Troubleshooting
(1) No lights on power unit
•Check the power unit is connected to the mains power
supply and that the mains switch is turned ON. (3)Service indicator remains illuminated
•
•
•Check that the connection between air tube and power unit
is tightly secured.
•Indicates that system is scheduled for an annual service.
Please contact Ultimate Healthcare or your local dealer to
arrange.
Check to ensure that the CPR valve is in the closed position.
Check whether power was suddenly shut down.
(2)Low Pressure indicator is flashing and sounding
•Check that all tubing connections along the mattress are
secured.
•If all of above steps have been checked. Press Alarm Mute
(4)Power failure alarm indicator is flashing and sounding
•Check the power unit is connected to the mains power
supply and that the mains switch is turned ON.
Mattress Setup
Cable Management System Cable management fixings are
located on each side of the mattress underneath the flap of the
cover. The mains power cable should be secured through the
cable management fixing as follows:
•Locate each cable management fixing.
•If necessary, open the press studs.
•Run the mains power cable along the side of the mattress
securing each fixing loop around the cable using the press studs.
Security Straps
There are security straps fitted to the base cover of the
mattress which should be fastened loosely to convenient
points on the bedframe patient surface.
Ensure when fixing the mattress to the bed the security straps
are only fitted to the moving parts of the mattress platform.
Connecting the Mattress to the Pump
Remove the cover of the air pipe connector and connect the air
pipe connector to the Pump. When turning the power on, the
pump goes directly into auto firm to inflate the mattress. When
fully inflated it reverts to the initial setting of 10 minutes. At this
point the comfort control can be set to the desired weight or
activate the auto detection.
Image showing 3 of 6 straps. All 6 straps should be fastened
Setting up the pump
Whilst holding the pump, unfold the
hanging hooks on the rear of the Pump
and hang it from the bed’s footboard.
www.ultimatehealthcare.co.uk | Tel: 0333 321 8996
This Quick Reference Guide (QRG) has been produced to assist users with only the functional aspects of the system
Transport Mode
Transport mode will maintain the
cells in their present state for up
to 12 - 24 hours.
1.
2.
To reset CPR:
1. Ensurethe dial of the CPR attachment is rotated until it
‘clicks’ into the CLOSED position
CPR valve is located at the head end on the left hand side
of the mattress.
Gently pull and rotate the dial of the CPR valve to ‘click’
into the OPEN position. At the same time, disconnect the
air pipe connector from the Pump to speed up the air
release. The mattress deflates in 10-15 seconds.
CPR
In the event of a patient suffering
a cardiac arrest and CPR needing
to be administered:
To activate the CPR:
•Simply disconnect air pipe connector
and place the connector cover over
it. The air pressures in the mattress
will now equalise, but maintain a
degree of comfort.
•It is important to restore the mattress as quickly as possible
by reconnecting the supply tubes to the power unit.
Ensure power cable the air output
pipe are routed in such a manner so
that they cannot be twisted, trapped, crushed or stressed.
Issue 4: 01/2021
UPRA367803A-3.QRG-4
Secure the mattress to the moving parts of the bedframe
using the securing straps, and when the system is fully
inflated it is ready for the patient to be positioned onto the
mattress.
Lay the mattress on the bed patient surface ensuring that the
air pipe connector is situated on the bottom left hand side of
the bed foot end (as viewed from the foot of the bed).