometer sensors, which will produce erroneous temperature readings.
• The GENIUS 2 Checker/Calibrator may only be used indoors, in an area free from drafts and wide ambient temperature swings.
• The USB port on the front panel is intended to be used with only a USB Flash memory drive. No other USB devices should be
inserted into the USB port. The USB flash memory drive should be installed prior to device power-up.
• In the event that the GENIUS 2 Checker/Calibrator is dropped, return the unit to the factory for re-calibration.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

3
GENIUS™2Checker/Calibrator Operation and Service Manual
Section III - Icon Identification
SN
Serial Number
CE MARK SYMBOLS
NON-CE MARK SYMBOLS
ELECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS
Corrugated
Recycles
STERILE
Keep Away
From Sunlight Keep Dry
Single Use
LOT
0123
FRAGILE
THIS SIDE UP
DEHP–FREE
0°
40°
Avoid Temperature
Extremes
Do not use if package is
opened or damaged.
STERILE R
A
STERILE
EOSTERILE
STERILE
STERILE R
A
STERILE
EOSTERILE
Humidity Limitation
Atmospheric pressure limitation
Use By
C US
LATEX–FREE
Caution, consult
accompanying
documents
Consult
Instructions
For Use
MR
MR UNSAFE NON-PYROGENIC
PROTECT FROM HEAT AND
RADIOACTIVE SOURCES
NON
STERILE
2
STERILIZE
DO NOT RESTERILIZE
POWDER–FREE
POWDER
Manufacturer
Consult Accompanying
Documents
CE MARK SYMBOLS
NON-CE MARK SYMBOLS
ELECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS
Corrugated
Recycles
STERILE
Keep Away
From Sunlight Keep Dry
Single Use
LOT
0123
FRAGILE
THIS SIDE UP
DEHP–FREE
0°
40°
Avoid Temperature
Extremes
Do not use if package is
opened or damaged.
STERILE R
A
STERILE
EOSTERILE
STERILE
STERILE R
A
STERILE
EOSTERILE
Humidity Limitation
Atmospheric pressure limitation
Use By
C US
LATEX–FREE
Caution, consult
accompanying
documents
Consult
Instructions
For Use
MR
MR UNSAFE NON-PYROGENIC
PROTECT FROM HEAT AND
RADIOACTIVE SOURCES
NON
STERILE
2
STERILIZE
DO NOT RESTERILIZE
POWDER–FREE
POWDER
Date of Manufacture
Standby Power Double or Reinforced
Insulation Protection
Replace with Same
Rated Fuse CE Mark
USB Port Dispose of as Electrical
and Electronic Waste
F/C˚
High Temperature Target
LISTED
(LABORATORY EQUIPMENT)
46EA
UL Listing Information
F/C˚
Low Temperature Target
Caution: Indoor Use Only
GENIUS 2 Thermometer
Connection
™
The GENIUS 2 Infrared Tympanic Thermometer is a reliable and accurate temperature-taking device. The reason for the
accuracy of the GENIUS 2 Thermometer is the design, controlled calibration methods and stringent manufacturing controls. One
of the most critical functional parts of a thermometer is the ACCUSYSTEM√ probe cover produced by Tyco Healthcare/Kendall. The
GENIUS 2 Thermometer ACCUSYSTEM√ probe cover, when placed on a GENIUS 2 Thermometer, serves as an infection control barrier
between the patient and the device and the medium for heat transmission from the patient to the thermometer. The functionality
of the GENIUS 2 Thermometer ACCUSYSTEM√ probe cover is extremely important for preventing device contamination and also for
allowing accurate patient temperature measurements.
The reason the GENIUS 2 Infrared Tympanic Thermometer and GENIUS 2 Thermometer ACCUSYSTEM√ probe covers have
satised our customers’ expectations is due to the integration of these two parts during the manufacturing process. The GENIUS 2
Thermometer ACCUSYSTEM√ probe cover molding process parameters are tightly controlled at Tyco Healthcare/Kendall to minimize
variation and produce consistent and reliable products. The special grades of thermoplastic materials are also tightly controlled and
are specically chosen to deliver consistent temperature measurements with GENIUS 2 Thermometers. The factory calibration and
nal determination of an acceptable GENIUS 2 Thermometer is dependent on the use of acceptable Tyco Healthcare/Kendall
GENIUS 2 Thermometer ACCUSYSTEM√ probe covers. The use of generic probe covers or other probe covers not produced by
Tyco Healthcare/Kendall is not supported or sanctioned by Tyco Healthcare/Kendall. The use of unauthorized GENIUS 2 Thermometer
probe covers could jeopardize the accuracy of the GENIUS 2 Thermometer. Check your probe cover carton for the ACCUSYSTEM√ logo
to be certain that your thermometer will deliver an accurate temperature every time it is utilized.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

4
GENIUS™2Checker/Calibrator Operation and Service Manual
Section IV - Required Equipment
1. GENIUS 2 Checker/Calibrator
2. Power module with AC adapter
3. Probe cable
4. Operator’s Manual
5. GENIUS 2 Thermometers to be tested
6. GENIUS 2 disposable probe covers
7. Small Phillips screwdriver
Section V - System Setup
The following actions should precede use of the GENIUS 2 Checker/Calibrator:
1. Connect the correct AC plug adapter for your country to the power module.
2. Connect the power module cable to the power jack on the front panel.
3. Connect the power module to an AC power outlet.
4. Turn the unit on via the switch on the front panel.
5. Let the GENIUS 2 Checker/Calibrator unit warm up for at least 15 minutes.
6. Set aside a cassette with new GENIUS 2 probe covers.
7. Let the GENIUS 2 Thermometers equilibrate to the room temperature.
8. Prepare the GENIUS 2 Thermometers by thoroughly cleaning the lenses prior to use. Refer to the GENIUS 2 Thermometer
operating manual for cleaning instructions.
9. Remove the battery compartment access panel by using a small Phillips screwdriver. Remove all three AAA batteries.
10. Connect the probe cable round connector to the front panel of the GENIUS 2 Checker/Calibrator.
11. Connect the other end of the probe connector into the battery compartment. The connector will snap into place where the
batteries were. Refer to Figure 2.
Figure 2 - Insertion of Probe Connector into GENIUS 2 Battery Compartment
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

5
GENIUS™2Checker/Calibrator Operation and Service Manual
Section VI - Operational Procedure
The following procedure will describe the steps required to operate the GENIUS 2 Checker/Calibrator. In order to facilitate familiarity
with the product, it would be advisable to use this section of the manual during the calibration check of the rst few thermometers.
Installation of USB Flash Drive (Optional)
Install a USB ash drive into the USB port located on the front panel if a report le written to the USB ash drive is desired. If the USB
ash drive is installed after performing a calibration check, an error message may occur when attempting to write the report le if
the operator does not wait for the USB ash drive to complete initialization. If the GENIUS 2 Checker/Calibrator is powered o, all
internal memory (not yet saved to the USB ash drive) will be lost.
Apply Power
Make sure the AC-adapter end of the power module is plugged into a grounded outlet. Connect the power module to the front panel
power connector. Apply power to the GENIUS 2 Checker/Calibrator by using the power switch on the front panel.
Start-up Screen
The rst screen on the LCD display is the Initialization Screen, which displays the software revision and date of the GENIUS 2
Checker/Calibrator. If there is an internal problem during initialization, an Error screen will be displayed (See Section VII).
Reminder Screen
The next screen on the LCD display is a reminder to let the GENIUS 2 Checker/Calibrator warm-up for at least 15 minutes after
applying power. This will let the infrared blackbody targets warm up to the proper temperatures. To exit this screen, press the
encoder knob.
Main Menu
The Main Menu screen is displayed next. This screen has four menu options:
Check Calibration
Preferences
Set User/Site Information
Set Time/Date
To enter any menu option, use the encoder knob to scroll over and highlight the desired menu option, then press the encoder knob.
Preferences
The Preferences menu is used to setup the language option. Press the encoder knob to enter the Language Selection screen.
Language Selection
The Language Selection screen lists the available languages for the LCD screen text. To select a language, scroll over the selection and
it will be highlighted. Press the encoder knob to complete the selection and return to the Main Menu screen. The selected language
will be stored in non-volatile memory and will be used next time the unit is powered on.
Set User/Site Information
The Set User/Site Information screen is used to enter the operator’s name and the organization and/or site information. This
information will be written to the test report le.
To enter the information, use the encoder knob to scroll through the alphabet for a given letter position in the operator name or site
information. When the correct letter is shown, press the encoder knob to select the letter and advance to the next letter position. To
enter a space, just press the encoder knob when a space or blank letter is shown in that letter position. The user name can be up to
20 letters long. Use the encoder knob to scroll to the organization eld and enter a name for the site in a similar manner. To exit the
menu, scroll to the DONE text and press the encoder knob.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

6
GENIUS™2Checker/Calibrator Operation and Service Manual
Set Date and Time
The Set Date and Time screen menu option is used to setup the date and time that is used on the test report. The time and date are
internally stored and automatically updated when the unit is powered up.
To enter time, use the encoder knob to select the hours and minutes. Note: the hours is setup using a 24 hour clock format, where
1PM is entered as 13:00 etc. When the value is correct, push the encoder knob to move to the next eld of the time and then to the
date eld. The date eld is setup using a MMDDYYYY format. After the last eld of the date is entered, the next encoder push will exit
this menu and return to the Main Menu. Note: the time and date are shown at the bottom of the Main Menu.
Check Calibration
This menu option performs the calibration check of a GENIUS 2 Thermometer, and if required, automatically recalibrates the
thermometer using a series of step-by-step screens. Before entering this menu option, make sure that the required equipment listed
in Section IV is available. This includes a cassette with unused (new) probe covers.
Step 1:
Warm up of infrared blackbody targets. This screen is displayed while the targets are checked for proper temperature. Both the Low
Target and the High Target are checked, and when they are at the correct temperature, the screen will change to “OK” for each target.
There is a prompt to press the encoder knob to continue to the next screen. If either target cannot reach the correct temperature,
an Error message screen will be displayed. If this happens, check to be sure the GENIUS 2 Checker/Calibrator is within the correct
ambient (room) temperature range, and has been given at least 15 minutes of warm up time since power was switched on.
Step 2:
Inspect the GENIUS 2 Thermometer that will be tested. Ensure that the thermometer lens is not scratched and is free from ear
wax and/or other contaminants. If the thermometer lens is “dirty”, clean the lens according to the instructions in the GENIUS 2
Thermometer operating manual. Once the thermometer lens is clean, connect the GENIUS 2 Thermometer to the Checker/Calibrator
by plugging the round end of the thermometer interface cable into the round connector socket on the GENIUS 2 Checker/Calibrator
front panel (refer to Figure 3). Open the battery door of the thermometer by using a small Phillips screwdriver. Remove the batteries
and set aside. Connect the other end of the thermometer probe cable into the battery compartment, making sure the connector
“snaps” into place securely. Refer to Figure 2 in Section V.
Note: A new probe cover must be used prior to every target insertion. Finger oils on previously used probe covers or re-used
probe covers that have been heated by GENIUS 2 Checker/Calibrator target can cause erroneous readings which may
result in a failed calibration check or recalibration. Therefore, it is extremely important that a new probe cover is used
each time.
Step 3:
Press the encoder knob to continue, then install a new GENIUS 2 probe cover onto the GENIUS 2 Thermometer. Make sure to use a
cover that is held in the cassette. Once the cover is on the thermometer, do not point the probe tip at any object, including hands,
ngers or LCD display, as this will cause an inaccurate temperature reading. Inspect the probe cover to make sure it is fully seated (no
space between the cover and the tip base) and no holes, tears, or wrinkles are present in the plastic lm. When the cover is on, insert
the probe tip into the Low temperature target, which will have a blinking LED indicator next to it. Make sure that the probe tip is fully
inserted into the target well. Refer to Figure 3 for proper insertion. In order to avoid a time-out error, insert the covered probe tip into
the target within 60 seconds from the time the encoder knob is pressed.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

7
GENIUS™2Checker/Calibrator Operation and Service Manual
™
GENIUS
F/C˚
F/C˚
Checker / Calibrator
™
GENIUS
™
º
F
C
º
Figure 3 - Proper Insertion of GENIUS 2 Probe Tip into Target
Step 4:
The insertion of the probe tip into the target will automatically be detected and the screen display will change to “Press GENIUS 2
Thermometer scan button.” Press the scan button on the GENIUS 2 Thermometer while it is inserted into the Low temperature target.
The GENIUS 2 Thermometer will output 3 short beeps as it takes the temperature of the blackbody target. After the temperature is
taken, the next screen will display.
Step 5:
Withdraw the thermometer from the Low temperature target and eject the probe cover. The screen display will change to “Install a
new probe cover and insert probe into the High target.” Again, use a new GENIUS 2 probe cover from the cassette and be careful not
to point the probe tip at any object, including hands, ngers or LCD display. Inspect the probe covers to make sure it is fully seated
(no space between the cover and the tip base) and no holes, tears, or wrinkles are present in the plastic lm. When the cover is on,
insert the probe tip into the High temperature target, which will have a blinking LED indicator next to it. Make sure that the probe tip
is fully inserted into the target well.
Step 6:
The insertion of the probe tip into the High target will automatically be detected and the screen display will change to “Press
GENIUS 2 Thermometer scan button.” Press the scan button on the GENIUS 2 Thermometer while it is inserted into the High
temperature target. The GENIUS 2 Thermometer will output 3 short beeps as it takes the temperature of the blackbody target. After
the temperature is taken, the next screen will display.
Step 7:
If the GENIUS 2 thermometer is within specied accuracy limits, a Results screen will be displayed showing the thermometer serial
number along with an indication that the thermometer passed calibration check. If the GENIUS 2 thermometer is found to be outside
the specied accuracy limits , the display will indicate to the user that the GENIUS 2 Checker/Calibrator is entering into a calibration
sequence that requires the user to repeat Steps 3 through 6 three more times in order to gather recalibration data. Following the
last pass through Steps 3 through 6, the GENIUS 2 Checker/Calibrator will attempt to recalibrate the GENIUS 2 thermometer. If the
recalibration is successful, a Results screen will be displayed showing the thermometer serial number along with an indication that
the thermometer passed calibration. If the recalibration is unsuccessful or the thermometer is unable to be calibrated, a Results
screen will be displayed showing the thermometer serial number along with an indication that the thermometer failed calibration.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

8
GENIUS™2Checker/Calibrator Operation and Service Manual
Step 8:
After the pass/fail Results screen has been displayed, press the encoder knob to display a second Results screen containing the target
temperatures, the GENIUS 2 readings, and the target versus thermometer variances.
Step 9:
The next screen will display “Write report to USB ash drive?” and display Yes and No below. If a report is desired, insert a ash
drive into the front panel USB port. Note: allow about 30 seconds after insertion for the USB ash drive to initialize. Use the encoder
knob to scroll to either Yes or No and press the encoder knob to make a selection. If No is selected, the screen returns to the Main
Menu. If Yes is selected, a report is written to the USB ash drive, using the thermometer serial number for lename and .TXT as the
le extension. The le contains the test results as well as other information such as serial number, date, time, user name, and
organization. This le is a simple ASCII text le and is only available in English. This le can be read by MicrosoftTM Windows
programs such as Microsoft™ Notepad and sent to a printer. After the le is written, the display will show “Push knob to continue.”
Press the encoder knob to return to the Main Menu.
Step 10:
After completing a calibration check/recalibration, remove the probe connector from the battery compartment, using a rocking
motion to free the connector. Then reinstall the 3 AAA batteries, observing the proper polarity orientation. Finally, reinstall
the battery compartment cover and tighten the screw.
Step 11:
Remove USB ash drive before closing the lid of the case. The lid should be closed when the GENIUS 2 Checker/Calibrator is not in use.
Section VII - Error Messages
The error messages will have a number associated with them, such as Error 1, etc. The following is an explanation of the errors:
Error 1: Internal Error
This error is an indication of a problem within the GENIUS 2 Checker/Calibrator. If this error persists after cycling power o and
back on, return the unit for factory servicing.
Error 2: Unable to communicate with GENIUS 2 Thermometer.
This error is an indication of either a lack of communication or a miscommunication between the GENIUS 2 Thermometer and
the GENIUS 2 Checker/Calibrator. Check both ends of the thermometer probe cable, especially the end that connects inside
the battery compartment. The pins in the thermometer probe cable can lose spring action if they are bent or damaged. After
inspecting the probe cable, re-attempt the calibration check. If the problem persists, try a dierent GENIUS 2 Thermometer, if
available. If this does not resolve the problem, return the unit for factory servicing, along with the thermometer cable and the
GENIUS 2 Thermometer that indicated this error.
Error 3: Timeout
This error is an indication that the operator took too long to perform an operation. This error might be seen when connecting
to the GENIUS 2 Thermometer, or while installing a probe cover on the thermometer. The timeout is set for one minute. Try
repeating the operation that gave this error.
Error 4: Target Out of Range
This error is an indication that one of the blackbody targets is not within the range of temperature allowed. Check that the
unit is being operated within the allowed ambient temperature operating range. Refer to Section VIII. If operating within the
specied ambient conditions does not eliminate the error, the blackbody target(s) may be damaged. Return the unit for factory
servicing.
Error 5: USB Flash Drive Write Error
This error is an indication of a problem when writing the report le to a USB ash drive. Check that the write protect switch on
the USB drive is in the “o ” position. Make sure that there is sucient free space on the drive, and that security or encryption
features are not in use.
Error 6: Ambient Temperature Out of Range
This error is an indication that the ambient temperature is not within the specied range for performing an accurate recalibration.
Check that the unit is being operated within the allowed ambient temperature operating range. Refer to Section VIII.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

9
GENIUS™2Checker/Calibrator Operation and Service Manual
Error 7: Wrong Target
This error is an indication that the GENIUS 2 Thermometer has been inserted into wrong target or was inserted into the target
prematurely. Please take care to follow the instructions on the screen and not perform steps prior to the instructions being
displayed.
The Error screen will be displayed for 60 seconds before returning to the Main Menu. To exit the Error screen sooner, press the
encoder knob.
Section VIII - Checker/Calibrator Specifications
Temperature Set Points
Low Target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32.2 °C +/- 0.3 °C (90.0 °F +/- 0.5 °F)
High Target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40.6 °C +/- 0.3 °C (105.0 °F +/- 0.5 °F)
Target Temperature Accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Initial Limit +/- 0.03 °C (0.05 °F)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Post year 1 +/- 0.06 °C (0.10 °F)
GENIUS 2 Thermometer Accuracy After Recalibration*
Readings between 36. 0 °C and 39.0°C(96.8 °F and 102.2 °F) . . . . . . . . . . . . . . . . . . . . . . . . +/- 0.2 °C (+/- 0.4 °F)
Readings less than 36. 0 °C (96.8 °F) or greater than 39.0°C (102.2 °F) . . . . . . . . . . . . . . . . . . . . +/- 0.3 °C (+/- 0.5 °F)
Warm-upTime Before Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 minutes (from non-extreme ambient)
Ambient Temperature Operating Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.1-26.7 °C (70-80 °F)
Relative Humidity Operating Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 +/- 20%, non-condensing
Altitude Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Up to 2000 meters (6563 feet )
Storage Temperature Range . . . . . . . . . . . . . . . . . . . . . . . . . -25 °C - 55 °C (-13°F - 131°F ) 85% RH non-condensing
Approximate Size
Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.1 cm (12.25” )
Depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15.9 cm (6.25” )
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26.0 cm (10.25” )
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3 kg (7.2 lbs )
Power Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100-240V 0.8-0.4A 47-63Hz
Fuse Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Schurter FSF 5 x 20mm Fast Blow
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Low Breaking Capacity 250 V, 3.15A
Recommended Checker/Calibrator Calibration Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . One Year
ESD Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Meets IEC 61000-4-2 Non-continuous Criteria C
Pollution Degree. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Installation Category. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Indoor Use Only
*Post recalibration accuracy using the GENIUS 2 Checker/Calibrator may not necessarily be equivalent to factory calibration
Section IX - Cleaning Instructions
CAUTION: DO NOT IMMERSE THE Checker/Calibrator, CABLES, OR POWER CORD IN WATER OR OTHER CLEANING SOLUTION; CLEAN
USING A DAMP (NOT WET) CLOTH. FAILURE TO FOLLOW CLEANING PROCEDURES DESCRIBED HEREIN COULD RESULT IN HAZARDS TO
USERS.
As with any AC powered electrical device, care must be taken to prevent liquid from entering the Checker/Calibrator to avoid
electrical shock hazard, re hazard, or damage to electrical components.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

10
GENIUS™2Checker/Calibrator Operation and Service Manual
General Cleaning Instructions
Cleaning of the GENIUS 2 Checker/Calibrator may be performed as follows:
• FIRST, UNPLUG THE GENIUS 2 Checker/Calibrator FROM POWER SOURCE. NEVER CLEAN THE DEVICE WHILE CONNECTED TO THE
MAINS SUPPLY.
• A mild detergent should be used for general cleaning of outside surfaces. Also, a cloth damped with isopropyl alcohol can be
used to clean the unit. DO NOT USE abrasives or solvents. Avoid excess moisture around the target assemblies and the electrical
connectors.
• Do not attempt to clean the target below the rubber flaps.
• Do not clean the target nests or flaps with alcohol.
If there is any doubt about the compatibility of a cleaning agent with part of this equipment or its materials contact
Tyco Healthcare/Kendall customer service (See Section XI).
Section X - Troubleshooting
Some of the common problems are as follows:
Problem Probable Cause Corrections
GENIUS 2 Checker / Calibrator will
not power up GENIUS 2 Checker / Calibrator not receiving AC power Check AC power connections
Blown fuse Replace fuse
Test report does not write to USB
flash drive USB flash drive not formatted Use PC or laptop to format flash drive
USB flash drive write protected Use PC or laptop to turn off write protection
USB flash drive does not contain enough free memory space Use PC or laptop to delete unnecessary items or use different USB flash drive
USB flash drive incompatible with GENIUS 2 Checker/Calibrator Use different USB flash drive
GENIUS 2 Checker / Calibrator has not had enough time to detect
the newly inserted USB flash drive Retry write process
Multiple GENIUS 2 thermometers
fail recalibration Probe cover stuck in one of the blackbody targets Remove probe cover from blackbody target using small needle-nose pliers or
hemostats. Do not scratch the interior surface of the target.
Dirty probe cover Use a new probe cover for each reading. Do not re-use a probe cover. Probe
covers should never be touched by hand.
Dirty thermometer lens Clean the thermometer lens per the instructions in the GENIUS 2 operating
manual.
GENIUS 2 thermometer not seated correctly Ensure that the GENIUS 2 thermometer is completely seated in the target and
at the proper angle.
GENIUS 2 thermometers, probe covers, and/or checker/calibrator
not at room temperature. Ensure that the GENIUS 2 thermometers, probe covers, and/or checker/
calibrator have had enough time to equilibrate to room temperature before
proceeding.
Error 6 repeatedly appears while
testing the same GENIUS 2
thermometer
Potential power consumption issue with GENIUS 2 thermometer Repeat calibration check / recalibration with a different GENIUS 2
thermometer and return problem thermometer for factory servicing.
GENIUS 2 thermometer seems
to read high or low following a
successful recalibration
Probe cover reused during calibration check / recalibration Repeat calibration check / recalibration using a new probe cover prior to
every target insertion during the process.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

11
GENIUS™2Checker/Calibrator Operation and Service Manual
Section XI - Customer Service
The GENIUS 2 Checker/Calibrator requires factory calibration and inspection once a year. Notify and return the device to Tyco
Healthcare/ Kendall on or before the calibration date for the Checker/Calibrator. The Checker/Calibrator will display a message one
month prior to the calibration date as a reminder.
CAUTION: THERE ARE NO SERVICEABLE PARTS INSIDE, RETURN TO FACTORY FOR SERVICE.
In the event that it is necessary to return a unit for repair, please observe the following:
1. Contact Tyco Healthcare/Kendall as shown below, or your local customer service representative for an Authorized Return Number
and shipping instructions.
2. Ship insured parcel to your local service contact or the appropriate location below.
United States Canada Outside US & Canada
Tyco Healthcare Group LP Tyco Healthcare Canada Tyco Healthcare Group LP
98.6 Faichney Drive 7300 Trans Canada Highway Unit 2 Talisman Business Center, London Road
Watertown, NY 13601 Pointe-Claire, QC Bicester, OX26 6HR, UK
1-800-448-0190 or H9R 1C7 +44-1869-328065
(315) 788-5246 1-877-664-8926 or
(514) 695-1220
Parts Listing
Please contact your local customer service center or sales representative for the parts listed below.
Description
GENIUS 2 Checker/Calibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303097
Refurbished GENIUS 2 Checker/Calibrator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303096
Section XII - Warranty
Tyco Healthcare/Kendall warrants the GENIUS 2 Checker/Calibrator to be free of defects in materials and workmanship under normal
use and service for a period of 1 year from the date of delivery by Tyco Healthcare/Kendall to the rst purchaser. Liability hereunder
is limited to repair or replacement, at Tyco Healthcare/Kendall’s option, at Tyco Healthcare/Kendall service facility for any product,
which shall under normal use and service, appear to Tyco Healthcare/Kendall to have been defective in material or workmanship. This
warranty shall not apply to, and Tyco Healthcare/Kendall shall not be responsible for, any loss arising in connection with the purchase
of any product that has been repaired by anyone other than Tyco Healthcare/Kendall or its authorized representative, or which has
been subject to misuse, neglect, or accident, or which has been used otherwise than in accordance with the instructions furnished by
Tyco Healthcare/Kendall. Tyco Healthcare/Kendall neither assumes nor authorizes any representative or other person to assume for it
any liability other than as expressly set forth herein.
Tyco Healthcare/Kendall DISCLAIMS ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OR APPLICATION OTHER THAN AS EXPRESSLY SET FORTH IN THE PRODUCT
LABELING. IN NO EVENT WILL Tyco Healthcare/Kendall BE LIABLE FOR ANY INCIDENTAL, INDIRECT, OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE PURCHASE OR USE OF THIS PRODUCT.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

12
GENIUS™2Checker/Calibrator Operation and Service Manual
Section XIII - Electromagnetic Conformity Declaration
The Genius 2 Checker/Calibrator has been built and tested according to IEC/EN61010-1 and EN61326-1 standards
The Checker/Calibrator is intended for use in the electromagnetic environment specied below. The user of the equipment should
assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment -Guidance
Conducted and Radiated
Emissions (EN61326-1) Group 1, Class B The Checker / Calibrator emissions are very low and are not likely
to cause interference in nearby electronic equipment.
Harmonic Current
(EN61000-3-2) Class A
Voltage Fluctuation and Flicker
(EN61000-3-3) Complies
Radiated Disturbance Immunity
(EN61000-4-3) Complies
Conducted Disturbance
Immunity (EN61000-4-6) Complies
Power Frequency Magnetic Field
Immunity (EN61000-4-8) Complies
Voltage Dips and Interrupts
Immunity (EN61000-4-11) Complies
Electrical Fast Transient/Bursts
Immunity (EN61000-4-4) Complies
Surge Immunity (EN61000-4-5) Complies
Electrostatic Discharge
(EN61000-4-2) Meets Criteria C,
Non-Continuous
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

JOB NO. ......................................................................................................16245
ARTWORK NO. ................................................................................... 182486-02
PACKAGE TYPE.....................................................................................CD Manual
SIZE..................................................................................................... 8.5" X 11"
PRINTED........................................................................................................ N/A
INK COLOR..................................................................................................BLACK
BAR CODE INK COLOR......................................................................................N/A
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

Table des matières
Page
Section I – Description fonctionnelle............................................................................................................................................. 1
Section II - Précautions et avertissements.................................................................................................................................... 2
Section III – Description des icônes................................................................................................................................................ 3
Section IV – Matériel requis ............................................................................................................................................................ 4
Section V - Configuration du système ............................................................................................................................................ 4
Section VI - Procédure opérationnelle........................................................................................................................................... 5
Brancher........................................................................................................................................................................................ 5
Ecran de démarrage ...................................................................................................................................................................... 5
Ecran Rappel ................................................................................................................................................................................. 5
Menu principal.............................................................................................................................................................................. 5
Préférences.................................................................................................................................................................................... 5
Sélection de la langue................................................................................................................................................................... 5
Dénir les informations sur l’utilisateur/site ................................................................................................................................. 5
Régler la date / heure.................................................................................................................................................................... 6
Contrôler l’étalonnage................................................................................................................................................................... 6
Section VII – Messages d’erreur....................................................................................................................................................... 8
Section VIII – Caractéristiques de l’appareil de vérification/calibrateur ................................................................................. 9
Section IX - Instructions de nettoyage......................................................................................................................................... 10
Section X — Dépannage................................................................................................................................................................ 10
Section XI – Service client.............................................................................................................................................................. 11
Section XII - Garantie...................................................................................................................................................................... 11
Section XIII - Déclaration de conformité électromagnétique .................................................................................................. 12
Ce produit contient un logiciel détenu uniquement par Tyco Healthcare Group LP. Tyco Healthcare Group LP accorde à l’utilisateur une
licence limitée et non exclusive lui permettant d’utiliser le logiciel en fonction des instructions de fonctionnement. Il est possible de se
procurer un exemplaire de cette licence auprès de Tyco Healthcare Group LP.
Français
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

1
GENIUS™2Appareil de vérification/Calibrateur Manuel d’utilisation et d’entretien
Section I - Description fonctionnelle
Ce manuel décrit le fonctionnement de l’appareil de vérication/calibrateur GENIUS 2. Cet instrument a été développé pour vérier
la précision du thermomètre auriculaire GENIUS 2 et réétalonne automatiquement le thermomètre, si nécessaire. De plus, l’appareil
de vérication/calibrateur GENIUS 2 peut générer un rapport de test pour chaque thermomètre testé et l’enregistrer sur une clé USB.
L’appareil de vérication/calibrateur GENIUS 2 fonctionnera avec toutes les versions logicielles de thermomètres GENIUS 2.
L’appareil de vérication/calibrateur GENIUS 2 contient deux cibles d’étalonnage infrarouges contrôlées indépendamment qui sont
similaires aux cibles d’étalonnage usine. Ces cibles « corps noir » sont conçues pour orir un transfert thermique radiatif ecace et
produire un rayonnement infrarouge proche de l’oreille humaine à des températures de 32,2° C et 40,6° C.
™
GENIUS
F/C˚
F/C˚
1 2 3 4
8910
5
6
7
11
Checker /Calibrator
F 3.15A 250V
™
Figure 1 - Disposition du panneau avant
Légende
1 Prise d’alimentation 7 Cible haute température
2 Interrupteur d’alimentation 8 DEL cible haute température
3 Porte-fusibles 9 Achage à cristaux liquides
4 DEL cible basse température 10 Bouton de commande
5 Cible basse température 11 Port USB
6 Connecteur du thermomètre GENIUS 2
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

2
GENIUS™2Appareil de vérification/Calibrateur Manuel d’utilisation et d’entretien
Section II - Précautions et avertissements
• L’appareil de vérification/calibrateur GENIUS 2 est conçu pour être utilisé comme indiqué dans ce manuel. La protection fournie
par les dispositifs de sécurité dans l’appareil de vérification/calibrateur GENIUS 2 peut être mise à mal si le dispositif n’est pas
utilisé conformément aux instructions contenues dans ce manuel.
• Le boîtier de l’appareil de vérification/calibrateur GENIUS 2 n’est pas étanche. Eviter de renverser du liquide, de quelque nature
que ce soit, sur le panneau avant de l’instrument. De même, l’appareil de vérification/calibrateur GENIUS 2 est alimenté depuis
une source c.a. Eviter tout contact potentiel entre du liquide et toute surface externe de l’instrument lorsque ce dernier est relié
au secteur.
• La qualité de la surface de l’intérieur des cibles corps noir de référence est essentielle à leur fonctionnement en tant que
radiateurs infrarouges efficaces. Lorsque l’on travaille à proximité de l’entrée des cibles corps noir, il est impératif de s’assurer
qu’aucun contact n’est réalisé avec la surface à l’intérieur des rabats en caoutchouc.
• Si un fusible vient à griller, le remplacer par un fusible correspondant exactement à la tension nominale, à l’intensité et aux
dimensions spécifiées dans la section Caractéristiques de l’appareil de vérification/calibrateur GENIUS 2 de ce manuel. Avant de
changer un fusible grillé, s’assurer que l’instrument est déconnecté de la source d’alimentation c.a.
• Cet instrument ne doit pas être utilisé en présence d’anesthésiques inflammables. Ces anesthésiques génèrent un risque
d’explosion.
• Aucune des pièces de cet instrument ne peut être réparée par l’utilisateur. Ne pas retirer le panneau avant. Confier l’entretien à
un technicien qualifié.
• Utiliser uniquement le module d’alimentation fourni avec cet instrument.
• Après avoir été invité à insérer le thermomètre GENIUS 2 dans une des cibles corps noir, l’utilisateur dispose de 60 secondes. La
procédure est interrompue si ce délai est dépassé.
• Avant de vérifier/d’étalonner un thermomètre GENIUS 2, s’assurer que la lentille du thermomètre est propre. Un nettoyage
efficace assurera que les tolérances d’étalonnage sont correctement vérifiées. Se reporter aux instructions de nettoyage du
manuel d’utilisation du thermomètre Genius 2.
• Lors de la mise en place d’une protection de sonde GENIUS 2 sur l’extrémité du thermomètre GENIUS 2, utiliser uniquement
des embouts se trouvant dans une cassette et non libérée. Essayer d’éviter tout contact manuel avec les embouts afin d’empêcher
toutes valeurs imprécises. Ne pas réutiliser les embouts. N’utiliser que des embouts GENIUS 2.
• Avant de l’utiliser, accorder au moins 15 minutes de préchauffage à l’appareil de vérification/calibrateur GENIUS 2 après sa mise
sous tension. S’assurer qu’un délai suffisant a été accordé aux thermomètres GENIUS 2, aux embouts et à l’appareil de vérification/
calibrateur GENIUS 2 pour s’équilibrer à la température ambiante avant de poursuivre. Accorder une période de réchauffement
supérieure si l’appareil ou les thermomètres se trouvaient aux limites extrêmes de la température ambiante autorisée. La
portabilité de l’appareil de vérification/calibrateur GENIUS 2 permettra un repositionnement de l’appareil dans des zones
répondant aux exigences de température ambiante.
• Pour maintenir le respect de la compatibilité électromagnétique (CEM), utiliser une clé USB marquée CE et conforme à la norme
EN55022.
• Ne jamais placer ou poser le thermomètre GENIUS 2 sur la surface de l’appareil de vérification/calibrateur GENIUS 2. Les capteurs
thermiques à l’intérieur de l’extrémité de la sonde du thermomètre détectent la chaleur transmise des cibles corps noir via
la surface du panneau avant. Cela entraîne une instabilité des capteurs du thermomètre, qui produisent alors des valeurs de
température erronées.
• L’appareil de vérification/calibrateur GENIUS 2 ne peut être utilisé qu’en intérieur, dans une zone à l’abri des courants d’air et
d’importantes variations de la température ambiante.
• Le port USB sur le panneau avant est conçu pour être utilisé uniquement avec une clé USB. Aucun autre périphérique USB ne doit
être inséré dans ce port USB. La clé USB doit être installée avant la mise sous tension de l’instrument.
• Dans l’éventualité où l’appareil de vérification/calibrateur GENIUS 2 tomberait, retourner l’unité à l’usine pour un nouvel
étalonnage.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

3
GENIUS™2Appareil de vérification/Calibrateur Manuel d’utilisation et d’entretien
Section III - Description des icônes
SN
Numéro de série
CE MARK SYMBOLS
NON-CE MARK SYMBOLS
ELECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS
Corrugated
Recycles
STERILE
Keep Away
From Sunlight Keep Dry
Single Use
LOT
0123
FRAGILE
THIS SIDE UP
DEHP–FREE
0°
40°
Avoid Temperature
Extremes
Do not use if package is
opened or damaged.
STERILE R
A
STERILE
EOSTERILE
STERILE
STERILE R
A
STERILE
EOSTERILE
Humidity Limitation
Atmospheric pressure limitation
Use By
C US
LATEX–FREE
Caution, consult
accompanying
documents
Consult
Instructions
For Use
MR
MR UNSAFE NON-PYROGENIC
PROTECT FROM HEAT AND
RADIOACTIVE SOURCES
NON
STERILE
2
STERILIZE
DO NOT RESTERILIZE
POWDER–FREE
POWDER
Fabricant
Consulter la
documentation jointe
CE MARK SYMBOLS
NON-CE MARK SYMBOLS
ELECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS
Corrugated
Recycles
STERILE
Keep Away
From Sunlight Keep Dry
Single Use
LOT
0123
FRAGILE
THIS SIDE UP
DEHP–FREE
0°
40°
Avoid Temperature
Extremes
Do not use if package is
opened or damaged.
STERILE R
A
STERILE
EOSTERILE
STERILE
STERILE R
A
STERILE
EOSTERILE
Humidity Limitation
Atmospheric pressure limitation
Use By
C US
LATEX–FREE
Caution, consult
accompanying
documents
Consult
Instructions
For Use
MR
MR UNSAFE NON-PYROGENIC
PROTECT FROM HEAT AND
RADIOACTIVE SOURCES
NON
STERILE
2
STERILIZE
DO NOT RESTERILIZE
POWDER–FREE
POWDER
Date de fabrication
Alimentation de secours Isolation double ou
renforcée
Remplacer par un fusible
de même intensité Marque CE
Port USB Remplacer par un
fusible de même
intensité
F/C˚
Cible haute température
LISTED
(LABORATORY EQUIPMENT)
46EA
Informations relatives
à la classication UL
F/C˚
Cible basse température
Mise en garde :
utilisation en interne
uniquement
Connexion de
thermomètre GENIUS 2
™
Le thermomètre infrarouge auriculaire GENIUS 2 est un dispositif able et précis de relevé de la température. Sa conception, ses
procédures d’étalonnage contrôlées et les vérications très strictes réalisées lors de sa fabrication font du thermomètre GENIUS 2
un appareil d’extrême précision. La qualité et la précision du thermomètre reposent sur un certain nombre de pièces essentielles,
notamment les embouts ACCUSYSTEM√ de Tyco Healthcare/Kendall. Lorsqu’ils sont placés sur un thermomètre GENIUS 2, ces
embouts servent de barrière anti-infection entre le patient et le thermomètre, mais aussi de vecteur de transmission thermique
entre eux. L’utilisation d’un embout ACCUSYSTEM√ pour thermomètre GENIUS 2 est extrêmement importante an d’éviter la
contamination de l’appareil et de garantir la prise de températures précises.
Si le thermomètre infrarouge auriculaire GENIUS 2 et les embouts ACCUSYSTEM√ pour thermomètre GENIUS 2 répondent bien aux
attentes de nos clients, c’est parce qu’ils sont associés dès le processus de fabrication.
Pour réduire au minimum les variations et assurer la fabrication de produits uniformes et ables, les paramètres de moulage des
embouts ACCUSYSTEM√ pour le thermomètre GENIUS 2 sont étroitement contrôlés par Tyco Healthcare/Kendall. Les matières
thermoplastiques employées sont de classes particulières. Elles sont strictement contrôlées et leur sélection rigoureuse permet
d’obtenir des mesures de température uniformes sur les thermomètres GENIUS 2. L’étalonnage en usine et la certication nale de
chaque thermomètre GENIUS 2 s’appuient sur l’utilisation d’embouts ACCUSYSTEM√ pour thermomètre GENIUS 2 de Tyco Healthcare/
Kendall acceptables. L’utilisation d’embouts génériques ou de marque autre que Tyco Healthcare/Kendall n’est pas recommandée ni
agréée par Tyco Healthcare/Kendall. L’utilisation d’embouts non certiés pour le thermomètre GENIUS 2 risque d’altérer la précision
de ses mesures. Pour être certain d’obtenir des températures précises à chaque utilisation, vérier sur le carton des protections de
sonde la présence du logo ACCUSYSTEM√.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

4
GENIUS™2Appareil de vérification/Calibrateur Manuel d’utilisation et d’entretien
Section IV - Matériel requis
1. Appareil de vérication/Calibrateur GENIUS 2
2. Module d’alimentation avec adaptateur c.a.
3. Câble de la sonde
4. Manuel d’utilisation
5. Thermomètres GENIUS 2 à tester
6. Embouts GENIUS 2 jetables
7. Petit tournevis cruciforme
Section V - Configuration du système
Avant d’utiliser l’appareil de vérification/calibrateur GENIUS 2, procéder comme suit :
1. Connecter sur le module d’alimentation la prise de courant à ches c.a. correspondant au pays d’utilisation.
2. Connecter le câble du module d’alimentation sur la prise sur le panneau avant.
3. Connecter le module d’alimentation sur une prise de courant c.a.
4. Mettre l’unité sous tension via le commutateur sur le panneau avant.
5. Accorder au moins 15 minutes de préchauage à l’appareil de vérication/calibrateur GENIUS 2.
6. Mettre de côté une cassette contenant des embouts GENIUS 2 neufs.
7. Attendre que les thermomètres GENIUS 2 s’équilibrent à la température ambiante.
8. Préparer les thermomètres GENIUS 2 en nettoyant soigneusement les lentilles avant utilisation. Consulter les instructions de
nettoyage dans le manuel d’utilisation du thermomètre GENIUS 2.
9. Retirer le panneau d’accès au compartiment des piles à l’aide d’un petit tournevis cruciforme. Retirer les trois piles AAA.
10. Connecter le connecteur rond du câble de la sonde sur le panneau avant de l’appareil de vérication/calibrateur GENIUS 2.
11. Connecter l’autre extrémité du connecteur de la sonde dans le compartiment des piles. Le connecteur se mettra en place là où les
piles se trouvaient. Voir Figure 2.
Figure 2 - Insertion du connecteur de sonde dans le compartiment des piles GENIUS 2
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
Table of contents
Languages:
Other Tyco Test Equipment manuals
Popular Test Equipment manuals by other brands

National Instruments
National Instruments NI PXI-5105 CALIBRATION PROCEDURE

Q3 Innovations
Q3 Innovations AlcoHAWK ABI owner's manual

barfield
barfield 2548H Instruction supplement

PIE
PIE PIECAL 334Plus operating instructions

MSG
MSG MS015 COM user manual

Commercial Electric
Commercial Electric MS8900H Use and care guide