Routine maintenance
Desiccant Replacement
The desiccant removes excess moisture from the optical head.
XDisconnect the power supply from the bottom of the head.
XRemove the screw securing the desiccant cover u, lift the cover off
and remove the old desiccant.
XInsert the new desiccant and replace the desiccant cover.
LED Arrays
XRemove the desiccant cover u.
XRemove the two crosshead screws that secure the LED array vand
remove the array complete with lters w.
XReassembly is the reverse of this procedure.
General care
XCover your Mantis with a dust cover when not in use.
XRemove dust with a soft brush or cleaning cloth.
XThe Mantis viewing screen and lenses should be cleaned with a lens cleaning cloth.
XKeep accessories in a dust-free environment when not in use.
Service
Service and repair work must only be carried out by service engineers authorised by Vision Engineering.
Specication
Objective Lenses
Objective Lenses Working Distance Field of View
x2 167mm 45.0mm
x4 96mm 27.5mm
x6 73mm 19.2mm
x8 58.5mm 14.3mm
Lighting Data
LUX intensity measured at subject plane with colour correction lters.
20 LEDs 9,400 LUX 10,000 hours
Substage illumination (bench stand only)
58 LEDs 2,700 LUX 10,000 hours
Warranty
This product is warranted to be free from defects in material and workmanship for a period of one year from the date of invoice to the original purchaser.
If during the warranty period the product is found to be defective, it will be repaired or replaced at facilities of Vision Engineering or elsewhere, all at the option of Vision
Engineering. However, Vision Engineering reserves the right to refund the purchase price if it is unable to provide replacement, and repair is not commercially practicable or
cannot be timely made. Parts not of Vision Engineering manufacture carry only the warranty of their manufacturer. Expendable components such as fuses carry no warranty.
This warranty does not cover damage in transit, damage caused by misuse, neglect, or carelessness, or damage resulting from either improper servicing or modication by
other than Vision Engineering approved service personnel. Further, this warranty does not cover any routine maintenance work on the product described in the user guide or
any minor maintenance work which is reasonably expected to be performed by the purchaser.
No responsibility is assumed for unsatisfactory operating performance due to environmental conditions such as humidity, dust, corrosive chemicals, deposition of oil or other
foreign matter, spillage, or other conditions beyond the control of Vision Engineering.
Except as stated herein, Vision Engineering makes no other warranties, express or implied by law, whether for resale, tness for a particular purpose or otherwise. Further,
Vision Engineering shall not under any circumstances be liable for incidental, consequential or other damages.
Power Supply
100-240VAC 50-60HZ 1.0A Max