
VR8500
8
5. Click the "Auto" button. The output level should slowly ramp up until the shaker approaches a 0.1-
inch peak-to-peak displacement (0.05 inches if using the Servo Hydraulic recommended set-up).
6. If the shaker does not move, click the "Stop" button to turn the output off. Verify that the Drive
output connector from the VR8500 is
connected to the shaker amplifier's input,
that the amplifier is turned on and that
the amplifier output is connected to the
shaker. Return to step 4.
7. If the shaker still does not move, connect
the Drive output connector of the
VR8500 to an oscilloscope or true RMS
voltmeter, set the output voltage level to
0.5 volts (0-to-peak) and verify that you
have a 0.5 volt (0-to-peak) amplitude
reading (0.35 volts RMS). If the Drive
output connector is not measuring 0.35
volts RMS, your system may require
calibration.
8. If the shaker vibrates, but the input
waveform remains flat, click the "Auto"
button a second time so that the text on
the "Auto" button is dark and the output stops ramping up. Check the accelerometer cables to verify
that they have good connections and that the accelerometer conditioning equipment is turned on. If
you are using the internal current source to power the accelerometer, verify that "Accel power
source" on Configuration..Inputs dialog box is checked. If there still is no input signal, connect the
conditioned accelerometer signal to an oscilloscope or voltmeter and verify that you see a voltage
reading. If there is no reading, click the "Stop" button to turn the output off, replace the
accelerometer cable, accelerometer and/or the accelerometer conditioning equipment with devices
known to be working and return to step 4. Note:
most problems are due to bad cable connections.
9. Once you get both an output and an input signal,
use a displacement measurement tool to verify that
the shaker peak-to-peak displacement matches the
value shown in the System Check Control Center.
If it does not match, select the
Configuration..Inputs menu command to verify
that the accelerometer sensitivity settings match
the calibrated values for the accelerometers you
are using.
Note: To use the V Scope Displacement Measurement Tool – tape or hold firmly on shaker head so
that it moves in the direction of vibration (see arrows). The lines will blur and intersect at the
0.1 inch mark on the measurement tool if the shaker is moving as directed above (0.05 for the
servo hydraulic set-up). If the lines don’t intersect there, check that your accelerometer
sensitivities are set correctly (and power source on, if required).