Entering the Abdomen: The ‘Half Closed’ Position
A feature of the Sterishot II Applicator is a white visual guide located within
the trigger housing, designed to help signal the ‘Half Closed’ position and
hence avoid prematurely closing or distorting the Filshie Clip so that it may
fail to close properly. To ensure the ‘Half Closed’ position is not exceeded, the
edge of the black polished section of the trigger should not pass beyond the
white visual guide located inside the black trigger housing.
Application of the Filshie Clip
• IdentifyandinspecttheFallopiantubethoroughly.
• EnsurethattheFilshieClipcanaccommodatethewholediameterofthe
Fallopiantube.
• LocatetheFilshieClipovertheisthmicportionoftheFallopiantube,1-2cm
fromthecornu.
• HavingestablishedthebestlocationfortheFilshieClipafterinsertion
throughthetrocar,theapplicatorshouldbere-openedandadvancedafew
millimeterstomovetheFallopiantubegentlytothebackoftheFilshieClip,
closetothehinge.
• ClosetheFilshieClipintopositionbyapplyingrm,butgentlepressureon
thetriggerinasmoothactionuntilthetriggerreachesitsmechanicalstop.
• WhentheFilshieClipissecuredinposition,gentlyreleasethetriggerandthe
FilshieClipwillautomaticallyfreeitselffromtheapplicator‘ClipTray’.
• Donotuseanabruptclosureactionorthetubemightbetransected.Should
thisoccur,applyasecondFilshieClipontheproximal(uterine)sideofthe
transection.
Digital photography and video recording of the closure process are encouraged to
support the patient record case file.
Removal of the Applicator from the Abdomen
BeforetheSterishotIIApplicatorcanbewithdrawnfromthe
abdomenitmustbereturnedtothe‘HalfClosed’positionsoit
canpassthroughthecannula.
CorrectClosure
Polished Trigger
Section
White Visual Guide
Trigger Housing
Clipwithapplicator
halfclosed
Clipopenovertube Clipclosedontube
Applicatorjawsin
‘HalfClosed’position
(clipnotprematurelyclosedordistorted)