Vega VEGATOR 112 User manual

Operating Instructions
Double channel signal conditioning
instrument for level detection according
to NAMUR (IEC 60947-5-6)
VEGATOR 112
Document ID: 46106

2
Contents
VEGATOR 112 •
46106-EN-140625
Contents
1 About this document
1.1 Function ........................................................................................................................... 4
1.2 Target group ..................................................................................................................... 4
1.3 Symbols used................................................................................................................... 4
2 For your safety
2.1 Authorised personnel ....................................................................................................... 5
2.2 Appropriate use................................................................................................................ 5
2.3 Warning about incorrect use............................................................................................. 5
2.4 General safety instructions............................................................................................... 5
2.5 CE conformity................................................................................................................... 6
2.6 Safety label on the instrument .......................................................................................... 6
2.7 SIL conformity (optional) .................................................................................................. 6
2.8 Safety instructions for Ex areas ........................................................................................ 6
2.9 Environmental instructions ............................................................................................... 6
3 Product description
3.1 Conguration.................................................................................................................... 7
3.2 Principle of operation........................................................................................................ 7
3.3 Operation ......................................................................................................................... 8
3.4 Packaging, transport and storage..................................................................................... 8
4 Mounting
4.1 Mounting instructions ....................................................................................................... 9
5 Connecting to power supply
5.1 Preparing the connection ............................................................................................... 10
5.2 Connection procedure.................................................................................................... 10
5.3 Wiring plan ..................................................................................................................... 11
6 Setup
6.1 Adjustment system......................................................................................................... 12
6.2 Adjustment elements...................................................................................................... 12
6.3 Function chart "Point level"............................................................................................. 15
6.4 Function chart, two-point control .................................................................................... 15
7 Maintenanceandfaultrectication
7.1 Maintenance .................................................................................................................. 18
7.2 Rectify faults................................................................................................................... 18
7.3 How to proceed if a repair is needed.............................................................................. 19
8 Dismount
8.1 Dismounting steps.......................................................................................................... 20
8.2 Disposal ......................................................................................................................... 20
9 Supplement
9.1 Technical data ................................................................................................................ 21
9.2 Dimensions .................................................................................................................... 23

3
Contents
VEGATOR 112 •
46106-EN-140625
Supplementary documentation
Information:
Supplementary documents appropriate to the ordered version come
withthedelivery.Yo u canndthemlistedinchapter"Product descrip-
tion".
Editing status:2014-06-04

4
1 About this document
VEGATOR 112 •
46106-EN-140625
1 About this document
1.1 Function
This operating instructions manual provides all the information you
need for mounting, connection and setup as well as important instruc-
tionsformaintenanceandfaultrectication.Pleasereadthisinforma-
tion before putting the instrument into operation and keep this manual
accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained specialist
personnel.The contents of this manual should be made available to
these personnel and put into practice by them.
1.3 Symbols used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
SIL applications
This symbol indicates instructions for functional safety which must be
taken into account particularly for safety-relevant applications.
•List
The dot set in front indicates a list with no implied sequence.
→ Action
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat-
teries and accumulators.

5
2 For your safety
VEGATOR 112 •
46106-EN-140625
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual must
be carried out only by trained specialist personnel authorised by the
plant operator.
During work on and with the device the required personal protective
equipment must always be worn.
2.2 Appropriate use
VEGATOR 112 is a universal signal conditioning instrument for con-
nection of level switches.
Yo u cannddetailedinformationabouttheareaofapplicationin
chapter "Product description".
Operational reliability is ensured only if the instrument is properly
usedaccordingtothespecicationsintheoperatinginstructions
manual as well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbi-
traryconversionsormodicationsareexplicitlyforbidden.
2.3 Warning about incorrect use
Inappropriate or incorrect use of the instrument can give rise to
application-specichazards,e.g.vesseloverllordamagetosystem
components through incorrect mounting or adjustment.
2.4 General safety instructions
This is a state-of-the-art instrument complying with all prevailing
regulations and guidelines.The instrument must only be operated in a
technicallyawlessandreliablecondition.Theoperatorisresponsible
for the trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new regula-
tions.
The safety instructions in this operating instructions manual, the na-
tional installation standards as well as the valid safety regulations and
accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbi-
traryconversionsormodicationsareexplicitlyforbidden.
The safety approval markings and safety tips on the device must also
be observed.

6
2 For your safety
VEGATOR 112 •
46106-EN-140625
2.5 CE conformity
ThedevicefulllsthelegalrequirementsoftheapplicableECguide-
lines.ByaxingtheCEmarking,weconrmsuccessfultestingofthe
product.
Yo u canndtheCECerticateofConformityinthedownloadsection
of our homepage.
2.6 Safety label on the instrument
The safety approval markings and safety tips on the device must be
observed.
2.7 SIL conformity (optional)
InstrumentswithSILoptionfullltherequirementsoffunctionalsafety
accordingtoIEC61508.Yo u canndfurtherinformationinthesup-
plied Safety Manual.
2.8 Safety instructions for Ex areas
PleasenotetheEx-specicsafetyinformationforinstallationandop-
eration in Ex areas.These safety instructions are part of the operating
instructions manual and come with the Ex-approved instruments.
2.9 Environmental instructions
Protection of the environment is one of our most important duties.
That is why we have introduced an environment management system
with the goal of continuously improving company environmental pro-
tection.Theenvironmentmanagementsystemiscertiedaccording
to DIN EN ISO 14001.
Pleasehelpusfulllthisobligationbyobservingtheenvironmental
instructions in this manual:
•Chapter "Packaging, transport and storage"
•Chapter "Disposal"

7
3 Product description
VEGATOR 112 •
46106-EN-140625
3 Product description
3.1 Conguration
The scope of delivery encompasses:
•VEGATOR 112 signal conditioning instrument
•Documentation
– this operating instructions manual
– Ex-specic"Safety instructions" (with Ex-version)
– Safety Manual "Functional safety (SIL) acc. to IEC 61508" (with
SIL version)
– ifnecessary,furthercerticates
Thetypelabelcontainsthemostimportantdataforidenticationand
use of the instrument:
•Article number
•Serial number
•Technical data
•Article numbers, documentation
The type label contains the serial number of the instrument. With it
youcanndthefollowingdataonourhomepage:
•Product code of the instrument (HTML)
•Delivery date (HTML)
•Order-specicinstrumentfeatures(HTML)
•Operating instructions at the time of shipment (PDF)
Go to www.vega.com, "VEGA Tools" and "Instrument search".
Alternatively, you can access the data via your smartphone:
•Download the smartphone app "VEGA Tools" from the "Apple App
Store" or the "Google Play Store"
•Scan the Data Matrix code on the type label of the instrument or
•Enter the serial number manually in the app
3.2 Principle of operation
The VEGATOR 112 is a double-channel signal conditioning instru-
ment for level detection with two level switches with NAMUR interface.
Simple monitoring and control functions can be realised via the inte-
grated relays.Typical applications are monitoring functions such as
overllanddryrunprotection.Asanalternativetothedouble-channel
level detection, a two-point control can be activated.
The VEGATOR 112 signal conditioning instrument powers connected
sensors and simultaneously processes their measuring signals.
When the medium reaches the switching point of the sensor, the sen-
sor current changes.This is measured and evaluated by VEGATOR
112.The change causes the output relays to switch according to the
set operating mode.
Wide-range power supply unit with 20 … 253 V AC/DC for worldwide
use.
Scope of delivery
Type label
Serial number
Application area
Functional principle
Voltage supply

8
3 Product description
VEGATOR 112 •
46106-EN-140625
Detailed information about the power supply can be found in chapter
"Technical data".
3.3 Operation
All adjustment elements are located under a hinged front cover. The
operating mode and the line monitoring can be set via a DIL switch
block. In addition, the correct function of the measuring system can
be checked with a test key.
3.4 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
based on ISO 4180.
The packaging of standard instruments consists of environment-
friendly, recyclable cardboard. For special versions, PE foam or PE
foil is also used. Dispose of the packaging material via specialised
recycling companies.
Transport must be carried out in due consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or con-
cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
•Not in the open
•Dry and dust free
•Not exposed to corrosive media
•Protected against solar radiation
•Avoiding mechanical shock and vibration
•Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
•Relative humidity 20 … 85 %
Packaging
Transport
Transport inspection
Storage
Storage and transport
temperature

9
4 Mounting
VEGATOR 112 •
46106-EN-140625
4 Mounting
4.1 Mounting instructions
VEGATOR 112 is designed for carrier rail mounting (top hat rail
35 x 7.5 according to DIN EN 50022/60715). Due to its protection
rating of IP 20, the instrument is suitable for mounting in switching
cabinets.Itcanbemountedhorizontallyandvertically.
Note:
When several instruments are mounted together without space in
between, the ambient temperature at the installation location of the
instrument must not exceed 60 °C.
The VEGATOR 112 in Ex version is an associated, intrinsically safe
instrumentandmustnotbeinstalledinhazardousareas.Safeopera-
tion is only ensured if the operating instructions and EC Type Approval
Certicateareobserved.VEGATOR112mustnotbeopened.A
certicationforExZone2isalsoavailableasanoption.

10
5 Connecting to power supply
VEGATOR 112 •
46106-EN-140625
5 Connecting to power supply
5.1 Preparing the connection
Always keep in mind the following safety instructions:
•Connect only in the complete absence of line voltage
•If overvoltage surges are expected, overvoltage arresters should
be installed
Inhazardousareasyoumusttakenoteoftherespectiveregulations,
conformityandtypeapprovalcerticatesofthesensorsandpower
supply units.
Thevoltagesupplycanbe20…253VAC/DC,50/60Hz.Yo u cannd
details on the voltage supply in the technical data.
The operating voltage of VEGATOR 112 is connected with standard
cable according to the national installation standards.
The sensors are connected with standard two-wire cable without
screen. If electromagnetic interference is expected which is above the
test values of EN 61326 for industrial areas, screened cable should
be used.
Connect the cable screen on both ends to ground potential. In the
sensor, the screen must be connected directly to the internal ground
terminal.The ground terminal on the outside of the sensor housing
must be connected to the potential equalisation (low impedance).
If potential equalisation currents are expected, the connection on the
processing side must be made via a ceramic capacitor (e. g. 1 nF,
1500 V).The low-frequency potential equalisation currents are thus
suppressed,buttheprotectiveeectagainsthighfrequencyinterfer-
ence signals remains.
Take note of the corresponding installation regulations for Ex applica-
tions. In particular, make sure that no potential equalisation currents
owoverthecablescreen.Incaseofgroundingonbothsidesthis
can be achieved by the use of a capacitor or a separate potential
equalisation.
5.2 Connection procedure
The pluggable terminals can be removed as needed to allow more
convenient connection. To make the electrical connection, proceed as
follows:
1. Mount the instrument as described in the previous chapter
2. Connect sensor cable 1 to terminal 1/2, and where applicable,
connect the screen
3. Connect sensor cable 2 to terminal 4/5, and where applicable,
connect the screen
4. Connectpowersupply(switchedo)toterminal16/17
5. Connect relay 1 to terminal 10/11/12
6. Connect relay 2 to terminal 13/14/15
Note safety instructions
Take note of safety
instructions for Ex
applications
Select power supply
Select connection cable
Cable screening and
grounding
Select connection
cable for Ex applica-
tions

11
5 Connecting to power supply
VEGATOR 112 •
46106-EN-140625
Theelectricalconnectionisnished.
5.3 Wiring plan
1
2
5
3
4
10 11 12
13 14 15
-+
4 5
-+
1 2
-+
16 17
OPEN
1 Sensor circuit, channel 1 (1.2/2.1 mA)
2 Sensor circuit, channel 2 (1.2/2.1 mA)
3 Relay output channel 1
4 Relay output channel 2
5 Voltage supply
Information:
The connection terminals can be detached towards the front, if
necessary. This can be useful when working in tight spaces or when
exchanging an instrument.

12
6 Setup
VEGATOR 112 •
46106-EN-140625
6 Setup
6.1 Adjustment system
1
2
3
4
Fig. 2: Display and adjustment elements
1 Test key
2 DIL switch block
3 Signal lamps (LEDs)
4 Hinged front cover
6.2 Adjustment elements
Control lamps (LED) in the front plate indicate operation, switching
status and fault signal.
•Green
– Operating control lamp
– Mains voltage on, instrument is operating
•Red
– Fault indicator
– Fault on the sensor circuit due to sensor failure or line break
– The relay deenergises in case of failure
Control lamps

13
6 Setup
VEGATOR 112 •
46106-EN-140625
•Yellow
– Relay control lamp
– The yellow relay control lamp reacts according to the set mode
– The relay control lamp generally indicates the activated (ener-
gized)conditionoftherelay
– An unlit relay control lamp means that the relay is deenergised
All adjustment elements are located under a hinged front cover. To
open it, use a small screwdriver in conjunction with the slot on the up-
per side of the front cover. To close it, insert the two retaining clips on
the lower end of the cover. Then press the upper end of the cover until
you hear it snap closed.
The DIL switch block is located behind the front cover. The individual
switches are assigned as follows:
3
4
5
maxmin
1
2
maxmin
CH1CH2
Fig. 3: DIL switch VEGATOR 112
1 Mode (min./max. adjustment), channel 1
2 LinemonitoringO/On,channel1
3 Mode (min./max. adjustment), channel 2
4 LinemonitoringO/On,channel2
5 Two-pointcontrolOn/O
The requested operating mode is set with the min./max. switch (min.
detectioni.e.dryrunprotectionormax.detectioni.e.overllprotec-
tion)
•Dry run protection:Relayisswitchedowhenthelevelfalls
below the min. level (safe currentless state), relay is switched on
again when the max. level is exceeded (switch-on point > switch-
opoint)
•Overowprotection:Relayisswitchedowhenthemax.levelis
exceeded (safe currentless state), relay is switched on again when
thelevelfallsbelowthemin.level(switch-onpoint<switch-o
point)
Note:
The mode selection on the signal conditioning instrument only func-
tions correctly if the mode switch on the sensor is set to "Max.".
The following functionalities can be selected via this switch:
•Single point control (double channel level detection with one sen-
soreachandthesameswitchon/switchopoint)
•Two-pointcontrol(leveldetectionwithtwosensorsanddierent
switchon/switchopoints)
Front cover
DIL switch block
Mode (min./max. adjust-
ment)
Two-point control

14
6 Setup
VEGATOR 112 •
46106-EN-140625
As opposed to single point control, two-point control allows the
switch-onandswitch-opointstobesetatdierentlocations(hyster-
esis).Through the installation or the distance of the two sensors, the
hysteresiscanbedenedindividually.
Thetwo-pointcontrolisusedforexamplewhenllingandemptying
iscarriedoutviaapump.Thellingofavesselcanbeforexample
switchedonat10%andswitchedoagainat90%.Theoutputof
thesecondchannelbehavesidenticallyliketherstchannelprovided
the same mode is set. By changing the mode, the function of the
relays can be inverted.
When NAMUR level switches are connected, the line monitoring func-
tion can check the input continuously for line break or short-circuit. If
amechanicalswitch,e.g.aoatswitch,istobeconnected,theline
monitoring must be deactivated because only the switching condi-
tions "open" or "closed" can be delivered.
When level switches VEGAVIB/VEGAWAVE in NAMUR version are
connected, a function test can be carried out.The test key is recessed
behind the front cover of the signal conditioning instrument. Press the
test key for at least one second with a suitable object (screwdriver,
pen, etc.).
Pushing the test key interrupts the circuit to the sensor and triggers
a restart of the sensor. In the process, the conditions fault signal, full
alarm (< 1.2 mA) and empty alarm (> 2.1 mA) are simulated one after
the other. Check if a fault signal is triggered when you push the test
key (red LED lights). The red LED does not light up when line monitor-
ing is deactivated.
After the test key is released, the sensor is again supplied with
voltage.Duringthewarm-upphaseofthesensor,thecurrentbriey
increases,whichcanenergizetherelayandtheyellowLED.Thisis
not relevant for the test assessment.
Once the sensor is ready for operation, a full alarm is simulated for
approx. 3 s, the yellow LED does not light and the relay is deener-
gized.Forapositivetestassessment,thetimemustbebetween2s
and 4 s.
Finally, an empty alarm is simulated by the sensor for approx. 1.5 s.
TheyellowLEDlightsandtherelayisenergized.Forapositivetest
assessment, this time must be at least 1 s.
Thetestprocedureisnownished,LEDandrelayreturntotheactual
operating condition.
When a VEGASWING is connected, it is only line break that is simu-
lated when the test key is pushed. Thus, only the correct behaviour
of VEGATOR and the downstream device in the event of fault can be
tested.
Note:
Keep in mind that downstream connected devices are activated dur-
ing the function test.This allows you to check the correct function of
the measuring system.
Line monitoring
Test key

15
6 Setup
VEGATOR 112 •
46106-EN-140625
6.3 Function chart "Point level"
The following chart provides an overview of the switching conditions
depending on the set mode and the level.
Note:
The switching conditions in the tables are only valid if the mode
switch on the sensor is set to "Max.".
Sensor Signal conditioning instrument
Level Sensor cur-
rent
LED yellow
(output)
LED red (fail-
ure)
Relay
> 2.1 mA ON
< 1.2 mA OFF
any < 0.35 mA
> 6.8 mA OFF
Sensor Signal conditioning instrument
Level Sensor cur-
rent
LED yellow
(output)
LED red (fail-
ure)
Relay
< 1.2 mA ON
> 2.1 mA OFF
any < 0.35 mA
> 6.8 mA OFF
6.4 Function chart, two-point control
The following charts provide an overview of the switching conditions
depending on the set mode and the level.
Note:
The switching conditions in the tables are only valid if the mode
switch on the sensor is set to "Max.".
Overllprotection,point
level
Dry run protection, point
level

16
6 Setup
VEGATOR 112 •
46106-EN-140625
Sensor Signal conditioning instrument
Level Sensor cur-
rent
LED yellow
(output)
LED red (fail-
ure)
Relay
> 2.1 mA
> 2.1 mA
ON
> 2.1 mA
< 1.2 mA
ON
< 1.2 mA
< 1.2 mA
OFF
> 2.1 mA
< 1.2 mA
OFF
> 2.1 mA
> 2.1 mA
ON
any < 0.35 mA
> 6.8 mA OFF
Sensor Signal conditioning instrument
Level Sensor cur-
rent
LED yellow
(output)
LED red (fail-
ure)
Relay
> 2.1 mA
> 2.1 mA
OFF
> 2.1 mA
< 1.2 mA
OFF
< 1.2 mA
< 1.2 mA
ON
Overllprotection,two-
point control
Dru run protection, two-
point control

17
6 Setup
VEGATOR 112 •
46106-EN-140625
Sensor Signal conditioning instrument
Level Sensor cur-
rent
LED yellow
(output)
LED red (fail-
ure)
Relay
> 2.1 mA
< 1.2 mA
ON
> 2.1 mA
> 2.1 mA
OFF
any < 0.35 mA
> 6,8 mA OFF

18
7Maintenanceandfaultrectication
VEGATOR 112 •
46106-EN-140625
7 Maintenanceandfaultrectication
7.1 Maintenance
If the instrument is used properly, no special maintenance is required
in normal operation.
7.2 Rectify faults
Maximum reliability is ensured. Nevertheless, faults can occur during
operation.These may be caused by the following, e.g.:
•Measured value from sensor not correct
•Voltage supply
•Interference in the cables
Therstmeasuretobetakenistochecktheinputandoutputsignals.
The procedure is described as follows. In many cases the causes can
bedeterminedthiswayandfaultscanbeeasilyrectied.
Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.
Thehotlineismanned7daysaweekround-the-clock.Sinceweoer
this service worldwide, the support is only available in the English
language.The service is free, only standard call charges are incurred.
Depending on the reason for the fault and the measures taken, the
steps described in chapter "Set up" may have to be carried out again.
Cause Rectication
Sensor not connected
correctly
– In Ex systems, make sure that the Ex protection
isnotinuencedbythemeasuringinstruments
used.
– Measure the current and voltage on the connec-
tion cable to the sensor
– Faults in the sensor that cause a current change
under 0.05 mA or over 6.8 mA lead to a fault
signal with signal conditioning instruments.
– In idle state, the terminal voltage (without sensor)
is at least 8.2 V
– The signal conditioning instrument has an inner
resistance of 1 kOhm.With this, the voltages for
thedierentcurrentscanbecalculated.
Causes of malfunction
Faultrectication
24 hour service hotline
Reaction after fault recti-
cation
Red fault LED lights up

19
7Maintenanceandfaultrectication
VEGATOR 112 •
46106-EN-140625
Cause Rectication
Sensor current
< 0.05 mA
– Check signal conditioning instrument
– Check the terminal voltage in idle state on the
signal conditioning instrument; if it is < 8 V, the
signal conditioning instrument is defective -> ex-
change signal conditioning instrument or return
it for repair
– If the terminal voltage is > 8 V, disconnect the
sensor cable on the signal conditioning instru-
mentandreplaceitwitha1kΩresistor.Ifthe
fault signal does not disappear, the signal con-
ditioning instrument is defective -> exchange
signal conditioning instrument or return it for
repair
– Check sensor or sensor cable
– Reconnect the sensor cable to the signal con-
ditiong instrument, disconnect the sensor and
replaceitwitha1kΩresistor.Ifthefaultsignal
does not disappear, then the sensor cable is
broken -> replace the sensor cable
– If there is no longer a fault signal on the line,
the sensor is defective -> exchange sensor or
return it for repair
Sensor current > 6.8 mA – Check signal conditioning instrument
– Disconnect sensor cable and replace it with
a1kΩresistor.Ifthefaultsignaldoesnot
disappear, the signal conditioning instrument
is defective -> exchange signal conditioning
instrument or return it for repair
– Check sensor or sensor cable
– Reconnect the sensor cable to the signal con-
ditiong instrument, disconnect the sensor and
replaceitwitha1kΩresistor.Ifthefaultsignal
does not disappear, then the sensor cable is
short-circuited -> eliminate the short-circuit or
replace the sensor cable
– If there is no longer a fault signal on the line,
the sensor is defective -> exchange sensor or
return it for repair
7.3 How to proceed if a repair is needed
Yo u canndarepairformaswellasdetailedinformationonhowto
proceed at www.vega.com/downloads and "Formsandcerticates".
By doing this you help us carry out the repair quickly and without hav-
ing to call back for needed information.
If a repair is necessary, please proceed as follows:
•Printandlloutoneformperinstrument
•Clean the instrument and pack it damage-proof
•Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
•Please contact the agency serving you to get the address for
thereturnshipment.Yo u canndtheagencyonourhomepage
www.vega.com.

20
8 Dismount
VEGATOR 112 •
46106-EN-140625
8 Dismount
8.1 Dismounting steps
Take note of chapters "Mounting" and "Connecting to power supply"
and carry out the listed steps in reverse order.
8.2 Disposal
The instrument consists of materials which can be recycled by spe-
cialised recycling companies. We use recyclable materials and have
designed the parts to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and
the respective national laws. Pass the instrument directly on to a spe-
cialised recycling company and do not use the municipal collecting
points.These may be used only for privately used products according
to the WEEE directive.
Correctdisposalavoidsnegativeeectsonhumansandtheenviron-
ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
Other manuals for VEGATOR 112
1
Table of contents
Other Vega Measuring Instrument manuals

Vega
Vega VEGACAP 35 User manual

Vega
Vega ASV-1 User manual

Vega
Vega VEGACAP 69 User manual

Vega
Vega VEGAFLEX 86 User manual

Vega
Vega VEGAPULS 62 User manual

Vega
Vega VEGAPULS 67 User manual

Vega
Vega VEGABAR 17 User manual

Vega
Vega VEGACAL 63 User manual

Vega
Vega VEGAVIB 61 User manual

Vega
Vega VEGABAR 12 User manual

Vega
Vega VEGAFLEX 80 User guide

Vega
Vega VEGACAL 63 User manual

Vega
Vega VEGAPULS 65 User manual

Vega
Vega VEGACAP 66 User manual

Vega
Vega VEGABAR 81 User manual

Vega
Vega VEGAMIP R61 User manual

Vega
Vega VEGAWAVE Wiring diagram

Vega
Vega VEGAPULS 69 User manual

Vega
Vega VEGACAL 65 User manual

Vega
Vega VEGAPULS 61 User manual