Vetario T40M User manual

HD453 GB Issue 08
1
www.vetario.co.uk
T40M and T50M
Thermal Life-support Cabinets
Medical intensive care unit for mammals, reptiles and birds.
Operating and Routine Service Manual
Contents
Section Subject Page
1 Important Safety Information 2
2 Intended Use 4
3 Introduction 4
4 Unpacking 5
5 Location and Assembly 6
6 Control and Alarm System Operation 9
7 Temperature 16
8 Humidity and Ventilation 17
9 Introducing Your Patients 18
10 Internal Lighting 18
11 Oxygen Administration 19
12 Nebuliser 22
13 Cleaning and Disinfection 23
14 Safety inspection, Maintenance and Calibration 25
15 Specification 33
16 Disposal 33
Read the instructions before use! Do not cover!
IMPORTANT NOTICE
Brinsea Products Ltd and its agents or distributors will not be responsible for loss of animals in
the event of failure however caused and the user is advised to arrange his own insurance cover
where loss of power or mechanical or electrical failure might result in unacceptable losses.
Original Instructions, English Language.

HD453 GB Issue 08
2
1 Important Safety Information
READ THE INSTRUCTIONS AND ALL SAFETY NOTICES BEFORE USE.
FAILURE TO FOLLOW THE INSTRUCTIONS MAY RESULT IN A HAZARDOUS SITUATION
CAUSING SERIOUS INJURY OR DEATH OF THE USER OR PATIENT.
All operatives must read and understand the precautions necessary when oxygen gas
administration is to be performed (see section 11). Oxygen concentrations above normal
atmospheric conditions create a greatly increased risk of fire, including immediately outside the
apparatus. Materials that do not burn in air may react and burn vigorously in an oxygen enriched
atmosphere.
MEANING OF SYMBOLS:
Warning –identifies situations or actions that may affect patient or user safety.
Disregarding this warning may result in patient or user injury.
Indicates a Fire Hazard
Do not cover the appliance
Read the operating manual
GENERAL SAFETY PRECAUTIONS:
This incubator / animal intensive care unit / brooder should only be used by appropriately trained
personnel and under the direction of qualified medical personnel familiar with currently known
risks and benefits of incubator use. (Incorrect temperature or oxygen concentration may cause
permanent harm or death of the patient).
Direct sunlight or other radiant heat sources can cause an increase in incubator temperature to
dangerous levels. (Sunlight may move around onto the incubator or heating appliances may start
unexpectedly).
Not suitable for use in the presence of a flammable anaesthetic mixture. (The incubator is not
intended to be used in a location where flammable anaesthetics may be present. This would
create a serious risk of fire or explosion).
Avoid alcohol based solutions and make sure that all disinfectant is thoroughly rinsed off plastic
parts with clean water. Disinfectant solutions may cause severe cracking of plastic if not
correctly rinsed off.
Perform routine safety service as detailed in section 14 of this document. Only use spare parts
approved by Brinsea Products, use of unapproved parts will invalidate the warrantee and may
create a safety hazard. Never bypass safety devices such as fuses or thermal cut-outs. Never
deliberately block ventilation holes or slots. The on-going safety of the incubator depends on
correct servicing using exactly specified parts.
The date of construction forms part of the serial number e.g. HDxxxx/130123yyy where xxxx is
the model code, 130123 is year, month, day and yyy is the unique device number. The date as
shown would be 23rd January 2013. The intensive care unit must be serviced and checked
regardless of frequency of use and date of first use as some items degrade over time (i.e. have a
limited shelf-life).
Do not cover the intensive care unit. (Blocking of ventilation slots/holes may lead to oxygen
enrichment of air in unintended areas leading to an increased risk of fire. It may also lead to
increased concentration of CO2in the animal cabinet. Covering the unit may lead to fire hazard
due to reduced dissipation of heat).

HD453 GB Issue 08
3
For indoor use only. (This device is only intended to be used in a dry location with ambient
temperature controlled between 20° - 25°C. Extremes of moisture or temperature may lead to
dangerous malfunction).
This apparatus must be earthed. (The incubator must be connected to the power supply using a
plug and socket with an earth termination. The mains wiring and the appliance should be
periodically checked to ensure earth continuity).
ADDITIONAL SAFETY PRECAUTIONS WHEN USED FOR OXYGEN THERAPY
No auxiliary equipment shall be placed in the incubator. Only use patient connections that are
designed for safe use in an oxygen enriched atmosphere. If in any doubt check with the
manufacturer of the auxiliary equipment. (Fire may be caused by sparking contacts or hot
surfaces in equipment not designed for oxygen service).
Even small quantities of flammable agents, such as ether and alcohol, left in the incubator can
cause fire in connection with oxygen. (Never use flammable cleaners or disinfectants and ensure
such agents are not introduced on a patient).
The incubator must only be used in a well-ventilated room away from potential sources of heat or
ignition. NO SMOKING. Do not use in a confined location, the incubator needs at least 30cm free
space above and to both sides. (Oxygen concentration in the air around the incubator may rise
significantly if ventilation isn’t provided or the space around the incubator is confined. This may
create a fire hazard).
Remove all electrical devices (including the water pump) that are within 90cm of the incubator.
Place the incubator at least 90cm away from power sockets. Ensure there are no power sockets
or electrical items directly under the shelf or table supporting the incubator. (Oxygen gas may
sink through air and cause fire in other equipment).
Before oxygen administration, disconnect any nebuliser and tube from the solution holder. (This
is to prevent oxygen flowing into the nebuliser pump and creating a fire hazard).
If fitted, disconnect the water pump control lead and water delivery tube. (The pump is not
intended for use during oxygen therapy and may create a fire hazard).
Always test the fan rotation and power failure alarms and check ventilation holes before using
with oxygen gas. If the alarm sounds during use disconnect the oxygen supply, disconnect
power and open the door to ventilate the care unit. Do not re-use until the fault has been
investigated and rectified. (Failure of the ventilation fan or blocking of ventilation slots/holes may
lead to oxygen enrichment of air in unintended areas leading to an increased risk of fire. It may
also lead to increased concentration of CO2in the animal cabinet).
Do not exceed 1lpm (one litre per minute) oxygen flow rate. (The maximum safe long-term
concentration of oxygen for canines and felines is achieved at this flow rate and increased flow
may lead to permanent injury or death. Greater flow rates also increase the risk of oxygen
leaking in the event of a fault and could create a fire risk).
The incubator contains no flow restriction or pressure regulator device. The oxygen supply must
be reliably restricted to 400kPa (50psi). Always use a medical grade oxygen pressure regulator
and flow meter that has been serviced in accordance with the manufacturer’s instructions.
Failure of an external device may create a fire or pressure hazard or prevent effective treatment.
It is assumed the operator will be trained in the safe and correct use of such devices.
After oxygen administration, animals and any associated cloth materials must be allowed
sufficient time in the chamber in normal air to allow trapped oxygen to disperse. This may take
30 minutes or more depending on the size of animal and amount of material. (Fabric and similar
materials that trap oxygen enriched air may burn vigorously if ignited by a spark or other source
of ignition).
Be sure to read and follow the material safety data available from your supplier of medical
oxygen gas. Local regulations may apply; your oxygen gas supplier will be able to offer further
guidance. (Some states may apply restrictions on sale or use of oxygen for medical purposes).
Do not obstruct ventilation holes. (Blocking of ventilation slots/holes may lead to oxygen
enrichment of air in unintended areas leading to an increased risk of fire. It may also lead to
increased concentration of CO2in the animal cabinet).

HD453 GB Issue 08
4
2 Intended Use
The Vetario T40M and T50M are designed solely for the care of small animals under the supervision of
trained medical staff in a Veterinarian Surgery or Hospital. The intensive care unit is to be attended in use
(alarms shall be audible) to monitor patient condition and to verify continued correct operation.
Animal patients vary greatly in strength between species and individual cases. Larger animals may be
placed in the unit but they must be closely supervised in case of becoming agitated. Although the cabinet
is made of highly robust, impact resistant materials a large, frightened animal may cause damage if left
unattended.
This manual provides instructions for the assembly, installation, use and operator maintenance of the
intensive care unit. Brinsea Products cannot be responsible for the performance of the care unit if the
user fails to follow the instructions and maintain the care unit in accordance with the instructions.
3 Introduction
The Vetario T40M and T50M provide the ideal veterinary intensive care environment for sick, injured or
postoperative animals and birds sensitive to hypothermia, shock and other complications. These products
give patients the best possible chance of a successful recovery.
These instructions detail the operation of your new Vetario intensive care unit. Please read them carefully
before setting up your unit to achieve best results and keep these instructions safe for future reference.
Your intensive care unit is designed to allow the user to vary the environmental conditions to suit recovery
of a wide range of species and the specific set-up for every recovery scenario is beyond the scope of
these instructions.
Fig. 1 Functional features of the T40M and T50M (T40M model shown)
Digital Display
Battery Compartment
(at rear)
Control Housing
Adjustable Vent
Water Reservoir
Pan and
Evaporating
Block
Water Fill Hole /
Oxygen Supply Port
Door Latch
Control Buttons
Inlet Air Filter
(circulation
filters are
inside)
Door Hinges
(Reversible)
Nebuliser
Solution
Holder
Fixed
Exhaust
Vent
I.V. tube slot

HD453 GB Issue 08
5
4 Unpacking
Your intensive care unit has been supplied in protective packaging. Please remove all tape, strapping and
packing from the unit and parts. Retain the carton and packing materials to enable the unit to be
repacked.
Your intensive care unit will include as standard:
Quantity Item
1 Cabinet top assembly (with air filters fitted)
1 Cabinet base
1 Door
1 Hinge socket
1 Fastener and tool kit
1 Water pan
1 Evaporating block
1 Water funnel
1 Nebuliser solution holder
1 Nebuliser bracket
1 Hose barb (for oxygen administration)
1 Brass nut for hose barb (for oxygen administration)
1 Battery
Fastener and Tool Kit Contents:
Quantity T40M Quantity T50M Item
2 2 M5 x 16mm countersunk screws
2 2 M5 flat washers
2 2 M5 domed nuts
12 16 M4 x 30mm cap-headed screws
12 16 M4 star washers
12 16 M4 nuts
1 1 3mm hex key
1 1 4mm hex key
1 1 7mm / 8mm spanner

HD453 GB Issue 08
6
Damaged appliances shall not be used. (Cracked or broken parts may create a risk of fire if
oxygen can leak out or electric shock if water can leak in. Missing or damaged parts may expose
live or hot parts).
The mains cable is specially prepared and may only be replaced by Brinsea or their authorised
agent. Do not use if the cable becomes damaged.
4.1 Please identify each part and check that they are all present and undamaged. If there are any parts
damaged or missing please contact your retailer or Brinsea Products (at the address at the end of
the document).
4.2 Note that if your intensive care unit has been ordered with additional options (such as the Brinsea
Advance Humidity Pump) separate instructions and component lists apply.
4.3 Check also that the electrical supply matches the machine’s requirements (marked on the technical
label on top of the cabinet).
4.4 To register your new Vetario product please visit www.vetario.co.uk and follow the link on the right
hand side of the home page to qualify for your free 3 year guarantee.
5 Location and Assembly
Ensure that the room temperature cannot drop on a cold night. Ideally thermostatically control
the room at between 20 and 25°C (68 and 77°F). Never allow the room temperature to drop below
15°C (59°F) and ensure that the unit cannot be exposed to direct sunlight.
The apparatus must only be used in a well-ventilated room away from potential sources of heat
or ignition. NO SMOKING.
Do not use in a confined location –see figure 5.0 below. There must be a minimum 30cm free
space to each side and above the unit. (Oxygen concentration in the air around the incubator
may rise significantly if ventilation isn’t provided or the space around the incubator is confined.
This may create a fire hazard).
Figure 5.0 –Minimum clearance to walls and other surfaces
Remove all electrical devices that are within 90cm of the incubator. Place the incubator at least
90cm away from power sockets. Ensure there are no power sockets or electrical items directly
under the shelf or table supporting the incubator. (Oxygen gas may sink through air and cause
fire in other equipment).
Minimum
30cm

HD453 GB Issue 08
7
Incorrect or incomplete assembly may allow oxygen gas to escape and create a fire hazard.
The security of the animal chamber is dependent on the correct and complete use of all
components and fasteners.
5.1 Assemble the cabinet using the tools and fasteners provided. Follow the diagrams. Do not over-
tighten fasteners.
5.2 The door is hinged at the left as supplied but may be reversed to provide better access if required.
Fit the top hinge socket to the cabinet using the M5 x 16mm countersunk screws, M5 flat washers
and M5 dome nuts.
5.3
5.4 Place the door in the lower hinge and close the latch. Lower the top cabinet down onto the other
parts.
5.5 Fit the M4 x 30mm cap-headed screws, star washers and do nuts into each set of holes around the
edge of the cabinet. Tighten enough to prevent them spinning. This creates an air tight seal around
the cabinet. Fit the nebuliser bracket as shown.
1
2
3
4
5
6
7
8

HD453 GB Issue 08
8
5.6 Place the white evaporating block upright in the clear water pan. Open the door and lift the pan into
position, it pushes up and across into two slots on the heater enclosure. The pad may need to be
softened with a little water to help it flatten while the pan is fitted. The pan is located in this manner
to help prevent accidental removal by animals.
5.7 Your intensive care unit is supplied with air filter media fitted at the air inlet (see fig 1) and at each
end of the heater enclosure (see picture 9 above).
5.8 Fit the nebuliser solution holder by inserting the "mouthpiece" through the hole in the cabinet and
locating the solution holder pipe in the slot in the bracket.
5.9 Place the intensive care unit on a scratch and moisture resistant surface. Worktop height is ideal.
Clear the area around the unit of any electrical equipment. Ensure the unit is not within 90cm of a
power socket as oxygen gas may spread and cause a fire in power sockets or other equipment.
5.10 Fit a PP3 type battery in the compartment on the rear of the control housing (see section 11 for
important safety information and details).
5.11 Connect the mains supply cord to a power socket that is easy to reach.
The supply cord shall be placed or protected so as not to be accessible by animals.
5.12 The intensive care unit is now ready for use. The next section of these instructions includes a
checklist that should be photocopied so that it may be completed for each use.
8
9
Air Filter

HD453 GB Issue 08
9
6 Control and Alarm System Operation
Confirm a menu option or toggle the interior lighting on/off
Scroll through the menu options, reduce a setting or silence the incubation temperature alarm
Scroll through the menu options or increase a setting
Start (switch on the heater control system)
Test the power fail and fan condition alarm system
Silence the power fail and fan condition alarm system
Please refer to the Start-up Check Sheet while familiarising yourself with these instructions.
Each step of the Start-up Check is to ensure the safety of the animal and the operator. Failure to
check each item could result in a hazard. If at any stage the behaviour of the control system is
not as stated then call Brinsea Products for more information and discontinue use of the unit.
Never ignore or attempt to bypass the alarm system, this may result in a serious hazard.
6.1 Check and sign off the first 13 steps of the Start-up Check Sheet which confirm the unit is in correct
condition and a safe location. The control system alarm and safety systems must now be tested:
6.2 Apply power to the unit.
The ventilation fan must operate but the heating control system should not function.
The display and both indicator lamps must be off.
T*30.0C
H*40%

HD453 GB Issue 08
10
6.3 Press the start button .
The green indicator lamp will illuminate to show the heating control system is now active.
The incubation temperature alarm will sound momentarily with a pulsed tone to show the alarm
sounder is functioning correctly.
After briefly showing the software version screen the display will revert to normal operation and
show the current temperature and humidity in the cabinet.
6.4 Test the power failure and fan function alarm by pressing the alarm test button .
The alarm must sound in a continuous tone.
The green heater control indicator lamp must go off.
The red alarm indicator lamp must illuminate.
Open the door and confirm the fan is now blowing air that is getting cooler (the heating element is
switched off).
6.5 Press the alarm silence button for at least 2 seconds .
6.6 Close the door and press the start button to reset the alarm and switch the heating control back
on.
6.7 Check the set temperature is appropriate for the patient and circumstance (see section 7). Allow
the appliance to run for at least 20 minutes to stabilise the temperature before introducing animals
or starting oxygen supply. The unit is now ready for use.
See section 11 for further checks when using the unit for administration of oxygen.
If any stage of the Start-Up Check results in a failed test or inspection do not use. Seek advice
from Brinsea Products.
The following part of this section describes the digital display modes and alarm system.
6.8 CHANGING SETTINGS –The Main Menu allows the various settings to be modified and saved. All
changes are retained in the event of a power cut.
To access the Main Menu press the and buttons simultaneously to unlock the display. For full
details of menu settings please refer to the guide on page 12.
6.9 NORMAL OPERATION –Temperature and Relative Humidity are continuously displayed. The
green heater control indicator is illuminated.
6.10 The asterisk “*” adjacent to the temperature reading shows when the heater power is on. When
warming the asterisk will be continuously on, once warmed up the asterisk will slowly flash as the
heater is pulsed to maintain the correct temperature. When reducing the temperature setting the
asterisk may go off, this is normal.
The asterisk “*” adjacent to the relative humidity display is only on when the pump control output is
on (see section 8) and is only applicable when using the optional Brinsea Advance Humidity Pump.

HD453 GB Issue 08
11
6.11 HIGH TEMPERATURE ALARM DISPLAY –If the measured temperature goes up by more than
the figure in the ALARM HI screen, the alarm will sound immediately and “+T” will be displayed.
Press the “-“ button to silence the alarm for 30 minutes.
Check the intensive care unit is not (and has not been) in direct sunlight or too near a heat source
such as a room heater. An animal's metabolic heat may also cause temperature to increase
beyond the set temperature if set close to room temperature conditions.
If the high temperature problem rectifies itself the “+T” remains on the display to show this has
happened. Press to clear the indicator.
If the cause is not immediately apparent discontinue use and contact Brinsea Products.
6.12 LOW TEMPERATURE ALARM DISPLAY –If the measured temperature goes down by more than
the figure in the ALARM LO screen, after 30 minutes “-T” will be displayed and the alarm will
sound. Press the “-“to silence the alarm for 30 minutes.
Check the intensive care unit is not (and has not been) in a cold draught or that the room
temperature has dropped significantly.
If the low temperature problem rectifies itself the “-T” remains on the display to show this has
happened. Press to clear the indicator.
If the cause is not immediately apparent discontinue use and contact Brinsea Products.
T 39.8C
H 23% +T
T*32.1C
H 45% -T

HD453 GB Issue 08
12
6.13 POWER FAILURE / HEATER CONTROL SYSTEM ALARM
If the alarm sounds and the red alarm indicator lamp is illuminated disconnect the oxygen
supply, and open the door to ventilate the care unit. Do not use until the fault has been
investigated and cleared:
Try another power supply;
replace fuse.
Is the mains power
on and fuse in plug
okay?
Press
alarm
reset.
Is the fan running?
DISCONTINUE USE
AND CONTACT
BRINSEA
PRODUCTS.
Press the
start
button.
Does the heater
control initiate and
the green indicator
illuminate?
Make sure the unit is not near a source
of heat. Wait for the air temperature to
stabilise at the set control temperature.
Is the air
temperature
correct and stable?
Continue use and reconnect oxygen
supply if required. Make a note of the
alarm situation on the Start-up Check
Sheet. A repeat alarm that cannot be
attributed to a power interruption must
be investigated. CONTACT BRINSEA
PRODUCTS FOR ADVICE ON
SERVICING THE UNIT.
No
No
No
No
Yes
Yes
Yes
Yes

HD453 GB Issue 08
13
MAIN MENU
TEMP
OK →
RH%
OK →
ALARM HI
OK →
ALARM LO
OK →
C/F
OK →
SAVE
OK →
CANCEL
OK →
30.0C
OK - +
RH 20%
OK - +
HI 2.0C
OK - +
LOW 3.0C
OK - +
DISP C
OK C F
TEMPERATURE.
RANGE 20.0 –38.0°C (68.0 –100.4°F).
DEFAULT 30.0°C (86.0°F).
RELATIVE HUMIDITY.
RANGE 20% –80%.
DEFAULT 20%.
ONLY FOR USE WITH HUMIDITY PUMP
–SEE SECTION 8.
HIGH TEMPERATURE ALARM. RANGE
1.0 –5.0°C (1.8 –9.0°F) ABOVE SET
INCUBATION TEMPERATURE.
DEFAULT 2.0°C (3.6°F).
LOW TEMPERATURE ALARM. RANGE
1.0 –5.0°C (1.8 –9.0°F) BELOW SET
INCUBATION TEMPERATURE.
DEFAULT 3.0°C (5.4°F).
CELSIUS / FAHRENHEIT DISPLAY.
SWITCHES ALL TEMPERATURE
FIGURES BETWEEN °C AND °F.
DEFAULT °C.
SAVE.
ALL CHANGES ARE SAVED. RETURN TO
NORMAL OPERATION SCREEN.
CANCEL.
ALL CHANGES ARE IGNORED. RETURN
TO NORMAL OPERATION SCREEN.
PRESS BOTH BUTTONS TO UNLOCK THE MAIN MENU
SELECT THE OPTION / RETURN TO THE MENU.
GO FORWARD ONE SCREEN / INCREASE THE VALUE / DISPLAY IN CELSIUS.
GO BACK ONE SCREEN / DECREASE THE VALUE / DISPLAY IN FAHRENHEIT.

HD453 GB Issue 08
14
Vetario T40M and T50M Start-Up Check Sheet. Issue 02, 19/06/2013
Date:
Sheet 1 of 2.
Checked by:
The following checks must be performed before every use to ensure the safety of patients and
operators. Failure to perform these checks may lead to serious injury or death of the patient or
operator.
Read the instruction manual before using the unit for the first time and periodically so that you
are familiar with all functions and safety precautions.
Oxygen concentrations above normal atmospheric conditions create a greatly increased risk of
fire, including immediately outside the apparatus. Materials that do not burn in air may react and
burn vigorously in an oxygen enriched atmosphere.
TICK the relevant box for a pass or fail. If any item is unclear or fails the inspection seek advice
from the responsible person or Brinsea Products and do not use the intensive care unit.
Check cabinet for cracks or gaps that could allow oxygen gas to escape.
Check the joint between the upper and lower cabinet halves for gaps. Check
the electrical enclosure protective cover is secure and undamaged.
Pass:
Fail:
Check the room is ventilated (open window or door) or oxygen could build up
and be a fire hazard. Do not use in a confined space, there must be 30cm of
free space to both sides and above the unit.
Pass:
Fail:
Check there are no sources of heat near the unit (e.g. direct sunlight or heating
appliance). Sunlight can move round a room. This could over-heat the animal
inside.
Pass:
Fail:
Ensure there are no electrical devices in the unit such as heat pads. These can
be a fire hazard if oxygen is being used.
Pass:
Fail:
Check there are no electrical devices within 90cm of the unit, including power
sockets. These can be a fire hazard if oxygen is used.
Pass:
Fail:
Check there are no electrical devices or sockets under the table or shelf the
unit stands on. Oxygen gas can sink through air and may cause fire in other
electrical devices beneath the unit or its shelf.
Pass:
Fail:
Do not cover the unit. Make sure there is nothing that can fall onto the unit.
This could cause a fire hazard.
Pass:
Fail:
Check there are no traces of liquid left in the cabinet. Clear fluids are difficult to
see, wipe inside the base with tissue if necessary. Even trace amounts of
flammable liquids such as ether or alcohol may cause fire or explosion in
connection with oxygen.
Pass:
Fail:
Check the hinges and door latch are secure. Loose parts may allow an animal
to escape.
Pass:
Fail:
Check all 3 filters are fitted and are clean. Blocked or missing filters will affect
the temperature and could create a fire hazard.
Pass:
Fail:
Clean the water pan and replace evaporation block if dirty.
Pass:
Fail:
Fill the water pan (if humidifier is required).
Pass:
Fail:
Apply power, the fan must start but the lamps and display must be off.
Pass:
Fail:
Press start , the temperature alarm must pulse for several seconds. (The
interior lights will briefly flash once).
Pass:
Fail:
The green indicator must be on. The red indicator must be off.
Pass:
Fail:
Continued on sheet 2

HD453 GB Issue 08
15
Vetario T40M and T50M Start-Up Check Sheet. Issue 02, 19/06/2013
Date:
Sheet 2 of 2
Checked by:
Test the fan and power fail alarm system by pressing the alarm test button .
The alarm must sound and red indicator lamp must illuminate.
Pass:
Fail:
Press the alarm silence button for 2 seconds. The alarm shall be silent.
Pass:
Fail:
Open the door and check the fan is now blowing air that is getting cooler
(heater is off).
Pass:
Fail:
Close the door, press start and check the set temperature is correct.
Pass:
Fail:
Does the air temperature stabilise at the set temperature after approximately
20 minutes? Do not use if temperature is unstable, check set temperature if air
temperature stabilises at unexpected figure. Note that the temperature may
briefly overshoot depending on room condition and settings.
Pass:
Fail:
The intensive care unit may now be used for general therapy. Check the door latch is fastened.
Perform the following additional checks before administration of oxygen gas:
Remove the tube from the nebuliser solution holder if fitted. This prevents
oxygen leaking back to a nebuliser pump and creating a fire hazard.
Pass:
Fail:
Disconnect the water pump control lead and water delivery tube if fitted.
Pass:
Fail:
Only use cotton based bedding as this reduces the risk of fire in oxygen.
Pass:
Fail:
Check the oxygen supply hose for splits or cracks that could leak and create a
fire risk.
Pass:
Fail:
Ensure the oxygen supply hose connection “barb” is tightly fitted and therefore
cannot leak or fall off.
Pass:
Fail:
Ensure the oxygen supply hose clamp is tight and that the hose cannot be
pulled off.
Pass:
Fail:
Set the pressure regulator to a maximum of 400kPa (50psi) –follow
instructions supplied with the regulator.
Pass:
Fail:
Set the oxygen flow meter to 1 lpm (One litre per minute) –follow instructions
supplied with the flow meter. A sustained higher flow rate will increase the
oxygen concentration to a level that may cause permanent injury to the animal
or its death. A lower flow rate will make therapy ineffective.
Pass:
Fail:
Use only patient connections (e.g. pulse-oximeter) that are approved for use in
an oxygen enriched atmosphere. Unapproved connections may cause a fire.
Pass:
Fail:
Monitor the animal frequently.
Ensure large, agitated animals do not damage the door or cabinet. Do not leave unattended until
the animal is settled. Do not move the unit with an animal inside.
Frequently check the exhaust ventilation hole on the door is not blocked.
Frequently check the oxygen flow rate and supply pressure.
If during use the alarm sounds or the red alarm indicator is illuminated disconnect the oxygen
supply, disconnect power and open the door to ventilate the care unit. Do not re-use until the
fault has been investigated and rectified (see section 6.13 in manual).

HD453 GB Issue 08
16
7 Temperature
The metabolic heat from the patient(s) will contribute heat to the intensive care unit.
The intensive care unit may not control properly if the room temperature is less than 3ºC (10°F)
lower than the temperature required inside the unit.
7.1 Note: your intensive care unit may not be set to the correct temperature from the factory and the
following procedure must be followed before introducing animals.
7.2 As the unit warms up and approaches its control setting the heater on asterisk “*” will change from
continuously on to flashing.
7.3 Press the and buttons simultaneously to unlock the main menu. Press to select the
temperature screen and adjust as necessary using the and buttons. Press to return to the
Main Menu and then scroll down to Save. Press to save the changes. When reducing
temperature the asterisk may go out while the intensive care unit cools –this is normal.
7.4 Refer to the digital temperature display to check temperature. The display shows the air
temperature in increments of 0.1°.
7.5 The Display can be switched to show all temperature settings in degrees Fahrenheit. Press the
and buttons simultaneously to unlock the main menu. Scroll to the C/F option and press to
select the C/F display screen. Press the button to select °F or the button to select °C. Press
to return to the Main Menu and then scroll down to Save. Press to save the changes.
7.6 For most applications involving intensive care the unit should be set to between 30 and 35ºC (86
and 95ºF). Depending on room temperature, it should take approximately 20 minutes to stabilise
from cold. Note that the temperature should be gradually reduced to room temperature (20 –25ºC
or 68 –77ºF) as the patient recovers to avoid sudden temperature change when the patient is
removed.
7.7 The Vetario T40M and T50M have a built-in temperature alarm which warns of high or low
incubation temperatures in the chamber. See section 6 for details.
7.8 In the unlikely event of a control failure the heater is automatically switched off by a safety
thermostat. If this operates the alarm will sound and the red alarm indicator LED will illuminate.
This must be investigated before further use. Switch off the oxygen supply, disconnect power and
press the alarm reset button to silence the alarm. If oxygen has been supplied allow time for
oxygen to disperse from the animals bedding and fur. Move the animal to an alternative unit and
contact Brinsea Products for advice on testing the intensive care unit.

HD453 GB Issue 08
17
8 Humidity and Ventilation
Check the water level at least daily to avoid the air becoming too dry (low relative humidity).
Avoid spillage of water near electrical parts, use the funnel provided. Clean up any spilled water
from the top of the unit immediately.
Inspect air filters before and after each use for dust / dirt and clean as necessary (see section
13). Blocked filters restrict air flow and may lead to increased CO2level, incorrect temperature
and incorrect oxygen concentration. Dirty filters may hold bacteria.
Do not use without the heater enclosure air filters in place. Dust and dirt will rapidly contaminate
the heater and sensor surfaces and create a possible fire hazard.
Inspect the evaporation block weekly and replace if dirty. The block can provide a breeding
ground for bacteria. In addition to the use of water-based disinfectant in the water, it is
recommended that the block is replaced every 2 months of use.
Empty the water pan after each use and disinfect to avoid bacterial growth.
8.1 Elevated air temperatures in the intensive care unit will reduce the relative humidity level (RH) and
can cause dehydration. A water reservoir is fitted to counteract this effect.
8.2 The care unit is fitted with a water reservoir (see fig.1) which humidifies air as it is drawn into the
heater enclosure. Use a solution of proprietary water-based disinfectant (diluted in accordance with
the manufacturer's instructions) in the water reservoir (pan) to inhibit bacterial build-up. It is
recommended that the reservoir is topped up with solution daily to reduce dehydration. This can be
a particular problem with very young birds.
8.3 The unit and its occupants need not be disturbed to fill the water pan. Use the funnel provided to
pour water through the water fill point in the top of the unit (see fig. 1) directly onto the evaporating
block and into the water pan. Push the funnel down gently into the hole to make sure the water
pours directly in. Clean up any spilled water from the top of the unit immediately.
8.4 To further increase humidity levels within the unit the water reservoir is fitted with an evaporating
block of absorbent paper mesh as standard. This block may be set across the water pan (instead
of upright) or removed entirely to give lower humidity levels if condensation forms. The block can
provide a breeding ground for bacteria. In addition to the use of water-based disinfectant in the
water, it is recommended that the block is replaced every 2 months of use.
8.5 An adjustable vent is fitted to the door which may be opened or closed to give greater control of
humidity (close vent to increase) and fresh air flow. The vent may be fully closed as fixed
ventilation is also provided.
8.6 The Brinsea Advance Humidity Pump is available as an option for the Vetario TM models. The
digital control system not only reads the humidity in the unit but it provides a control signal to
operate the water pump and accurately maintain the humidity level at the desired level.

HD453 GB Issue 08
18
9 Introducing your patients
The intensive care unit relies on mains power. Monitor the intensive care unit frequently and
switch off the oxygen supply if electrical power fails. Remove the animal if the power failure
continues for a prolonged period. In critical situations use an uninterruptible power supply (UPS)
with “true sine wave”output suitable for use with medical devices. Always verify the UPS and
intensive care unit operate correctly together as some UPS devices may cause poor temperature
regulation.
Only introduce patients once the correct temperature has been established and is stable.
Be sure to check the door latch is secured correctly. An unsecured latch may allow the animal to
escape and lead to a consequent fall or other injury.
Do not lift the unit with an animal inside. This may damage the cabinet and allow oxygen to leak,
or may cause the animal distress.
A large animal that is very agitated may cause damage to the cabinet parts or door. Do not leave
unattended until the animal has settled. Broken or cracked parts may allow oxygen gas to
escape and create a fire hazard. Broken or cracked door hinges or latch may allow the animal to
escape and lead to a consequent fall or other injury.
Monitor patients and device settings / displays frequently so that incorrect operation or
deterioration of patient condition is detected promptly. Remain within hearing distance of
alarms.
Wipe up fluid spillages in the animal chamber promptly to avoid splashed liquid entering the
heater enclosure.
9.1 Patients of similar age and size can be placed together directly in the base of the cabinet and
benefit from the warmth and comfort.
9.2 If disparity in size is too great a smaller patient is at risk of being crushed or smothered and it may
be necessary to place patients in separate tubs within the cabinet.
9.3 In order to maintain maximum air-flow the air filters should be checked once a week for dust or
down, and cleaned if necessary.
10 Internal Lighting
10.1 The Vetario TM brooders are equipped with gentle, internal LED lighting for night-time inspection of
animals and birds. The LED lamps are energy efficient, do not affect temperature and would not
normally require replacement. Amber colour LED's do not produce UV light often associated with
"white" LED's.
10.2 The LED lights are located to the left of the fan diffuser in the T40M size model and on both sides
of the fan diffuser in the larger T50M model.
10.3 The lights may be switched on and off by pressing the button. This has a toggle action.

HD453 GB Issue 08
19
11 Oxygen Administration
WARNING. RISK OF SERIOUS INJURY OR DEATH OF OPERATOR OR PATIENT:-
VETARIO TM INTENSIVE CARE UNITS HAVE BEEN SPECIFICALLY DESIGNED AND TESTED
FOR SAFE OPERATION WITH INCREASED OXYGEN CONCENTRATION OF AIR WITHIN THE
CABINET. ALL OPERATIVES MUST READ AND UNDERSTAND THE SAFETY INSTRUCTIONS.
Complete a Start-Up Check Sheet each time the unit is used for oxygen service.
Oxygen concentrations above normal atmospheric conditions create a greatly increased risk of
fire, including immediately outside the apparatus. Materials that do not burn in air may react and
burn vigorously in an oxygen enriched atmosphere.
No auxiliary equipment shall be placed in the incubator. Only use patient connections that are
designed for safe use in an oxygen enriched atmosphere. If in any doubt check with the
manufacturer of the auxiliary equipment. (Fire may be caused by sparking contacts or hot
surfaces in equipment not designed for oxygen service).
Even small quantities of flammable agents, such as ether and alcohol, left in the incubator can
cause fire in connection with oxygen. (Never use flammable cleaners or disinfectants and ensure
such agents are not introduced on a patient).
The incubator must only be used in a well-ventilated room away from potential sources of heat or
ignition. NO SMOKING. Do not use in a confined location, the incubator needs at least 30cm of
free space above and to both sides. (Oxygen concentration in the air around the incubator may
rise significantly if ventilation isn’t provided or the space around the incubator is confined. This
may create a fire hazard).
Remove all electrical devices (including the water pump) that are within 90cm of the incubator.
Place the incubator at least 90cm away from power sockets. Ensure there are no power sockets
or electrical items directly under the shelf or table supporting the incubator. (Oxygen gas may
sink through air and cause fire in other equipment).
Always test the fan rotation and power failure alarms and check ventilation holes before using
with oxygen gas. If the alarm sounds during use disconnect the oxygen supply, disconnect
power and open the door to ventilate the care unit. Do not re-use until the fault has been
investigated and rectified. (Failure of the ventilation fan or blocking of ventilation slots/holes may
lead to oxygen enrichment of air in unintended areas leading to an increased risk of fire. It may
also lead to increased concentration of CO2in the animal cabinet).
LED lamp
location
(T40M and
T50M)
LED lamp
location
(T50M
only)

HD453 GB Issue 08
20
Not suitable for use in the presence of a flammable anaesthetic mixture. (The incubator is not
intended to be used in a location where flammable anaesthetics may be present. This would
create a serious risk of fire or explosion).
Before oxygen administration, disconnect any nebuliser and tube from the solution holder. (This
is to prevent oxygen flowing into the nebuliser pump and creating a fire hazard).
If fitted, disconnect the water pump control lead and water delivery tube. (The pump is not
intended for use during oxygen therapy and may create a fire hazard).
Do not cover the intensive care unit. (Blocking of ventilation slots/holes may lead to oxygen
enrichment of air in unintended areas leading to an increased risk of fire. It may also lead to
increased concentration of CO2in the animal cabinet. Covering the unit may lead to fire hazard
due to reduced dissipation of heat).
Do not obstruct ventilation holes. (Blocking of ventilation slots/holes may lead to oxygen
enrichment of air in unintended areas leading to an increased risk of fire. It may also lead to
increased concentration of CO2in the animal cabinet).
Do not exceed 1 lpm (litre per minute) oxygen flow rate. (The maximum safe long-term
concentration of oxygen for canines and felines is achieved at this flow rate and increased flow
may lead to permanent injury or death. Greater flow rates also increase the risk of oxygen
leaking in the event of a fault and could create a fire risk).
The incubator contains no flow restriction or pressure regulator device. The oxygen supply must
be reliably restricted to 400kPa (50psi). Always use a medical grade oxygen pressure regulator
and flow meter that has been serviced in accordance with the manufacturer’s instructions.
Failure of an external device may create a fire or pressure hazard or prevent effective treatment.
It is assumed the operator will be trained in the safe and correct use of such devices.
After oxygen administration, animals and any associated cloth materials must be allowed
sufficient time in the chamber in normal air to allow trapped oxygen to disperse. (Fabric and
similar materials that trap oxygen enriched air may burn vigorously if ignited by a spark or other
source of ignition). Use only pure cotton materials to reduce the risk of static electricity sparks.
Perform the following tests and safety checks before EVERY use for oxygen administration and fill in the
Start-Up Check Sheet.
11.1 Check the cabinet, door and hinges/lock for cracks, damage or loose parts that may allow oxygen
gas to escape and create a fire hazard. Check for gaps between the upper and lower halves of the
white cabinet.
11.2 If fitted, disconnect the flexible tube from the nebuliser solution holder to prevent any flow of
oxygen to a nebuliser pump. This could create a fire hazard.
11.3 Remove the water pan and evaporating block. Disconnect the Brinsea Advance Humidity Pump
tube and electrical cable if fitted. Please note that it is not practical to use the pump during oxygen
administration due to capacity restraints. A hydrated oxygen supply is commonly used and helps
humidify the animal chamber.
11.4 Fit the oxygen hose barb to the inlet hole using the locknut supplied. Check for tightness if already
in place, there must be no gap or oxygen gas could leak and create a fire hazard. The hose barb is
suitable for hoses with a 6mm (1/4 inch) bore size. Always use a hose clamp device and ensure it
is tightly fitted so that the hose cannot possibly fall off.
Hose Barb. (Fit
locknut on inside of
cabinet).
This manual suits for next models
1
Table of contents
Other Vetario Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Organ recovery systems
Organ recovery systems LifePort Kidney Transporter LKT100P Operator's manual

Dermlite
Dermlite Lumio 2 instructions

StayClear
StayClear Handpiece quick start guide

KYRA MEDICAL
KYRA MEDICAL k-tek KTEK7300 Instructions for use

Braun
Braun Acculan 3Ti Instructions for use

EVER Pharma
EVER Pharma D-mine Care manual