Ultimate Healthcare Tamora III Advance User manual

Tamora III Advance
Quick Reference Guide
Please refer to the product User Manual for full instructions for use.
Control Panel Lock Out
To unlock panel:
•Press & hold for 3 seconds to unlock control panel.
To lock panel:
•To lock out the control panel press the Lock Out
button and all functions will be locked.
•If the control panel is not used for a period of 30
seconds it will automatically lock out and inhibit
the use of the functions.
Alternating Mode
Static Mode
•Pressing the Mode button until the Static indicator
illuminates puts the system into Static mode.
•The system will automatically return to
Alternating mode after 20 minutes.
Alternating Cycle Time selection
•Choice of 10/15/20/25/30 minutes.
•To select press Cycle button until required cycle
time is displayed in LCD screen.
Auto Firm Mode
•
•
Pressing the Auto Firm button quickly inflates
all cells to a single pressure setting.
The system will automatically return to
Last Mode of Operation after 20 minutes.
Comfort Control
The Soft and Firm buttons allow adjustment of
pressures within a safe pre-set range.
•When pressing the Firm button, the output
pressure will increase to provide a higher
pressure output and thus increased support.
•When pressing the Soft button, the output
pressure will be decreased to provide a lower
pressure output and thus increased comfort.
•Alternatingmode is the default mode for the
and deflate in an alternating 1-in-3 cell cycle.
Function Mode Button
•Pressing theModebutton allows you
to cycle between the different modes
available such as Alternating, Static and
Constant Low Pressure.
Alarms
Alarm Mute
The Alarm Mute button temporarily resets the
audible Low Pressure/Power Failure/Service
alarms. Should the situation not be resolved and
the fault condition continues the alarm will resume.
Low Pressure Alarm
•Upon detection of low pressure, an audible alarm will be
heard and the Low Pressure indicator will illuminate.
Power Failure Alarm
•If mains power is removed from power unit or the power cable is
unplugged without turning the power unit OFF, an audible alarm
will sound and the Power Failure indicator will illuminate.
Service Alarm
•Should your system develop a fault condition whilst in use, an
audible alarm will be heard and the Service indicator will illuminate.
Constant Low Pressure (CLP) Mode
www.ultimatehealthcare.co.uk | Tel: 0333 321 8996
Auto-Detection -
detecting patient weight (distribution)
•Pressing the Mode button until the CLP indicator
illuminates puts the system into Constant Low Pressure
mode.
•Within this mode the cells are not alternating and the
internal cell pressure within the mattress is set at 15mmhg
to provide a comfortable pressure reducing surface.
•The Alternating cycle will continue at the selected
cycle time until another mode is selected.
•
•
Simultaneously pressing the Soft and Firm buttons on the
control panel will manually activate Auto-Detection.
2QFHDFWLYDWHGWKHLQGLFDWRUOLJKWZLOOIODVKDQGWKHV\VWHP
will start detecting the patients weightGLVWULEXWLRQ to set
internal pressuresappropriately. The system will
continuously check this every 2 hours.

Troubleshooting
(1) No lights on power unit
•Check the power unit is connected to the mains power
supply and that the mains switch is turned ON. (3)Service indicator remains illuminated
•
•
•Check that the connection between air tube and power unit
is tightly secured.
•Indicates that system is scheduled for an annual service.
Please contact Ultimate Healthcare or your local dealer to
arrange.
Check to ensure that the CPR valve is in the closed position.
Check whether power was suddenly shut down.
(2)Low Pressure indicator is flashing and sounding
•Check that all tubing connections along the mattress are
secured.
•If all of above steps have been checked. Press Alarm Mute
(4)Power failure alarm indicator is flashing and sounding
•Check the power unit is connected to the mains power
supply and that the mains switch is turned ON.
Mattress Setup
Cable Management System Cable management fixings are
located on each side of the mattress underneath the flap of the
cover. The mains power cable should be secured through the
cable management fixing as follows:
•Locate each cable management fixing.
•If necessary, open the press studs.
•Run the mains power cable along the side of the mattress
securing each fixing loop around the cable using the press studs.
Security Straps
There are security straps fitted to the base cover of the
mattress which should be fastened loosely to convenient
points on the bedframe patient surface.
Ensure when fixing the mattress to the bed the security straps
are only fitted to the moving parts of the mattress platform.
Connecting the Mattress to the Pump
Remove the cover of the air pipe connector and connect the air
pipe connector to the Pump. When turning the power on, the
pump goes directly into auto firm to inflate the mattress. When
fully inflated it reverts to the initial setting of 10 minutes. At this
point the comfort control can be set to the desired weight or
activate the auto detection.
Image showing 3 of 6 straps. All 6 straps should be fastened
Setting up the pump
Whilst holding the pump, unfold the
hanging hooks on the rear of the Pump
and hang it from the bed’s footboard.
www.ultimatehealthcare.co.uk | Tel: 0333 321 8996
This Quick Reference Guide (QRG) has been produced to assist users with only the functional aspects of the system
Transport Mode
Transport mode will maintain the
cells in their present state for up
to 12 - 24 hours.
1.
2.
To reset CPR:
1. Ensurethe dial of the CPR attachment is rotated until it
‘clicks’ into the CLOSED position
CPR valve is located at the head end on the left hand side
of the mattress.
Gently pull and rotate the dial of the CPR valve to ‘click’
into the OPEN position. At the same time, disconnect the
air pipe connector from the Pump to speed up the air
release. The mattress deflates in 10-15 seconds.
CPR
In the event of a patient suffering
a cardiac arrest and CPR needing
to be administered:
To activate the CPR:
•Simply disconnect air pipe connector
and place the connector cover over
it. The air pressures in the mattress
will now equalise, but maintain a
degree of comfort.
•It is important to restore the mattress as quickly as possible
by reconnecting the supply tubes to the power unit.
Ensure power cable the air output
pipe are routed in such a manner so
that they cannot be twisted, trapped, crushed or stressed.
Issue 4: 01/2021
UPRA367803A-3.QRG-4
Secure the mattress to the moving parts of the bedframe
using the securing straps, and when the system is fully
inflated it is ready for the patient to be positioned onto the
mattress.
Lay the mattress on the bed patient surface ensuring that the
air pipe connector is situated on the bottom left hand side of
the bed foot end (as viewed from the foot of the bed).