
19
7Maintenanceandfaultrectication
VEGATOR 112 •
46106-EN-140625
Cause Rectication
Sensor current
< 0.05 mA
– Check signal conditioning instrument
– Check the terminal voltage in idle state on the
signal conditioning instrument; if it is < 8 V, the
signal conditioning instrument is defective -> ex-
change signal conditioning instrument or return
it for repair
– If the terminal voltage is > 8 V, disconnect the
sensor cable on the signal conditioning instru-
mentandreplaceitwitha1kΩresistor.Ifthe
fault signal does not disappear, the signal con-
ditioning instrument is defective -> exchange
signal conditioning instrument or return it for
repair
– Check sensor or sensor cable
– Reconnect the sensor cable to the signal con-
ditiong instrument, disconnect the sensor and
replaceitwitha1kΩresistor.Ifthefaultsignal
does not disappear, then the sensor cable is
broken -> replace the sensor cable
– If there is no longer a fault signal on the line,
the sensor is defective -> exchange sensor or
return it for repair
Sensor current > 6.8 mA – Check signal conditioning instrument
– Disconnect sensor cable and replace it with
a1kΩresistor.Ifthefaultsignaldoesnot
disappear, the signal conditioning instrument
is defective -> exchange signal conditioning
instrument or return it for repair
– Check sensor or sensor cable
– Reconnect the sensor cable to the signal con-
ditiong instrument, disconnect the sensor and
replaceitwitha1kΩresistor.Ifthefaultsignal
does not disappear, then the sensor cable is
short-circuited -> eliminate the short-circuit or
replace the sensor cable
– If there is no longer a fault signal on the line,
the sensor is defective -> exchange sensor or
return it for repair
7.3 How to proceed if a repair is needed
Yo u canndarepairformaswellasdetailedinformationonhowto
proceed at www.vega.com/downloads and "Formsandcerticates".
By doing this you help us carry out the repair quickly and without hav-
ing to call back for needed information.
If a repair is necessary, please proceed as follows:
•Printandlloutoneformperinstrument
•Clean the instrument and pack it damage-proof
•Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
•Please contact the agency serving you to get the address for
thereturnshipment.Yo u canndtheagencyonourhomepage
www.vega.com.