Vega VEGABAR 82 User manual

Quick setup guide
Pressure transmitter with ceramic
measuring cell
VEGABAR 82
Slave sensor for electronic differential pres-
sure
Document ID: 46332

2
Contents
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
Contents
1 For your safety
1.1 Authorised personnel ....................................................................................................... 3
1.2 Appropriate use................................................................................................................ 3
1.3 Warning about incorrect use............................................................................................. 3
1.4 General safety instructions............................................................................................... 3
1.5 CE conformity................................................................................................................... 3
1.6 Permissible process pressure .......................................................................................... 4
1.7 Environmental instructions ............................................................................................... 4
2 Product description
2.1 Conguration.................................................................................................................... 5
3 Mounting
3.1 General instructions for use of the instrument .................................................................. 6
3.2 Ventilation and pressure compensation............................................................................ 6
3.3 Combination Master - Slave ............................................................................................. 6
4 Connecting to power supply
4.1 Connecting....................................................................................................................... 9
4.2 Single chamber housing................................................................................................. 10
4.3 Connection example ...................................................................................................... 11
5 Set up with the display and adjustment module
5.1 Parameter adjustment - Quick setup .............................................................................. 12
5.2 Parameter adjustment - Extended adjustment................................................................ 15
6 Supplement
6.1 Technical data ................................................................................................................ 19
Information:
This quick setup guide enables quick setup and commissioning of
your instrument.
You can nd supplementary information in the corresponding, more
detailed Operating Instructions Manual as well as the Safety Manual
that comes with instruments with SIL qualication. These manuals are
available on the supplied DVD or in the download area of "www.vega.
com".
Operating instructions VEGABAR 82 - Slave sensor for elec-
tronic dierential pressure: Document-ID 45050
Editing status of the quick setup guide: 2016-02-04

3
1 For your safety
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
1 For your safety
1.1 Authorised personnel
All operations described in this operating instructions manual must
be carried out only by trained specialist personnel authorised by the
plant operator.
During work on and with the device the required personal protective
equipment must always be worn.
1.2 Appropriate use
VEGABAR 82 is a slave sensor for electronic dierential pressure
measurement.
You can nd detailed information about the area of application in
chapter "Product description".
Operational reliability is ensured only if the instrument is properly
used according to the specications in the operating instructions
manual as well as possible supplementary instructions.
1.3 Warning about incorrect use
Inappropriate or incorrect use of the instrument can give rise to
application-specic hazards, e.g. vessel overll or damage to system
components through incorrect mounting or adjustment. Also the pro-
tective characteristics of the instrument can be inuenced.
1.4 General safety instructions
This is a high-tech instrument requiring the strict observance of stand-
ard regulations and guidelines. The user must take note of the safety
instructions in this operating instructions manual, the country-specic
installation standards as well as all prevailing safety regulations and
accident prevention rules.
The instrument must only be operated in a technically awless and
reliable condition. The operator is responsible for trouble-free opera-
tion of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new regula-
tions.
1.5 CE conformity
The device fullls the legal requirements of the applicable EC guide-
lines. By axing the CE marking, we conrm successful testing of the
product.
You can nd the CE Certicate of Conformity in the download section
of our homepage.

4
1 For your safety
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
1.6 Permissible process pressure
The permissible process pressure is specied on the type label with
"Process pressure", see chapter "Conguration". For safety reasons,
this range may not be exceeded. This applies even if a measuring cell
with a measuring range (order-related) higher than the permissible
pressure range of the process tting is installed.
1.7 Environmental instructions
Protection of the environment is one of our most important duties.
That is why we have introduced an environment management system
with the goal of continuously improving company environmental pro-
tection. The environment management system is certied according
to DIN EN ISO 14001.
Please help us fulll this obligation by observing the environmental
instructions in this manual:
• Chapter "Packaging, transport and storage"
• Chapter "Disposal"

5
2 Product description
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
2 Product description
2.1 Conguration
The type label contains the most important data for identication and
use of the instrument:
2
1
13
14
12
10
11
5
3
6
4
7
8
9
Fig. 1: Layout of the type label (example)
1 Instrument type
2 Product code
3 Field for approvals
4 Power supply and signal output, electronics
5 Protection rating
6 Measuring range
7 Permissible process pressure
8 Material, wetted parts
9 Order number
10 Serial number of the instrument
11 Data-Matrix-Code for smartphone app
12 Symbol of the device protection class
13 ID numbers, instrument documentation
14 Reminder to observe the instrument documentation
The type label contains the serial number of the instrument. With it
you can nd the following instrument data on our homepage:
• Product code (HTML)
• Delivery date (HTML)
• Order-specic instrument features (HTML)
• Operating instructions and quick setup guide at the time of ship-
ment (PDF)
• Order-specic sensor data for an electronics exchange (XML)
• Test certicate (PDF) - optional
Go to www.vega.com "VEGA Tools" and "Instrument search". Enter
the serial number.
Alternatively, you can access the data via your smartphone:
• Download the smartphone app "VEGA Tools" from the "Apple App
Store" or the "Google Play Store"
• Scan the Data Matrix code on the type label of the instrument or
• Enter the serial number manually in the app
Type label
Serial number - Instru-
ment search

6
3 Mounting
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
3 Mounting
3.1 General instructions for use of the instrument
Protect your instrument against moisture ingress through the following
measures:
• Use the recommended cable (see chapter "Connecting to power
supply")
• Tighten the cable gland
• When mounting horizontally, turn the housing so that the cable
gland points downward
• Loop the connection cable downward in front of the cable gland
This applies mainly to outdoor installations, in areas where humidity is
expected (e.g. through cleaning processes) and on cooled or heated
vessels.
3.2 Ventilation and pressure compensation
The lter element is mounted into the electronics housing. It has the
following functions:
• Ventilation of the electronics housing
• Atmospheric pressure compensation (with relative pressure meas-
uring ranges)
→ Turn the housing so that the lter element points downward after
the instrument is installed. This provides better protection against
buildup.
1 2 3
4 4 4
Fig. 2: Position of the lter element - non-Ex, Ex-ia version
1 Housing plastic, stainless steel precision casting
2 Housing aluminium
3 Housing stainless steel, electropolished
4 Filter element
With the following instruments a blind plug is installed instead of the
lter element:
• Instruments in protection IP 66/IP 68 (1 bar) - ventilation via capil-
laries in non-detachable cable
• Instruments with absolute pressure
3.3 Combination Master - Slave
In principle, any sensor combination within the VEGABAR 80 series is
allowed. The following requirements must be fullled:
Protection against mois-
ture
Instruments in non-Ex
and Ex-ia version

7
3 Mounting
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
• Conguration, Master sensor suitable for electronic dierential
pressure
• Pressure type is identical for both sensors, i.e. relative pressure/
relative pressure or absolute pressure/absolute pressure
• Master sensor measures the higher pressure
• Measurement setup as shown in the following chapters
The measuring range of each sensor is selected such that it ts the
measuring loop. For this, the max. recommended turn down must be
noted. See chapter "Technical data". It is absolutely necessary the the
measuring ranges of Master and Slave correspond.
Measurement result = Measured value of Master (total pres-
sure) - measured value of Slave (static pressure)
Depending on the application, individual combinations can result, see
following examples:
Data
Application: Level measurement
Medium: Water
Vessel height: 12 m, hydrostatic pressure = 12 m x 1000 kg/m3x
9.81 m/s2 = 117.7 kPa = 1.18 bar
Superimposed pressure: 1 bar
Total pressure: 1.18 bar + 1 bar = 2.18 bar
Instrument selection
Nominal measuring range Master: 2.5 bar
Nominal measuring range Slave: 1 bar
Turn Down: 2.5 bar/1.18 bar = 2.1 : 1
Data
Application: Level measurement
Medium: Water
Vessel height: 500 mm, hydrostatic pressure = 0.50 m x 1000 kg/m3x
9.81 m/s2 = 4.9 kPa = 0.049 bar
Superimposed pressure: 350 mbar = 0.35 bar
Total pressure: 0.049 bar + 0.35 bar = 0.399 bar
Instrument selection
Nominal measuring range Master: 0.4 bar
Nominal measuring range Slave: 0.4 bar
Turn Down: 0.4 bar /0.049 bar = 8.2 : 1
Data
Application: Dierential pressure measurement
Medium: Gas
Static pressure: 0.8 bar
Dierential pressure on orice: 50 mbar bar = 0.050 bar
Total pressure: 0.8 bar bar + 0.05 bar = 0.85 bar
Example - large vessel
Example - small vessel
Example - orice in
pipeline

8
3 Mounting
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
Instrument selection
Nominal measuring range Master: 1 bar
Nominal measuring range Slave: 1 bar
Turn Down: 1 bar/0.050 bar = 20 : 1
The measuring result (level, pressure dierence) as well as measured
value Slave (static or superimposed pressure) are outputted by the
sensor. Depending on the instrument version, output as 4 … 20 mA
signal or digitally via HART, Probus PA or Foundation Fieldbus.
Output measured values

9
4 Connecting to power supply
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
4 Connecting to power supply
4.1 Connecting
The connection to the Master sensor is carried out through spring-
loaded terminals in the respective housing. For this, use the supplied,
confectioned cable. Solid cores as well as exible cores with cable
end sleeves are inserted directly into the terminal openings.
In case of exible cores without end sleeves, press the terminal from
above with a small screwdriver, the terminal opening is then free.
When the screwdriver is released, the terminal closes again.
Information:
The terminal block is pluggable and can be removed from the
electronics. To do this, lift the terminal block with a small screwdriver
and pull it out. When reinserting the terminal block, you should hear it
snap in.
You can nd further information on the max. wire cross-section under
"Technical data/Electromechanical data"
Proceed as follows:
1. Unscrew the housing lid
2. Loosen compression nut of the cable gland and remove blind
plug
3. Remove approx. 10 cm (4 in) of the cable mantle, strip approx.
1 cm (0.4 in) insulation from the individual wires or use supplied
connection cable
4. Insert the cable into the sensor through the cable entry
Fig. 3: Connection steps 5 and 6
5. Insert the wire ends into the terminals according to the wiring plan
6. Check the hold of the wires in the terminals by lightly pulling on
them
Connection technology
Connection procedure

10
4 Connecting to power supply
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
7. Connect the screen to the internal ground terminal, connect the
external ground terminal to potential equalisation
8. Tighten the compression nut of the cable entry gland. The seal
ring must completely encircle the cable
9. Unscrew the blind plug on the Master, screw in the supplied cable
gland
10. Connection cable to Master, see steps 3 to 8
11. Screw the housing lid back on
The electrical connection is nished.
4.2 Single chamber housing
The following illustration applies to the non-Ex, Ex-ia and Ex-d-ia
version.
5678 4
2
1
connect to Master
Fig. 4: Wiring plan VEGABAR 82 Slave sensor
1 To the Master sensor
2 Ground terminal for connection of the cable screen1)
Electronics and terminal
compartment
1) Connect screen here. Connect ground terminal on the outside of the housing
to ground as prescribed. The two terminals are galvanically connected.

11
4 Connecting to power supply
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
4.3 Connection example
5678
4
2
3
1
4
connect to Master
5
12
+
( ) (-)
678
4...20mA
Fig. 5: Connection example, electronic dierential pressure
1 Master sensor
2 Slave sensor
3 Connection cable
4 Supply and signal circuit, Master sensor
Connection between master and sensor is carried out according to
the table:
Master sensor Slave sensor
Terminal 5 Terminal 5
Terminal 6 Terminal 6
Terminal 7 Terminal 7
Terminal 8 Terminal 8
Connection example,
electronic dierential
pressure

12
5 Set up with the display and adjustment module
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
5 Set up with the display and adjustment
module
5.1 Parameter adjustment - Quick setup
To quickly and easily adapt the sensor to the application, select
the menu item "Quick setup" in the start graphic on the display and
adjustment module.
Carry out the following steps in the sequence specied below. The
presettings apply to all applications.
You can nd "Extended adjustment" in the next sub-chapter.
1. Measurement loop name
In the rst menu item you assign a suitable measurement loop name.
Permitted are names with max. 19 characters.
2. Application
In this menu item you activate/deactivate the slave sensor for elec-
tronic dierential pressure and select the application.
Note:
It is absolutely necessary to activate the slave sensor in advance to
have the applications displayed in the electronic dierential pressure
measurement menus.
3. Units
In this menu item you determine the adjustment and temperature
units of the instrument. Depending on the selected application in the
menu item "Application", dierent adjustment units are available.
4. Unit, static pressure
In this menu item, you determine the unit of the static, i.e. superim-
posed pressure.
5. Position correction
In this menu item you compensate the inuence of the installation
position of the instrument (oset) on the measured value.
6. Min. adjustment
In this menu item you carry out the min. adjustment for level.
Enter the percentage value and the corresponding pressure value for
the min. level.
7. Max. adjustment
In this menu item you carry out the max. adjustment for level.
Presettings
Quick setup - Level meas-
urement

13
5 Set up with the display and adjustment module
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
Enter the percentage value and the corresponding pressure value for
the max. level.
The quick setup for level measurement is nished.
4. Position correction
In this menu item you compensate the inuence of the installation
position of the instrument (oset) on the measured value.
5. Min. adjustment
In this menu item you carry out the min. adjustment for ow.
Enter the percentage value and the corresponding pressure value for
the min. ow.
6. Max. adjustment
In this menu item you carry out the max. adjustment for ow.
Enter the percentage value and the corresponding pressure value for
the max. ow.
7. Linearization
In this menu item, you select the characterics of the output signal.
The quick setup for ow measurement is nished.
4. Unit, static pressure
In this menu item, you determine the unit of the static, i.e. superim-
posed pressure.
5. Position correction
In this menu item you compensate the inuence of the installation
position of the instrument (oset) on the measured value.
6. Zero adjustment
In this menu item you carry out the Zero adjustment for the dierential
pressure.
Enter the corresponding pressure value for 0 %.
7. Span adjustment
In this menu item you carry out the Span adjustment for the dieren-
tial pressure
Enter the corresponding pressure value for 100 %.
Quick setup - Flow meas-
urement
Quick setup - Dierential
pressure measurement

14
5 Set up with the display and adjustment module
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
The quick setup for dierential pressure measurement is nished.
4. Unit, static pressure
In this menu item, you determine the unit of the static, i.e. superim-
posed pressure.
5. Position correction
In this menu item you compensate the inuence of the installation
position of the instrument (oset) on the measured value.
6. Distance
In this menu item, you enter the installation distance between master
and slave sensor.
7. Min. adjustment
In this menu item you carry out the min. adjustment for density.
Enter the percentage value and the corresponding density value for
the min. density.
8. Max. adjustment
In this menu item you carry out the max. adjustment for density.
Enter the percentage value and the corresponding density value for
the max. density.
The quick setup for density measurement is nished.
4. Unit, static pressure
In this menu item, you determine the unit of the static, i.e. superim-
posed pressure.
5. Position correction
In this menu item you compensate the inuence of the installation
position of the instrument (oset) on the measured value.
6. Distance
In this menu item, you enter the installation distance between master
and slave sensor.
7. Min. adjustment
In this menu item, you carry out the adjustment for the min. height of
the interface.
Enter the percentage value and the corresponding height of the
interface.
8. Max. adjustment
In this menu item, you carry out the adjustment for the max. height of
the interface.
Enter the percentage value and the corresponding height of the
interface.
Quick setup - Density
measurement
Quick setup - Interface
measurement

15
5 Set up with the display and adjustment module
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
The quick setup for interface measurement is nished.
4. Unit, static pressure
In this menu item, you determine the unit of the static, i.e. superim-
posed pressure.
5. Position correction
In this menu item you compensate the inuence of the installation
position of the instrument (oset) on the measured value.
6. Distance
In this menu item, you enter the installation distance between master
and slave sensor.
7. Min. adjustment
In this menu item you carry out the min. adjustment for level.
Enter the percentage value and the corresponding pressure value for
the min. level.
8. Max. adjustment
In this menu item you carry out the max. adjustment for level.
Enter the percentage value and the corresponding pressure value for
the max. level.
The quick setup for density-compensated level measurement is
hence nished.
5.2 Parameter adjustment - Extended adjustment
For technically demanding measuring points, you can carry out
extended settings in "Extended adjustment".
The main menu is divided into ve sections with the following func-
tions:
Setup: Settings, e.g., for measurement loop name, application, units,
position correction, adjustment, signal output
Quick setup - density-
compensated level meas-
urement
Main menu

16
5 Set up with the display and adjustment module
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
Display: Settings, e.g., for language, measured value display, lighting
Diagnosis: Information, e.g. on instrument status, pointer, measure-
ment reliability, simulation
Additional adjustments: PIN, date/time, reset, copy function
Info: Instrument name, hardware and software version, date of manu-
facture, sensor features
In the main menu item "Setup", the individual submenu items should
be selected one after the other and provided with the correct param-
eter values.
The following submenu points are available:
In the following section, the menu items from the menu "Setup" for
electronic dierential pressure measurement are described in detail.
Depending on the selected application, dierent sections are relevant.
Information:
Further menu items of the menu "Setup" as well as the complete
menus "Display", "Diagnosis", "Additional adjustments" and "Info"
are described in the operating instructions of the respective master
sensor.
Menu - Setup
Menu item Parameter Default setting
Measurement
loop name
Sensor
Application Application Level
Slave for elec-
tronic dierential
pressure
Deactivated
Units Unit of measure-
ment
mbar (with nominal measuring range
≤400 mbar)
bar (with nominal measuring rang-
es ≥1 bar)
Static pressure bar
Position correc-
tion
0.00 bar
Adjustment Distance (with
density and inter-
face)
1.00 m
Zero/Min. adjust-
ment
0.00 bar
0.00 %
Span/Max. adjust-
ment
Nominal measuring range in bar
100.00 %
Menu and parameter
overview

17
5 Set up with the display and adjustment module
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
Menu item Parameter Default setting
Damping Integration time 0.0 s
Linearization Linear
Current output Current output -
Mode
Output characteristics
4 … 20 mA
Reaction when malfunctions occur
≤ 3.6 mA
Current output -
Min./Max.
3.8 mA
20.5 mA
Lock adjustment Last setting
Menu - Display
Menu item Default setting
Menu language Order-specic
Displayed value 1 Current output in %
Displayed value 2 Ceramic measuring cell: Measuring cell tempera-
ture in °C
Metallic measuring cell: Electronics temperature
in °C
Display format 1 and 2 Number of positions after the decimal point, auto-
matically
Backlight Switched on
Reset - Diagnosis
Menu item Parameter Default setting
Sensor status -
Peak value Pressure Actual measured value
Temperature Actual temperature values from meas-
uring cell, electronics
Simulation -
Reset - Additional settings
Menu item Parameter Default setting
PIN 0000
Date/Time Actual date/Actual time
Copy in-
strument
settings
-
Special pa-
rameters
-
Scaling Scaling size Volume in l
Scaling format 0 % corresponds to 0 l
100 % corresponds to 0 l

18
5 Set up with the display and adjustment module
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
Menu item Parameter Default setting
Current out-
put
Current output - Meas. vari-
able
Lin. percent - Level
Current output - Adjustment 0 … 100 % correspond to
4 … 20 mA
HART mode Address 0
DP ow ele-
ment
Unit m3/s
Adjustment 0.00 % correspond to 0.00
m3/s
100.00 %, 1 m3/s
Menu - Info
Menu item Parameter
Device name VEGABAR 8.
Instrument version Hardware and software version
Factory calibration
date
Date
Sensor characteristics Order-specic characteristics

19
6 Supplement
VEGABAR 82 • Slave sensor for electronic dierential pressure
46332-EN-160212
6 Supplement
6.1 Technical data
Electromechanical data - version IP 66/IP 67
Options of the cable entry
ƲCable entry M20 x 1.5, ½ NPT
ƲCable gland M20 x 1,5; ½ NPT (cable ø see below table)
Ʋ Blind plug M20 x 1.5; ½ NPT
Ʋ Closing cap ½ NPT
Material cable
gland
Material seal
insert
Cable diameter
4.5 … 8.5 mm 5 … 9 mm 6 … 12 mm 7 … 12 mm 10 … 14 mm
PA black NBR – ● ● –●
PA blue NBR – ● ● –●
Brass, nickel-
plated
NBR ● ● ● – –
Stainless steel NBR – ● ● –●
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, stranded wire 0.2 … 2.5 mm² (AWG 24 … 14)
Ʋ Stranded wire with end sleeve 0.2 … 1.5 mm² (AWG 24 … 16)
Interface to the Master sensor
Data transmission Digital (I²C-Bus)
Connection cable to the master sensor
Ʋ Conguration four wires, one suspension wire, one breather capillary,
screen braiding, metal foil, mantle
ƲWire cross-section 0.5 mm² (AWG 20)
ƲWire resistance < 0.036 Ω/m (0.011 Ω/ft)
ƲStandard length 5 m (16.40 ft)
ƲMax. length 25 m (82.02 ft)
Ʋ Min. bending radius at 25 °C/77 °F 25 mm (0.985 in)
Ʋ Diameter approx. 8 mm (0.315 in)
ƲColour Blue
Voltage supply for the complete system through Master
Operating voltage
ƲUB min 12 V DC
ƲUB min - lighted display and adjustment
module
12 V DC
ƲUB max Depending on signal output and version

Printing date:
VEGA Grieshaber KG
Am Hohenstein 113
77761 Schiltach
Germany
46332-EN-160212
All statements concerning scope of delivery, application, practical use and operat-
ing conditions of the sensors and processing systems correspond to the information
available at the time of printing.
Subject to change without prior notice
© VEGA Grieshaber KG, Schiltach/Germany 2016
Phone +49 7836 50-0
Fax +49 7836 50-201
E-mail: [email protected]
www.vega.com
Other manuals for VEGABAR 82
22
Table of contents
Other Vega Accessories manuals

Vega
Vega VEGAPULS 6X User manual

Vega
Vega VEGAPULS 65 User manual

Vega
Vega VEGAPULS 61 User manual

Vega
Vega VEGAPULS C 22 User manual

Vega
Vega VEGAPULS 61 User manual

Vega
Vega B-LIFT 8120 200 Evo Series User manual

Vega
Vega VEGAPULS 66 User manual

Vega
Vega VEGAPULS 65 User manual

Vega
Vega VEGAPULS 61 User manual

Vega
Vega VEGAFLEX 81 User manual

Vega
Vega VEGABAR 38 User manual

Vega
Vega VEGASON 61 User manual

Vega
Vega VEGABAR 39 User manual

Vega
Vega vegaflex 83 User manual

Vega
Vega VEGAKON 61 User manual

Vega
Vega ATEX Installation instructions

Vega
Vega VEGAPULS 62 User manual

Vega
Vega vegaflex 83 User manual

Vega
Vega VEGAVIB Wiring diagram

Vega
Vega VEGAPULS 67 User manual