PATIENT
QUICK START GUIDE
TRIDENT™NANO 3-LEAD LANYARD CABLE
For clinical support, please contact your medical provider.
For technical support related to your cardiac monitoring device, please call (971) 250-4237 or email
clearbeatsuppor[email protected]. Include your name and preferred contact method (phone, email, other), and a
support representative will respond as soon as possible.
Step 4: Insert Device
Slide the device into the patient
lanyard cable. Hold the device in
place until the green light turns off.
Push
Step 9: Charging
When the device requires charging
(alternating G-Y-G-Y-G light),
remove the device from the lanyard
lead set and insert into the
charging dock. A solid
green light will indicate
that the device is fully
charged.
Step 3: Application
Remove backing from the adhesive
electrodes and apply to the prepped
and dried sites. Make sure to press
the border firmly and match the
locations shown below.
Step 1: Skin Prep
Remove all hair from electrode sites
with a razor and scrub the electrode
application sites with alcohol wipe,
or warm soapy water. Dry area
thoroughly.
Step 5: Check Status
Check the recording status by briefly
pressing the button. The device will
blink 5 times with the color of light
indicating the status as shown to the
right.
Step 6: Status Lights
Recording Electrode has poor
connectivity
Recording is
complete
Blinks green 5 times Blinks yellow 5 times
Blinks red 5 times
Device requires
charging
Blinks green & yellow
Step 7: Record Event
If you feel a symptom while wearing
the monitor, hold the button for 4
seconds to record the event. The
light will blink green and an “Event
Recorded” tone will play.
Step 8: Remove Device
Eject the device by firmly pressing
the front edge near the
status/event button and slide the
device upward.
Push
Step 2: Electrodes
Connect color coded snap
connectors to the adhesive
electrodes. Place lanyard around
the neck.
TRIDENT™NANO
What you need to know
while monitoring...
TZ Medical Inc.
17750 SW Upper Boones Ferry Rd. #150
Portland, OR 97224
www.tzmedical.com
19919 R0.4
Allow electrode sites to dry completely and remove all residue before applying adhesive
electrodes.
Avoid exposing the device and lanyard cable lead set to water and/or excessive amounts of
moisture. Disconnect all leads from the electrodes adhered to the chest and completely
remove the lanyard cable prior to showering. Once the electrodes and surrounding areas are
completely dry, place the lanyard around the neck, and reconnect the leads to the
electrodes to continue monitoring.
While inserting the recorder, firmly press and hold the exposed edge of the device closest
to the LED into the lanyard cable until the green light turns off to ensure the device begins
recording properly.
It may be difficult to observe the LED while wearing the device. Viewing the LED can easily
be done by pressing the button while standing in front of a mirror.
If you need to apply new electrodes for an extended study, recharge the device and
repeat the instructions for application.
It is recommended to take a warm shower during the removal process of the electrodes.
If available, apply mineral oil as you peel the electrodes away to loosen the adhesive.

Other TZ Medical Medical Equipment manuals

TZ Medical TRIDENT NANO User manual

TZ Medical

TZ Medical TRIDENT NANO User manual

TZ Medical ARC User manual

TZ Medical

TZ Medical ARC User manual

TZ Medical Trident Pro User manual

TZ Medical

TZ Medical Trident Pro User manual

TZ Medical Aera CT 3G User manual

TZ Medical

TZ Medical Aera CT 3G User manual

Popular Medical Equipment manuals by other brands

Gaumard S230.40 user guide

Gaumard

Gaumard S230.40 user guide

Dräger Medical Isolette C2000 operating instructions

Dräger Medical

Dräger Medical Isolette C2000 operating instructions

axess vision Screeni 30030001 user manual

axess vision

axess vision Screeni 30030001 user manual

lenstar LS 900 Instructions for use

lenstar

lenstar LS 900 Instructions for use

Bowa SHE SHA operating manual

Bowa

Bowa SHE SHA operating manual

Beurer medical PO 30 instruction manual

Beurer medical

Beurer medical PO 30 instruction manual

Ocutech VES II Fittings guide

Ocutech

Ocutech VES II Fittings guide

AURICAL SpeechLink 100 Reference manual

AURICAL

AURICAL SpeechLink 100 Reference manual

Great Lakes Kinesia 360 user guide

Great Lakes

Great Lakes Kinesia 360 user guide

IMP De Mayo D2 Knee Positioner Instructions for use

IMP

IMP De Mayo D2 Knee Positioner Instructions for use

Power Plate Pro7M Instructions for use

Power Plate

Power Plate Pro7M Instructions for use

GE Vivid E9 Service manual

GE

GE Vivid E9 Service manual

Fritonex PowerTube QuickZap instruction manual

Fritonex

Fritonex PowerTube QuickZap instruction manual

Dräger Fabius GS SW 3.n Reprocessing guide

Dräger

Dräger Fabius GS SW 3.n Reprocessing guide

Medline Microban MDS86960KDMB troubleshooting guide

Medline

Medline Microban MDS86960KDMB troubleshooting guide

Zimmer Biomet Tricera instruction manual

Zimmer Biomet

Zimmer Biomet Tricera instruction manual

Graham Field intensa IN587WD Series Assembly and operating instructions

Graham Field

Graham Field intensa IN587WD Series Assembly and operating instructions

IngMar Medical ASL 5000 Quick setup guide

IngMar Medical

IngMar Medical ASL 5000 Quick setup guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.