
6FINESSE+ Service Manual
Accessories. Any reusable accessories should be periodically tested for function and safety in
accordance with their manufacturers' instructions. Use only accessories whose connectors match
those on the generator. Adaptors should not be used unless they are approved by Utah Medical
Products as being compatible with the FINESSE+.
The FINESSE+ system should only be used with the accessories that are offered by Utah Medical
Products. However, it should be noted that certain accessories may appear to be physically com-
patible with the FINESSE+ system, but may not comply with quality and performance regulations,
and as such may limit or impede the intended performance and safety features of the FINESSE+
system.
To ensure that the FINESSE+ system complies with electromagnetic emissions and immunity regu-
lations, users should use only the switchpens and dispersive pads listed in the Accessories section
of this manual. Under no circumstances should switchpen cables exceed 10 ft (3m) in length, and
dispersive pad cables should never exceed 10 ft (3m) in length. The use of accessories with longer
cable lengths may result in increased emissions or decreased immunity of the FINESSE+ system.
Only use active electrodes and pencils that have Rated Accessory Voltage greater than the
Maximum Output Voltage for the selected output setting and mode (see Table 1 and Figure 8 in
the Technical Specifications section).
Power Deficiencies. The output setting selected should be appropriate for the intended purpose
(see the section titled "Electrosurgical Procedure Guidelines"). Always use the lowest possible
setting which achieves the desired excision quality and speed. An apparent power deficiency
in cutting or coagulation may indicate faulty application of the dispersive electrode or failure of a
patient lead. It may also result from attempting to push the electrode through the tissue faster
than the natural speed of the electrode for the output level set inside the generator. If a slower
cut does not produce the desired results, then the patient circuit, including the active and disper-
sive electrodes, should be checked before increasing the output settings.
Sparks. The sparks generated in electrosurgical cutting or coagulation can easily ignite flammable
substances at the surgical site. The use of flammable anesthetics or oxidizing gases such as
nitrous oxide (N2O) and oxygen should be avoided if a surgical procedure is carried out in the
region of the thorax or the head, unless these agents are drawn away. Non-flammable agents
should be used for cleaning and disinfection wherever possible. Flammable agents used for clean-
ing or disinfection or as solvents of adhesives should be allowed to evaporate before the applica-
tion of the electrosurgical device. There is a risk of pooling of flammable solution under the
patient or in body depressions such as the umbilicus and body cavities such as the vagina. Any
fluid pooled in these areas should be removed before the electrosurgical equipment is used.
Attention should be called to the danger of ignition of endogenous gases. Some materials, for
example cotton wool and gauze, when saturated with oxygen may be ignited by sparks produced
in normal use of the electrosurgical generator.
Filters. After use, the external disposable filter can be a source of odor and possible viral con-
tamination. Also, using this filter for too many procedures may compromise its particle or odor
removal efficiency. For optimum performance, this filter should be discarded either daily or after
15 procedures if more than 15 procedures are performed in a single day. It should be removed