Topography Pachymetry Tonometry
For more detailed instructions
and special features please refer
to the specific user manuals on
https://academy.visionix.com/
VER.2.00
Document non contractuel. ©kosept/Visionix - ©Photos : Visionix - 12/22 - Features and specifications are subject to change without prior notice
www.visionix.com
2
VX 120+ Dry eye
How To Retake Measurements How To Add Measurements
You can only retake/add measurements before you save results
Quick start
For more detailed instructions
and special features please refer
to the specific user manuals on
https://academy.visionix.com/
Manual Focus
These should be as
clear/dark as possible
This should align with
center of corneal apex.
Moves focus farther
from eye Lowers apex
Moves
focus closer
to eye
Raises
apex
«Acquire»
Initiates
measure-
ment
«Acquire»
Initiates
measure-
ment
Misaligned circles
Tap center of gray
circle to align
circles.
Blurry gray circle
Tap «+» button
repeatedly until
focused
Step 1: Center/focus rings &
tap «Acquire»
Step 1: Center/focus rings
(as in topography manual
mode)
Step 2: Align corneal apex
and tap «Acquire»
NOTE: If you have trouble with manual focus, tap to restart.
«Manual Focus» activates if the device is unable to autofocus
Tap «Measure» on
summary screen
Tap «Measure» on
summary screen
Select test
to retake
Select
diagnostic
Ta p
«OK»
Retake
measurement
Perform
measurements
NOTE: Repeating a measurement erases previous results. NOTE: A new diagnostic is created in patient record.
Note: Stand to the side when using manual focus to ensure device does not touch the patient.
General Troubleshooting
Problem Possible Causes Suggested Solutions
• Images are not centered
• Windows password expired
• Tonometer unable to center
• Tickets do not print
• Measured pupil size different from
expected/actual OR
High-order aberrations
• Patient/camera alignment
• Lights or windows near unit
• Password expires every 40 days
• Initial patient position not correct
• Paper jam
• Out of paper
• External lights/windows near unit
• Large patches of light near unit
• Direct light falling on unit/patient
• Realign and redo measurement
• Minimize lights or windows near unit
• Cover window or patient if necessary
• Reset password to «user» - Don’t
change!
NOTE: You must enter the password
3 times to reset and will need to scroll
down to see the third entry field
• Readjust chinrest or restart measure-
ment
Open paper compartment:
• Remove any loose/jammed paper
• Replace paper, feeding printer from
the top of roll
• Ensure that there are no bright lights
or windows near the unit/patient
• Cover window or patient if necessary
NOTE: If you have trouble with manual focus, tap to restart.