ENG
19Page of 28
09140 Issue 1
In2itive e-Diary - Instructions for Use
supplying 500mA must be used or it will not charge the device
fully or charging will take longer than 5 hours.
16. The Vitalograph In2itive e-Diary owhead is intended to be a
single patient use device. If, for any reason, a device is used on
more than one subject (e.g. for training demonstrations) it is
intended that a new single use bacterial viral lter is used per
subject to protect both the device and the subject from cross
contamination.
17. To isolate the Vitalograph In2itive e-Diary from the mains
power supply, remove the 5V Power Supply from the mains
wall socket. When charging the device via the 5V Power Supply
do not position the 5V Power Supply where it is dicult to
unplug from the socket.
18. The device contains a Lithium coin cell battery and a Lithium
ion main battery, neither of which are accessible by the user
and are not serviceable parts. Any suspected battery faults
should be reported to the manufacturer.
19. The internal Lithium ion battery is not user accessible,
replaceable or serviceable. In the unlikely event that any
issues are noted with the power or battery such as swelling or
leaking, stop using the device immediately, do not charge the
device and contact the site administrator or the manufacturers’
support. In case of leaking ensure the electrolyte does not get
in the eyes or touch skin as it may cause damage. If electrolyte
gets into the eyes ush the area immediately with water for 15
minutes and seek medical attention. If it gets on the skin wash
the affected area immediately and seek medical attention. Do
not inhale the leaked material, leave the area immediately and
allow the battery to cool and any material to dissipate.
20. Keep the device away from thermal ignition sources including
but not limited to, res, cookers and heaters. The ISO 26782
spirometry standard species a temperature of 17-37oC
for spirometry testing. The Vitalograph In2itive e-Diary also
contains a lithium ion battery and excessive heat may cause
damage or swelling.
21. The power supply supplied with the device has a long cable
which may be a strangulation hazard and therefore should be
kept out of the reach of children and pets.